PHILIPS MICRODOSE

K120255 · Philips Digital Mammography Sweden AB · MUE · Apr 19, 2012 · Radiology

Device Facts

Record IDK120255
Device NamePHILIPS MICRODOSE
ApplicantPhilips Digital Mammography Sweden AB
Product CodeMUE · Radiology
Decision DateApr 19, 2012
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1715
Device ClassClass 2
AttributesReal-World Evidence

Real-World Evidence

SubmissionDeviceSponsorRWD SourcesRWE Use SummaryKey Tags
K120255 · Apr 19, 2012PHILIPS MICRODOSEPhilips Digital Mammography Sweden ABPublished clinical literature; National mammography screening programsThe sponsor utilized published clinical evidence from national screening programs to demonstrate consistent levels of dose reduction with the MicroDose system compared to other FFDM and film-screen systems in routine clinical use.Dose reduction; Screening programs; Literature review

Clinical Evidence

Study DesignPopulationComparatorKey Endpoints
Published studies on various national mammography screening programs; Retrospective analysis of national screening program dataPatients undergoing routine mammography screeningOther FFDM and film screen systemsDose reduction

Indications for Use

The Philips MicroDose L30 is indicated for generating mammographic images that can be used for screening and diagnosis of breast cancer. The Philips MicroDose L30 is intended to be used in the same clinical applications as traditional film/screen systems.

Device Story

Full-field digital mammography system; utilizes photon counting technology with crystalline silicon strip detectors; incorporates multi-slit scanning technique to reject scattered radiation. System components include x-ray tube (tungsten target, aluminum filtration), high voltage generator, compression paddles, and acquisition workstation. Operated by clinicians in clinical settings. Provides three exposure modes: manual, automatic (based on breast thickness), and SmartAEC (continuous adjustment based on transmission). Captures x-ray photons, converts to electrical signals, and generates digital images for PACS transfer. Enables dose-efficient imaging; supports screening and diagnostic workflows. Benefits include reduced radiation dose compared to traditional systems while maintaining image quality.

Clinical Evidence

Clinical evidence includes published studies on national mammography screening programs demonstrating consistent dose reduction compared to other FFDM and film-screen systems. Non-clinical bench testing (Sensitometric response, spatial resolution, noise analysis, DQE, dynamic range, AEC performance, phantom testing, radiation dose) confirmed performance equivalent or superior to predicate.

Technological Characteristics

Photon counting technology using crystalline silicon strip detectors; multi-slit scanning; tungsten target x-ray tube with aluminum filtration. System includes gantry and acquisition workstation. Operates via manual, automatic, or SmartAEC exposure modes. Testing performed per IEC standards.

Indications for Use

Indicated for screening and diagnosis of breast cancer in patients requiring mammographic imaging. Intended for use in clinical applications equivalent to traditional film/screen systems.

Regulatory Classification

Identification

A full-field digital mammography system is a device intended to produce planar digital x-ray images of the entire breast. This generic type of device may include digital mammography acquisition software, full-field digital image receptor, acquisition workstation, automatic exposure control, image processing and reconstruction programs, patient and equipment supports, component parts, and accessories.

Special Controls

*Classification.* Class II (special controls). The special control for the device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Full-Field Digital Mammography System.”*See* § 892.1(e) for the availability of this guidance document.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ 120255 # 510(k) Summary (21 CFR 807.92) APR 1 9 2012 #### A. SUBMITTER INFORMATION · Philips Digital Mammography Sweden AB Submitter's name Address Smidesvägen 5 SE-171 41 Solna Sweden Establishment registration 3009307584 number Contact person Gustav Lins Telephone +46-8-623 52 00 FAX +46-8-623 52 01 Title Manager Regulatory Affairs Date of the summary preparation 2012-01-24 #### B. DEVICE IDENTIFICATION Device trade name Device common name Classification name Classification product code Device class Regulation code Philips MicroDose L30 Full-Field Digital Mammography X-ray System Full-Field Digital Mammography X-ray System MUE = 21 CFR 892.1715 ## C. DEVICE DESCRIPTION The Philips MicroDose L30 is a full-field digital mammography system comprised of an image acquisition system, a gantry and an acquisition workstation computer equipped with a keyboard, a keypad, a mouse, and a monitor. The image acquisition system includes a digital detector of photon counting technology, x-ray tube (with tungsten target and aluminum filtration), high voltage generator, compression paddle(s), and multi-slit collimator. The acquisition is the user interface for preparing and initiating image acquisition, image processing, and image transfer to the desired destination (e.g. PACS) for diagnosis and archiving. The Philips MicroDose L30 detector is based on photon counting technology and consists of a large number of crystalline silicon strip detectors. The technology enables high detection efficiency of photons and efficient rejection of electronic noise. The Philips MicroDose 130 uses a multi-slit scanning technique that prevents image degradation caused by scattered radiation by removing photons scattered in the breast and not directed towards the detector. These factors combine into a dose efficient system. The Philips MicroDose 130 provides three exposure modes; manual, automatic (parameters predefined based on compressed breast thickness), and SmartAEC continuously adjusts the exposure based on measured transmission from the leading detector edge. {2}------------------------------------------------ #### D. INDICATIONS FOR USE The Philips MicroDose L30 is indicated for generating mammographic images that can be used for screening and diagnosis of breast cancer. The Philips MicroDose L30 is intended to be used in the same clinical applications as traditional film/screen systems. #### E. PREDICATE DEVICE | | Predicate device | |-------------------------------|----------------------------------------------------| | Device Classification<br>Name | Full Field Digital, System, X-ray,<br>Mammographic | | 510(k) Number | K110025 | | Device Name | Sectra MicroDose Mammography | | Model | L30 | | Applicant | Sectra Imtec AB | | Product Code | MUE | | Advisory Committee | Radiology | ### E TECHNICAL CHARACTERISTICS As described in K110025, the Philips MicroDose L30 System employs photon counting technology; x-ray photons are captured in the detector and ultimately converted to an electrical signal, forming a digital image. All system components, hardware, and system features of the subject device are identical to the predicate MicroDose System. This 510(k) premarket notification is to obtain FDA clearance for the quantitative reduced dose claim for the MicroDose system based on the supporting clinical data. There is no change in the indications for use of the MicroDose system. The principle of operations of the MicroDose system is not affected, since there is no change to the device including image acquisition, X-ray generation, user interface, etc. As described in the original 510(k), K110025, performance data from non-clinical testing of the Sectra MicroDose Mammography L30 covering Sensitometric response, Spatial resolution, Noise analysis. Signal-to-noise-ratio transfer - DQE. Dynamic range, Automatic exposure control performance, Phantom testing, Patient radiation dose was compared with data from the predicate devices. This comparison showed that the Sectra MicroDose Mammography L30 device performed as well as or better than the predicate devices in all relevant areas. The testing was performed in accordance with generally accepted test methods, e.g. using IEC standards, published factors for dose calculations etc. In addition, clinical evidence including published studies on various national mammography screening programs demonstrated consistent levels of dose reduction with the use of the MicroDose system compared to other FFDM and film screen systems. #### G. CONCLUSION The Philips MicroDose L30 described in this submission is substantially equivalent to the predicate device in indication for use, principle of operation, technology and materials. Performance data demonstrate that the devices are substantially equivalent. {3}------------------------------------------------ Image /page/2/Picture/0 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three stripes forming its body and wings. The eagle's head is facing left. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the eagle. # DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002 Mr. Gustav Lins Manager Regulatory Affairs Philips Digital Mammography Sweden AB Smidesvagen 5 SE-171 41 SOLNA SWEDEN Re: K120255 Trade/Device Name: Philips MicroDose L30 Regulation Number: 21 CFR 892.1715 Regulation Name: Full-field digital mammography system Regulatory Class: II Product Code: MUE Dated: January 24, 2012 Received: January 27, 2012 APR 1 9 2012 ## Dear Mr. Lins: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28. 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of {4}------------------------------------------------ Page 2 medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely Yours, Janine M. Morris Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # Indications for Use 510(k) Number (if known): _ K |2025 Device Name: Philips MicroDose L30 Indications for Use: The Philips MicroDose L30 is indicated for generating mammographic images that can be used for screening and diagnosis of breast cancer. The Philips MicroDose L30 is intended to be used in the same clinical applications as traditional film/screen systems. X Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety K120255 510K Page 4-1
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...