ASPIRE HD PLUS, ASPIRE HD-S

K121674 · Fujifilm Medical System U.S.A., Inc. · MUE · Sep 21, 2012 · Radiology

Device Facts

Record IDK121674
Device NameASPIRE HD PLUS, ASPIRE HD-S
ApplicantFujifilm Medical System U.S.A., Inc.
Product CodeMUE · Radiology
Decision DateSep 21, 2012
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1715
Device ClassClass 2

Indications for Use

The Fujifilm Digital Mammography Systems, Aspire HD Plus (FDR MS-2500) and Aspire HD-s (FDR MS-2000), generate full-field digital mammography images that can, as other full-field digital mammography systems, be used for screening and diagnosis of breast cancer and are intended for use in the same clinical applications as traditional screen-film mammography systems.

Device Story

Integrated full-field digital mammography (FFDM) system; utilizes Fujifilm X-ray generator and amorphous selenium (a-Se) detector to capture X-ray energy; detector employs dual-layer a-Se scintillator with direct optical switching circuitry. System includes acquisition workstation (AWS) for automated post-processing, technologist preview, and quality control. Operated by technologists in clinical settings; images transmitted to diagnostic workstations, printers, or archiving systems. Provides automated breast compression and three automatic exposure control (AEC) modes. AWS interface manages patient information, exposure menu selection, and study viewing. Radiologists review digital images to support breast cancer screening and diagnostic decision-making.

Clinical Evidence

Clinical image attribute review conducted by independent mammographic radiologists. Six image sets per exposure mode (screening and diagnostic) were evaluated. Review concluded that the Aspire HD Plus provides acceptable quality for mammographic use. Results apply to both Aspire HD Plus and Aspire HD-s due to identical imaging geometry and components.

Technological Characteristics

Full-field digital mammography system; amorphous selenium (a-Se) detector; dual-layer a-Se scintillator; direct optical switching circuitry. Includes X-ray stand, tube, generator, and acquisition workstation (AWS) with 5-megapixel portrait monitor. Connectivity via hub to control cabinet and exposure stand. Software-based automated post-processing and AEC. Complies with FFDM 510(k) guidance for sensitometric response, spatial resolution, DQE, and phantom testing.

Indications for Use

Indicated for screening and diagnosis of breast cancer in patients requiring mammographic imaging, consistent with clinical applications of traditional screen-film mammography.

Regulatory Classification

Identification

A full-field digital mammography system is a device intended to produce planar digital x-ray images of the entire breast. This generic type of device may include digital mammography acquisition software, full-field digital image receptor, acquisition workstation, automatic exposure control, image processing and reconstruction programs, patient and equipment supports, component parts, and accessories.

Special Controls

*Classification.* Class II (special controls). The special control for the device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Full-Field Digital Mammography System.”*See* § 892.1(e) for the availability of this guidance document.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## FUJKFILM 419 WEST AVENUE STAMFORD, CT 06902 203-324-2000 PHONE: TOLL-FREE: 800-431-1850 203-353-0926 FAX: #### 510(k) Summary k 121674 Date Prepared: June 1, 2012 #### Submitter's Information FUJIFILM Medical Systems U.S.A., Inc. 419 West Avenue Stamford, Connecticut 06902 Telephone: (203) 602-3576 Facsimile: (203) 353-0296 Contact: Peter Altman #### Device Name and Classification: | Product Name: | Aspire HD Plus and Aspire HD-s | |-----------------------|---------------------------------------------| | Model Number: | FDR MS-2500 and FDR MS-2000 | | Classification Name: | Full-Field Digital Mammography X-ray System | | Classification Panel: | Radiology | | CFR Section: | 21 CFR 892.1715 | | Device Class: | Class II | | Product Code: | MUE | #### Substantial Equivalence/Predicate Devices: FUJIFILM Aspire HD (FDR MS-1000) (K110729) The predicate and proposed devices have the same Indication For Use (IFU). All three devices generate digital mammographic images that are intended for screening and diagnosis of breast cancer and are intended for use in the same clinical applications as traditional screen-film mammography systems. All systems employ the same digital amorphous selenium x-ray detector integrated into gantry (stand) based x-ray systems. The x-ray stand, tube and generator of the Aspire HD Plus and Aspire HD-s are essentially identical. Note that the Aspire HD's X-ray system was manufactured by Siemens, whereas the Aspire HD Plus and Aspire HD-s X-ray systems are manufactured by Fujifilm. The technological characteristics of the devices are similar as demonstrated by the comparison of imaging characteristics such as MTF, Noise Analysis, DQE, CNR, Phantom testing, etc. measured during non-clinical testing. Non-clinical testing was conducted in accordance with the FFDM 510(k) Guidance document covering Sensitometric Response, Spatial Resolution, Noise Analysis, Signal-to-Noise Ratio Transfer - DQE, Dynamic Range, Image Erasure and Fading, Repeated Exposure Test, AEC Performance, ACR MAP Phantom Testing, Contrast Detail Phantom Testing, Patient Radiation Dose Testing, and Breast Compression system Testing. A clinical image attribute review was conducted by independent mammographic radiologists in accordance with the FFDM 510(k) Guidance document. The mammographic attributes of six (6) image sets per exposure mode of screening and diagnostic cases were reviewed concluding that the Aspire HD Plus provides sufficiently acceptable quality for mammographic use. Since the Aspire HD Plus and Aspire HD-s have exactly the same imaging geometry and components (detector, and X-ray system), the results of the image attribute review apply to both systems. SEP 21 2012 {1}------------------------------------------------ # FUJKFILM 419 WEST AVENUE STAMFORD, CT 06902 203-324-2000 PHONE: TOLL-FREE: 800-431-1850 203-353-0926 FAX: #### Description of the Device: The Aspire HD Plus and Aspire HD-s are integrated FFDM systems combining an X-ray system made by Fujifilm with Fujifilm's a-Se detector and Acquisition Workstation (AWS). The Aspire HD Plus and Aspire HD-s create digital mammography images by direct capture of x-ray energy using the a-Se detector. The detector is a Fujifilm design utilizing an exclusive dual layer a-Se scintillator with Direct Optical switching circuitry to acquire image data and transfer images to the AWS for automated post processing, technologist preview and QC, and subsequent transmission to hard copy printers, diagnostic workstations and archiving systems. The Aspire HD Plus and Aspire HD-s provide automated compression and three AEC modes. The Aspire HD Acquisition Workstation (FDR 2000AWS) includes an off the shelf personal computer, the application software, Windows 7 Operating System, a 5megapixel portrait type monitor, and a hub. The hub transmits signals between the personal computer and control cabinet, and between the personal computer and exposure stand. The AWS display primarily consists of three windows: - . Patient Information Input window - Exposure Menu Selection window . - . Study window. The user may switch between these windows depending on the operation being performed. The X-ray control panel, which controls and observes the exposure stand, is always displayed in the lower part of each window. This allows setting the exposure conditions and confirming the radiation conditions on a single view. #### Intended Use: The Fujifilm Digital Mammography Systems, Aspire HD Plus (FDR MS-2500) and Aspire HD-s (FDR MS-2000), generate full-field digital mammography images that can, as other full-field digital mammography systems, be used for screening and diagnosis of breast cancer and are intended for use in the same clinical applications as traditional screen-film mammography systems. #### Safety Information: The Aspire HD Plus and Aspire HD-s introduce no new safety or efficacy issues other than those already identified with the predicate device. The results of the Hazard Analysis combined with the appropriate preventive measures taken indicate that the devices are of moderate concern as per the May 11, 2005 issue of the "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" and is consistent with the level of concern indicated in the "Class II Special Controls Guidance Document: Full-Field Digital Mammography System" document issued on: November 5, 2010. #### Conclusion: This 510(k) premarket notification submission has demonstrated Substantial Equivalence as defined and understood in the Federal Food Drug and Cosmetic Act and various guidance documents issued by the Center for Devices and Radiological Health. We conclude the subject devices to be as safe and effective as the predicate device based upon the clinical and non-clinical data summarized above. {2}------------------------------------------------ Image /page/2/Picture/0 description: The image is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the edge. In the center of the seal is a stylized image of a bird or eagle with its wings spread. The bird is made up of three curved lines that form the wings and body. The seal is black and white. #### DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002 SEP 21 2012 Mr. Peter Altman Regulatory Consultant FUJIFILM Medical Systems, USA Inc. 419 West Avenue STAMFORD CT 06902 Re: K121674 Trade/Device Name: Aspire HD Plus (FDR MS-2500) and Aspire HD-s (FDR MS-2000) Regulation Number: 21 CFR 892.1715 Regulation Name: Full-field digital mammography system Regulatory Class: II Product Code: MUE Dated: September 21, 2012. Received: September 21, 2012 #### Dear Mr. Altman: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of {3}------------------------------------------------ #### Page 2 medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html Sincerely Yours, Janine M. Morris Director Bivision of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ### Indications for Use 510(k) Number (if known): K121674 Device Name: _________________________________________________________________________________________________________________________________________________________________ Indications for Use: The Fujifilm Digital Mammography Systems, Aspire HD Plus (FDR MS-2500) and Aspire HD-s (FDR MS-2000), generate full-field digital mammography images that can, as other full-field digital mammography systems, be used for screening and diagnosis of breast cancer and are intended for use in the same clinical applications as traditional screen-film mammography systems. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of Raciological D Offica of litro Diagnostic Devica Evaluation and Salety 510K 121674 Page 1 of 1 Page 22 of 2288
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...