K153614 · Hitachi, Ltd., Health Care Company · LHN · Apr 21, 2016 · Radiology
Device Facts
Record ID
K153614
Device Name
PROBEAT-V and accessories
Applicant
Hitachi, Ltd., Health Care Company
Product Code
LHN · Radiology
Decision Date
Apr 21, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.5050
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The PROBEAT-V is a medical device designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation. The PROBEAT-V includes optional accessories to assist the radiation oncologist in the delivery of proton radiation to defined target volumes while sparing surrounding normal tissue and critical organs from excess radiation.
Device Story
PROBEAT-V is a proton beam therapy system delivering radiation to localized tumors. System comprises beam delivery system and proton beam generation equipment. Modifications include reduced minimum monitor units (MU) for precise dose distribution and a high-accuracy camera system for patient positioning. Camera system uses infrared (IR) optical tracker and multi-sided probes (MSP) with IR LEDs mounted on the treatment couch; couch position data is transmitted to the Patient Positioning System (PPS) to provide feedback for motion control and position correction. Operated by radiation oncologists in clinical settings to deliver prescribed dose distributions while sparing healthy tissue. Output is a therapeutic proton beam; clinical benefit is precise tumor irradiation.
Clinical Evidence
Bench testing only. Testing evaluated dose accuracy, tolerance confirmation for reduced minimum MU, and position accuracy for the PPS upgrade. All tests successfully met pre-defined acceptance criteria.
Technological Characteristics
Proton beam irradiation system. Components: beam generation equipment and beam delivery system. Positioning system utilizes IR camera (optical tracker) and multi-sided probes (MSP) with IR LEDs. Connectivity involves data transmission from camera system to PPS couch motion control. Software-controlled dose delivery and positioning.
Indications for Use
Indicated for patients with localized tumors and other conditions susceptible to treatment by radiation.
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
K181676 — Mini Ridge Filter and auxiliary functions for PROBEAT-V · Hitachi Ltd. Healthcare Hitachi Works · Aug 15, 2018
K161163 — PROBEAT-V Proton Beam Therapy System with X-ray Limiting Accessory · Hitachi , Ltd. · Jul 13, 2016
K053280 — PROBEAT · Hitachi, Ltd., Power Systems Group · Mar 9, 2006
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other, forming a wing-like shape.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 21, 2016
Hitachi Ltd., Health Care Company % Mr. Jonathan Kahan Regulatory Counsel Hogan Lovells US LLP 555 13th Street NW WASHINGTON DC 20016
Re: K153614
Trade/Device Name: PROBEAT-V and Accessories Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: II Product Code: LHN Dated: March 28, 2016 Received: March 28, 2016
Dear Mr. Kahan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael D. O'Hara
For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### 510(k) Number (if known)
K153614
Device Name
PROBEAT-V and Accessories
Indications for Use (Describe)
The PROBEAT-V is a medical device designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation.
The PROBEAT-V includes optional accessories to assist the radiation oncologist in the delivery of proton radiation to defined target volumes while sparing surrounding normal tissue and critical organs from excess radiation.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.qov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) SUMMARY
## Accessories to the PROBEAT-V Proton Beam Therapy Systems
# Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared
Hitachi, Ltd., Health Care Company 4-14-1 Sotokanda, Chiyodaku,Tokyo, 101-8010, Japan Telephone: +81 (3) 45643563 Facsimile: +81 (3) 45642882
Contact Person: Naoya Nishimura
Date Prepared: March 28, 2016
## Name of Device and Name/Address of Sponsor
Accessories to the PROBEAT-V Proton Beam Therapy Systems
Hitachi, Ltd. Power Systems Company 3-1-1 Saiwai-cho, Hitachi-shi, Ibaraki-ken, 317-8511 Japan
## Common or Usual Name
Proton beam therapy system
#### Classification Name
Medical Charged-Particle Radiation Therapy System, 21 CFR 892.5050, Product Code LHN
#### Predicate Devices
Hitachi Ltd. PROBEAT-V (K151132, K152592)
#### Intended Use / Indications for Use
The PROBEAT-V is a medical device designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation.
The PROBEAT-V includes optional accessories to assist the radiation oncologist in the delivery of proton radiation to defined target volumes while sparing surrounding normal tissue and critical organs from excess radiation.
#### Technological Characteristics
The PROBEAT-V is a proton beam irradiation system, which provides a therapeutic proton beam for clinical treatment. It is designed to deliver a proton beam with the prescribed dose,
{4}------------------------------------------------
dose distribution and directed to the prescribed patient treatment site. The equipment to perform the above work is composed of two main components: (1) a beam delivery system whose primary responsibility is to ensure that the desired prescription parameters are properly delivered and (2) equipment necessary to generate the proton beam and direct it to the beam delivery system for patient treatment.
# (1) Change to Minimum MU
The minimum MU of the PROBEAT-V 510(k) (K151132 and K152592) has been reduced to allow for more precise dose distribution utilizing the exact same spot size as previously cleared.
# (2) High Accuracy Camera for Patient Positioning System
The high accuracy camera system is a supplemental system for the Patient Positioning System (PPS) which enables enhanced PPS positioning accuracy. The camera system used for enhanced PPS positioning accuracy provides additional feedback to the couch motion control and position correction functions to improve the positioning accuracy. The system comprises an Infrared (IR) camera (optical tracker) and multi-sided probes (MSP) with IR LEDs, and the MSPs are mounted on the couch. The couch position is measured by the camera and the data are then transmitted to the PPS.
# Performance Data
The company performed testing to evaluate the dose accuracy, tolerance confirmation for the reduction of minimum MU and the position accuracy for PPS accuracy upgrade. All the tests were successfully conducted and confirmed that the device meets the pre-defined acceptance criteria.
# Substantial Equivalence
The PROBEAT-V with the alternate settings and optional accessories for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation is substantially equivalent to the cleared PROBEAT-V system as both the current and cleared devices have the same intended use, and substantially similar indications for use and technological characteristics. The addition of the optional settings and accessories to the system do not alter the intended therapeutic effect of the device system, and do not raise new or different questions of safety or efficacy. Specifically, the options are simply minor modifications to components of the cleared PROBEAT-V device to allow for greater precision in the positioning of the patient and the delivery of the treatment dose to the patient. Accordingly, the modified PROBEAT-V is substantially equivalent to the cleared PROBEAT-V.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.