The PROBEAT-CR is a medical device designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation.
Device Story
Proton beam therapy system; generates/delivers proton beams for tumor treatment. Components: synchrotron accelerator, beam transport system, treatment rooms (rotating gantry/fixed beam). Inputs: patient positioning data (CBCT, orthogonal X-ray), fiducial tracking data (via integrated Real Time Image Gating System). Operation: beam delivery system scans proton beam (discrete spot scanning) to match prescribed dose/distribution. Used in clinical oncology settings by trained medical staff. Integrates Real Time Image Gating System (K171049) to track implanted fiducials and gate beam delivery based on patient motion. Optional patient couch top extension allows expanded positioning angles (±90 degrees). Output: controlled proton beam irradiation. Benefits: precise dose delivery to localized tumors while sparing surrounding healthy tissue.
Clinical Evidence
Bench testing only. No clinical data provided. Validation included design verification and validation testing for the optional couch top extension and software verification and validation for the updated Real Time Image Gating System (RGPT).
Indicated for patients with localized tumors and other conditions susceptible to radiation therapy.
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
Hitachi Ltd., Healthcare Hitachi Works, Real Time Image Gating System for Proton Beam Therapy Systems (K171049)
Submission Summary (Full Text)
{0}------------------------------------------------
July 13, 2020
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG" in blue and "ADMINISTRATION" in a smaller font size below.
Hitachi Ltd. % Mr. Jonathan Kahan Partner Hogan Lovells US LLP 555 Thirteenth Street, NW WASHINGTON DC 20004
Re: K201042
Trade/Device Name: PROBEAT-CR Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: Class II Product Code: LHN Dated: April 20, 2020 Received: April 20, 2020
Dear Mr. Kahan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting
{1}------------------------------------------------
combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
510(k) Number (if known)
K201042
## Device Name
### PROBEAT-CR
Indications for Use (Describe)
The PROBEAT-CR is a medical device designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
区 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(k) SUMMARY
### Hitachi Ltd.'s PROBEAT-CR Proton Beam Therapy Device
### Submitter
K201042
Hitachi Ltd., Healthcare Radiation Oncology Systems, Kashiwa 2-1 Shintoyofuta, Kashiwa-shi Tiba-ken, 277-0804, Japan Telephone: +81 (4) 7131-4280 Facsimile: +81 (4) 7132-6837
Contact Person: Tomoyuki Seino
Date Prepared: April 20, 2020
### Name of Device:
PROBEAT-CR Proton Beam Therapy System
#### Common or Usual Name:
Proton Beam Therapy Device
### Classification Name:
System, Radiation Therapy, Charged-Particle, Medical
### Requlatory Class:
Class II
#### Product Code:
LHN
#### Predicate Devices
Hitachi Ltd., Healthcare Hitachi Works, PROBEAT-CR Proton Beam Therapy Device (K191801)
#### Reference Devices
Hitachi Ltd., Healthcare Hitachi Works, Real Time Image Gating System for Proton Beam Therapy Systems (K171049)
{4}------------------------------------------------
## Device Description
The PROBEAT-CR is a proton beam irradiation system. which provides a therapeutic proton beam for clinical treatment. It is designed to deliver a proton beam with the prescribed dose. dose distribution and directed to the prescribed patient treatment site.
### Intended Use / Indications for Use
The PROBEAT-CR is a medical device designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation.
### Summary of Technological Characteristics
The PROBEAT-CR has two main subsystems: (1) equipment necessary to generate the proton beam and direct it to the beam delivery system for patient treatment, and (2) a beam delivery system whose primary responsibility is to ensure that the desired prescription parameters are properly delivered. The PROBEAT-CR comprises the following components and subsystems:
- . Beam production system
- Accelerator system (LINAC. Synchrotron). o
- Beam transport system (Low/High Energy Beam Transport systems). o
- Beam delivery system in 4 separate treatment rooms. Each of 3 rooms will have a ● rotating gantry and 1 room will have a fixed beam.
- o Gantry Room
- . Scanning Nozzle
- . Rotating Gantry
- l Patient Positioning System
- l Orthogonal X-ray system
- . Cone Beam CT
- Fixed Beam Room o
- . Patient Positioning System
- . Orthogonal X-ray system
- . Treatment Control and Safety System
The subject PROBEAT-CR is a modification to the cleared PROBEAT-CR to include the incorporation of the previously cleared Real Time Image Gating System for Proton Beam Therapy Systems ("RGS" or "RGPT") (K171049) for tracking implanted fiducials to qate the delivery of the proton beam, and the addition of an optional patient couch top extension as an accessory to allow for different patient positioning configurations.
A table comparing the key features of the subject and predicate devices is provided below.
{5}------------------------------------------------
# Performance Data
The following testing was performed to validate the modifications to the device:
- Design verification and validation testing for the addition of the optional top couch . extension
- Software verification and validation for the updated RGS (RGPT) software .
# Conclusions
The company's subject PROBEAT-CR is as safe and effective as the predicate PROBEAT-The subject PROBEAT-CR has the same intended use and indications as the CR. previously cleared PROBEAT-CR. In addition, the subject PROBEAT-CR has very similar technological characteristics and principles of operation as its predicate. Although there are minor differences between the PROBEAT-CR and the predicate, namely the incorporation of the cleared RGS (RGPT) software system and the addition of an optional patient couch top extension, those differences do not raise new questions of safety or efficacy. The performance testing demonstrates that the device is as safe and effective as the predicate. Thus, the PROBEAT-CR is substantially equivalent.
# Substantial Equivalence Table
| | Hitachi PROBEAT-CR (Subject<br>Device) | Hitachi PROBEAT-CR (K191801) |
|------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | The PROBEAT-CR is a medical<br>device designed to produce and<br>deliver a proton beam for the<br>treatment of patients with localized<br>tumors and other conditions<br>susceptible to treatment by<br>radiation. | The PROBEAT-CR is a medical<br>device designed to produce and<br>deliver a proton beam for the<br>treatment of patients with localized<br>tumors and other conditions<br>susceptible to treatment by<br>radiation. |
| Accelerator | Synchrotron | Synchrotron |
| Particle | Protons | Protons |
| Variable Energy | 70-230 MeV | 70-230 MeV |
| Nozzles | Discrete Spot Scanning | Discrete Spot Scanning |
| Support for Patient<br>Positioning | Gantry built-in type Cone Beam<br>CT is used to verify patient<br>positioning.<br>A conventional x-ray system can<br>be also equipped in the same<br>treatment room.<br>A computer assisted patient<br>position system, (PIAS) is<br>available for use. | Gantry built-in type Cone Beam CT<br>is used to verify patient positioning.<br>A conventional x-ray system can<br>be also equipped in the same<br>treatment room.<br>A computer assisted patient<br>position system, (PIAS) is available<br>for use. |
| | Hitachi PROBEAT-CR (Subject<br>Device) | Hitachi PROBEAT-CR (K191801) |
| Treatment Room | Typically 3 rotating gantry rooms<br>and 1 fixed beam room (Maximum<br>5 total).<br><Note>All rooms are dedicated<br>to the DSSS | Typically 3 rotating gantry rooms<br>and 1 fixed beam room (Maximum<br>5 total).<br><Note>All rooms are dedicated to<br>the DSSS |
| Gantry rotating angle | 360 degrees<br>(-180-180degrees) | 360 degrees<br>(-180-180degrees) |
| Patient Positioner | Rotating gantry room and fixed<br>beam room,<br>(a) Patient Couch<br>Swing type Robotic Patient<br>Positioning System with 6 Degrees<br>of freedom<br>Isocentric rotating angle of the<br>couch is ±45 degrees for standard<br>base and extension configuration.<br>With the optional couch top<br>extension, isocentric rotating angle<br>can be extended to ±90 degrees<br>for short base and overlay<br>configuration, with a ±5 degree<br>movement range for the rolling and<br>pitching angles.<br>(b) Laser Alignment System | Rotating gantry room and fixed<br>beam room,<br>(a) Patient Couch<br>Isocentric rotating angle of the<br>couch is ±45 degrees for standard<br>base and extension configuration.<br>(b) Laser Alignment System |
| | | |
| | | |
| | | |
| Patient Positions<br>Supported | Head first-supine ("HFS"), head<br>first-prone ("HFP"), feet first-<br>supine ("FFS"), and feet first-prone<br>( | |
{6}------------------------------------------------
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.