PROBEAT-V Proton Beam Therapy System with X-ray Limiting Accessory

K161163 · Hitachi , Ltd. · LHN · Jul 13, 2016 · Radiology

Device Facts

Record IDK161163
Device NamePROBEAT-V Proton Beam Therapy System with X-ray Limiting Accessory
ApplicantHitachi , Ltd.
Product CodeLHN · Radiology
Decision DateJul 13, 2016
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.5050
Device ClassClass 2
AttributesTherapeutic

Intended Use

The PROBEAT-V with X-ray Limiting Accessory is a medical device designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation.

Device Story

PROBEAT-V is a proton beam therapy system for treating localized tumors. The system includes a beam delivery system and proton beam generation equipment. The new X-ray Limiting Accessory consists of four motorized leaves integrated into the existing system. During patient positioning using Cone Beam CT (CBCT), the imaging software controls these leaves to restrict the X-ray field, thereby minimizing patient radiation exposure. The device is operated by clinical staff in a radiation therapy environment. The output is a controlled proton beam for therapy and limited X-ray exposure for imaging, which facilitates precise tumor targeting while reducing unnecessary radiation dose to the patient.

Clinical Evidence

Bench testing only. Evaluated slice pitch accuracy, spatial accuracy, spatial resolution, isocenter accuracy, and integration with the proton beam therapy positioning system. Electrical safety and electromagnetic compatibility (EMC) testing were performed. All tests met pre-specified success criteria.

Technological Characteristics

Proton beam irradiation system comprising beam generation and delivery components. Accessory features four motorized leaves for X-ray field limitation. Controlled via PROBEAT-V imaging software. System integration includes electrical and electromagnetic compatibility compliance.

Indications for Use

Indicated for patients with localized tumors and other conditions susceptible to treatment by radiation.

Regulatory Classification

Identification

A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is a stylized design featuring three human profiles that are connected to each other. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 July 13, 2016 Hitachi, Ltd. % Mr. Jonathan Kahan Regulatory Counsel Hogan Lovells US LLP 555 13th Street NW WASHINGTON DC 20016 Re: K161163 Trade/Device Name: PROBEAT-V Proton Beam Therapy System with X-ray Limiting Accessory Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: II Product Code: LHN Dated: April 25, 2016 Received: April 25, 2016 Dear Mr. Kahan: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours. Michael D'Hara For Robert Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ 510(k) Number (if known) K161163 Device Name PROBEAT-V Proton Beam Therapy System with X-ray Limiting Accessory Indications for Use (Describe) The PROBEAT-V with X-ray Limiting Accessory is a medical devices designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|-----------------------------------------------| | X Prescription Use (Part 21 CFR 801 Subpart D) | □ Over-The-Counter Use (21 CFR 801 Subpart C) | ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number {3}------------------------------------------------ {4}------------------------------------------------ ### 510(k) SUMMARY ## PROBEAT-V Proton Beam Therapy System with X-ray Limiting Accessory # Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared Hitachi, Ltd. 2-16-1 Higashi-Ueno, Taito-ku, Tokyo, 110-0015, Japan Telephone: +81 (3) 62843741 Facsimile: +81 (3) 62843657 Contact Person: Naoya Nishimura Date Prepared: June 30, 2016 ## Name of Device and Name/Address of Sponsor PROBEAT-V Proton Beam Therapy System with X-ray Limiting Accessory Hitachi, Ltd. Healthcare, Hitachi Works 3-1-1 Saiwai-cho, Hitachi-shi, Ibaraki-ken, 317-8511 Japan ### Common or Usual Name Accessory to proton beam therapy system #### Classification Name Medical Charged-Particle Radiation Therapy System, 21 C.F.R. § 892.5050, Product Code LHN #### Predicate Devices PROBEAT-V (K152592) #### Intended Use / Indications for Use The PROBEAT-V with X-ray Limiting Accessory is a medical device designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation. #### Technological Characteristics The PROBEAT-V with X-ray Limiting Accessory is a proton beam irradiation system, which provides a therapeutic proton beam for clinical treatment. It is designed to deliver a proton {5}------------------------------------------------ beam with the prescribed dose, dose distribution and directed to the prescribed patient treatment site. The equipment to perform the above work is composed of two main components: (1) a beam delivery system whose primary responsibility is to ensure that the desired prescription parameters are properly delivered and (2) equipment necessary to generate the proton beam and direct it to the beam delivery system for patient treatment. The X-ray Beam Limiting accessory for use in conjunction with the cleared PROBEAT-V system is designed to limit the patient's X-ray exposure during for patient positioning with the Cone Beam CT (CBCT). The device has 4 motorized leaves which are controlled by the PROBEAT-V imaging software to position the leaves in order to limit the patient's radiation exposure during patient positioning. ## Performance Data Testing was performed to evaluate that the integration of the X-ray Beam Limiting Device with the previously cleared PROBEAT-V does not impact the performance of the system. This testing evaluated slice pitch accuracy, spatial accuracy, spatial resolution, and isocenter accuracy as well as testing with the proton beam therapy positioning system. Electrical safety was evaluated and electromagnetic compatibility testing was performed with the integrated system. In all cases, the device performed as intended and met all pre-specified success criteria. ## Substantial Equivalence The PROBEAT-V with integrated X-ray beam limiting device accessory for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation is substantially equivalent to the cleared PROBEAT-V system. Both the current and cleared devices have the exact same intended use and substantially similar indications for use and technological characteristics. The addition of the X-ray beam limiting accessory to the system does not alter the intended therapeutic effect of the device system, and does not raise new or different questions of safety or efficacy. The addition of this accessory simply allows for minimizing radiation exposure during image acquisition and patient positioning. Thus, the PROBEAT-V Proton Beam Therapy System with X-ray Limiting Accessory is substantially equivalent to its predicate.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...