0.6CC IONIZATION CHAMBER - WATERPROOF 0.6CC IONIZATION CHAMBER

K961208 · Ptw-New York Corp. · IYE · Jun 17, 1996 · Radiology

Device Facts

Record IDK961208
Device Name0.6CC IONIZATION CHAMBER - WATERPROOF 0.6CC IONIZATION CHAMBER
ApplicantPtw-New York Corp.
Product CodeIYE · Radiology
Decision DateJun 17, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.5050
Device ClassClass 2

Indications for Use

The PTW T30004, T30005, T30001 and T30002 0.6 cc Farmer type ionization chambers are intended to be used for the collection of beam data in water, air, or other suitable solid state phantom material from radiation therapy treatment machines. This data is used to completely document the beam characteristics of treatment machines and to establish and maintain an on going treatment machine quality assurance program.

Device Story

0.6 cc Farmer type ionization chambers collect radiation beam data in water, air, or solid state phantoms. Used in radiation therapy clinics to document treatment machine beam characteristics and maintain quality assurance programs. Chambers connect to an electrometer (e.g., PTW-UNIODS) via 1 or 10-meter cables with various standard connectors (BNT, TNC, M, BNC banana). Thimble-vented versions are non-waterproof; connector-vented versions are waterproof. T30004 features graphite thimble/aluminum stem; T30005 features solid water thimble/aluminum stem. Data collected by the chamber is processed by the electrometer to inform clinical decision-making regarding machine calibration and treatment accuracy, benefiting patients by ensuring precise radiation delivery.

Clinical Evidence

Bench testing only. No clinical data provided.

Technological Characteristics

0.6 cc Farmer type ionization chamber. Materials: graphite or solid water thimbles, aluminum stems. Connectivity: wired connection to electrometer via BNT, TNC, M, or BNC banana connectors. Form factor: 0.6 cc volume. Venting: thimble (non-waterproof) or connector (waterproof).

Indications for Use

Indicated for use by medical physicists or radiation therapy personnel to collect beam data from radiation therapy treatment machines for documentation of beam characteristics and quality assurance programs.

Regulatory Classification

Identification

A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0} JUN 17 1996 PTW-New York Corporation 2437 Grand Avenue Bellmore, New York (P) 1-516-221-4708 (F) 1-516-221-4329 K961208 # PTW 0.6 cc Ionization Chambers ## 510(k) Submission ### Manufacturer's 510(k) Summary, 21 CFR 807.92: #### 1. Company: PTW-New York Corporation 2437 Grand Avenue Bellmore, New York 11710 (P) 1-516-221-4708 (F) 1-516-221-4329 **Contact:** Stephen R. Szeglin General Manager PTW-New York Corporation (P) 1-516-221-4708 (F) 1-516-221-4329 **Date of Submission:** March 22, 1996 #### 2. Trade/Proprietary Name: PTW T30004 0.6 cc Farmer type ionization chamber, thimble vent - not waterproof, connector vent - waterproof, PTW T30005 0.6 cc Farmer type ionization chamber, thimble vent - not waterproof, connector vent - waterproof, PTW T30001 0.6 cc Farmer type ionization chamber, connector vent - waterproof, PTW T30002 0.6 cc Farmer type ionization chamber, connector vent - waterproof. **Common/Usual Name:** 0.6 cc Farmer type ionization chamber #### 3. Predicate Device(s): PTW T30001 and T30002 0.6 cc Farmer type ionization chamber, K951764. #### 4. Description of Device(s): The PTW T30004 and T30005 are classical 0.6 cc Farmer type ionization chambers. These chambers differ only in the type of material their respected thimbles are made of. The T30004 has a graphite thimble and an aluminum stem and the T30005 has a solid water thimble and an aluminum stem. The chambers can be thimble vented, which is not waterproof, or connector vented, which is water proof. Page 1 {1} PTW-New York Corporation 2437 Grand Avenue Bellmore, New York (P) 1-516-221-4708 (F) 1-516-221-4329 K961208 # PTW 0.6 cc Ionization Chambers ## 510(k) Submission ### 4. Description of Device(s) (con't): Both chambers offer 1 or 10 meter cables for connection to an electrometer, and a variety of standard electrometer connectors (BNT, TNC, M, BNC banana). The connector vented version of the PTW T30001 and T30002 Farmer type ionization chambers are simply a waterproof version of the thimble vented chambers that were cleared to market by the FDA under K951764. The Farmer type ionization chamber, when connected to an appropriate electrometer like the PTW-UNIODS, K951764, is used to collect beam data from radiation therapy treatment machines. ### 5. Statement of Intended Use: The PTW T30004, T30005, T30001 and T30002 0.6 cc Farmer type ionization chambers are intended to be used for the collection of beam data in water, air, or other suitable solid state phantom material from radiation therapy treatment machines. This data is used to completely document the beam characteristics of treatment machines and to establish and maintain an on going treatment machine quality assurance program. ### 6. Comparison of Technological Characteristics to the Predicate Devices: The indications for use are exactly the same as the predicate devices, PTW T30001 and T30002 0.6 cc Farmer type ionization chambers, which were cleared to market by the FDA under K951764. The designs are exactly the same. The manufacturing and testing, process and procedures are exactly the same. The materials used are the same as in the predicate devices with the exception of the solid water used for the T30005 thimble. For this particular device and its application, solid water is universally accepted as equivalent to PMMA or graphite, which are the materials used in the thimbles of the predicate chambers. The specifications are the same as the predicate devices. The indications for use, design, materials, manufacturing, and specifications of the PTW T30004, T30005, T30001, and T30002 do not raise any issues with regard to safety and effectiveness. PTW considers all of these 0.6 cc Farmer type ionization chambers equivalent to the predicate devices for radiation therapy beam data acquisition. **Note:** Any statement made in conjunction with this Summary regarding substantial equivalence to another product was made in relation to the 510(k) premarket approval process and should not be interpreted as an admission or used as evidence in patient infringement litigation. Page 2
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...