The PTW T10004 MULTIDOS as a stand alone dosemeter is intended to be used to measure the dose delivered to a patient during a radiation therapy treatment. Dose measurements during treatment may be accomplished by using suitable ionization chambers or solid state detectors. Data acquired by MULTIDOS is used to compare and to verify treatment dose to prescribed dose. The PTW T42007 QC6PLUS cable connects to MULTIDOS and is intended to be used to measure the beam characteristics of radiation therapy freatment machines. Data acquired with the MULTIDOS and the QC6PLUS is used to compile radiation beam data over time as part of a quality assurance program.
Device Story
MULTIDOS (T10004) is a multi-channel electrometer measuring dose, dose rate, charge, and current for photons/electrons; used with ionization chambers or solid-state detectors. QC6PLUS (T42007) is a therapy beam evaluation system consisting of 6 or 10 ionization chambers embedded in a 5 cm PMMA slab; connects to MULTIDOS to monitor beam parameters. Used in clinical radiation therapy settings by medical physicists/technicians to verify treatment dose against prescription and compile longitudinal beam data for quality assurance. Provides accurate, precise dosimetry data to ensure therapeutic radiation delivery; enables routine monitoring and documentation of machine performance.
Clinical Evidence
No clinical data; bench testing only.
Technological Characteristics
Electrometer (MULTIDOS) and ionization chamber array (QC6PLUS). QC6PLUS utilizes 0.54 cm³ ionization chambers embedded in 5 cm thick PMMA slab. Configurable for 2, 6, or 12 channels. Measures dose, dose rate, charge, current, flatness, symmetry, and energy stability. Standalone or connected system.
Indications for Use
Indicated for use in radiation therapy environments to measure patient dose during treatment and to monitor beam characteristics (flatness, symmetry, energy stability, central beam dose) of radiation therapy treatment machines for quality assurance purposes.
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
Nuclear Associates, 37-702 and 37-703, PC Rainbow (K900932)
Keithley 35300, Tracker™, Therapy Beam Evaluation System (K874893)
Submission Summary (Full Text)
{0}------------------------------------------------
SEP 1 0 1997
## PTW-New York Corporation 2437 Grand Avenue Bellmore, New York 11710
## 510(k) Premarket Notification for PTW T10004 and T42007 MULTIDOS and QC6PLUS
q72211
## Manufacturer's 510(k) Summary Certification, 21 CFR 807.92:
- 1. Company:
PTW-New York Corporation 2437 Grand Avenue Bellmore, New York 11710 (P) 1-516-221-4708 (F) 1-516-221-4329
### Contact:
Stephen R. Szeglin General Manager PTW-New York Corporation (P) 1-516-221-4708 (F) 1-516-221-4329
### Date of Submission:
June 9, 1997
#### 2. Trade/Proprietary Name:
PTW T10004, MULTIDOS, Multiple Channel Dosemeter, PTW T42007, QC6PLUS, Therapy Beam Evaluation System.
### Common/Usual Name:
Multiple channel electrometer, Therapy beam constancy check device.
#### Predicate Device(s): 3.
MULTIDOS:
PTW T10006, MULTIDOS ME48, 48 channel electrometer, K954165, and Nuclear Associates, 37-702 and 37-703, PC Rainbow, K900932.
### QC6PLUS:
Keithley 35300, Tracker™, Therapy Beam Evaluation System, K874893
#### Description of Device(s): 4.
1:
The PTW T10004 MULTIDOS is a field class multiple-purpose multiple-channel electrometer. This device is designed to accurately measure dose, dose rate, charge, and current for photons and electrons in the therapeutic energy ranges. MULTIDOS may be configured as a stand alone 2, 6, or 12 channel electrometer and can be used with suitable ionization chambers and solid state detectors.
The PTW T42007 QC6PLUS is a therapy beam constancy device which may configured with 6 or 10 individual 0.54 cm³ ionization chambers orthogonally arranged and imbedded in a 5 cm thick PMMA slab. When the QC6PLUS is connected to MULTIDOS, therapy beam parameters like flatness, symmetry, energy stability, and central beam dose are easily monitored.
{1}------------------------------------------------
## PTW-New York Corporation 2437 Grand Avenue Bellmore, New York 11710
## 510(k) Premarket Notification for PTW T10004 and T42007 MULTIDOS and QC6PLUS
### 5. Statement of Intended Use:
: 1
The PTW T10004 MULTIDOS as a stand alone dosemeter is intended to be used to measure the dose delivered to a patient during a radiation therapy treatment. Dose measurements during treatment may be accomplished by using suitable ionization chambers or solid state detectors. Data acquired by MULTIDOS is used to compare and to verify treatment dose to prescribed dose.
The PTW T42007 QC6PLUS cable connects to MULTIDOS and is intended to be used to measure the beam characteristics of radiation therapy freatment machines. Data acquired with the MULTIDOS and the QC6PLUS is used to compile radiation beam data over time as part of a quality assurance program.
#### 6. Comparison of Technological Characteristics to the Predicate Devices:
The indications for use of the MULTIDOS and the QC6PLUS are exactly the same as the predicate devices.
The designs are essentially the same as the predicate devices.
The manufacturing and testing, process and procedures are the same.
The materials used are the same as in the predicate devices.
The specifications are the same as the predicate devices.
The indications for use, design, materials, manufacturing, and specifications of the PTW T10004 MULTIDOS and T42007 QC6PLUS do not raise any issues with regard to safety and effectiveness.
PTW considers the T10004 MULTIDOS and T42007 QC6PLUS equivalent in all respects to the predicate devices for radiation therapy beam measurements.
- Any statement made in conjunction with this Summary regarding substantial equivalence Note: to another product was made in relation to the 510(k) premarket approval process and should not be interpreted as an admission or used as evidence in patient infringement litigation.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the text "DEPARTMENT OF HEALTH & HUMAN SERVICES". The text is in all caps and is in a serif font. The text is black and the background is white. The text is slightly slanted.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# SEP 1 0 1997
Re:
Stephen R. Szeglin General Manager PTW-New York Corporation 2437 Grand Avenue Bellmore, New York 11710
K972211 Multidos Multi Channel Electrometer and QC6PLUS Constancy Check Device Dated: June 9, 1997 Received: June 12, 1997 Regulatory class: II 21 CFR 892.5050/Procode: 90 LHN
Dear Mr. Szeglin:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Fedgral Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in yitre diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for question and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours
H.J.Liau Yin
Lillian Yin, Ph.D. Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
{3}------------------------------------------------
510(k) Number (if known): K9722//
Device Name: MULTIDOS AND QC6 PLUS
PTW T10004 PTW T42007
Indications For Use:
The PTW T10004 MULTIDOS is a multiple-purpose, multiple-channel electrometer inlended to perform a range of measuring tasks in radiation therapy. As a stand alone dosemeter, MULTIDOS is intended to be used to measure the dose delivered to a palient during a radiation therapy treatment. Dose measurements during treatment may be accomplished by using suitable ionization chambers or solid state detectors and are used to compare and to verify treatment dose to prescribed dose.
When connected to the optional QC6PLUS, MULTIDOS is intended to be used to measure the beam characteristics of radiation therapy treatment machines. Data acquired with the MULTIDOS and the QC6PLUS is used to compile radiation beam data over time as part of a quality assurance program.
High quality dosimetry devices like the PTW T10004 MULTIDOS and the optional T42007 QC6PLUS make it possible to routinely measure, record, document and verify therapeutic amounts of ionizing radiation with an exceptionally high degree of accuracy and precision. High quality multiple purpose dosimetry devices are essential if accurate data from trealment machines that produce therapeutic amounts of ionizing radiation are to be properly monitored.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
David h. Sygmm
(Division Sign-Off) Division of Reproductive. Abdominal. EN and Radiological Device 510(k) Number
Prescription Use
(Per 21 CFR 801.109)
OR
Over-The-Counter Use
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.