0.015CC PINPOINT IONIZATION CHAMBER, 0.005 CC PINPOINT IONIZATION CHAMBER
Applicant
Ptw-New York Corp.
Product Code
LHN · Radiology
Decision Date
Sep 10, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.5050
Device Class
Class 2
Indications for Use
The PTW T31006 and T31007 Pinpoint type ionization chambers are intended to be used for the collection of beam data in water, air, or other suitable solid state phantom material from radiation therapy treatment machines. This data is used to completely document the beam characteristics of treatment machines and to establish and maintain an on going treatment machine quality assurance program.
Device Story
PTW T31006 (0.015 cc) and T31007 (0.005 cc) are waterproof ionization chambers; used for radiation therapy beam data collection in water, air, or solid phantoms. Devices connect to electrometers (e.g., PTW-UNIDOS) via 1m or 10m cables; collect ionizing radiation signals; output data used by medical physicists to document beam characteristics and maintain quality assurance programs for treatment machines. Devices are passive sensors; no internal processing; clinical benefit includes accurate monitoring of therapeutic radiation doses.
Clinical Evidence
Bench testing only. No clinical data provided.
Technological Characteristics
Waterproof ionization chambers; active volumes 0.015 cc (T31006) and 0.005 cc (T31007); active lengths 5 mm and 2 mm respectively; vented to atmosphere via connectors; cable lengths 1m or 10m; connectors include BNT, TNC, PTW M, or BNC banana; passive sensing principle.
Indications for Use
Indicated for use by medical physicists or radiation therapy professionals to measure, record, and document therapeutic ionizing radiation beam characteristics from treatment machines in water, air, or solid-state phantoms for quality assurance and baseline establishment.
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
Predicate Devices
PTW T31002 0.125 cc waterproof ionization chamber (K954165)
PTW T31003 0.03 cc waterproof ionization chamber (K954165)
{0}------------------------------------------------
K972212
# SEP 1 0 1997
## PTW-New York Corporation 2437 Grand Avenue Bellmore, New York 11710
## 510(k) Premarket Notification for PTW T31006 and T31007 Pinpoint type ionization chambers
#### Manufacturer's 510(k) Summary Certification, 21 CFR 807.92:
- 1. Company:
PTW-New York Corporation 2437 Grand Avenue Belimore, New York 11710 (P) 1-516-221-4708 (F) 1-516-221-4329
#### Contact:
Stephen R. Szeglin General Manager PTW-New York Corporation (P) 1-516-221-4708 (F) 1-516-221-4329
Date of Submission: June 9, 1997
- 2. Trade/Proprietary Name:
PTW T31006. 0.015 cc Pinpoint type ionization chamber. waterproof. PTW T31007, 0.005 cc Pinpoint type ionization chamber, waterproof.
### Common/Usual Name:
0.015 cc waterproof ionization chamber, 0.005 cc waterproof ionization chamber.
#### 3. Predicate Device(s):
PTW T31002. 0.125 cc waterproof ionization chamber, K954165, and PTW T31003, 0.03 cc waterproof ionization chamber, K954165.
#### 4. Description of Device(s):
The PTW T31006 and T31007 are classical waterproof ionization chambers that differ only in the lengths of their respective active volumes. The T31006 is a 0.015 cc ionization chamber with a 5 mm active length and the T31007 is a 0.005 cc ionization chamber with a 2 mm active length. Both chamber volumes are open to the atmosphere, they are waterproof, and they are vented via their connectors. These chambers are constructed of precisely the same materials, are available with a 1 m or 10 m cable for connection to an electrometer, and may be terminated in a variety of standard electrometer connectors (BNT, TNC, PTW M, or BNC banana.
The Pinpoint type ionization chamber, when connected to an appropriate electrometer like the PTW-UNIDOS, K951764, are used to collect beam data from radiation therapy treatment machines.
{1}------------------------------------------------
### PTW-New York Corporation 2437 Grand Avenue Bellmore, New York 11710
### 510(k) Premarket Notification for PTW T31006 and T31007 Pinpoint type ionization chambers
#### 5. Statement of Intended Use:
The PTW T31006 and T31007 Pinpoint type ionization chambers are intended to be used for the collection of beam data in water, air, or other suitable solid state phantom material from radiation therapy treatment machines. This data is used to completely document the beam characteristics of treatment machines and to establish and maintain an on going treatment machine quality assurance program.
#### Comparison of Technological Characteristics to the Predicate Devices: 6.
The indications for use are exactly the same as the predicate devices, the PTW T31002 - 0.125 cc waterproof ionization chamber, and the PTW T31003 - 0.03 cc waterproof ionization chamber, which were both cleared to market by the FDA under K954165, 5 December 1995, as accessories to the PTW MP3 and MP3S Automatic Water Phantom.
The designs are exactly the same.
The manufacturing and testing, process and procedures are exactly the same.
The materials used are the same as in the predicate devices with the exception of the steel electrode.
The specifications are the same as the predicate devices.
The indications for use, design, materials, manufacturing, and specifications of the PTW T31006 and T31007 Pinpoint ionization chambers do not raise any issues with regard to safety and effectiveness.
PTW considers the T31006 and T31007 Pinpoint ionization chambers equivalent in all respects to the predicate devices for radiation therapy beam data acquisition.
- Note: Any statement made in conjunction with this Summary regarding substantial equivalence to another product was made in relation to the 510(k) premarket approval process and should not be interpreted as an admission or used as evidence in patient infringement litigation.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure with three curved lines representing its wings or body. The logo is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter of the circle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
SEP 1 0 1997
Stephen R. Szeglin General Manager PTW-New York Corporation 2437 Grand Avenue Bellmore, N.Y. 11710
Re: K972212
PTW T31006 0.015 cc Pinpoint Type Ionization Chamber PTW T31007 0.005 cc Pinpoint Type Ionization Chamber Dated: June 9, 1997 Received: June 12, 1997 Regulatory class: II 21 CFR 892.5050/Procode: 90 LHN
Dear Mr. Szeglin:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranion.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for question and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsmamain.html".
Sincerely vours
W.Liau Yu
Lillian Yin, Ph.D. Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
{3}------------------------------------------------
## PTW-New York Corporation 2437 Grand Avenue Bellmore, New York 11701 (P) 1-516-221-4708 (F) 1-516-221-4329
# 510(k) Premarket Notification for PTW T31006 and T31007 Pinpoint type ionization chambers
June 9, 1997
# Premarket Notification Indication For Use Statement
510(k) Number (if known): Not assigned as of June 9, 1997
0.015 cc Pinpoint type ionization chamber, Device Name: 0.005 cc Pinpoint type ionization chamber.
Indications For Use:
High quality ionization chambers, like the PTW T31006 and T31007 Pinpoint type ionization chambers make it possible to routinely measure, record, and document therapeutic amounts of ionizing radiation with an exceptionally high degree of accuracy and precision.
This type of lonization chamber is intended to be used for the collection of radiation beam data from radiation therapy treatment machines in water, air, or other suitable solid state phantoms. Since the T31006 and T31007 Pinpoint chambers are waterproof, they are ideally suited for measurements of high energy photons and electrons in a water phantom.
The data acquired with this type of chamber can be used to completely document the beam characteristics of treatment machines, compile radiation beam data over time as part of a quality assurance program, and to establish an initial baseline for the radiation therapy beams generated by the treatment unit.
High quality ionization chambers, like the PTW Pinpoint type chambers, are essential if accurate data from treatment machines that produce therapeutic amounts of ionizing radiation are to be properly monitored.
tgl.R.Sgl.
Signature:
Date:
Typed Name:
Stephen R. Szeglin
JUNE 9, 199
David A. Seaman
Prescription Use
(Per 21 CFR 801.109)
(Division Sign-Off) Division of Reproductive, Abdominal, ENT. and Radiological Dev 510(k) Number
Attachment B
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.