K192405 · Jiangsu Rayer Medical Technology Co., Ltd. · IYE · May 1, 2020 · Radiology
Device Facts
Record ID
K192405
Device Name
Image Guidance System for Radiotherapy
Applicant
Jiangsu Rayer Medical Technology Co., Ltd.
Product Code
IYE · Radiology
Decision Date
May 1, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.5050
Device Class
Class 2
Indications for Use
The Image Guidance System for Radiotherapy is intended to be used in combination with the radiotherapy device, for image-guided positioning verification and real-time tracking of target motion for patient radiation therapy. The Image Guidance System for Radiotherapy sends the positioning data or real-time tracking data to the radiotherapy device but does not control any part of the radiotherapy device.
Device Story
Image Guidance System for Radiotherapy (IGTS/IGPS) provides image-guided positioning and real-time tracking for radiotherapy. System inputs include X-ray images from flat panel detectors (CARERAY CareView 1800IF or PerkinElmer XRD 0840AN) and respiratory motion data. Software processes these inputs using bony structure or fiducial marker registration algorithms to calculate 6-degree-of-freedom positioning corrections (3 translations, 3 rotations). Output consists of positioning or tracking data sent to compatible radiotherapy systems (Varian, Elekta, MASEP). Device is used in clinical radiotherapy settings by trained personnel. It does not control the radiotherapy device. By providing accurate target localization and motion tracking, it assists clinicians in verifying patient position, potentially improving treatment accuracy and sparing healthy tissue.
Clinical Evidence
No clinical data. Bench testing only. Accuracy verified via end-to-end tests with anthropomorphic phantoms, demonstrating positioning accuracy ≤1.0 mm and rotational accuracy ≤1.0°. Tracking accuracy for respiratory motion verified to be ≤1.5 mm. Compatibility testing confirmed integration with Varian, Elekta, and MASEP systems did not degrade their performance.
Technological Characteristics
Medical electrical equipment; includes flat panel detectors (CARERAY CareView 1800IF or PerkinElmer XRD 0840AN). Sensing via X-ray imaging and optical tracking. Connectivity: data export to radiotherapy systems. Standards: IEC 60601-1, IEC 60601-1-3, IEC 60601-2-68, IEC 60601-2-28, IEC 60601-2-54, IEC 60601-1-2, IEC 60601-1-6, IEC 62366-1. Non-sterile, no patient contact.
Indications for Use
Indicated for image-guided positioning verification and real-time tracking of target motion for patients undergoing radiation therapy.
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
ELEKTA SYNERGY, ELEKTA SYNERGY S AND XVIR3.5 (K051932)
Related Devices
K072506 — EXACTRAC 5.5 · Brainlab AG · Oct 26, 2007
K040585 — EXACTRAC 4.0 AND EXACTRAC X-RAY 6D · Brainlab AG · May 11, 2004
K122451 — EXACTRAC VERO · Brainlab AG · Sep 28, 2012
K132847 — I2C · Ion Beam Applications S.A. · Nov 20, 2013
K203173 — ExacTrac Dynamic · Brainlab AG · Nov 12, 2020
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. Food & Drug Administration".
Jiangsu Rayer Medical Technology Co., Ltd. % Peng Li Quality Director No. 99 Furongzhongsan Road, Xishan Economic Development Zone Wuxi, Jiangsu 214192 CHINA
May 1, 2020
# Re: K192405
Trade/Device Name: Image Guidance System for Radiotherapy Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: Class II Product Code: IYE Dated: April 6, 2020 Received: April 9, 2020
Dear Peng Li:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K192405
## Device Name
Image Guidance System for Radiotherapy
### Indications for Use (Describe)
Image Guidance System for Radiotherapy is intended to be used in combination with the radiotherapy device, for imageguided positioning verification and real-time tracking of target motion therapy. The Image Guidance System for Radiotherapy sends the positioning data or real-time tracking data to the radiotherapy device but does not control any part of the radiotherapy device.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------|
| <div> <span> × Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
This 510(k) Summary is being submitted in accordance with requirements of Title 21, CFR Section 807.92.
The assigned 510(k) Number: K192405
- Date of Preparation 1.
4/29/2020
- 2. Sponsor Identification
# Jiangsu Rayer Medical Technology Co.,Ltd.
No.99 Furongzhongsan Road, Xishan Economic Development Zone, 214192 Wuxi, PEOPLE'S
REPUBLIC OF CHINA
Establishment Registration Number: Not yet registered.
Contact Person: Peng Li
Position: Quality Director
Tel: +86-0510-83781608
Fax:+86-0510-83781708
Email:pli(@rayertech.com
- Designated Submission Correspondent 3.
Peng Li (Quality Director)
# Jiangsu Rayer Medical Technology Co.,Ltd.
No.99 Furongzhongsan Road, Xishan Economic Development Zone, 214192 Wuxi, PEOPLE'S REPUBLIC OF CHINA
Tel: +86-0510-83781608
{4}------------------------------------------------
#### 510(k) Summary
# ▼ RAYER江苏瑞尔医疗科技有限公司
Fax:+86-0510-83781708 Email:pli(@rayertech.com
- 4. Identification of Subject device
Trade Name: Image Guidance System for Radiotherapy Common Name: Patient Positioning System with respiratory Gating, Radiation Therapy, Charged-Particle, Medical Models: IGTS, IGPS
### Regulatory Information
Classification Name: System, Radiation Therapy, Charged-Particle, Medical Classification: II Product Code: IYE Regulation Number: 21 CFR 892.5050 Review Panel: Radiology
#### Indications for Use:
The Image Guidance System for Radiotherapy is intended to be used in combination with the radiotherapy device, for image-guided positioning verification and real-time tracking of target motion for patient radiation therapy. The Image Guidance System for Radiotherapy sends the positioning data or real-time tracking data to the radiotherapy device but does not control any part of the radiotherapy device.
### Device Description:
The product is a medical electrical device which is composed of hardware and software, and is jointly used with the radiation therapy device. Image Guidance System for Radiotherapy is intended to be used in combination with the radiotherapy device, for image-guided positioning verification and real-time tracking of target motion for patient radiation therapy.
{5}------------------------------------------------
- న్. Identification of Predicate Device 510(k) Number: K120789 Product Name: ExacTrac Classification Name: System, Radiation Therapy, Charged-Particle, Medical Classification: II Product Code: IYE Regulation Number: 21 CFR 892.5050 Review Panel: Radiology
- 6. Compatibility
The positioning feature of the subject device is compatible with Varian and Elekta linear accelerators and MASEP gamma ray stereotactic radiotherapy system. The real-time tracking feature is compatible with MASEP gamma ray stereotactic radiotherapy system. The subject device sends the positioning data or real-time tracking data to the radiotherapy device but does not control any part of the radiotherapy device.
The information of the compatible radiotherapy systems is provided as follows:
- 1) Gamma ray stereotactic radiotherapy
Device Name: ROTATING GAMMA SYSTEM INFINI (INFINI)
Model: INFINI
Manufacturer: MASEP Medical Science & Technology Development (Shenzhen) Co., Ltd.
510(k) Number: K102533
- 2) Linear accelerator
Device Name: TRILOGY SYSTEM WITH RAPIDARC
Model: Trilogy
Manufacturer: VARIAN MEDICAL SYSTEMS INC
510(k) Number: K072916
- 3) Linear accelerator
Device Name: ELEKTA SYNERGY, ELEKTA SYNERGY S AND XVIR3.5
{6}------------------------------------------------
Model: Synergy Manufacturer: ELEKTA INC 510(k) Number: K051932
#### 7. Couch Corrections for 6 Patient Offset Parameters
The compatible external couches used in most radiotherapy systems have no capability of rotation corrections. In the case of ignoring rotation angle couch correction parameters, the uncorrected rotation shifts may affect the clinical results. Therefore, it is recommended that the clinicians carefully evaluate the impact and make the right decision. It is also strongly recommended that the large rotations shall be corrected manually if the couch only receive and perform the 3 translation correction and ignore the 3 rotation parameters.
#### 8. X-ray Exposure Technique for Radiation Dose Control
The subject device doesn't have ACE function. Instead, it has its own exposure technique. The imaging system in subject devises an imaging scenario to achieve optimum image quality while minimizing radiation dose. The subject device also meets the safety requirement terms of IEC 60601-1-3 and IEC 60601-2-68 for controlling excessive radiation dose.
#### Non-Clinical Test Conclusion 9.
Non clinical tests were conducted to verify that the subject device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the subject device complies with the following standards:
- A IEC 60601-1:2005 And A1:2012 Medical Electrical Equipment-Part 1: General Requirements for Basic Safety and Essential Performance
- > IEC 60601-1-3:2013 Medical Electrical Equipment - Part 1-3: General Requirements for Basic Safety and Essential Performance -Collateral Standard: Radiation Protection in Diagnostic X-ray Equipment
- A IEC 60601-2-68: 2014 Medical Electrical Equipment - Part 2-68: Particular Requirements For
{7}------------------------------------------------
The Basic Safety And Essential Performance Of X-ray-Based Image-Guided Radiotherapy Equipment For Use With Electron Accelerators, Light Ion Beam Therapy Equipment And Radionuclide Beam Therapy Equipment
- > IEC 60601-2-28:2010 Medical Electrical Equipment-Part 2-8: Particular Requirements For The Basic Safety And Essential Performance of X-Ray Tube Assemblies For Medical Diagnosis
- > IEC 60601-2-54:2009 Medical Electrical Equipment- Part 2-54: Particular Requirements For The Basic Safety And Essential Performance Of X-ray Equipment For Radiography And Radioscopy
- A IEC 60601-1-2:2014 Medical Electrical Equipment - Part 1-2: General Requirements For Basic Safety And Essential Performance-Collateral Standard: Electromagnetic Compatibility -Requirements And Tests
- > IEC 60601-1-6:2013 Medical electrical equipment Part 1-6: General requirements for safety -Collateral Standard: Usability
- A IEC 62366-1:2015 Medical devices-Part 1: Application of usability engineering to medical devices
### Bench testing summary
- A Compatibility Testing with ROTATING GAMMA SYSTEM INFINI(INFINI)
The subject device is integrated with Masep's gamma ray head SRT system. The performance of the SRT system before and after the integration is compared to access the safety and effectiveness of the gamma ray SRT system.
The targeting accuracy of the integrated gamma ray SRT system is measured using end-to-end test. The targeting accuracy should be no larger than 1.0 mm.
The anthropomorphic target phantom is used to simulate the clinical process of real patient treatment, and test the targeting accuracy.
According to the test results, the targeting accuracy of Masep's gamma ray head SRT system after integration well meets the requirements for SRS. Its clinical process remains as effective as before integration. In summary, the integration of the subject device with Masep's gamma ray SRT system does not affect the performance of the gamma ray SRT system.
{8}------------------------------------------------
#### > Compatibility Testing with TRILOGY SYSTEM WITH RAPIDARC
The subject device is integrated with Varian linear accelerator system. The performance of the linear accelerator system before and after the integration is compared to access the safety and effectiveness of the linear accelerator system.
The targeting accuracy of the integrated linear accelerator system is measured using end-to-end test. The targeting accuracy should be no larger than 1.0 mm.
Positioning accuracy comparison between the product and CBCT of the linear accelerator; targeting accuracy measurement for the integrated linear accelerator system.
According to the test results, the positioning results of the subject devic and Varian OBI CBCT are highly consistent. The targeting accuracy of Varian Trilogy accelerator after integration well meets the requirements for SRS. Its clinical process remains as effective as before integration. In summary, the integration of the subject device with Varian Trilogy accelerator system does not affect the performance of the Varian Trilogy accelerator system.
#### > Compatibility Testing with ELEKTA SYNERGY, ELEKTA SYNERGY S AND XVIR3.5
The subject device is integrated with Elekta linear accelerator system. The performance of the linear accelerator system before and after the integration is compared to access the safety and effectiveness of the linear accelerator system.
The targeting accuracy of the integrated linear accelerator system is measured using end-to-end test. The targeting accuracy should be no larger than 1.0 mm.
Positioning accuracy comparison between the product and CBCT of the linear accelerator; targeting accuracy measurement for the integrated linear accelerator system.
According to the test results, the positioning results of the subject device and Elekta XVI CBCT are highly consistent. The targeting accuracy of Elekta Synergy accelerator after integration well meets the requirements for SRS. Its clinical process remains as effective as before integration. In summary, the integration of the subject device with Elekta Synergy accelerator system does not affect the performance of the Elekta Synergy accelerator system.
- > Optical Guided Tracking Accuracy Testing
The objective is to evaluate the optical tracking accuracy of the NDI unit under the worst
{9}------------------------------------------------
environmental conditions.
The 3D RMS volumetric accuracy acceptance criterion is 0.15 mm for extended pyramid volume.
The optical guided tracking accuracy is measured using a treatment couch to provide the translational shift as a reference.
According to the test results of the tests performed under the worst environmental conditions show that the 3D RMS volumetric accuracy is 0.11 mm.
- A System Positioning Accuracy Testing
The subject device provides two image registration approaches using bony structures and fiducial markers, respectively. The software exports 6 degrees of freedom positioning correction results, including three translational shifts and three rotational angles.
The overall positioning accuracy should be no larger than 1.0 mm, and the individual rotational angle accuracy should be no larger than 1.0°.
The subject device is installed with two types of flat panel detectors, CARERAY CareView 1800IF and PerkinElmer XRD 0840AN, respectively. Positioning accuracy of the subject device is tested using fiducial registration or bony structure registration algorithm with an anthropomorphic head phantom and a thoracic spine phantom. Measurements are performed using a treatment couch to provide the translational shifts as reference.
According to the test results, the subject device with either CARERAY CareView 1800F or PerkinElmer XRD 0840AN flat panel detectors meet the testing criteria that the overall translational error is no larger than 1.0 mm, and the maximum of rotational errors is no larger than 1.0°.
- A System Tracking Accuracy Testing
The objective is to evaluate the tracking accuracy for regular and sporadic respiratory motion under the worst environmental conditions.
The target tracking accuracy is better than 1.5 mm.
The tests are performed on the respiratory motion with moving inserts, the phantom used for this test provides respiratory waveforms collected from clinical patient data.
{10}------------------------------------------------
According to the test results, the overall tracking error of the subject device, configured with either CARERAY CareView 1800IF or PerkinElmer XRD 0840AN flat panel detectors, is no larger than 1.5 mm, testing with both typical and irregular respiratory motion waveforms.
- > System Performance Testing
The objective is to evaluate the specifications defined for the subject device.
- The mechanical accuracy of the imaging center shall not exceed 0.5 mm. a)
- The positioning accuracy shall not exceed 1.0 mm. b)
- The positioning stability shall not exceed 0.3 mm. c)
- For regular respiratory movement, the detection accuracy of the target movement caused by d) respiration shall not exceed 1.5 mm.
The performance testing described in this submission was performed according to existing manufacturer Standard Operation Procedure (SOP), protocols and pre-determined performance criteria.
According to the test results, all specifications of the subject device meets the requirement.
- 10. Clinical Test Conclusion
No clinical study is included in this submission.
### 11. Substantially Equivalent (SE) Comparison
Table 1 Comparison of Technology Characteristics
| Item | Subject device | Predicate Device |
|----------------------|-----------------|------------------|
| Product Code | IYE | IYE |
| Regulation<br>Number | 21 CFR 892.5050 | 21 CFR 892.5050 |
| Class | II | II<br>K120789 |
{11}------------------------------------------------
| Intended Use | The Image Guidance System for Radiotherapy is intended to be used in combination with the radiotherapy device, for image-guided positioning verification and real-time tracking of target motion for patient radiation therapy. The Image Guidance System for Radiotherapy sends the positioning data or real-time tracking data to the radiotherapy device but does not control any part of the radiotherapy device. | ExacTrac is intended to be used to place patients at an accurately defined pointed within the treatment beam of a medical accelerator for stereotactic radiosurgery or radiotherapy procedures, in order to treat lesions, tumors and conditions anywhere in the body when radiation treatment is indicated. ExacTrac may also be used to monitor the patient position during treatment. |
|------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Single Use | No | No |
| Sterility<br>Condition | Non-sterile | Non-sterile |
| Patient-contact<br>component | None | None |
| Positioning<br>accuracy | ≤1.0 mm | ≤1.0 mm |
The subject device and predicate device have the same functionalities, image-guided positioning and real-time tracking of target motion. Both devices have intended use. Therefore, the subject device is substantially equivalent to the predicate device.
## 12. Substantially Equivalent (SE) Conclusion
Based on the comparison and analysis above, the subject device is determined to be Substantially Equivalent (SE) to the predicate devices.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.