EXACTRAC 5.5

K072506 · Brainlab AG · IYE · Oct 26, 2007 · Radiology

Device Facts

Record IDK072506
Device NameEXACTRAC 5.5
ApplicantBrainlab AG
Product CodeIYE · Radiology
Decision DateOct 26, 2007
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.5050
Device ClassClass 2

Indications for Use

ExacTrac is a system that is intended to be used to place patients at an accurately defined point within the treatment beam of a medical accelerator for stereotactic radiosurgery or radiotherapy procedures. The ExacTrac system uses optical tracking of infrared markers and xray registration to determine and correct the position of the patient. With the additional ExacTrac gating module, the treatment beam is gated in accordance with the patient's breathing level.

Device Story

ExacTrac 5.5 is an image processing system for patient positioning and respiratory gating in radiation therapy. Inputs include infrared marker tracking data, X-ray images, and reference CT data (DRRs). The system compares real-time X-ray images or implanted radioopaque marker positions against reference CT data to calculate patient position corrections. For gating, it monitors 3D marker positions relative to breathing levels. Outputs include patient position correction values and a gating signal for the linear accelerator. Used in clinical radiation therapy settings by trained personnel; assists clinicians in verifying and correcting patient alignment to ensure accurate dose delivery. Benefits include improved targeting precision and motion management during treatment.

Clinical Evidence

Bench testing only. Validation performed according to manufacturer's design and development procedures to prove safety and effectiveness.

Technological Characteristics

Image processing system for patient positioning; utilizes optical tracking of infrared markers and X-ray registration; integrates with linear accelerators for radiation therapy; includes respiratory gating module; software-based image comparison (DRR vs. X-ray).

Indications for Use

Indicated for patients undergoing stereotactic radiosurgery or radiotherapy procedures requiring precise positioning within a medical accelerator treatment beam.

Regulatory Classification

Identification

A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # 77506 510 (k) Summary of Safety and Effectiveness for ExacTrac 5.5 #### Manufacturer Address | BrainLAB AG | |--------------------------| | Kapellenstr. 12 | | 85622 Feldkirchen | | Germany | | Phone: +49 89 99 15 68 0 | | Fax: +49 89 99 15 68 33 | | Mr. Rainer Birkenbach | | July, 2007 | OCT 26 2007 #### Device Name Contact Person Summary Date Device Name Common Name Classification Name Classification Number Regulatory Class FDA Establishment Registration Number ExacTrac 5.5, will also be marketed under the name ExacTrac X-Ray 6D or Novalis Body Patient Positioning System with respiratory Gating, Radiation Therapy, Charged-Particle, Medical System, Radiation Therapy, Charged-Particle, Medical 892.5050 Class II 804 39 33 #### Predicate Device ExacTrac 4.0(K040585) ExacTrac Gating (K033287) X-ray Generator Device Classification Name: Generator, High Voltage X-Ray, Diagnostic Regulatory Class: Class I Exempt Accession Number: 0110172-00 X-ray Tubes Device Classification Name: Assembly, Tube Housing, X-Ray, Diagnostic Regulatory Class: Class | Exempt Accession Number: 7410266-15 {1}------------------------------------------------ # Intended Use ExacTrac is a system that is intended to be used to place patients at an accurately defined point within the treatment beam of a medical accelerator for stereotactic radiosurgery or radiotherapy procedures. The ExacTrac system uses optical tracking of infrared markers and xray registration to determine and correct the position of the patient. With the additional ExacTrac gating module, the treatment beam is gated in accordance with the patient's breathing level. ## Device Description ExacTrac 5.5 is an Image Processing System for patient positioning with the ability of a gated treatment. It is based on an imported isocenter from a planning system or on an isocenter imported from a simulator. It allows verification and, if necessary, correction of the patient's position. The correction of the patient's position is based on a comparison of digital reconstructed radiographs (DRR) calculated from a corresponding CT data set (reference image). Alternatively implanted radioopaque markers may be used for localization by comparing their position in the x-ray with their original position in the reference CT data set. Within the Gating module the 3D position of the implanted markers is compared, for at least one breathing level of the patient, to their expected position based on the CT data With the resultant correction values and target movement information the patient is repositioned and the system generates a signal to gate the linear accelerator depending on the patients breathing. ## Substantial Equivalence ExacTrac 5.5 has been verified and validated according to BrainLAB's procedures for product design and development. The validation proves the safety and effectiveness of the system. The information provided by BrainLAB in this 510 (k) application was found to be substantially equivalent with the predicate device ExacTrac 4.0 (K040585) combined with ExacTraction (K033287). {2}------------------------------------------------ #### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image contains the words "Public Health Service" in a simple, sans-serif font. The text is black against a white background. The words are stacked on top of each other, with "Public Health" on the top line and "Service" on the bottom line. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 # OCT 26 2007 Mr. Rainer Birkenbach Executive Vice President BrainLAB AG Kapellenstraße 12 85622 Feldkirchen GERMANY Re: K072506 Trade/Device Name: ExacTrac 5.5 Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: II Product Code: IYE Dated: July 31, 2007 Received: September 6, 2007 #### Dear Mr. Birkenbach: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter. | 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 | |-----------------|----------------------------------|--------------| | 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 | | 21 CFR 892.xxxx | (Radiology) | 240-276-0120 | | Other | | 240-276-0100 | Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely vours. Nancy C. Brogdon Nancy C. Brogdon Director, Division of Reproductive. Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Indications for Use 510(k) Number (if known): K072506 Device Name: ExacTrac 5.5 Indications For Use: ExacTrac is a system that is intended to be used to place patients at an accurately defined point within the treatment beam of a medical accelerator for stereotactic radiosurgery or radiotherapy procedures. The ExacTrac system uses optical tracking of infrared markers and x-ray registration to determine and correct the position of the patient. With the additional ExacTrac gating module, the treatment beam is gated in accordance with the patient's breathing level. Amyu Th Thx (Division Sián-Off) Division of Reproductive, Abdominal and Radiological Devices 510(k) Number Prescription Use × (Per 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...