K132847 · Ion Beam Applications S.A. · LHN · Nov 20, 2013 · Radiology
Device Facts
Record ID
K132847
Device Name
I2C
Applicant
Ion Beam Applications S.A.
Product Code
LHN · Radiology
Decision Date
Nov 20, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.5050
Device Class
Class 2
Attributes
Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K132847 · Nov 20, 2013
I2C
Ion Beam Applications S.A.
Anonymized patient data from IBA treatment centers
Anonymized patient data collected from routine clinical practice at IBA treatment centers were used as part of the performance verification and validation testing for the I2C system.
Performance verification; Anonymized patient data; Retrospective clinical data
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Performance verification testing; Retrospective analysis of clinical data
Patients treated at IBA treatment centers
Not applicable for this study
System performance and matching accuracy
Indications for Use
I2C is used with a charged particle or photon radiation therapy system for localization of the patient position with respect to the therapy equipment and to provide correction feedback to the radiation therapy device.
Device Story
I2C (IBA Image Guided Therapy Console) integrates with radiation therapy systems to verify patient positioning. Input: 2D, 2D stereoscopic, or 3D X-ray images acquired via OEM detectors (Varian/Thales). Operation: Software registers acquired images against reference DRRs (from planning CT) or 3D datasets using intensity-based matching, anatomical/fiducial markers, or manual drag-and-drop. Output: 6-degree-of-freedom correction vector exported to third-party radiation therapy systems. Used in clinical settings by therapists to ensure target volume alignment at isocenter. Benefits: Enables submillimeter matching accuracy, improving precision of radiation delivery. Connectivity: Exchanges data with Oncology Information Systems via DICOM.
Clinical Evidence
Bench testing only. Validation performed in simulated clinical environments using phantoms and real hardware configurations. Testing included workflow scenarios, exception cases, and communication verification with third-party systems (Elekta Mosaiq, Varian Aria). Performance evaluated using datasets from simulated treatments, phantom radiographs, and anonymized patient data. Usability assessed by reference users in proton therapy.
Indicated for patient position localization and correction feedback in charged particle or photon radiation therapy systems.
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
{0}------------------------------------------------
# NOV 2 0 2013
K132847
- Page 1 of 8
iba
M-Id 39876 Rev. B Page 2
510(k) Summary 12C
# 510(k) SUMMARY lon Beam Applications S.A.
06 November 2013
Submitter
lon Beam Applications S.A. Chemin du Cyclotron, 3 B-1348 Louvain-la-Neuve Belgium Contact person : Mr Baelen Michel 32-10-47-58-45 Phone : 32-10-47-58-10 Facsimile: E-mail: Michel Baelen@iba-group.com
# Contact and Agent for lon Beam Applications S.A. in the US
Bruce D. Armon Saul, Ewing, Remick & Saul Centre Square West 38th Floor Philadelphia, PA 19102-2186 Phone: (215) 972-7124 Facsimile: (215) 972-1906 E-mail: barmon@saul.com
ಗಿ
{1}------------------------------------------------
# Name of the Device
I₂C (IBA Image Guided Therapy Consol) (brand name: adapt Insight)
## Classification Name
Medical charged-particle radiation therapy systems. (21 C.F.R. §892.5050)
### Predicate Devices
I2C is claimed to be equivalent to the predicate devices listed in table 1 below, based on the fact that they have the same intended use and principles of operation, and are substantially equivalent in terms of performance and technological characteristics.
| Device | Indication for use |
|-------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Verisuite<br>(K080742, Medcom) | The VeriSuite patient position verification system is used for<br>verification and correction of the patient's position during a<br>radiotherapy treatment with external beams or charged particles. It is<br>based on stereoscopic X-ray images and DRRs calculated from a CT<br>image series of the treatment region of the patient and information<br>from the treatment planning. |
| ExacTrac 3rd Party<br>(K072046, Brainlab) | The ExacTrac 3rd Party system is intended to be used in conjunction<br>with the MHI-TM2000 radiation therapy linear accelerator system<br>manufactured by Mitsubishi Heavy Industries, Ltd.<br>ExacTrac 3rd Party uses the images received from the MH1-TM2000<br>linear accelerator for analyzing the current patient position and<br>calculating - when applicable - a necessary correction shift. The<br>correction shift is then exported to the MHI-TM2000 linear<br>accelerator.<br>The ExacTrac 3rd Party system uses stereoscopic x-ray or cone<br>beam CT registration and optical tracking of infrared reflective<br>markers in order to localize and correct the patient position before<br>and during treatment. |
| OBI<br>(K040192 &<br>K042720, Varian) | The On-Board Imager device is used for verification of correct patient<br>position in relation to isocenter and verification of the treatment fields<br>in relation to anatomical and/or fiducial landmarks |
Table 1: List of the identified predicate devices and their indication for use
# Indication for Use of the device
IzC is used with a charged particle or photon radiation therapy system for localization of the patient position with respect to the therapy equipment and to provide correction feedback to the radiation therapy system.
{2}------------------------------------------------
### Description of the device
For clinical use, l2C must be integrated into a radiation therapy system. I2C will interact with components of the radiation therapy center.
I2C supports the acquisition of 2D, 2D stereoscopic and 3D images using 2D detectors.
I2C will be used by the clinical therapist to verify by imaging that the treatment target position received from the treatment control applicative laver is 'valid', i.e. that it brings the center of the treatment target volume at the isocenter of the therapy equipment with required accuracy. If it is not, InC will propose a correction shift - or correction vector - that will be exported to the radiation therapy system.
### Technological Characteristics of the device
#### 1. Image acquisition
The I2C software supports the acquisition of 2D, 2D stereoscopic and 3D images using 2D detectors.
The 2D and 2D stereoscopic images can be acquired using one or multiple imaqing devices arranqed in various geometrical configurations.
The 3D images are created using a single 2D imaging device which rotates around the volume to be imaged: The 3D images are then reconstructed using the acquired 2D images and the associated spatial information (angle of acquisition).
The image acquisition supports a range of OEM devices1, such as the following X-ray tubes, generators and detectors:
| | 2D | 2D/3D |
|------------|----------------------|----------------------|
| X-ray tube | Varian B130H | Varian GS2075 |
| Generator | Sedecal SHF835 | Sedecal SHF845RF |
| Detector | Varian Paxscan 4030E | Thales Pixium RF4343 |
13
<sup>1</sup> The conformity of these devices and their proper integration into the I-C system were ensured by the verification and validation activities followed during the development of the system and described in 12C VnV plan - MID 37685.
{3}------------------------------------------------
### 2. Image registration
Image registration is used to verify, based on the reference and acquired images. that the position of the patient in regard to the equipment matches the treatment plan geometry and eventually compute a correction vector that can be applied to ensure correct patient positioning. It has been designed such that a submillimeter matching accuracy can be reached.
The registration provides a correction vector of up to 6 degrees of freedom that can be sent to third party systems.
### 2D image registration
The 2D or 2D stereoscopic image registration uses either:
- Intensity based information between the acquired X-ray planar images and the Digital Reconstructed Radiograph (DRR) (generated from the treatment planning reference CT) data or provided by Treatment Planning System - TPS -, ... ), this is done automatically, or
- Markers that have been identified on the acquired X-ray planar images and the DRR generated from the CT data, or
- Manual registration of the images by drag and drop of one image to the other one.
#### 3D image registration
The 3D image reqistration uses either
- Intensity based information between the acquired 3D data set and the a CT data, or
- Manual registration of the images by drag and drop of one image to the other one.
- ・ The registration provides a correction vector of 6 degrees of freedom that can be sent to third party system.
### 3. Connectivity to third party systems
The device exchanges data with 30 party systems such as an Oncology Information System using either a push or a pull method, or any third party system supporting DICOM standards.
{4}------------------------------------------------
## Comparison to the predicate devices
Like the predicate devices in Table 1, 12C is designed to be used with a charged particle or photon radiation therapy system for localization of the patient position with respect to the therapy equipment and to provide correction feedback to the radiation therapy device-
The predicate devices also provide the same or substantially equivalent functions, characteristics, and accessories.
Based on the purpose of the 12C system, its environment of use, its main components and features, the following elements have been used for comparison:
- . Intended use
- Radiation type: clinical relevance, patient safety, known technology .
- . Acquisition geometry
- Detector type .
- Image used as reference for matching .
- o Data compared during matching
- . Type of matching approach
- Communication to 3rd party software .
#### Table 2: Comparison table of I2C with identified predicate devices
| Characteristics | I₂C | Verisuite<br>(K080742,<br>Medcom) | ExacTrac<br>3rd Party<br>(K072046,<br>Brainlab) | OBI<br>(K040192 &<br>K042720,<br>Varian) |
|----------------------------------------------------------------|-----|-----------------------------------|-------------------------------------------------|------------------------------------------|
| Indication for use: | | | | |
| • Verification of patient<br>setup position | ✓ | ✓ | ✓ | ✓ |
| • Used with a charge<br>particle radiation therapy<br>system | ✓ | ✓ | | |
| • Used with a photon<br>radiation therapy system | ✓ | ✓ | ✓ | ✓ |
| Image acquisition: | | | | |
| • High accuracy kV X-Ray<br>imaging with digital flat<br>panel | ✓ | ✓ | ✓ | ✓ |
{5}------------------------------------------------
iba
| Characteristics | I₂C | Verisuite<br>(K080742,<br>Medcom) | ExacTrac<br>3rd Party<br>(K072046,<br>Brainlab) | OBI<br>(K040192 &<br>K042720,<br>Varian) |
|---------------------------------------------------------------------------------------|-----|-----------------------------------|-------------------------------------------------|------------------------------------------|
| • Stereoscopic acquisition | ✓ | ✓ | ✓ | ✓ |
| • Flexible X-Ray setup:<br>various arrangements of<br>panels, generators and<br>tubes | ✓ | ✓ | | |
| Image matching: | | | | |
| • Registration using<br>anatomical landmarks | ✓ | ✓ | ✓ | ✓ |
| • Registration using fiducial<br>landmarks | ✓ | ✓ | ✓ | ✓ |
| • Registration using CBCT<br>with planning CT | ✓ | | | ✓ |
| • Registration of<br>radiographs with DRR<br>generated from planning<br>CT | ✓ | ✓ | ✓ | ✓ |
| o 2D/2D (5 DOF<br>registration) | ✓ | ✓ | ✓ | ✓ |
| o 2D/3D (6 DOF<br>registration) | ✓ | ✓ | ✓ | - |
| Interface to 3rd party applications: | | | | |
| • Transfer of correction shift<br>(correction vector) | ✓ | ✓ | ✓ | - |
| • DICOM exchange of<br>clinical data | ✓ | ✓ | ✓ | ✓ |
Table 3 provides a comparison of 12C with predicate devices for some of characteristic specification.
Table 3: Indication of performance and technological specification of I-C and its predicate devices
| Performance / technological<br>specification | I2C | Verisuite<br>(K080742,<br>Medcom) | ExacTrac<br>3rd Party<br>(K072046,<br>Brainlab) | OBI<br>(K040192 &<br>K042720,<br>Varian) |
|----------------------------------------------|-----|-----------------------------------|-------------------------------------------------|------------------------------------------|
|----------------------------------------------|-----|-----------------------------------|-------------------------------------------------|------------------------------------------|
16
{6}------------------------------------------------
iba
| Performance / technological<br>specification | I2C | Verisuite<br>(K080742,<br>Medcom) | ExacTrac<br>3rd Party<br>(K072046,<br>Brainlab) | OBI<br>(K040192 &<br>K042720,<br>Varian) |
|----------------------------------------------|-----------------------|-----------------------------------|-------------------------------------------------|------------------------------------------|
| Generator operating range<br>(radiographic) | 40-150 kVp | 40-150 kVp | - | 40-150 kVp |
| Generator operating range<br>(CBCT) | 40-125 kVp | N/A | N/A | 60-140 kVp |
| Flat panel pixel size | 148 µm | 127 µm | - | 194 µm |
| Flat panel pixel matrix | > 2880x2880<br>pixels | 3200x3200<br>pixels | - | 3200x2304<br>pixels |
| CBCT scale & distance<br>accuracy | 1% | N/A | N/A | 1% |
| CBCT spatial resolution | At least 5<br>lp/cm | N/A | N/A | 4-7 lp/cm |
| CBCT low contrast resolution | 15mm@1% | N/A | N/A | 15mm@1% |
| CBCT numbers accuracy | +/- 40 HU | N/A | N/A | +/- 40 HU |
| CBCT Uniformity | +/- 40 HU | N/A | N/A | +/- 40 HU |
| Achievable matching accuracy | < 1 mm | < 1 mm | 1 - 2 mm | - |
Based on the previous analysis and discussion, we can state that for each of its functionalities reqarding its intended use. I2C always has the same technological characteristics and is as safe and effective as at least one of the identified predicate devices. I2C performs as well or better as the identified predicate devices while not providing or using new functionality or technology.
### Nonclinical tests for determination of substantial equivalence
When verifying the effectiveness and safety of the design and its implementation, part of the activities was done in a simulated clinical environment with the actual hardware and 30 party software applications. The X-Ray imaging equipment was installed on a test bench in a vault appropriate for radiologic operation. The qeometry of the system was adapted to represent different configuration setups. The rotation of the gantry was simulated by a rotation of the phantom on a turn table driven by a real gantry controller. A configuration of 30 party software applications was setup in order to test the whole treatment workflow scenarios, including exception cases.
{7}------------------------------------------------
K132897
Page 8 of 8
iba
18
A second test environment was used to verify communication with different 30 party software configurations (Elekta Mosaiq, Varian Aria) installed on their dedicated workstations.
Third, additional performance tests were done on a stand-alone system with appropriate datasets collected from simulated treatments and radiographs of phantom acquired in IBA treatment centres, from test bench data acquired with phantoms, and from anonymised patient data provided by IBA treatment centers.
As a fourth verification and validation strategy axis, a process of evaluation by a group of a-users has been put in place to assess the usability of the software. Throughout the development, intermediate releases are distributed to a group of reference users in proton therapy. A process has been put in place to collect a user evaluation of the system, especially reqarding usability.
For a more elaborate description of the V&V activities, please refer to M-ID 37685 (I2C v1: V&V Plan).
### Conclusions
In conclusion, the verification and validation activities insure that the device is as safe, as effective, and performs as well as the legally marketed device identified in Table 1.
{8}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/8/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight, composed of three curved lines.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 20. 2013
lon Beam Applications S.A. % Mr. Bruce D. Armon, Partner Saul Ewing LLP Center Square West 1500 Market Street, 38th Floor PHILADELPHIA PA 19102
Re: K132847 Trade/Device Name: I2C Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: II Product Code: LHN Dated: September 9. 2013 Received: September 23, 2013
Dear Mr. Armon:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you. however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting vour device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA`s issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act s requirements, including, but not limited to: registration and listing (2 l CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803): good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{9}------------------------------------------------
Page 2-Mr. Armon
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.ida.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Michael O. Husa
for
Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{10}------------------------------------------------
# Indications for Use
510(k) Number (if known): K132847
Device Name: 12C
Indications For Use:
Indications For Use: 12C is used with a charged particle or photon radiation therapy system for localization of the patient position with respect to the therapy equipment and to provide correction feedback to the radiation therapy device.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of Center for Devices and Radiological Health (CDRH)
Michart. D. OHara
(Division Sign-Off) Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health .510(k) K132847
Page 1 of _ 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.