The device is intended for intracavitary brachytherapy for cancer treatment of the cervix and the endometrium. Optional needles can be placed for interstitial brachytherapy.
Device Story
Geneva is a gynecological applicator for intracavitary brachytherapy; used for cervix and endometrium cancer treatment. Device consists of tubes and ovoids; guides radioactive source from afterloader to target volume. Optional needles allow interstitial brachytherapy. Used in hospitals by qualified physicians; compatible with Nucletron remote afterloader systems. Operates in MR, CT, X-ray, and US environments. Provides path for isotope source to reach target volume; facilitates radiation delivery to tumor sites. Benefits patients by enabling precise, localized radiation therapy.
Clinical Evidence
Bench testing only. Validation performed at a hospital site under simulated clinical conditions with clinical personnel. Testing included usability validation, reprocessing process validation, and biological evaluation. Results demonstrated suitability for intended use, clinical acceptance, and performance meeting requirements in MR, CT, X-ray, and US environments.
Technological Characteristics
Tandem and ovoid type applicator with compatible interstitial needles. Materials are identical to the predicate device. Compatible with Nucletron remote afterloader systems. Designed for use in MR, CT, X-ray, and US imaging environments.
Indications for Use
Indicated for intracavitary brachytherapy for cancer treatment of the cervix and endometrium; optional interstitial needles for interstitial brachytherapy.
Regulatory Classification
Identification
A remote controlled radionuclide applicator system is an electromechanical or pneumatic device intended to enable an operator to apply, by remote control, a radionuclide source into the body or to the surface of the body for radiation therapy. This generic type of device may include patient and equipment supports, component parts, treatment planning computer programs, and accessories.
Predicate Devices
Utrecht Interstitial CT/MR Applicator Set (K091154)
K040704 — MODIFICATION TO HENSCHKE HDR CERVIX APPLICATOR AND HILARIS/NORI HDR CERVIX APPLICATOR · Mick Radio-Nuclear Instruments, Inc. · Jun 18, 2004
K160516 — Universal Cylinder Applicator Family, Universal Segmented Cylinder Applicator Set, Universal Stump Applicator Set, Universal Cervix Probe Set, odd lengths, Universal Cervix Probe Set, even lengths · Varian Medical Systems, Inc. · Aug 10, 2016
K132889 — FLETCHER CT/MR SHIELDED APPLICATOR SET (6 MM) FLETCHER CT/MR SHIELDED APPLICATOR SET (4 MM) · Nucletron B.V. · Feb 26, 2014
K222775 — Aarhus Applicator Set · Varian Medical Systems, Inc. · Mar 13, 2023
Submission Summary (Full Text)
{0}------------------------------------------------
July 16, 2020
Image /page/0/Picture/1 description: The image contains the logo for the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Nucletron BV % Rigo Meens Head of QA Waardgelder 1 Veenendaal, Utrecht 3905 TH THE NETHERLANDS
Re: K201272
Trade/Device Name: Geneva Regulation Number: 21 CFR 892.5700 Regulation Name: Remote controlled radionuclide applicator system Regulatory Class: Class II Product Code: JAO Dated: March 20, 2020 Received: May 12, 2020
Dear Rigo Meens:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for Elekta. The logo consists of a circular graphic on the left and the word "Elekta" on the right. The circular graphic is made up of a larger circle with three smaller circles inside of it. The word "Elekta" is written in a sans-serif font.
#### Indications for Use Statement A4.
See next page
{3}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known)
K201272
Device Name Geneva
Indications for Use (Describe)
The device is intended for intracavitary for cancer treatment of the cervix and the endometrium. Optional needles can be placed for interstitial brachytherapy.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
O Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda. hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Elekta, a company that specializes in precision radiation medicine. The logo consists of a circular icon with three smaller circles inside, followed by the word "Elekta" in a bold, sans-serif font. The logo is teal in color and is set against a white background.
#### 510(k) Summary A5.
See next pages
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Elekta, a company that specializes in precision radiation medicine. The logo consists of a circular graphic with three smaller circles inside, followed by the word "Elekta" in a sans-serif font. The color of the logo is a teal or cyan shade, giving it a modern and professional appearance.
#### K201272
20 March 2020
# Submitter of 510(k):
510(k) Summary
| Submitter of 510(k): | |
|----------------------|----------------------------------------------------|
| Company name: | Nucletron B.V. |
| Registration number: | 611894 |
| Address: | Waardgelder 1, 3905 TH Veenendaal, The Netherlands |
| Phone: | +31 318 557 133 |
| Correspondent: | Leidimar Guararima |
#### Device Name:
Trade/Proprietary Name: Common/Usual Name: Classification Name: Classification: Product code:
Geneva Brachytherapy Applicator Remote controlled radionuclide applicator system 21CFR892.5700, Class II JAQ
# Legally Marketed Device(s)
Our device is substantially equivalent to the following legally marketed predicate device:
| Manufacturer | Device | 510(K)# | Device classification |
|----------------|----------------------------------------------|---------|-----------------------|
| Nucletron B.V. | Utrecht Interstitial CT/MR<br>Applicator Set | K091154 | Class II |
The Geneva is a gynecological applicator for intracavitary brachytherapy with the option for interstitial brachytherapy treatment. It can be used for the treatment of the cervix and endometrium. The applicator consists of tubes and ovoids, guiding the radioactive source of the afterloader to the target volume. Compatible components can be combined to reach a wide range of target areas.
## Indications for use:
The device is intended for intracavitary brachytherapy for cancer treatment of the cervix and the endometrium. Optional needles can be placed for interstitial brachytherapy.
The indications for use for the Geneva are identical to the indications for use of the legally marketed above-mentioned predicate device.
## Summary of technological considerations:
The subject device and the predicate devices are intended to provide a path for the isotope source to travel to the target volume. The devices are used in the hospital by qualified physicians. Geneva is used in the same anatomical sites as the predicate device and is made of the same materials.
The devices are compatible with Nucletron remote afterloader systems and accessories. Section C provides an overview of the similarities and differences of Geneva with the predicate device.
The subject device is tandem and ovoid type applicator with compatible interstitial needles. Geneva is identical in principle technology, clinical function and technical and biological characteristics to the predicate device design fits in the same
{6}------------------------------------------------
clinical workflow as the predicate device. As a result, it is determined that the Geneva is substantially equivalent to the legally marketed predicate device.
# Summary of testing:
Bench testing was performed at a hospital site, under simulated clinical conditions and with the involvement of clinical personnel but excluding the delivery of treatment to patients. Experienced users reviewed the device design and executed validation tests. The test results demonstrated the suitability of the device to its intended use and user needs and demonstrated clinical acceptance of the device.
Usability testing demonstrated that the device can be used by the intended users, under simulated clinical conditions, without serious use errors or problems. Validation of reprocessing processes and biological evaluation was performed. The device can be used in the MR, CT, X-ray and US environments. Bench testing (similar to bench testing done to the Legally Marketed Device) shows that the device meets its performance requirements.
The results of the testing provided in this submission adequately demonstrate that the Geneva performs as defined in the requirements, meets clinical expectations and is safe and effective for clinical use.
## Conclusion:
As a result, it was determined that Geneva is substantially equivalent in intended use, function, design and technological characteristics to the legally marketed predicate device.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.