FLETCHER CT/MR SHIELDED APPLICATOR SET (6 MM) FLETCHER CT/MR SHIELDED APPLICATOR SET (4 MM)

K132889 · Nucletron B.V. · JAQ · Feb 26, 2014 · Radiology

Device Facts

Record IDK132889
Device NameFLETCHER CT/MR SHIELDED APPLICATOR SET (6 MM) FLETCHER CT/MR SHIELDED APPLICATOR SET (4 MM)
ApplicantNucletron B.V.
Product CodeJAQ · Radiology
Decision DateFeb 26, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.5700
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Fletcher CT/MR Shielded Applicator Set is intended for gynecological brachytherapy treatment of the cervix and endometrium.

Device Story

Gynecological brachytherapy applicator set; used for treatment of cervix and endometrium cancer. Consists of one intrauterine tube and two ovoid tubes; guides radioactive source from afterloader to target tissue. Incorporates tungsten alloy shielding in ovoids to reduce radiation dose to rectum and bladder. Shielding position is adjustable to minimize CT imaging artifacts; compatible with MR and X-ray imaging without shield adjustment. Used in clinical brachytherapy settings by trained medical professionals. Enables 3D treatment planning by allowing CT/MR imaging compatibility. Benefits patient by facilitating precise radiation delivery while protecting healthy surrounding tissues.

Clinical Evidence

Bench testing only. Validation of sterilization processes, biological evaluation, and performance testing in MR and CT environments were conducted. Results demonstrate that the modified device meets performance requirements and is equivalent to the legally marketed predicate device.

Technological Characteristics

Intracavitary remote afterloading applicator. Materials: Tungsten alloy shielding. Dimensions: Comparable to predicate. Imaging: CT/MR compatible. Energy: Radionuclide source (afterloader). Sterilization: Validated process.

Indications for Use

Indicated for gynecological brachytherapy treatment of cervical and endometrial cancer.

Regulatory Classification

Identification

A remote controlled radionuclide applicator system is an electromechanical or pneumatic device intended to enable an operator to apply, by remote control, a radionuclide source into the body or to the surface of the body for radiation therapy. This generic type of device may include patient and equipment supports, component parts, treatment planning computer programs, and accessories.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K132889/ Page 1 of 2 Image /page/0/Picture/1 description: The image shows the word "Nucletron" in a bold, sans-serif font. To the left of the word is a logo consisting of a large, dark circle with two smaller circles overlapping it. The smaller circles are white with a dark outline, creating a visual effect of atomic structure. NUCLETRON B.V. Waardgelder 1 3905 TH Veenendaal P.O.Box 930 3900 AX Veenendaal The Netherlands Phone +31 318 557 133 Fax +31 318 557 118 FEB 2 6 2014 Department of Health and Human Services Centre of Device and Radiological Health Office of Device Evaluation Traditional 510(k) section ## 510(k) Summary September 13, 2013 ## Submitter of 510(k): Company name: Nucletron BV Registration number: 611894 Address: Waardgelder 1, 3905 TH Veenendaal, The Netherlands Phone: +31 318 557 133 Fax: +31 318 557 118 Correspondent: Rudolf Vos ### New Device Name: | Trade/Proprietary Name: | Fletcher CT/MR Shielded Applicator Set | |-------------------------|-----------------------------------------------------| | Common/Usual Name: | Intracavitary Remote Afterloading Applicator | | Classification Name: | system, applicator, radionuclide, remote-controlled | | Classification: | 21 CFR 892.5700, Class II | ## Legally Marketed Device(s) Our device is substantially equivalent to the legally marketed predicate device cited in the table below: | Manufacturer Device | | 510(k) # | |-----------------------|----------------------------------------------------|----------| | | Nucletron BV Fletcher Williamson Applicator Set· | K983341 | ## Description: The Fletcher CT/MR Shielded Applicator Set is a modification of the Fletcher Williamson Applicator Set (K983341). It is a gynecological applicator for brachytherapy procedures. The applicator uses one intrauterine tube and two ovoid tubes for treatment of the cervix and endometrium cancer. The tubes quide the radioactive source of the afterloader to the location where treatment is to be applied. {1}------------------------------------------------ K132889 Page 2 of 2 Tungsten allov shielding is incorporated in the ovoid to lower the dose in organs like rectum and bladder during treatment. The shielding position in the ovoid is adjustable, enabling CT imaging with few artifacts. For MR data acquisition and X-ray it is not necessary to move the shields. The Fletcher CT/MR Shielded Applicator Set is available in two variations and is compatible with Nucletron afterloaders and accessories. #### Intended use: The Fletcher CT/MR Shielded Applicator Set is intended for gynecological brachytherapy treatment of the cervix and endometrium. #### Summary of technological considerations: The operating principle is the same as the marketed device. The modified design combines the characteristics of the marketed device (shielding in the ovoids) with materials that allow CT and MR imaging, enabling 3D treatment planning. The shielding position in the ovoid is adjustable, enabling CT imaging with few artifacts. For MR data acquisition and X-ray it is not necessary to move the shields. Ovoid and intrauterine dimensions of the modified device are comparable to the marketed devices. #### Summary of testing: Validation of sterilization processes and biological evaluation was performed. The device was tested for use in the MR and CT environment. Bench testing (similar to bench testing done to the Legally Marketed Device) shows that the device meets its performance requirements, and that the modified device performance is equivalent to the marketed devices. #### Conclusion: Nucletron considers the Fletcher CT/MR Shielded Applicator Set to be substantially equivalent to legally marketed predicate devices through the data and information presented. No safety or effectiveness issues were identified. {2}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's emblem, which is a stylized representation of a human figure embracing a globe. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" are arranged in a circular pattern around the emblem. Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002 February 26, 2014 Nucletron B.V. % Mr. Rudolf Vos QA/RA Engineer Waardgelder 1 Veenendaal, 3905 TH THE NETHERLANDS Re: K132889 Trade/Device Name: Fletcher CT/MR Shielded Applicator Set (6 mm/4 mm) Regulation Number: 21 CFR 892.5700 Regulation Name: Remote Controlled Radionuclide Applicator System Regulatory Class: II Product Code: JAQ Dated: January 30, 2014 Received: February 3, 2014 Dear Mr. Vos: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ Page 2- Mr. Vos If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. 1 6 Sincerely yours, Michael D. O'Hara for Janine M. Morris Director. Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Indications for Use 510(k) Number (if known): Device Name: Indications for Use: K132889 Fletcher CT/MR Shielded Applicator Set The Fletcher CT/MR Shielded Applicator Set is intended for gynecological brachytherapy treatment of the cervix and endometrium. Prescription Use _ × (Part 21 CFR 801 subpart D) AND/OR Over-The-Counter Use (Part 21 CFR 801 subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of Center for Devices and Radiological Health (CDRH) Michael D. O'Hara (Division Sign-Off) Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health 510(k) _______________________________________________________________________________________________________________________________________________________________________ Page 1 of 1
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...