Sentinel

K200600 · C-Rad Positioning AB · IYE · Apr 20, 2020 · Radiology

Device Facts

Record IDK200600
Device NameSentinel
ApplicantC-Rad Positioning AB
Product CodeIYE · Radiology
Decision DateApr 20, 2020
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 892.5050
Device ClassClass 2

Indications for Use

The system is intended for use in radiation therapy clinics together with diagnostic or treatment equipment and provides: · accurate and reproducible patient positioning. • a respiratory signal to be supplied to diagnostic imaging equipment (primarily CTs) for prospectively (aka 4DCT) gated imaging and reconstruction. The system cannot directly determine the intended treatment target, since only the patient external surface is detected. The actual target position must therefore, whenever deemed necessary by qualified personnel, be verified using other systems such as CBCT or EPID.

Device Story

Sentinel system uses ceiling-mounted laser and camera scanner to acquire 3D patient surface data via laser line triangulation. Operated by qualified radiation therapy or diagnostic personnel in CT rooms. Software (c4D) reconstructs 3D surface; user marks respiratory motion measurement points on surface. System measures respiratory signal at ~16Hz during 4DCT imaging; provides signal to CT equipment for gating. Supports dual detection points for simultaneous thoracic and abdominal monitoring. No markers required on patient/couch; no additional radiation exposure. Output used by clinicians to facilitate gated imaging/reconstruction; aids in tumor localization. Does not identify internal targets; requires verification via CBCT or EPID.

Clinical Evidence

Bench testing only. No clinical data provided. Verification testing confirmed that the 64-bit software architecture update meets design inputs and maintains performance equivalent to the predicate.

Technological Characteristics

Ceiling-mounted laser and camera scanner; laser line triangulation for 3D surface reconstruction. Connectivity: interfaces with CT systems. Software: c4D version 6.0.0, 64-bit architecture. Sampling frequency: ~16Hz. No markers required.

Indications for Use

Indicated for patients undergoing radiation therapy or diagnostic CT imaging requiring accurate patient positioning and respiratory motion monitoring for 4DCT gated imaging and reconstruction. No specific age or gender contraindications stated.

Regulatory Classification

Identification

A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. C-Rad Positioning AB % Mr. Thomas Matzen Director Quality and Regulatory Affairs Bredgränd 18 Uppsala, Uppsala 75320 SWEDEN April 20, 2020 Re: K200600 Trade/Device Name: Sentinel Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: Class II Product Code: IYE Dated: March 23, 2020 Received: March 23, 2020 Dear Mr. Matzen: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for {1}------------------------------------------------ devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. For Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K200600 Device Name Sentinel Indications for Use (Describe) The system is intended for use in radiation therapy clinics together with diagnostic or treatment equipment and provides: · accurate and reproducible patient positioning. • a respiratory signal to be supplied to diagnostic imaging equipment (primarily CTs) for prospectively (aka 4DCT) gated imaging and reconstruction. The system cannot directly determine the intended treatment target, since only the patient external surface is detected. The actual target position must therefore, whenever deemed necessary by qualified personnel, be verified using other systems such as CBCT or EPID. X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for C-RAD. The logo consists of a stylized letter "N" in orange, followed by the text "C-RAD". The "C" is also in orange, while the "RAD" is in gray. The logo is simple and modern, and the colors are eye-catching. ## 510(k) Summary K200600 ### Owner: C-Rad Positioning Ab Bredgränd 18 753 20 Uppsala Phone no. +46 (0)18 66 69 30 Fax no. +46 (0)18 12 69 30 #### Establishment registration no: 3006621300 500021500 ## Contact person: Ellinor Nami Regulatory Affairs Specialist Phone no: +46 (0)18 410 82 36 ## Date of preparation: February 24, 2020 # Trade name of device: Sentinel Sentinel 4DCT #### Common name: Radiotherapy positioning system #### Classification name: Medical charged-particle radiation therapy system Regulation: 21 CFR 892.5050 Product code: IYE ## Predicate marketed device(s): Sentinel (K120668) #### Device description: The Sentinel hardware consists of a single scanner unit containing the laser and camera. mounted in the ceiling in front of the CT. The scanner is connected to the PC running the c4D software. During patient surface acquisition, a laser line is swept along the patient while the camera records a number of images. From the data acquired, a complete 3D surface of the patient can be reconstructed using laser line triangulation. On this surface the user, a health care professional, virtually marks where on the patient the respiratory motion signal shall be measured and in the next step, when the 4DCT imaging starts, the signal is measured with a frequency of around 16Hz. In general, the system is capable of acquiring more than 50 contours per second. The acquired 3D surface can also be exported and used as a reference image in the treatment room. The Sentinel system for the CT room is used in radiation therapy clinics to perform prospective or retrospective gated imaging (4DCT) prior to treatment. The system Sentinel 510(k) application Vol 006 510(k) Summary Page 1 (3) {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the logo for C-RAD. The logo consists of an orange geometric shape resembling an abstract letter "N" or a stylized mountain range. To the right of the shape is the text "C-RAD," with the "C" in orange and the "RAD" in gray. The logo has a clean and modern design. provides information about a patient's respiratory motion during the localization of the tumour in CT imaging. The system shall only be used by hospital personnel, qualified to work in radiation therapy or diagnostics departments. The Sentinel platform is based on advanced laser technology with multipurpose software modules covering different tasks in the diagnostic procedure. The software is user friendly and requires a minimum of user interaction in the daily clinical workflow. The software is designed to integrate with existing CT systems at the clinic. The Sentinel system does not require any markers to be placed on the patient or the couch and doesn't subject the patient to any additional radiation. This also means that the personnel can stay in the CT room during the whole set up procedure. Sentinel for the CT room includes the cRespiration module for respiratory gating during diagnostic CT imaging, so called 4D CT studies. #### cRespiration The Sentinel system allows you to perform prospectively or retrospectively gated imaging (4DCT) and with an interface to the CT system the user can easily integrate cRespiration to the clinical work flow with improved accuracy in the CT room. cRespiration allows you to have two detection points which enables both thoracic and abdominal breathing motions to be detected in parallel. #### Intended use: The system is intended for use in radiation therapy clinics together with diagnostic or treatment equipment and provides: - accurate and reproducible patient positioning. - a respiratory signal to be supplied to diagnostic imaging equipment (primarily CTs) . for prospectively and retrospectively (aka 4DCT) gated imaging and reconstruction. The system cannot directly determine the location of the intended treatment target, since only the patient external surface is detected. The actual target position must therefore, whenever deemed necessary by qualified personnel, be verified using other systems such as CBCT or EPID. #### Technological comparison: The Sentinel system is substantially equivalent to its predicate device Sentinel (K120668) in terms of their technological characteristics and its safety and efficacy profile. The Sentinel system has been used in both treatment room and CT room in radiotherapy clinics. However, since 2017 it has only been used and supported in the CT room. The change we made to the intended use reflects the actual usage of the system to show that we no longer support patient motion supervision in treatment rooms. Based on a business decision this feature has been withdrawn for new sales and hence the intended use has been updated by removing the option of patient motion supervision in treatment room. This change does not have any effect on the validation and verification activities performed on the system. The c4D software version 6.0.0 is supporting 64-bit architecture. In earlier versions the software was only supporting 32-bit architecture. In c4D 6.0.0 all program components have been compiled for 64-bit architecture. The functionality of the software with respect to the 64-bit support is unchanged. Complete re-verification has been performed to verify that the design outputs meets the design inputs after the change from 32-bit to 64-bit support. {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the logo for C-RAD. The logo consists of an orange geometric shape on the left, followed by the text "C-RAD" in gray. The "C" is orange, while the "RAD" is in a lighter shade of gray. ## Safety and effectiveness: Performance data has been submitted to show that Sentinel achieves its intended use and that the software update and rephrasing of intended use raise no new issues of safety or efficacy.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...