K233622 · Vision Rt, Ltd. · IYE · Mar 29, 2024 · Radiology
Device Facts
Record ID
K233622
Device Name
AlignRT Plus
Applicant
Vision Rt, Ltd.
Product Code
IYE · Radiology
Decision Date
Mar 29, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.5050
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The AlignRT Plus system is indicated for: - Tracking respiratory motion throughout the simulation process to facilitate subsequent 4DCT reconstruction and coaching the patient in breathing techniques required for Deep-Inspiration Breath Hold (DIBH). - Verification of patient identity for their radiation treatment session. - Positioning and monitoring of patients during radiation delivery, relative to the setup isocenter and/or the prescribed treatment isocenter. - Withholding the beam automatically during radiation delivery, as well as gating the patient's respiratory motion. - Performing quality assurance on MV, kV imagers, room lasers, and the treatment couch. - Visualizing the Cherenkov signal associated with the radiation beam on entry and exit from the patient. - Passing and receiving information to/from other systems associated with the radiotherapy treatment.
Device Story
Video-based 3D surface imaging system; monitors patient position relative to reference surface; displays offsets to user. Tracks respiratory motion for 4DCT reconstruction and DIBH coaching. Monitors patient during radiation delivery; automatically withholds beam if patient moves out of position. Uses up to 6 cameras to project patterns on patient skin to acquire 3D images. Includes DoseRT feature for visualizing Cherenkov light emitted by radiation beam (6-18MV) on patient skin. Used in hospital/clinic radiotherapy environments by trained radiation oncology personnel. Output visualized by clinician to guide patient positioning and treatment delivery. Benefits include improved positioning accuracy, respiratory gating, and real-time visualization of radiation beam entry/exit.
Clinical Evidence
Bench testing only. No clinical investigations performed. Verification tests confirmed module functionality against pass/fail criteria. Validation testing performed using summative evaluation techniques per IEC 62366-1:2015. All test cases passed.
Technological Characteristics
Video-based 3D surface imaging; near infra-red pattern projection. Components: PC workstation, cables, video cameras, calibration tools, DoseRT camera (anodised aluminium/polyester powder coating), DoseRT reference board (ABS plastic). Positioning accuracy <1mm (0.5mm) target registration error. Respiratory tracking accuracy ≤0.5mm (≥95% confidence). Cherenkov imaging for 6-18MV beams at 100-2400MU/Min. Compliant with IEC 60601-1 and IEC 60601-1-2. Software-based processing.
Indications for Use
Indicated for any individual (adult or child) undergoing radiotherapy. Used for respiratory motion tracking, patient identity verification, patient positioning/monitoring during radiation delivery, beam gating/withholding, quality assurance of imaging/lasers/couches, and visualization of Cherenkov signals during radiation therapy.
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
{0}------------------------------------------------
March 29, 2024
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Vision RT Ltd % Chelsey Ludlow Regulatory Affairs Specialist Dove House. Arcadia Avenue London, N3 2JU UNITED KINGDOM
#### Re: K233622
Trade/Device Name: AlignRT Plus Regulation Number: 21 CFR 892.5050 Regulation Name: Medical Charged-Particle Radiation Therapy System Regulatory Class: Class II Product Code: IYE Dated: November 10, 2023 Received: November 13, 2023
Dear Chelsey Ludlow:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrb/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming
{1}------------------------------------------------
product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Locon Weidner
Lora D. Weidner, Ph.D. Assistant Director Radiation Therapy Team DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K233622
Device Name AlignRT Plus
#### Indications for Use (Describe)
The AlignRT Plus system is indicated for:
- · Tracking respiratory motion throughout the simulation process to facilitate subsequent 4DCT reconstruction and coaching the patient in breathing techniques required for Deep-Inspiration Breath Hold (DIBH).
- · Verification of patient identity for their radiation treatment session.
- · Positioning and monitoring of patients during radiation delivery, relative to the setup isocenter and/or the prescribed treatment isocenter.
- · Withholding the beam automatically during radiation delivery, as well as gating the patient's respiratory motion.
- · Performing quality assurance on MV, kV imagers, room lasers, and the treatment couch.
- · Visualizing the Cherenkov signal associated with the radiation beam on entry and exit from the patient.
- · Passing and receiving information to/from other systems associated with the radiotherapy treatment.
| Type of Use (Select one or both, as applicable) | | | | |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|---|--|
| 1 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - | | | . | |
X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# visionrt
### 510(k) Premarket Notification - AlignRT Plus 7.3
### 510(k) Summary
The information below is provided for modifications to AlignRT Plus following the requirements of the Safe Medical Device Act (SMDA) of 1990, and the content is provided in conformance with 21 CFR 807.92.
#### Submitter's information
| Submitter's name: | Chelsey Ludlow |
|----------------------------|------------------------------------------------------------------|
| Company: | Vision RT Ltd. |
| Address: | Dove House<br>Arcadia Avenue<br>London, N3 2JU<br>United Kingdom |
| Contact person: | Chelsey Ludlow<br>Regulatory Affairs Specialist |
| Phone: | +44 (0)20 8346 4300 |
| Fax: | N/A |
| Email: | regulatory@visionrt.com |
| Date summary was prepared: | 10-Nov-2023 |
### Primary Predicate device information
Device name:
Premarket notification: Manufacturer: Common name:
Classification: Classification name:
Regulation number:
#### Reference device information
- Device name: Premarket notification: Manufacturer: Common name: Classification:
AlignRT Plus (incorporating AlignRT, AlignRT InBore, AlignRT Offline, GateCT, SimRT, GateRT) K212583 Vision RT Limited Surface Guided Radio Therapy System (Patient Positioning System - accessory to a linear accelerator) Class II Accessory to Medical charged-particle radiation therapy system 892.5050
BeamSite K212606 DoseOptics LLC Radiotherapy visualization system Class II
{4}------------------------------------------------
# visionrt
### 510(k) Premarket Notification - AlignRT Plus 7.3
Classification name:
Regulation number:
Medical charged-particle radiation therapy system 892.5050
### Device information
| Trade name: | AlignRT Plus (incorporating AlignRT, AlignRT<br>InBore, DoseRT, GateCT, GateRT, SimRT) |
|-----------------------|------------------------------------------------------------------------------------------------------------|
| Common name: | Surface Guided Radio Therapy System (Patient<br>Positioning System - accessory to a linear<br>accelerator) |
| Classification: | Class II |
| Classification name: | Accessory to Medical charged-particle radiation<br>therapy system |
| Regulation number: | 892.5050 |
| Product code: | IYE |
| Classification panel: | Radiology |
### Device description
The AlignRT Plus system (K212583) is a combination of the devices AlignRT InBore, AlignRT Offline, DoseRT, GateCT, GateRT and SimRT.
AlignRT Plus is a video-based three-dimensional (3D) surface imaging system used to monitor the patient's position in 3D before and during radiotherapy treatment. During each treatment session the patient's position is compared to the reference surface and offsets are displayed to the user. The system can be used to track pation during tumour localization in the CT scanner in order to facilitate subsequent 4D CT reconstruction. The system can be used to track the breath hold level consistency throughout the simulation process. The system can also be used to track patient motion during treatment delivery for radiation therapy procedures and hold the beam when the patient is not in position. Use of optional accessories allows the user to:
- monitor the patient's position inside bore-based linacs; and/or —
- to view the Cherenkov light emitted by the radiation beam as it enters and exits the patient's skin; and/or
- verify the patient's identity.
The system is non-invasive, does not require the use of body markers and produces no irradiation during the imaging process. The system mainly consists of advanced software. 3D cameras (≤6 cameras) and calibration tools. Each camera pod monitoring the pation will, however, project a pattern on the patient to acquire a 3D image of the patient.
This 510(k) notification is to obtain clearance for the following changes to the cleared device:
- 1. Introduction of new software, version 7.3. This is an updated version of the existing cleared software, to support the:
- a. New Cherenkov imaging feature, called DoseRT.
- b. Updated respiratory gating feature, called Respiratory Module.
{5}------------------------------------------------
# visionrt
### 510(k) Premarket Notification – AlignRT Plus 7.3
- 2. Introduction of new hardware:
- a. DoseRT camera to enable Cherenkov imaging.
- 3. Update to indications for use to include the new Cherenkov imaging feature and clarify existing indications.
Throughout this submission, the device will be referred to as AlignRT Plus as the family name for all configurations. No changes have been made to the currently cleared GateCT, GateRT and SimRT configurations.
### Indications for use
The AlignRT Plus system is indicated for:
- . Tracking respiratory motion throughout the simulation process to facilitate subsequent 4DCT reconstruction and coaching the patient in breathing techniques required for Deep-Inspiration Breath Hold (DIBH).
- Verification of patient identity for their radiation treatment session.
- Positioning and monitoring of patients during radiation delivery, relative to the setup isocenter • and/or the prescribed treatment isocenter.
- . Withholding the beam automatically during radiation delivery, as well as gating the beam based on the patient's respiratory motion.
- . Performing quality assurance on MV, kV imagers, room lasers, and the treatment couch.
- . Visualizing the Cherenkov signal associated with the radiation beam on entry and exit from the patient.
- . Passing and receiving information to/from other systems associated with the radiotherapy treatment.
{6}------------------------------------------------
Image /page/6/Picture/2 description: The image shows the word "visionrt" in a stylized font. The first part of the word, "vision", is in black, while the "rt" is in blue. The "o" in "vision" is replaced with a target symbol with a vertical line extending through it.
### Primary Predicate and subject device technological characteristics
The features of AlignRT Plus 7.3 are substantially equivalent to the predicate device, AlignRT Plus (K212583), as discussed in the table below:
| Technological<br>characteristics | Primary Predicate device<br>(AlignRT Plus, K212583) | Subject device<br>(AlignRT Plus 7.3) | Comments |
|------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | Vision RT Ltd | Vision RT Ltd | The manufacturer remains the same as that<br>of the predicate device. |
| Indications for<br>Use | The AlignRT Plus system is indicated for<br>use to position and monitor patients<br>relative to the prescribed treatment<br>isocenter, and to withhold the beam<br>automatically during radiation delivery.<br><br>For cranial treatments, a manual head<br>adjuster is included which can be used in<br>concert with AlignRT Plus to provide fine<br>corrections for pitch, roll and yaw<br>rotations.<br><br>AlignRT Plus is also used to track the<br>patient's respiratory pattern for respiratory<br>synchronized image acquisition, and<br>radiation therapy treatment. Patient<br>contour data can be extracted and<br>exported from the data acquired for the<br>purpose of treatment planning.<br><br>AlignRT Plus can be calibrated directly to<br>the treatment beam isocenter and in turn<br>assists in performing quality assurance on<br>MV, kV imagers, room lasers and the<br>treatment couch. | The AlignRT Plus system is indicated for:<br>Tracking respiratory motion<br>throughout the simulation process<br>to facilitate subsequent 4DCT<br>reconstruction and coaching the<br>patient in breathing techniques<br>required for Deep-Inspiration<br>Breath Hold (DIBH). Verification of patient identity for<br>their radiation treatment session. Positioning and monitoring of<br>patients during radiation delivery,<br>relative to the setup isocenter<br>and/or the prescribed treatment<br>isocenter. Withholding the beam<br>automatically during radiation<br>delivery, as well as gating the<br>beam based on the patient's<br>respiratory motion. Performing quality assurance on<br>MV, kV imagers, room lasers, and<br>the treatment couch. Visualizing the Cherenkov signal | The subject device indications for use have<br>been updated to include visualization of the<br>Cherenkov signal, verification of the<br>patient's identity prior to radiation treatment<br>and to clarify the existing indications.<br><br>The addition of Cherenkov imaging and<br>verification of the patient's identity and the<br>clarification of the cleared indications does<br>not affect the safety or effectiveness of the<br>device. AlignRT Plus works as intended.<br>Sufficient performance data has been<br>collated and shows the device functions<br>effectively to achieve its intended use. |
| Technological<br>characteristics | Primary Predicate device<br>(AlignRT Plus, K212583) | Subject device | Comments |
| | AlignRT Plus may be used during<br>simulation, setup and stereotactic<br>radiosurgery and precision radiotherapy<br>for lesions, tumors and conditions<br>anywhere in the body where radiation is<br>indicated. | (AlignRT Plus 7.3)<br>on entry and exit from the patient.<br>Passing and receiving information<br>to/from other systems associated<br>with the radiotherapy treatment. | |
| Product code | IYE | IYE | Equivalent to the predicate device. |
| Device<br>Classification | Class II | Class II | Equivalent to the predicate device. |
| Classification<br>Name | Accessory to Medical charged-particle<br>radiation therapy accessories | Accessory to Medical charged-particle<br>radiation therapy accessories | Equivalent to the predicate device. |
| Requlation<br>Number | 21 CFR 892.5050 | 21 CFR 892.5050 | Equivalent to the predicate device. |
| Contra-<br>indications | None known. | None known. | Equivalent to the predicate device. |
| Intended users | Trained clinically qualified radiation<br>oncology personnel. | Trained clinically qualified radiation<br>oncology personnel. | Equivalent to the predicate device. |
| OTC/Rx | Rx (Prescription) | Rx (Prescription) | Equivalent to the predicate device. |
| Principles of<br>operation | Video based imaging of 3D skin surface<br>data using surface matching software. | Video based imaging of 3D skin surface<br>data using surface matching software. | Equivalent to the predicate device. |
| Target<br>Population | Any individual (adult or child) undergoing<br>radiotherapy. | Any individual (adult or child undergoing<br>radiotherapy). | Equivalent to the predicate device. |
| Technological<br>characteristics | Primary Predicate device<br>(AlignRT Plus, K212583) | Subject device<br>(AlignRT Plus 7.3) | Comments |
| Materials | PC workstation, cables, video cameras.<br>Block Polystyrene (calibration phantom),<br>carbon fibre laminate material (head<br>adjuster). | PC workstation, cables, video cameras.<br>Block Polystyrene (calibration phantom),<br>carbon fibre laminate material (head<br>adjuster).<br>Anodised aluminium and polyester<br>powder coating (DoseRT camera casing).<br>ABS plastic (DoseRT reference board). | The component materials in the subject<br>device are identical to the primary<br>predicate.<br>The introduction of the new components<br>(DoseRT camera casing and DoseRT<br>reference board) and materials have not<br>been evaluated for biocompatibility as they<br>do not come in contact with the patient, nor<br>effect patient safety. |
| System<br>Performance<br>and Accuracy | Positioning accuracy: Target registration<br>errors (as measured using calibration<br>phantom) < 1mm (0.5mm) for all couch<br>angles.<br>Respiratory tracking: Tracks respiratory<br>signal from imaged surface data and<br>sends to CT (4D CT) or to Linac or<br>imaging device (gating).<br>Surface displacements can be tracked<br>with RMS errors < 0.5mm over 10 or more<br>breathing cycles. | Positioning accuracy: Target registration<br>errors (as measured using calibration<br>phantom) < 1mm (0.5mm) for all couch<br>angles.<br>Respiratory tracking: Tracks respiratory<br>signal from imaged surface data and<br>sends to CT (4D CT) or to Linac or<br>imaging device (gating).<br>The relative respiratory position within<br>respiratory cycles is calculated to an<br>accuracy of ≤0.5mm with a confidence<br>level of ≥95%.<br>The location of the radiation dose on the<br>patient surface during treatment can be<br>visualized using the Cherenkov Effect for<br>treatment at treatment beam energies of<br>≥6MV and ≤18MV and at treatment beam<br>delivery rates of between ≥100MU/Min<br>and ≤2400MU/Min. | As a result of the updated features, such as<br>the visualization of the Cherenkov signal<br>and respiratory tracking associated with the<br>Respiratory Module software module, the<br>subject's performance, and accuracy differ<br>slightly from the predicate device.<br>In support of the substantial equivalence<br>determination, the validation met the<br>predetermined specifications and<br>acceptance criteria.<br>The performance data demonstrate that the<br>device is as safe and effective as the<br>predicate devices |
| Technological<br>characteristics | Primary Predicate device<br>(AlignRT Plus, K212583) | Subject device<br>(AlignRT Plus 7.3) | Comments |
| Biocompatibility | The AlignRT Plus product requires no<br>direct contact with the patient.<br><br>The only interactions between the user<br>and the system are with:<br>• the PC (in the control room) or remote<br>workstation (in the vault),<br>• the Remote Control (in the vault),<br>• the Real Time Coach (RTC) (in the<br>vault), or<br>• the Head Adjuster (in the vault).<br>• Calibration plate (in the vault)<br>• Calibration cube (in the vault)<br>• Calibration levelling plate (in the vault)<br><br>The materials of the devices (which are<br>commonly used in light-industrial, | The AlignRT Plus product requires no<br>direct contact with the patient.<br><br>The only interactions between the user<br>and the system are with:<br>• the PC (in the control room) or remote<br>workstation (in the vault),<br>• the Remote Control (in the vault),<br>• the Real Time Coach (RTC) (in the<br>vault), or<br>• the Head Adjuster (in the vault).<br>• Calibration plate (in the vault)<br>• Calibration cube (in the vault)<br>• Calibration levelling plate (in the vault)<br>• DoseRT reference board (in the vault)<br><br>The materials of the devices (which are | Biocompatibility tests conducted on the<br>predicate device can be applied to AlignRT<br>Plus 7.3 since the surface contacting<br>components and materials are the same<br>except for the reference board.<br><br>No additional testing was required since the<br>reference board does not contact patients.<br>These differences do not raise new or<br>different concerns of safety and<br>effectiveness. |
| | commercial and home use) and that the<br>application only involves intermittent<br>external contact with intact skin. | commonly used in light-industrial,<br>commercial, and home use) and that the<br>application only involves intermittent<br>external contact with intact skin. | |
| Mechanical<br>Safety | Cameras are ceiling mounted and do not<br>contact patient or user.<br><br>The InBore camera solution is fixed to the<br>inside of the bore-based linac. | Cameras are ceiling mounted and do not<br>contact patient or user.<br><br>The InBore camera solution is fixed to the<br>inside of the bore-based linac. | Equivalent to the predicate device. |
| | Head adjuster is clamped to the treatment<br>couch through universal base plate. | Head adjuster is clamped to the treatment<br>couch through universal base plate. | |
| Anatomical<br>treatment sites | Entire body surface. | Entire body surface. | Equivalent to the predicate device. |
| Human factors | Imaging process is fully automatic as is<br>estimation of new couch position: 3D | Imaging process is fully automatic as is<br>estimation of new couch position: 3D | The subject device continues to meet FDA<br>guidance on "Applying Human Factors and |
| Technological<br>characteristics | Primary Predicate device<br>(AlignRT Plus, K212583) | Subject device<br>(AlignRT Plus 7.3) | Comments |
| | visual display provided to show any<br>discrepancy in patient position. | visual display provided to show any<br>discrepancy in patient position. | Usability Engineering to Medical Devices"<br>and IEC 62366-1: 2015 Medical Devices-<br>Part 1: Application of Usability Engineering<br>to Medical Devices. |
| | For respiratory tracking, user selects<br>region of interest or tracking point(s)<br>during first session. These are detected<br>automatically during subsequent sessions. | For respiratory tracking, user selects<br>region of interest or tracking point(s)<br>during first session. These are detected<br>automatically during subsequent sessions. | This is equivalent to the predicate device.…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.