K212583 · Vision Rt, Ltd. · IYE · Nov 10, 2021 · Radiology
Device Facts
Record ID
K212583
Device Name
AlignRT Plus
Applicant
Vision Rt, Ltd.
Product Code
IYE · Radiology
Decision Date
Nov 10, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.5050
Device Class
Class 2
Attributes
Pediatric
Intended Use
The AlignRT Plus system is indicated for use to position and monitor patients relative to the prescribed treatment isocentre, and to withhold the beam automatically during radiation delivery. For cranial treatments, a manual head adjuster is included which can be used in concert with AlignRT Plus to provide fine corrections for pitch, roll and yaw rotations. AlignRT Plus is also used to track the patient's respiratory pattern for respiratory synchronized image acquisition, and radiation therapy treatment. Patient contour data can be extracted and exported from the data acquired for the purpose of treatment planning. AlignRT Plus can be calibrated directly to the treatment beam isocentre and in turn assists in performing quality assurance on MV, kV imagers, room lasers and the treatment couch. AlignRT Plus is indicated for use during simulation, setup and stereotactic radiosurgery and precision radiotherapy for lesions, tumors and conditions anywhere in the body where radiation is indicated.
Device Story
AlignRT Plus is a video-based 3D surface imaging system for radiotherapy. It uses stereo camera pairs to project near-infrared patterns onto the patient, reconstructing 3D skin surface models without ionizing radiation or markers. The system tracks patient position and respiratory patterns in real-time, comparing them against a reference surface to detect movement. If movement exceeds user-defined tolerances, the system automatically withholds the radiation beam. It integrates with CT scanners for respiratory-synchronized imaging and assists in quality assurance for MV/kV imagers and treatment couches. Used in clinical settings by radiotherapy professionals, the system provides visual feedback for patient positioning and enables fine manual adjustments for cranial treatments via an optional head adjuster. It aids in precision radiotherapy and SRS by ensuring accurate patient alignment relative to the treatment isocenter, potentially improving treatment accuracy and patient outcomes.
Clinical Evidence
Bench testing only. Performance verified through target registration error measurements using a calibration phantom and RMS error analysis for surface displacement tracking over breathing cycles.
Technological Characteristics
Video-based 3D surface imaging system using near-infrared pattern projection. Components include PC workstation, cables, and 3D camera pods (HD and Horizon models). Materials include block polystyrene (calibration phantom) and carbon fiber laminate (head adjuster). Connectivity via 10BaseT internet behind local firewall. Compliant with IEC 60601-1 (electrical safety) and IEC 60601-1-2 (EMC).
Indications for Use
Indicated for positioning and monitoring patients relative to treatment isocenter, respiratory tracking for synchronized imaging/treatment, and patient contour extraction for treatment planning. Used during simulation, setup, stereotactic radiosurgery, and precision radiotherapy for lesions, tumors, and conditions anywhere in the body. Applicable to any individual (adult or child) undergoing radiotherapy.
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
K181218 — AlignRT Plus · Vision Rt, Ltd. · Aug 7, 2018
K203387 — AlignRT Plus · Vision Rt, Ltd. · Mar 4, 2021
K193431 — AlignRT Plus · Vision Rt, Ltd. · Jan 3, 2020
K123371 — ALIGNRT PLUS · Vision RT Limited · Jan 8, 2013
K253012 — AlignRT Plus (8.0) · Vision Rt, Ltd. · Mar 18, 2026
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Vision RT Ltd. % Jade Dunphy Regulatory Affairs Manager Dove House Arcadia Avenue London, N3 2JU United Kingdom
Re: K212583
Trade/Device Name: AlignRT Plus Regulation Number: 21 CFR 892.5050 Regulation Name: Medical Charged-Particle Radiation Therapy System Regulatory Class: Class II Product Code: IYE Dated: August 13, 2021 Received: August 16, 2021
Dear Jade Dunphy:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Thalia T. Mills, Ph.D Director. Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K212583
Device Name AlignRT Plus
Indications for Use (Describe)
The AlignRT Plus system is indicated for use to position and monitor patients relative to the prescribed treatment isocentre, and to withhold the beam automatically during radiation delivery.
For cranial treatments, a manual head adjuster is included which can be used in concert with AlignRT Plus to provide fine corrections for pitch, roll and yaw rotations.
AlignRT Plus is also used to track the patient's respiratory pattern for respiratory synchronized image acquisition, and radiation therapy treatment.
Patient contour data can be extracted and exported from the data acquired for the purpose of treatment planning.
AlignRT Plus can be calibrated directly to the treatment beam isocentre and in turn assists in performing quality assurance on MV, kV imagers, room lasers and the treatment couch.
AlignRT Plus is indicated for use during simulation, setup and stereotactic radiosurgery and precision radiotherapy for lesions, tumors and conditions anywhere in the body where radiation is indicated.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## visionrt
## 510(k) Premarket Notification – AlignRT Plus
### 510(k) Summary
The information below is provided for the modifications to AlignRT following the format of 21 CFR 807.92.
#### Submitter:
Vision RT Ltd. Dove House Arcadia Avenue London, N3 2JU United Kingdom Contact Name: Jade Dunphy Tel: +44 (0)20 8346 4300 Fax: +44 (0)20 8346 4634
Date Summary was prepared: 17 December 2020
### PREDICATE DEVICE INFORMATION:
| Name of the Device: | AlignRT Plus |
|-------------------------|----------------------------------------------------------------------------------------------------------------------|
| Premarket Notification: | K203387 |
| Trade/Proprietary Name: | AlignRT Plus (incorporating AlignRT, GateCT, GateRT), may be<br>branded as OSMS (Optical Surface Monitoring System). |
| Manufacturer: | Vision RT Limited |
| Common or Usual Name: | Surface Guided Radio Therapy System (Patient Positioning System<br>- accessory to a linear accelerator) |
| Device Class: | II |
| Classification Name: | Accessory to Medical charged-particle radiation therapy accessories,<br>IYE, (per 21 CFR section 892.5050) |
| Review Panel: | Radiology |
### SUBJECT (MODIFIED) DEVICE:
| Name of the Device: | AlignRT Plus (incorporating AlignRT, GateCT, SimRT, GateRT), may<br>be branded as OSMS (Optical Surface Monitoring System) |
|-----------------------|----------------------------------------------------------------------------------------------------------------------------|
| Device Manufacturer: | Vision RT Limited |
| Common or Usual Name: | Surface Guided Radio Therapy System (Patient Positioning System<br>accessory to a linear accelerator) |
| Device Class: | II |
| Classification Name: | Accessory to Medical charged-particle radiation therapy accessories,<br>IYE, (per 21 CFR section 892.5050) |
| Review Panel: | Radiology |
{4}------------------------------------------------
## 510(k) Premarket Notification - AlignRT Plus
### Description of Device:
The AlignRT Plus system (K203387) is a combination of the devices AlignRT, GateCT and GateRT.
AlignRT Plus is a video-based three-dimensional (3D) surface imaging system, which is used to image the skin surface of a patient in 3D before and during radiotherapy treatment. The system consists of advanced software, a computer workstation, and one, two or three 3D camera units (each camera unit comprising a stereo pair of sensors to allow 3D surface reconstruction). The system is non-invasive, does not require the use of body markers and produces no ionizing irradiation during the imaging process.
AlignRT Plus is also able to perform both respiratory synchronised CT imaging and treatment delivery. In both instances, the system acquires a gated 3D surface model of the patient. User selected points are then tracked in real time in order to provide gating and position monitoring signals.
Real-time imaging and surface matchinq of the patient is possible during both setup and the treatment delivery to determine any patient movement. During treatment delivery, AlignRT Plus is also able to withhold the beam automatically, should the patient move outside user-defined tolerances.
Patient contour data may be extracted from surface data acquired by the system and exported for the purpose of treatment planning by radiotherapy professionals.
AlignRT Plus may be calibrated directly to the treatment beam isocentre using an optional custom designed calibration phantom and image processing software. It can analyse MV and kV digital imaging data acquired by other cleared devices. This in turn assists the user in performing quality assurance on MV, kV imagers, room lasers and the treatment couch.
The AlignRT Plus system includes the optional Head Adjuster for cranial treatments to allow for the manual, fine correction of pitch, roll and yaw in the patient's head position.
Precise isocenter calibration and the optional Head Adjuster provide improved frameless Stereotactic Radiosurgery (SRS). This is provided with the brand name "AlignRT SRS module".
The AlignRT Plus system is also provided under the brand OSMS (Optical Surface Monitoring System). This product is identical to AlignRT.
This 510(k) is requested for a modifications to FDA 510(k) cleared product "AlignRT Plus" (K203387).
This 510(k) is to obtain clearance for the following changes to the cleared device:
- 一 Introduction of new software, SimRT 7.2. This is an updated version of the cleared software, GateCT, with a new user interface similar to the cleared AlignRT 6.2/6.3/6.4 software;
- 2. Introduction of new camera pods (hardware). These function in the same way as the currently cleared hardware but are aesthetically different and have upgraded capabilities to support future product development;
- 3. Improve the product's accuracy claim for Source-Surface Distancing (SSD) from <2.0mm to <0.2mm.
{5}------------------------------------------------
## visionrt
## 510(k) Premarket Notification – AlignRT Plus
Addition of a new medical device software application for patient and accessory identification, SafeID.
#### Indications for Use:
The AlignRT Plus system is indicated for use to position and monitor patients relative to the prescribed treatment isocentre, and to withhold the beam automatically during radiation delivery.
For cranial treatments, a manual head adjuster is included which can be used in concert with AlignRT Plus to provide fine corrections for pitch, roll and yaw rotations.
AlignRT Plus is also used to track the patient's respiratory pattern for respiratory synchronized image acquisition, and radiation therapy treatment.
Patient contour data can be extracted and exported from the data acquired for the purpose of treatment planning.
AlignRT Plus can be calibrated directly to the treatment beam isocentre and in turn assists in performing quality assurance on MV, kV imagers, room lasers and the treatment couch.
AlignRT Plus is indicated for use during simulation, setup and stereotactic radiosurgery and precision radiotherapy for lesions, tumors and conditions anywhere in the body where radiation is indicated.
#### No changes are made to the intended use and indications for use.
#### Summary of the Technological Characteristics:
The Substantial Equivalence Comparison Table shown below provides a comparison of the technological characteristics of AlignRT Plus to those of the predicate device:
{6}------------------------------------------------
| | PREDICATE DEVICE | MODIFIED DEVICE | COMMENTS |
|------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------|
| Device name | AlignRT Plus (K203387)<br>Incorporating AlignRT, GateCT, GateRT | AlignRT Plus<br>Incorporating AlignRT, GateCT, SimRT,<br>GateRT | Addition of SimRT v7.2 software. |
| Manufacturer | Vision RT Ltd | Vision RT Ltd | The manufacturer has not<br>changed. |
| Indications for<br>Use | The AlignRT Plus system is indicated for use<br>to position and monitor patients relative to the<br>prescribed treatment isocenter, and to<br>withhold the beam automatically during<br>radiation delivery.<br><br>For cranial treatments, a manual head<br>adjuster is included which can be used in<br>concert with AlignRT Plus to provide fine<br>corrections for pitch, roll and yaw rotations.<br><br>AlignRT Plus is also used to track the<br>patient's respiratory pattern for respiratory<br>synchronized image acquisition, and<br>radiation therapy treatment. Patient contour<br>data can be extracted and exported from the<br>data acquired for the purpose of treatment<br>planning.<br><br>AlignRT Plus can be calibrated directly to the<br>treatment beam isocenter and in turn assists<br>in performing quality assurance on MV, kV<br>imagers, room lasers and the treatment<br>couch. | The AlignRT Plus system is indicated for use<br>to position and monitor patients relative to the<br>prescribed treatment isocenter, and to<br>withhold the beam automatically during<br>radiation delivery.<br><br>For cranial treatments, a manual head<br>adjuster is included which can be used in<br>concert with AlignRT Plus to provide fine<br>corrections for pitch, roll and yaw rotations.<br><br>AlignRT Plus is also used to track the<br>patient's respiratory pattern for respiratory<br>synchronized image acquisition, and<br>radiation therapy treatment. Patient contour<br>data can be extracted and exported from the<br>data acquired for the purpose of treatment<br>planning.<br><br>AlignRT Plus can be calibrated directly to the<br>treatment beam isocenter and in turn assists<br>in performing quality assurance on MV, kV<br>imagers, room lasers and the treatment<br>couch. | The indications for use are the<br>same as those of the previously<br>cleared device. |
{7}------------------------------------------------
| | PREDICATE DEVICE | MODIFIED DEVICE | COMMENTS |
|---------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | AlignRT Plus may be used during simulation,<br>setup and stereotactic radiosurgery and<br>precision radiotherapy for lesions, tumors<br>and conditions anywhere in the body where<br>radiation is indicated. | AlignRT Plus may be used during simulation,<br>setup and stereotactic radiosurgery and<br>precision radiotherapy for lesions, tumors<br>and conditions anywhere in the body where<br>radiation is indicated. | |
| Principles of<br>operation | Video based imaging of 3D skin surface data<br>using surface matching software. | Video based imaging of 3D skin surface data<br>using surface matching software. | The principles of operation are<br>the same as those of the<br>previously cleared device. |
| Target<br>Population | Any individual (adult or child) undergoing<br>radiotherapy. | Any individual (adult or child) undergoing<br>radiotherapy. | The target population is the same<br>as that of the previously cleared<br>device. |
| Materials | PC workstation, cables, video cameras.<br>Block Polystyrene (calibration phantom),<br>carbon fibre laminate material (head<br>adjuster). | PC workstation, cables, video cameras.<br>Block Polystyrene (calibration phantom),<br>carbon fibre laminate material (head<br>adjuster). | Although the Horizon camera is<br>manufactured to a different<br>design specification, the materials<br>used are the same as those of<br>the previously cleared device. |
| | PREDICATE DEVICE | MODIFIED DEVICE | COMMENTS |
| System<br>Performance<br>and Accuracy | Positioning accuracy: Target registration<br>errors (as measured using calibration<br>phantom) < 1mm (0.5mm) for all couch<br>angles. | Positioning accuracy: Target registration<br>errors (as measured using calibration<br>phantom) < 1mm (0.5mm) for all couch<br>angles. | The system performance and<br>accuracy have not changed since<br>previous clearances. |
| | Respiratory tracking: Tracks respiratory<br>signal from imaged surface data and sends<br>to CT (4D CT) or to Linac or imaging device<br>(gating). | Respiratory tracking: Tracks respiratory<br>signal from imaged surface data and sends<br>to CT (4D CT) or to Linac or imaging device<br>(gating). | |
| | Surface displacements can be tracked with<br>RMS errors < 0.5mm over 10 or more<br>breathing cycles. | Surface displacements can be tracked with<br>RMS errors < 0.5mm over 10 or more<br>breathing cycles. | |
| Biocompatibility | The AlignRT Plus product requires no direct<br>contact with the patient. | The AlignRT Plus product requires no direct<br>contact with the patient. | Although the Horizon camera is<br>manufactured to a different<br>design specification, the materials |
| | The only interactions between the user and<br>the system are with: | The only interactions between the user and<br>the system are with: | used are the same as those of<br>the previously cleared device. |
| | the PC (in the control room) or<br>remote workstation (in the vault), the Remote Control (in the vault), the Real Time Coach (RTC) (in the<br>vault), or the Head Adjuster (in the vault). Calibration plate (in the vault) Calibration cube (in the vault) Calibration levelling plate (in the<br>vault) | the PC (in the control room) or<br>remote workstation (in the vault), the Remote Control (in the vault), the Real Time Coach (RTC) (in the<br>vault), or the Head Adjuster (in the vault). Calibration plate (in the vault) Calibration cube (in the vault) Calibration levelling plate (in the<br>vault) | |
| | PREDICATE DEVICE | MODIFIED DEVICE | COMMENTS |
| | The materials of the devices (which are<br>commonly used in light-industrial, commercial<br>and home use) and that the application only<br>involves intermittent external contact with<br>intact skin. | The materials of the devices (which are<br>commonly used in light-industrial, commercial<br>and home use) and that the application only<br>involves intermittent external contact with<br>intact skin. | |
| Mechanical<br>Safety | Cameras are ceiling mounted and do not<br>contact patient or user. | Cameras are ceiling mounted and do not<br>contact patient or user. | The mechanical safety of the<br>device is the same as that of the |
| | The InBore camera solution is fixed to the<br>inside of the bore-based linac. | The InBore camera solution is fixed to the<br>inside of the bore-based linac. | previously cleared device. |
| | Head adjuster is clamped to the treatment<br>couch through universal base plate. | Head adjuster is clamped to the treatment<br>couch through universal base plate. | |
| Anatomical<br>treatment sites | Entire body surface. | Entire body surface. | The anatomical treatment sites of<br>the device are the same as those<br>of the previously cleared device. |
| Human factors | Imaging process is fully automatic as is<br>estimation of new couch position; 3D visual<br>display provided to show any discrepancy in<br>patient position.<br>For respiratory tracking, user selects region<br>of interest or tracking point(s) during first<br>session. These are detected automatically<br>during subsequent sessions.<br>For cranial treatments, a manual head | Imaging process is fully automatic as is<br>estimation of new couch position; 3D visual<br>display provided to show any discrepancy in<br>patient position.<br>For respiratory tracking, user selects region<br>of interest or tracking point(s) during first<br>session. These are detected automatically<br>during subsequent sessions.<br>For cranial treatments, a manual head | The human factors are the same<br>as that of the previously cleared<br>device. |
| | adjuster may be used by turning designated<br>dials to provide fine corrections for pitch, yaw<br>and roll rotations in concert with real time<br>visual feedback provided to the user by<br>AlignRT Plus. | adjuster may be used by turning designated<br>dials to provide fine corrections for pitch, yaw<br>and roll rotations in concert with real time<br>visual feedback provided to the user by<br>AlignRT Plus. | |
| | PREDICATE DEVICE | MODIFIED DEVICE | COMMENTS |
| | There is an extra non-clinical workflow for the<br>InBore camera. A customer must remove the<br>camera in order to service the bore-based<br>linear accelerator. | There is an extra non-clinical workflow for the<br>InBore camera. A customer must remove the<br>camera in order to service the bore-based<br>linear accelerator. | |
| Optical pattern | Optical (near infra-red) pattern is projected to<br>patient. | Optical (near infra-red) pattern is projected to<br>patient. | The optical pattern of the device<br>is the same as that of the<br>previously cleared device. |
| Compatibility<br>with the<br>environment and<br>other devices | For use in hospital and clinic environments. | For use in hospital and clinic environments. | The compatibility with the<br>environment is the same as that<br>of the previously cleared device. |
| General<br>Electrical safety<br>standards | IEC 60601-1 compliant. | IEC 60601-1 compliant. | The modified device has been<br>tested and continues to comply to<br>IEC 60601-1. |
| EMC standards | IEC 60601-1-2 compliant. | IEC 60601-1-2 compliant. | The modified device has been<br>tested and continues to comply to<br>IEC 60601-1-2: 2014 (4th Ed.). |
| Size | The camera (key part of the system) has the<br>following dimensions:<br>Cameras (each) – 470 x 220 x 70 – 4.5kg<br>InBore optional camera accessory - Ring<br>499mm diameter (plus 1mm protrusions) x<br>177mm wide (max) – 3.2kg | The camera (key part of the system) has the<br>following dimensions:<br>HD Cameras (each) – 470 x 220 x 70 – 4.5kg<br>Horizon cameras (each) – 480 x 140 x 127 –<br>2.75kg<br>InBore optional camera accessory - Ring<br>499mm diameter (plus 1mm protrusions) x<br>177mm wide (max) - 3.2kg | The new Horizon camera is<br>slightly different in size, but lighter<br>in weight to that of the previously<br>cleared cameras. |
{8}------------------------------------------------
{9}------------------------------------------------
{10}------------------------------------------------
{11}------------------------------------------------
| | PREDICATE DEVICE | MODIFIED DEVICE | COMMENTS |
|------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Packaging | The system is packaged in a variety of boxes<br>and then packaged within palletised crate. | The system is packaged in a variety of boxes<br>and then packaged within palletised crate. | The packaging of the device is<br>the same as that of the previously<br>cleared device. |
| Environmental<br>range | AlignRT Plus is intended for use at altitudes<br>below 2000m (6,561ft).<br>The operating temperature is +16°C to +30°C<br>(60.8° to 86° Fahrenheit).<br>The shipping and storage conditions are -<br>20°C to +50°C (-4° to 122° Fahrenheit). | AlignRT Plus is intended for use at altitudes<br>below 3000m (9,842ft).<br>The operating temperature is +16°C to +30°C<br>(60.8° to 86° Fahrenheit).<br>The shipping and storage conditions are -<br>20°C to +50°C (-4° to 122° Fahrenheit). | The device is a solid-state<br>product and is not affected by<br>altitude.<br>The modified device is<br>substantially equivalent to its<br>predicate device. |
| Workstation<br>Operating<br>System | Windows 10 | Windows 10 | No change to operating system. |
| Number of<br>cameras | 1-3<br>Optional InBore camera | 1-3<br>Optional InBore camera | The number of cameras used are<br>the same as those of the<br>previously cleared device. |
| Power<br>requirements | 110-230V 50/60Hz | 110-230V 50/60Hz | The power requirements of the<br>device are the same as those of the<br>previously cleared device. |
| Network<br>requirements | 10BaseT internet connection behind a local<br>firewall. | 10BaseT internet connection behind a local<br>firewall. | The network requirements of the<br>device are the same as those of the<br>previously cleared device. |
| Service<br>provision | All service to be performed by swap-out and<br>return-to-base by Vision RT engineers.<br>Remote support provided by Vision RT<br>engineers through secure remote internet<br>software. | All service to be performed by swap-out and<br>return-to-base by Vision RT engineers.<br>Remote support provided by Vision RT<br>engineers through secure remote internet<br>software. | The service approach of the<br>device is the same as that of the<br>previously cleared device. |
{12}------------------------------------------------
Image /page/12/Picture/0 description: The image shows the word "visionrt" in a stylized font. The first part of the word, "vision", is in black, with the "o" replaced by a target symbol. The letters "rt" are in blue. The font is sans-serif and modern.
### 510(k) Premarket Notification - AlignRT Plus
#### Conclusion
The proposed changes are made by the legal manufacturer and submitted by the legal manufacturer authorized to market he existing device (AlignRT Plus). The changes made in subject device were test method and acceptance criteria as the predicate device and the subject device is substantially equivalent to the prodication and/or validation demonstrate that the device is as safe, as effective, and performs as well as or better than the predicate device. The conces intended use, indications for use, fundamental technological characteristics, principle of operation and performance specifications.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.