AlignRT Plus

K193431 · Vision Rt, Ltd. · IYE · Jan 3, 2020 · Radiology

Device Facts

Record IDK193431
Device NameAlignRT Plus
ApplicantVision Rt, Ltd.
Product CodeIYE · Radiology
Decision DateJan 3, 2020
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 892.5050
Device ClassClass 2
AttributesPediatric

Indications for Use

The AlignRT Plus system is indicated for use to position and monitor patients relative to the prescribed treatment isocentre, and to withhold the beam automatically during radiation delivery. For cranial treatments, a manual head adjuster is included which can be used in concert with AlignRT Plus to provide fine corrections for pitch, roll and yaw rotations. AlignRT Plus is also used to track the patient's respiratory pattern for respiratory synchronized image acquisition, and radiation therapy treatment. Patient contour data can be extracted and exported from the data acquired for the purpose of treatment planning. AlignRT Plus can be calibrated directly to the treatment beam isocentre and in turn assists in performing quality assurance on MV, kV imagers, room lasers and the treatment couch. AlignRT Plus is indicated for use during simulation, setup and stereotactic radiosurgery and precision radiotherapy for lesions, tumors and conditions anywhere in the body where radiation is indicated.

Device Story

AlignRT Plus is a video-based 3D surface imaging system for patient positioning and monitoring during radiotherapy. Input: 3D skin surface data acquired via stereo camera pairs (near-infrared). Operation: System performs real-time surface matching to track patient position relative to treatment isocenter; automatically withholds radiation beam if patient movement exceeds user-defined tolerances; tracks respiratory patterns for gated imaging/treatment. Output: Real-time visual feedback of patient position; exported patient contour data for treatment planning; QA data for MV/kV imagers and lasers. Used in clinical radiotherapy environments by radiation therapy professionals. Benefits: Enables precise, non-invasive patient positioning and motion management without ionizing radiation or body markers. Includes optional manual head adjuster for cranial SRS fine-tuning.

Clinical Evidence

Bench testing only. No clinical data presented. Verification of software modifications (frame rate, postural setup) performed using established test methods and acceptance criteria consistent with the predicate device.

Technological Characteristics

3D surface imaging system; near-infrared optical projection; stereo camera sensors. Materials: PC workstation, cables, video cameras, block polystyrene (calibration phantom), carbon fibre laminate (head adjuster). Connectivity: 10BaseT Ethernet. Standards: IEC 60601-1 (electrical safety), IEC 60601-1-2 (EMC). Software: Windows 10-based application.

Indications for Use

Indicated for any adult or child patient undergoing radiation therapy, including simulation, setup, stereotactic radiosurgery, and precision radiotherapy for lesions, tumors, and conditions anywhere in the body.

Regulatory Classification

Identification

A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left, and the FDA logo on the right. The FDA logo includes the letters "FDA" in a blue box, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. January 3, 2020 Vision RT Ltd. % Arun Mahendran Regulatory Affairs Manager Dove House. Arcadia Avenue London, N3 2JU UNITED KINGDOM #### Re: K193431 Trade/Device Name: AlignRT Plus Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: Class II Product Code: IYE Dated: December 9, 2019 Received: December 10, 2019 Dear Arun Mahendran: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see {1}------------------------------------------------ https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ### Indications for Use 510(k) Number (if known) K193431 Device Name AlignRT Plus Indications for Use (Describe) The AlignRT Plus system is indicated for use to position and monitor patients relative to the prescribed treatment isocentre, and to withhold the beam automatically during radiation delivery. For cranial treatments, a manual head adjuster is included which can be used in concert with AlignRT Plus to provide fine corrections for pitch, roll and yaw rotations. AlignRT Plus is also used to track the patient's respiratory pattern for respiratory synchronized image acquisition, and radiation therapy treatment. Patient contour data can be extracted and exported from the data acquired for the purpose of treatment planning. AlignRT Plus can be calibrated directly to the treatment beam isocentre and in turn assists in performing quality assurance on MV, kV imagers, room lasers and the treatment couch. AlignRT Plus is indicated for use during simulation, setup and stereotactic radiosurgery and precision radiotherapy for lesions, tumors and conditions anywhere in the body where radiation is indicated. | Type of Use (Select one or both, as applicable) | |-------------------------------------------------------------------------------------------------------------------------------| | <div> <span> <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | | <div> <span>Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # visi�nrt ## Special 510(k) Premarket Notification – AlignRT Plus Incorporating AlignRT V6.2 Tab 6 - 510(k) Summary The information below is provided for the modifications to AlignRT following the format of 21 CFR 807.92. | Submitter: | Vision RT Ltd.<br>Dove House<br>Arcadia Avenue<br>London, N3 2JU<br>United Kingdom<br>Contact Name: Arun Mahendran<br>Tel: +44 (0)77 3763 7807<br>Fax: +44 (0)208 436 4634<br>Date Summary was prepared: 09 December 2019 | |------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| |------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| ### PREDICATE DEVICE INFORMATION: | Name of the Device: | AlignRT Plus | |-------------------------|----------------------------------------------------------------------------------------------------------------------| | Premarket Notification: | K181218 | | Trade/Proprietary Name: | AlignRT Plus (incorporating AlignRT, GateCT, GateRT), may be<br>branded as OSMS (Optical Surface Monitoring System). | | Manufacturer: | Vision RT Limited | | Common or Usual Name: | Surface Guided Radio Therapy System (Patient Positioning System -<br>accessory to a linear accelerator) | | Device Class: | II | | Classification Name: | Medical charged-particle radiation therapy system, IYE, (per 21 CFR<br>section 892.5050) | | Review Panel: | Radiology | #### SUBJECT (MODIFIED) DEVICE: | Name of the Device: | AlignRT Plus | |-----------------------|----------------------------------------------------------------------------------------------------| | Device Manufacturer: | Vision RT Limited | | Common or Usual Name: | Surface Guided Radio Therapy System (Patient Positioning System accessory to a linear accelerator) | {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the word "visionrt" with a unique design. The letters "v," "i," "s," "i," "o," and "n" are in black, while the letters "r" and "t" are in blue. The "i" in "vision" is dotted with a small black circle, and the "o" in "vision" is replaced with a target symbol, adding a distinctive visual element to the logo. ### Special 510(k) Premarket Notification – AlignRT Plus Incorporating AlignRT V6.2 ll Device Class: Classification Name: Medical charged-particle radiation therapy system, IYE, (per 21 CFR section 892.5050) Radiology Review Panel: #### Description of Device: AlignRT Plus is a video-based three-dimensional (3D) surface imaging system, which is used to image the skin surface of a patient in 3D before and during radiotherapy treatment. The system consists of advanced software, a computer workstation, and one, two or three 3D camera units (each camera unit comprising a stereo pair of sensors to allow 3D surface reconstruction). The system is noninvasive, does not require the use of body markers and produces no ionizing irradiation during the imaging process. AlignRT Plus is also able to perform both respiratory synchronised CT imaging and treatment delivery. In both instances, the system acquires a gated 3D surface model of the patient. User selected points are then tracked in real time in order to provide gating and position monitoring signals. Real-time imaging and surface matching of the patient is possible during both setup and the treatment delivery to determine any patient movement. During treatment delivery, AlignRT Plus is also able to withhold the beam automatically, should the patient move outside user-defined tolerances. Patient contour data may be extracted from surface data acquired by the system and exported for the purpose of treatment planning by radiotherapy professionals. AlignRT Plus may be calibrated directly to the treatment beam isocentre using an optional custom designed calibration phantom and image processing software. It can analyse MV and kV digital imaging data acquired by other cleared devices. This in turn assists the user in performing quality assurance on MV, kV imagers, room lasers and the treatment couch. The AlignRT Plus system includes the optional Head Adjuster for cranial treatments to allow for the manual, fine correction of pitch, roll and yaw in the patient's head position. Precise isocenter calibration and the optional Head Adjuster provide improved frameless Stereotactic Radiosurgery (SRS). This is provided with the brand name "AlignRT SRS module". {5}------------------------------------------------ ### Special 510(k) Premarket Notification – AlignRT Plus Incorporating AlignRT V6.2 The AlignRT Plus system is also provided under the brand OSMS (Optical Surface Monitoring System). This product is identical to AlignRT. This Special 510(k) is requested only for software modification for an existing 510(k) cleared product "AlignRT Plus" (K181218). No hardware or material changes have been made since its last 510(k) clearance. The two key changes made in the existing AlignRT v6.0 software which are the subject of this Special 510(k) are listed below; - 1. System Frame Rate improvement - 2. Addition of Video Postural Setup Alignment feature #### Indications for Use: The AlignRT Plus system is indicated for use to position and monitor patients relative to the prescribed treatment isocentre, and to withhold the beam automatically during radiation delivery. For cranial treatments, a manual head adjuster is included which can be used in concert with AlignRT Plus to provide fine corrections for pitch, roll and yaw rotations. AlignRT Plus is also used to track the patient's respiratory pattern for respiratory synchronized image acquisition, and radiation therapy treatment. Patient contour data can be extracted and exported from the data acquired for the purpose of treatment planning. AlignRT Plus can be calibrated directly to the treatment beam isocentre and in turn assists in performing quality assurance on MV, kV imagers, room lasers and the treatment couch. AlignRT Plus is indicated for use during simulation, setup and stereotactic radiosurgery and precision radiotherapy for lesions, tumors and conditions anywhere in the body where radiation is indicated. No changes are made to the intended use and indications for use. #### Summary of the Technological Characteristics: The Substantial Equivalence Comparison Table shown below provides a comparison of the technological characteristics of AlignRT Plus to those of the predicate device: {6}------------------------------------------------ | | PREDICATE DEVICE | MODIFIED DEVICE | COMMENTS | |--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | DEVICE NAME | AlignRT Plus (K181218)<br>Incorporating AlignRT, GateCT, GateRT | AlignRT Plus<br>Incorporating AlignRT, GateCT, GateRT | The device name is the same.<br>The AlignRT application has been<br>up-issued from V6.0 to V6.2. | | MANUFACTURER | Vision RT Ltd | Vision RT Ltd | The manufacturer has not<br>changed. | | Indications for<br>Use | The AlignRT Plus system is indicated for use<br>to position and monitor patients relative to<br>the prescribed treatment isocenter, and to<br>withhold the beam automatically during<br>radiation delivery. For cranial treatments, a<br>manual head adjuster is included which can<br>be used in concert with AlignRT Plus to<br>provide fine corrections for pitch, roll and<br>yaw rotations. AlignRT Plus is also used to<br>track the patient's respiratory pattern for<br>respiratory synchronized image acquisition,<br>and radiation therapy treatment. Patient<br>contour data can be extracted and exported<br>from the data acquired for the purpose of<br>treatment planning. AlignRT Plus can be<br>calibrated directly to the treatment beam<br>isocenter and in turn assists in performing<br>quality assurance on MV, kV imagers, room<br>lasers and the treatment couch. AlignRT Plus<br>may be used during simulation, setup and | The AlignRT Plus system is indicated for use<br>to position and monitor patients relative to<br>the prescribed treatment isocenter, and to<br>withhold the beam automatically during<br>radiation delivery. For cranial treatments, a<br>manual head adjuster is included which can<br>be used in concert with AlignRT Plus to<br>provide fine corrections for pitch, roll and<br>yaw rotations. AlignRT Plus is also used to<br>track the patient's respiratory pattern for<br>respiratory synchronized image acquisition,<br>and radiation therapy treatment. Patient<br>contour data can be extracted and exported<br>from the data acquired for the purpose of<br>treatment planning. AlignRT Plus can be<br>calibrated directly to the treatment beam<br>isocenter and in turn assists in performing<br>quality assurance on MV, kV imagers, room<br>lasers and the treatment couch. AlignRT Plus<br>may be used during simulation, setup and | The indications for use are<br>exactly the same and have not<br>changed since previous<br>clearances. | | | PREDICATE DEVICE | MODIFIED DEVICE | COMMENTS | | | stereotactic radiosurgery and precision<br>radiotherapy for lesions, tumors and<br>conditions anywhere in the body where<br>radiation is indicated. | stereotactic radiosurgery and precision<br>radiotherapy for lesions, tumors and<br>conditions anywhere in the body where<br>radiation is indicated. | | | Principles of<br>operation | Video based imaging of 3D skin surface data<br>using surface matching software. | Video based imaging of 3D skin surface data<br>using surface matching software. | The principles of operation are<br>exactly the same and have not<br>changed since previous<br>clearances. | | Target<br>Population | Any individual (adult or child) undergoing<br>radiotherapy. | Any individual (adult or child) undergoing<br>radiotherapy. | The target population is exactly<br>the same and have not changed<br>since previous clearances. | | Materials | PC workstation, cables, video cameras. Block<br>Polystyrene (calibration phantom), carbon<br>fibre laminate material (head adjuster). | PC workstation, cables, video cameras. Block<br>Polystyrene (calibration phantom), carbon<br>fibre laminate material (head adjuster). | The materials used to<br>manufacture the device are<br>exactly the same and have not<br>changed since previous<br>clearances. | | System<br>Performance and<br>Accuracy | Positioning accuracy: Target registration<br>errors (as measured using calibration<br>phantom) < 1mm (0.5mm) for all couch<br>angles.<br><br>Respiratory tracking: Tracks respiratory<br>signal from imaged surface data and sends<br>to CT (4D CT) or to Linac or imaging device<br>(gating). | Positioning accuracy: Target registration<br>errors (as measured using calibration<br>phantom) < 1mm (0.5mm) for all couch<br>angles.<br><br>Respiratory tracking: Tracks respiratory<br>signal from imaged surface data and sends<br>to CT (4D CT) or to Linac or imaging device<br>(gating). | These system performances have<br>not changed since previous<br>clearances. | | | PREDICATE DEVICE | MODIFIED DEVICE | COMMENTS | | | Surface displacements can be tracked with<br>RMS errors $ < 0.5mm $ over 10 or more<br>breathing cycles. | Surface displacements can be tracked with<br>RMS errors $ < 0.5mm $ over 10 or more<br>breathing cycles. | | | Biocompatibility | The AlignRT Plus product requires no direct<br>contact with the patient.<br>The only interactions between the user and<br>the system are with:<br>the PC (in the control room) or<br>remote workstation (in the vault), the Remote Control (in the vault), the Real Time Coach (RTC) (in the<br>vault), or the Head Adjuster (in the vault). Calibration plate (in the vault) Calibration cube (in the vault) Calibration levelling plate (in the<br>vault) | The AlignRT Plus product requires no direct<br>contact with the patient.<br>The only interactions between the user and<br>the system are with:<br>the PC (in the control room) or<br>remote workstation (in the vault), the Remote Control (in the vault), the Real Time Coach (RTC) (in the<br>vault), or the Head Adjuster (in the vault). Calibration plate (in the vault) Calibration cube (in the vault) Calibration levelling plate (in the<br>vault) | The materials used to<br>manufacture the device are<br>exactly the same and have not<br>changed since the previous<br>clearances. | | PREDICATE DEVICE | MODIFIED DEVICE | COMMENTS | | | The materials of the devices (which are<br>commonly used in light-industrial,<br>commercial and home use) and that the<br>application only involves intermittent<br>external contact with intact skin. | The materials of the devices (which are<br>commonly used in light-industrial,<br>commercial and home use) and that the<br>application only involves intermittent<br>external contact with intact skin. | | | | Mechanical<br>Safety | Cameras are ceiling mounted and do not<br>contact patient or user.<br>Head adjuster is clamped to the treatment<br>couch through universal base plate. | Cameras are ceiling mounted and do not<br>contact patient or user.<br>Head adjuster is clamped to the treatment<br>couch through universal base plate. | The mechanical safety of the<br>device is exactly the same and<br>has not changed since the<br>previous clearances. | | Anatomical<br>treatment sites | Entire body surface. | Entire body surface. | The anatomical treatment sites<br>of the device are exactly the<br>same and have not changed since<br>the previous clearances. | | Human factors | Imaging process is fully automatic as is<br>estimation of new couch position; 3D visual<br>display provided to show any discrepancy in<br>patient position.<br>For respiratory tracking, user selects region<br>of interest or tracking point(s) during first<br>session. These are detected automatically<br>during subsequent sessions.<br>For cranial treatments, a manual head<br>adjuster may be used by turning designated<br>dials to provide fine corrections for pitch,<br>yaw and roll rotations in concert with real | Imaging process is fully automatic as is<br>estimation of new couch position; 3D visual<br>display provided to show any discrepancy in<br>patient position.<br>For respiratory tracking, user selects region<br>of interest or tracking point(s) during first<br>session. These are detected automatically<br>during subsequent sessions.<br>For cranial treatments, a manual head<br>adjuster may be used by turning designated<br>dials to provide fine corrections for pitch,<br>yaw and roll rotations in concert with real | Minor changes made to the UI<br>without functional change. These<br>changes do not raise different<br>questions of safety and<br>effectiveness because the results<br>of usability study are same as the<br>previous study cleared under<br>K181218. The modified device is<br>substantially equivalent to its<br>predicate device.<br>The modified device continues to<br>meet FDA guidance on "Applying | | | PREDICATE DEVICE | MODIFIED DEVICE | COMMENTS | | | time visual feedback provided to the user by<br>AlignRT Plus. | time visual feedback provided to the user by<br>AlignRT Plus. | Human Factors and Usability<br>Engineering to Medical Devices"<br>and IEC 62366-1: 2015 Medical<br>Devices-Part 1: Application of<br>Usability Engineering to Medical<br>Devices. | | Optical pattern | Optical (near infra-red) pattern is projected<br>to patient. | Optical (near infra-red) pattern is projected<br>to patient. | The optical pattern of the device<br>is exactly the same and has not<br>changed since the previous<br>clearances. | | Compatibility<br>with the<br>environment and<br>other devices | For use in hospital and clinic environments. | For use in hospital and clinic environments. | The compatibility with the<br>environment is exactly the same<br>as those of the previous<br>clearances. | | General Electrical<br>safety standards | IEC60601-1 compliant. | IEC60601-1 compliant. | The hardware remains same.<br>No changes were made to the<br>hardware in AlignRT v6.2 | | EMC standards | IEC60601-1-2 compliant. | IEC60601-1-2 compliant. | The hardware remains same.<br>No changes were made to the<br>hardware in AlignRT v6.2<br>Device complies with the current<br>FDA recognised standard IEC<br>60601-1-2: 2014 (4th Ed.).<br>Supporting information was<br>provided during the last 510(k) | | | PREDICATE DEVICE | MODIFIED DEVICE | COMMENTS | | Size | The camera (key part of the system) has the<br>following dimensions:<br>Cameras (each) – 470 x 220 x 70 – 4.5kg | The camera (key part of the system) has the<br>following dimensions:<br>Cameras (each) – 470 x 220 x 70 – 4.5kg…
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...