SENTINEL, MODEL SP-001

K082582 · C-Rad Positioning AB · IYE · Mar 30, 2009 · Radiology

Device Facts

Record IDK082582
Device NameSENTINEL, MODEL SP-001
ApplicantC-Rad Positioning AB
Product CodeIYE · Radiology
Decision DateMar 30, 2009
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.5050
Device ClassClass 2

Indications for Use

The Sentinel system is intended for use in radiation therapy clinics to accurately position patients in a reproducible way, prior to treatment and to monitor the patient continuously during treatment. The system provides information about a patient's position and the adjustments required in order to position the patient as close as possible to a reference setup. During monitoring, the system reports deviations in the patient's position during treatment. The system shall only be used by hospital personnel, qualified to work in radiation therapy or diagnostics departments.

Device Story

Sentinel is a radiotherapy positioning and monitoring system; uses ceiling-mounted laser scanner and camera to acquire patient surface data via laser line triangulation; reconstructs 3D surface models. Operates via c4D software; integrates with clinic CT, LINAC, and R&V systems. cPosition module compares patient surface to DICOM reference data; calculates 6-degree-of-freedom corrections for couch alignment. cMotion module monitors real-time patient movement; triggers alerts if deviations exceed tolerances. Used in radiotherapy clinics by qualified personnel; facilitates accurate, reproducible patient setup and treatment delivery; reduces manual intervention through automated synchronization with treatment hardware.

Clinical Evidence

No clinical data provided. Substantial equivalence supported by performance testing.

Technological Characteristics

Ceiling-mounted laser scanner and camera unit; laser line triangulation for 3D surface reconstruction; PC-based c4D software platform; DICOM-compliant; interfaces with LINAC and R&V systems; motorized couch integration.

Indications for Use

Indicated for patients undergoing radiation therapy requiring accurate positioning and continuous motion monitoring during treatment. Intended for use by qualified hospital personnel in radiation therapy or diagnostic departments.

Regulatory Classification

Identification

A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ MAR 3 0 2009 # C-RAD K082582 # 510(k) SUMMARY ### Owner C-RAD Positioning AB Bredgränd 14 753 20 Uppsala Phone no: +46 18 666930 Fax no: # Contact person Cristina Svensson Chief Executive Officer Phone no: +46 18 666936 # Date of preparation September 1, 2008 # Trade name of device Sentinel ### Common name Radiotherapy positioning system ## Classification name Medical charged-particle radiation therapy system Regulation: 21 CFR 892.5050 # Predicate marketed devices AlignRT (K052682) - Vision RT Ltd. {1}------------------------------------------------ ### Device description The Sentinel system is a new version of the Positioner system (previously registered with 510(k) Number K063839), updated with an application module for patient monitoring, cMotion (the design control documents are included in Appendices A and B). Sentinel is an advanced system for surface contour localization and patient monitoring during the radiotherapy treatment process. The Sentinel platform is based on advanced laser technology with multipurpose software modules covering different tasks in the treatment procedure. The c4D multi-application software supports all modes of operation in one integrated package. The software is user friendly and requires a minimum of user interaction in the daily clinical workflow, while providing the advanced user with sophisticated data management, analysis and reporting functionalities. The software is designed to integrate with existing systems at the clinic, such as CT, linacs and R&V systems, and with motorized couch tops. Sentinel includes two application modules, cPosition for fast and accuracy patient positioning and cMotion for motion detection during the treatment delivery procedure. The Sentinel hardware consists of a single scanner unit containing the laser and camera, mounted in the ceiling in front of the gantry. The scanner is connected to the PC running the c4D software. During patient surface acquisition, a laser line is swept along the patient while the camera records a number of images. From the data acquired, a complete 3D surface of the patient can be reconstructed using laser line triangulation. For patient positioning, the acquired surface is captured in a few seconds and can contain several hundred contours. For motion detection the number of contours is typically lowered so that the desired frame rate is achieved. #### cPosition Once the treatment planning has been performed, the resulting plan can be transferred to the Sentinel system through import from the industry-standard DICOM format, creating the reference data necessary for patient positioning. Reference data can also be by created using the Sentinel laser scanner. In the treatment room, synchronization with the LINAC or R & V (Record and Verify) system ensures that the correct reference data is called up automatically when the patient is selected for treatment, and also eliminates the need for any manual selection of the patient in the Sentinel system. By advanced surface registration algorithms the actual patient position is compared to the predefined reference, suggesting within seconds a correction in six degrees of freedom of the patient's position. With interface to major accelerator vendors the suggested patient position is transferred to the respective couch control system and fast and accurate alignment is achieved. ### cMotion cMotion monitors the movement of the patient during treatment delivery and automatically warns if the patient moves outside the allowed tolerances. {2}------------------------------------------------ #### Intended use The Sentinel system is intended for use in radiation therapy clinics to accurately position patients in a reproducible way, prior to treatment and to monitor the patient continuously during treatment. The system provides information about a patient's position and the adjustments required in order to position the patient as close as possible to a reference setup. During monitoring, the system reports deviations in the patient's position during treatment. #### Technological comparison The Sentinel system is substantial equivalent to the predicate devices in terms of their intended use and technological characteristics. There are differences between the Sentinel system and predicate devices in terms of their principles of operation, materials, performance, human factors and energy delivered by the system. However, performance data has been submitted to show that Sentinel achieves its intended use and that these technological differences raise no new efficacy or safety concerns. {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES (USA)" around the perimeter. Inside the circle is a stylized image of an eagle with three lines representing its wings. #### DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service MAR \$ 0 2003 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Ms. Cristina Svensson CEO C-RAD Positioning AB Bredgränd 14 Uppsala 75320 SWEDEN Re: K082582 Trade/Device Name: Sentinal Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: II Product Code: IYE Dated: February 23, 2009 Received: February 26, 2009 #### Dear Ms. Svensson: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding os substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter. | 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | (240) 276-0115 | |----------------|----------------------------------|----------------| | 21 CFR 884.xxx | (Obstetrics/Gynecology) | (240) 276-0115 | | 21 CFR 892.xxx | (Radiology) | (240) 276-0120 | | Other | | (240) 276-0100 | Also, please note the regulation entitled, "Misbranding by reference to premarket notification'' (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrics' (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufactures. International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrb/industry.suppot/index.html. Sincerely yours, Lorina M. Morris anine M. Morris Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # INDICATIONS FOR USE 510(k) Number: K082582 Device Name: Sentinel # Indications for use: The Sentinel system is intended for use in radiation therapy clinics to accurately position patients in a reproducible way, prior to treatment and to monitor the patient continuously during treatment. The system provides information about a patient's position and the adjustments required in order to position the patient as close as possible to a reference setup. During monitoring, the system reports deviations in the patient's position during treatment. The system shall only be used by hospital personnel, qualified to work in radiation therapy or diagnostics departments. Helbert Reuner (Division Sign-Off) Division of Reproductive, Abdominal and Radiological Devices 510(k) Number Prescription Use
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...