K201777 · Lg Electronics.Inc · PGY · Oct 8, 2020 · Radiology
Device Facts
Record ID
K201777
Device Name
31HN713D
Applicant
Lg Electronics.Inc
Product Code
PGY · Radiology
Decision Date
Oct 8, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Intended Use
This Medical Monitor is indicated for use in displaying radiological images (including full-field digital mammography and digital breast tomosynthesis) for review, analysis, and diagnosis by trained medical practitioners.
Device Story
The 31HN713D is a 31-inch medical-grade TFT LCD monitor with LED backlight, designed for use in clinical environments by radiologists and trained medical practitioners. It receives digital radiological image data (including mammography and breast tomosynthesis) via DisplayPort or HDMI inputs from a PACS or radiology system. The device transforms these inputs into high-resolution (4,200 x 2,800 pixels) color or grayscale visual outputs. By providing high-fidelity, calibrated medical imaging, the monitor assists clinicians in the review, analysis, and diagnosis of patient images. It supports 10-bit color drivers to ensure accurate representation of diagnostic data. The device is intended to facilitate clinical decision-making by providing reliable, standardized visual information to the healthcare provider.
Clinical Evidence
No clinical studies were performed. Substantial equivalence is supported by bench testing, including electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), and performance testing per FDA guidance 'Display Devices for Diagnostic Radiology' (e.g., spatial resolution, luminance, contrast, and color tracking).
Technological Characteristics
31-inch TFT LCD module (LB310FTM) with LED backlight; 4,200 x 2,800 resolution; 10-bit color support; DisplayPort and HDMI connectivity. Complies with IEC 60601-1 and IEC 60601-1-2. Software is classified as moderate level of concern.
Indications for Use
Indicated for trained medical practitioners to display, review, analyze, and diagnose radiological images, including full-field digital mammography and digital breast tomosynthesis.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
K221061 — 21HQ613D · Lg Electronics.Inc · Jun 3, 2022
K223423 — 32HQ713D · Lg Electronics.Inc · Mar 1, 2023
K192925 — 32HL512D · Lg Electronics.Inc · Mar 13, 2020
K231026 — 12MP Color LCD Monitor C1216W, C12*** (* = 0 to 9, A to Z or blank, and the difference among models means the product is named according to different appearance colors and customer models) · Shenzhen Beacon Display Technology Co., Ltd. · Aug 18, 2023
K240130 — Medical Monitor (21HQ613D) · Lg Electronics.Inc · Feb 15, 2024
Submission Summary (Full Text)
{0}------------------------------------------------
October 8, 2020
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left side of the logo is the Department of Health & Human Services logo. To the right of that is a blue square with the letters FDA in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
LG Electronics, Inc. % Jinhwan Jun Chief Research Engineer 222, LG-ro, Cheongho-ri, Jinwi-myeon Pyeongtaek-Si. Gyeonggi-do 17709 REPUBLIC OF KOREA
Re: K201777
Trade/Device Name: 31HN713D Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: Class II Product Code: PGY Dated: September 7, 2020 Received: September 10, 2020
Dear Jinhwan Jun:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K201777
Device Name 31HN713D
Indications for Use (Describe)
This Medical Monitor is indicated for use in displaying radiological images (including full-field digital mammography and digital breast tomosynthesis) for review, analysis, and diagnosis by trained medical practitioners.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <span style="font-size:16px">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|------------------------------------------------------------------------------------|
| ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
[As Required by 21 CFR 807.92]
# 1. Date Prepared [21 CFR 807.92(a)(a)]
K201777
June 25, 2020
# 2. Submitter's Information [21 CFR 807.92(a)(1)]
| Name of Sponsor: | LG Electronics Inc. |
|-----------------------|-----------------------------------------------------------------------------------------------|
| Address: | 222, LG-ro, Cheongho-ri, Jinwi-myeon, Pyeongtaek-si,<br>Gyeonggi-do, 17709, Republic of Korea |
| Contact Name: | Jinhwan Jun / Chief Research Engineer |
| Telephone No.: | +82-31-8066-5641 |
| Email Address: | jinhwan.jun@lge.com |
| Name of Manufacturer: | LG Electronics Inc. |
| Address: | 77, Sanho-daero, Gumi-si, Gyeongsangbuk-do, 39381, |
### 3. Trade Name, Common Name, Classification [21 CFR 807.92(a)(2)]
| • Trade Name: | 31HN713D |
|-----------------|-----------------|
| • Company Name: | Medical Monitor |
- . Common Name:
| Medical Monitor | |
|-----------------|--|
| | |
Republic of Korea
- Classification: ●
| Classification Name | Picture archiving and communications system |
|-----------------------|---------------------------------------------|
| Classification Number | 21 CFR 892.2050 |
| Product Code | PGY |
| Device Class | II |
| Review Panel | Radiology |
{4}------------------------------------------------
## 4. Identification of Predicate Device(s) [21 CFR 807.92(a)(3)]
The identified predicate devices within this submission are shown as follow;
### Predicate Device
| • 510(k) Number: | K141428 |
|------------------------|---------------------------------------------|
| • Applicant: | BARCO N.V. |
| • Classification Name: | Picture archiving and communications system |
| • Trade Name: | MDMC-12133 |
### 5. Description of the Device [21 CFR 807.92(a)(4)]
The Medical monitor is intended to provide high resolution color and grayscale medical imaging for PACS and Radiology system. This Medical Monitor is intended to be used by trained medical practitioners for displaying, reviewing, and analysis of medical images
#### 6. Indications for use [21 CFR 807.92(a)(5)]
This Medical Monitor is indicated for use in displaying radiological images (including full-field digital mammography and digital breast tomosynthesis) for review, analysis, and diagnosis by trained medical practitioners.
{5}------------------------------------------------
#### 7. Technological Characteristics (Equivalence to Predicate Device) [21 CFR 807.92(a)(6)]
The table below presents comparisons between the subject device (31HN713D) and the legally marketed predicate device (K141428):
| | Proposed Device | Predicate Device |
|-----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| K Number | Not known | K141428 |
| Manufacturer | LG Electronics Inc. | BARCO N.V. |
| Model Name | 31HN713D | MDMC-12133 |
| Classification Name | Picture archiving and communications system | Picture archiving and communications system |
| Classification Number | 21 CFR 892.2050 | 21 CFR 892.2050 |
| Indications for Use | This Medical Monitor is indicated for use<br>in displaying radiological images<br>(including full-field digital mammography<br>and digital breast tomosynthesis) for<br>review, analysis, and diagnosis by trained<br>medical practitioners. | The MDMC- 12133 is intended to be used<br>in displaying arid viewing digital images,<br>including standard and multi-frame<br>digital mammography. for review,<br>analysis and diagnosis by trained medical<br>practitioners. It is especially designed for<br>breast tomosynthesis applications. |
| Display Technology | The 31HN713D has a LB310FTM<br>module for displaying. The LB310FTM is<br>a 31" TFT Liquid Crystal Display module<br>with LED Backlight unit. This module<br>supports 4,200 x 2,800 pixels and can<br>display color driven by 10bit drivers. | Coronis Uniti Barco is a color LCD<br>monitor for viewing radiology,<br>mammography, and breast<br>tomosynthesis images. The color panel<br>employs in-plane switching (IPS)<br>technology allowing wide viewing angles<br>and the matrix size (or resolution) is<br>4,200 x 2,800 pixels. |
| Power Consumption | MAX. 150W<br>Sleep Mode ≤ 0.5W<br>Off Mode ≤ 0.3W | 190W (nominal)<br><0.5 W (hibernate) |
| Screen size | 676.9 x 459.7 mm | 853.44mm / 33.6" |
| LCD Screen | TFT LCD | TFT LCD |
| Pixel Pitch | 0.1554 x 0.1554 mm | 0.1686 x 0.1686 mm |
| Resolution | 4,200 x 2,800 pixels | 4,200 x 2,800 pixels |
| Horizontal Frequency | 30 kHz to 135kHz | 25-85 Hz |
| Vertical Frequency | 56 Hz to 61 Hz | 25-85 Hz |
| Viewing angle | View angle free (R/L 178(Typ.), U/D<br>178(Typ.)) | 178° |
| Input video signals | DisplayPort x 2<br>HDMI x 1 | DisplayPort x 2 |
#### [Table 1. Comparison of Proposed Device to Predicate Device]
The comparison table shows that the subject device (31HN713D) has the similar indications for use the predicate device. Although the devices have some different technological characteristics (screen size, pixel pitch, horizontal/vertical frequency and input video signals), these differences do not make the subject device less safe and reliable, so the subject device fits for diagnostic use as the predicate device does. There are no significant differences in the technological characteristics of the subject device. All the differences between the subject and predicate device do not raise different questions of safety and effectiveness. It is substantially equivalent to a predicate device in indications for use and technology characteristics.
{6}------------------------------------------------
#### 8. Non-Clinical Test summary
- 1) Electrical Safety and Electromagnetic Compatibility
The test results demonstrated that the proposed device complies with the following standards:
- · Electrical Basic Safety and Essential Performance requirements in accordance with IEC 60601-1:2005/AMD1:2012
- Electromagnetic Compatibility Testing in accordance with IEC 60601-1-2 Edition 4.0:2014
- · Medical Electrical Equipment Part 1-6: General Requirements for Basic Safety and Essential Performance - Collateral Standard: Usability in accordance with IEC 60601-1-6:2010/A1:2013
- 2) Software Validation
The 31HN713D contains MODERATE level of concern software. The software was designed and developed according to a software development process and was verified and validated.
The software information is provided in accordance with FDA guidance: The content of premarket submissions for software contained in medical devices, on May 11, 2005.
- 3) Guidance
Display Devices for Diagnostic Radiology – Guidance for Industry and Food and Drug Administration Staff, issued on October 2, 2017
| | Measurements | Description | SE Note |
|------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| a. | Spatial resolution | Measurements of the transfer of<br>information from the image data to the<br>luminance fields at different spatial<br>frequencies of interest typically done by<br>reporting the modulation transfer<br>function. Non-isotropic resolution<br>properties should be characterized<br>properly by providing two-dimensional<br>measurements or measurements along<br>at least two representative axes.<br>(Using TG18 QC Test Pattern) | Equivalent |
| b. | Pixel defects | Measurements (count, types (e.g., sub-<br>pixel or entire pixel, always-on, always-<br>off), and locations (map) of pixel<br>defects. This is typically provided as a<br>tolerance limit. Pixel defects can<br>interfere with the visibility of small<br>details in medical images. | Equivalent |
| c. | Artifacts | Evaluate for image artifacts such as<br>ghosting and/or image sticking from<br>displaying a fixed test pattern for a<br>period of time.<br>(Using 5x5 mosaic pattern, 64Gray /<br>127 Gray judgment) | Same |
| d. | Temporal response | Measurements of the temporal behavior<br>of the display in responding to changes<br>in image values from frame to frame.<br>Since these transitions are typically not<br>symmetric, rise and fall time constants | Equivalent |
| Measurements | Description | SE Note | |
| | are needed to characterize the system.<br>Slow displays can alter details and<br>contrast of the image when large image<br>stacks are browsed or in video,<br>panning, and zooming modes. | | |
| e. Luminance | Measurements of the maximum and<br>minimum luminance that the device<br>outputs as used in the application under<br>recommended conditions and the<br>achievable values if the device is set to<br>expand the range to the limit. | Same | |
| f. Conformance to a<br>grayscale-to-luminance<br>function | Measurements of the mapping between<br>image values and the luminance output<br>following a target model response for<br>256 or more levels. | Equivalent | |
| g. Luminance at 30° and 45°<br>in diagonal, horizontal,<br>and vertical directions at<br>center and four corners | Measurements of the luminance<br>response at off-normal viewing related<br>to the target model for the luminance<br>response. | Equivalent | |
| h. Luminance uniformity or<br>Mura test | Measurements of the uniformity of the<br>luminance across the display screen. | Equivalent | |
| i. Stability of luminance and<br>chromaticity response<br>with temperature and time<br>of operation (on-time) | Measurements of the change in<br>luminance and chromaticity response<br>with temperature and use time. | Same | |
| j. Spatial noise | Measurements of the spatial noise level<br>as represented by the noise power<br>spectrum using an appropriate ratio of<br>camera and display pixels. Spatial noise<br>and resolution affect the way images<br>are presented to the viewer and can<br>alter features that are relevant to the<br>interpretation process of the physician<br>or radiologist. | Equivalent | |
| k. Reflection coefficient | Measurements of the reflection<br>coefficients of the display device.<br>Specular and diffuse reflection<br>coefficients can be used as surrogates<br>for the full bidirectional reflection<br>distribution function. | Same | |
| l. Veiling glare or small-spot<br>contrast | Measurements of the contrast obtained<br>for small targets. | Equivalent | |
| m. Color tracking | Chromaticity at different luminance<br>levels of primary colors as indicated by<br>the color coordinates in an appropriate<br>units system (e.g., CIE u'v') and the<br>color gamut enveloped by the primary<br>colors. | Equivalent | |
| n. Gray tracking | Chromaticity at different luminance<br>levels of gray shades, including the<br>white point, as indicated by the color<br>coordinates in an appropriate units<br>system | Equivalent | |
According to the quidance, we test following performance items:
{7}------------------------------------------------
{8}------------------------------------------------
According to the above test results, there are no significant performance differences between 31HN713D and the predicate device that would adversely affect the use of the product. It has substantially equivalent performance compared to the predicate device.
#### Clinical Test Summary:
No clinical studies were considered necessary and performed.
### 9. Conclusion [21 CFR 807.92(b)(3)]
In according with the Federal Food & Drug and cosmetic Act, 21 CFR Part 807, and based on the information provided in this premarket notification LG Electronics, concludes that the 31HN713D is substantially equivalent in safety and effectiveness to the predicate devices as described herein.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.