K221061 · Lg Electronics.Inc · PGY · Jun 3, 2022 · Radiology
Device Facts
Record ID
K221061
Device Name
21HQ613D
Applicant
Lg Electronics.Inc
Product Code
PGY · Radiology
Decision Date
Jun 3, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Indications for Use
This Medical Monitor is indicated for use in displaying radiological images (including full-field digital mammography and digital breast tomosynthesis) for review, analysis, and diagnosis by trained medical practitioners.
Device Story
The 21HQ613D is a medical-grade color TFT LCD monitor designed for displaying high-resolution radiological images in PACS and radiology environments. It receives video input signals (DVI, DP) from a workstation and transforms them into high-resolution (2048x2560) grayscale or color images using a 10-bit driver. The device is operated by trained medical practitioners (radiologists) to review and diagnose medical images, including digital mammography and breast tomosynthesis. The monitor provides high-resolution visual output, enabling clinicians to perform detailed image analysis, which supports diagnostic decision-making and patient care. It is intended for clinical use in professional healthcare settings.
Clinical Evidence
No clinical studies were performed. Substantial equivalence is supported by non-clinical bench testing, including electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), usability (IEC 60601-1-6), and display performance metrics (spatial resolution, luminance, contrast, temporal response, etc.) per FDA guidance for diagnostic radiology display devices.
Technological Characteristics
TFT LCD module (TM213XDGP31) with LED backlight; 2048x2560 resolution; 10-bit color driver; 0.165 x 0.165 mm pixel pitch; input interfaces: DVI x 1, DP x 2, DP out x 1; max power 120W. Complies with IEC 60601-1, IEC 60601-1-2, and IEC 60601-1-6.
Indications for Use
Indicated for trained medical practitioners to display, review, analyze, and diagnose radiological images, including full-field digital mammography and digital breast tomosynthesis.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" stacked on top of each other.
LG Electronics Inc. % Jonghyun Kim Chief Consultant GMS Consulting 4th Floor, Digital Cube, 34, Sangamsan-ro Seoul, Mapo-gu 03909 KOREA
Re: K221061
June 3, 2022
Trade/Device Name: 21HO613D Regulation Number: 21 CFR 892.2050 Regulation Name: Medical image management and processing system Regulatory Class: Class II Product Code: PGY Dated: April 7, 2022 Received: April 11, 2022
Dear Jonghyun Kim:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Michael D. O'Hara, Ph.D. Deputy Director DHT 8C: Division of Radiological Imaging and Radiation Therapy Devices OHT 8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K221061
Device Name 21HQ613D
Indications for Use (Describe)
This Medical Monitor is indicated for use in displaying radiological images (including full-field digital mammography and digital breast tomosynthesis) for review, and diagnosis by trained medical practitioners.
Type of Use (Select one or both, as applicable)
| <span></span> | <span></span> Prescription Use (Part 21 CFR 801 Subpart D) |
|---------------|------------------------------------------------------------|
| <span></span> | <span></span> Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
[As Required by 21 CFR 807.92]
#### 1. Date Prepared [21 CFR 807.92(a)(a)]
April 07, 2022
# 2. Submitter's Information [21 CFR 807.92(a)(1)]
| • Name of Sponsor: | LG Electronics Inc. |
|-------------------------|-----------------------------------------------------------------------------------------------|
| - Address: | 222, LG-ro, Cheongho-ri, Jinwi-myeon, Pyeongtaek-si,<br>Gyeonggi-do, 17709, Republic of Korea |
| • Contact Name: | Jinhwan Jun / Chief Research Engineer |
| - Telephone No.: | +82-31-8066-5641 |
| - Email Address: | jinhwan.jun@lge.com |
| • Name of Manufacturer: | LG Electronics Inc. |
| - Address: | 77, Sanho-daero, Gumi-si, Gyeongsangbuk-do, 39381, |
#### 3. Trade Name, Common Name, Classification [21 CFR 807.92(a)(2)]
| Trade Name: | 21HQ613D |
|-------------|----------|
|-------------|----------|
- Common Name: ●
Medical Monitor
Republic of Korea
- Classification: ●
| Classification Name | Medical image management and processing system. |
|-----------------------|-------------------------------------------------|
| Classification Number | 21 CFR 892.2050 |
| Product Code | PGY |
| Device Class | II |
| Review Panel | Radiology |
{4}------------------------------------------------
## 4. Identification of Predicate Device(s) [21 CFR 807.92(a)(3)]
The identified predicate devices within this submission are shown as follow;
K172738
RX560
EIZO Corporation
Medical image management and processing system.
#### Primary Predicate Device
| | 510(k) Number: |
|--|----------------|
|--|----------------|
- Applicant:
- Classification Name: ●
- Trade Name: ●
#### Predicate Device
●
●
●
- K201777 510(k) Number: Applicant: LG Electronics Classification Name: Medical image management and processing system. Trade Name: 31HN713D
- ●
## 5. Description of the Device [21 CFR 807.92(a)(4)]
The Medical monitor is intended to provide high resolution color and grayscale medical imaging for PACS and Radiology system. This Medical Monitor is intended to be used by trained medical practitioners for displaying, reviewing, and analysis of medical images
#### 6. Indications for use [21 CFR 807.92(a)(5)]
This Medical Monitor is indicated for use in displaying radiological images (including full-field digital mammoqraphy and digital breast tomosynthesis) for review, analysis, and diagnosis by trained medical practitioners.
{5}------------------------------------------------
#### 7. Technological Characteristics (Equivalence to Predicate Device) [21 CFR 807.92(a)(6)]
The table below presents comparisons between the subject device (21HQ613D) and the legally marketed predicate devices (K172738 and K201777):
| | Proposed Device | Primary Predicate Device |
|-----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| K Number | Not known | K172738 |
| Manufacturer | LG Electronics Inc. | EIZO Corporation |
| Model Name | 21HQ613D | RX560 |
| Classification Name | Medical image management and<br>processing system. | Medical image management and<br>processing system. |
| Classification Number | 21 CFR 892.2050 | 21 CFR 892.2050 |
| Indications for Use | This Medical Monitor is indicated for use<br>in displaying radiological images<br>(including full-field digital mammography<br>and digital breast tomosynthesis) for<br>review, analysis, and diagnosis by trained<br>medical practitioners. | This product is intended to be used in<br>displaying and viewing digital images,<br>including standard and multiframe digital<br>mammography, for review, analysis and<br>diagnosis by trained medical<br>practitioners. It is specially designed for<br>breast tomosynthesis applications. |
| Display Technology | The 21HQ613D has a TM213XDGP31<br>module for displaying. The<br>TM213XDGP31 is a " TFT Liquid Crystal<br>Display module with LED Backlight unit<br>and one 30pin eDP port and 4ch-LVDS<br>interface. This module supports<br>2048x2560 5M Pixel mode and can<br>display color driven by 10bit drivers. The<br>LCD module includes built-in converter<br>for Backlight | This is a color LCD monitor for viewing<br>digital images, including standard and<br>multiframe digital mammography, for<br>review, analysis and diagnosis. The color<br>panel employs in-plane switching (IPS)<br>technology allowing wide viewing angles<br>and the matrix size (or resolution) is<br>2048x2560 pixels. |
| Power Consumption | MAX. 120W<br>Off Mode ≤ 0.3W | 87 W / Less than 1 W |
| Screen size | 473.4 x 364.5 mm | 337.9 mm x 422.4 mm |
| LCD Screen | TFT LCD | TFT LCD |
| Pixel Pitch | 0.165 x 0.165 mm | 0.165 x 0.165 mm |
| Resolution | 2,048 x 2,560 pixels | 2,048 x 2,560 |
| Horizontal Frequency | 30 kHz to 135kHz | 31 kHz to 135 kHz |
| Vertical Frequency | 24 Hz to 61 Hz | 59 Hz to 61 Hz |
| Input video signals | DVI IN x 1,<br>DP IN x 2<br>DP OUT x 1 | DVI x 1<br>DP IN x 1<br>DP OUT x 1 |
[Table 1. Comparison of Proposed Device to Primary Predicate Device]
The comparison table shows that the subject device (21HQ613D) has the similar indications for use the primary predicate device. Although the devices have some different technological characteristics (screen size, pixel pitch, Resolution, horizontal/vertical frequency and input video signals), these differences do not make the subject device less safe and reliable, so the subject device fits for diagnostic use as the predicate device does. There are no significant differences in the technological characteristics of the subject device. All the differences between the subject and predicate device do not raise different questions of safety and effectiveness. It is substantially equivalent to a predicate device in indications for use and technology characteristics.
{6}------------------------------------------------
| Proposed Device | | Predicate Device |
|-----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| K Number | Not known | K201777 |
| Manufacturer | LG Electronics Inc. | LG Electronics Inc. |
| Model Name | 21HQ613D | 31HN713D |
| Classification Name | Medical image management and processing system. | Medical image management and processing system. |
| Classification Number | 21 CFR 892.2050 | 21 CFR 892.2050 |
| Indications for Use | This Medical Monitor is indicated for use in displaying radiological images (including full-field digital mammography and digital breast tomosynthesis) for review, analysis, and diagnosis by trained medical practitioners. | This Medical Monitor is indicated for use in displaying radiological images (including full-field digital mammography and digital breast tomosynthesis) for review, analysis, and diagnosis by trained medical practitioners. |
| Display Technology | The 21HQ613D has a TM213XDGP31 module for displaying. The TM213XDGP31 is a " TFT Liquid Crystal Display module with LED Backlight unit and one 30pin eDP port and 4ch-LVDS interface. This module supports 2048x2560 5M Pixel mode and can display color driven by 10bit drivers. The LCD module includes built-in converter for Backlight | The 31HN713D has a LB310FTM module for displaying. The LB310FTM is a 31" TFT Liquid Crystal Display module with LED Backlight unit. This module supports 4,200 x 2,800 pixels and can display color driven by 10bit drivers. |
| Power Consumption | MAX. 120W<br>Off Mode ≤ 0.3W | MAX. 150W<br>Sleep Mode ≤ 0.5W<br>Off Mode ≤ 0.3W |
| Screen size | 473.4 x 364.5 mm | 676.9 x 459.7 mm |
| LCD Screen | TFT LCD | TFT LCD |
| Pixel Pitch | 0.165 x 0.165 mm | 0.1554 x 0.1554 mm |
| Resolution | 2,048 x 2,560 pixels | 4,200 x 2,800 pixels |
| Horizontal Frequency | 30 kHz to 135kHz | 30 kHz to 135kHz |
| Vertical Frequency | 24 Hz to 61 Hz | 56 Hz to 61 Hz |
| Input video signals | DVI IN x 1,<br>DP IN x 2<br>DP OUT x 1 | DisplayPort x 2<br>HDMI x 1 |
| [Table 2. Comparison of Proposed Device to Predicate Device] | | | | |
|--------------------------------------------------------------|--|--|--|--|
|--------------------------------------------------------------|--|--|--|--|
The comparison table shows that the subject device (21HQ613D) has the similar indications for use the predicate device. Although the devices have some different technological characteristics (screen size, pixel pitch, Resolution, horizontal/vertical frequency and input video signals), these differences do not make the subject device less safe and reliable, so the subject device fits for diagnostic use as the predicate device does. There are no significant differences in the technological characteristics of the subject device. All the differences between the subject and predicate device do not raise different questions of safety and effectiveness. It is substantially equivalent to a predicate device in indications for use and technology characteristics.
{7}------------------------------------------------
#### 8. Non-Clinical Test summary
- 1) Electrical Safety and Electromagnetic Compatibility
The test results demonstrated that the proposed device complies with the following standards:
- · Electrical Basic Safety and Essential Performance reguirements in accordance with IEC 60601-1:2005/AMD1:2012
- · Electromagnetic Compatibility Testing in accordance with IEC 60601-1-2 Edition 4.0:2014
- · Medical Electrical Equipment Part 1-6: General Requirements for Basic Safety and Essential Performance - Collateral Standard: Usability in accordance with IEC 60601-1-6:2010/A1:2013
- 2) Software Validation
The 21HQ613D contains MODERATE level of concern software. The software was designed and developed according to a software development process and was verified and validated.
The software information is provided in accordance with FDA guidance: The content of premarket submissions for software contained in medical devices, on May 11, 2005.
- 3) Guidance
Display Devices for Diagnostic Radiology – Guidance for Industry and Food and Drug Administration Staff, issued on October 2, 2017
| | Measurements | Description | SE Note |
|--------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| a. | Spatial resolution | Measurements of the transfer of<br>information from the image data to the<br>luminance fields at different spatial<br>frequencies of interest typically done by<br>reporting the modulation transfer<br>function. Non-isotropic resolution<br>properties should be characterized<br>properly by providing two-dimensional<br>measurements or measurements along<br>at least two representative axes.<br>(Using TG18 QC Test Pattern) | Equivalent |
| b. | Pixel defects | Measurements (count, types (e.g., sub-<br>pixel or entire pixel, always-on, always-<br>off), and locations (map) of pixel<br>defects. This is typically provided as a<br>tolerance limit. Pixel defects can<br>interfere with the visibility of small<br>details in medical images. | Equivalent |
| c. | Artifacts | Evaluate for image artifacts such as<br>ghosting and/or image sticking from<br>displaying a fixed test pattern for a<br>period of time.<br>(Using 5x5 mosaic pattern, 64Gray /<br>127 Gray judgment) | Same |
| d. | Temporal response | Measurements of the temporal behavior<br>of the display in responding to changes<br>in image values from frame to frame.<br>Since these transitions are typically not<br>symmetric, rise and fall time constants | Equivalent |
| Measurements | Description | SE Note | |
| | are needed to characterize the system.<br>Slow displays can alter details and<br>contrast of the image when large image<br>stacks are browsed or in video,<br>panning, and zooming modes. | | |
| e. | Luminance | Measurements of the maximum and<br>minimum luminance that the device<br>outputs as used in the application under<br>recommended conditions and the<br>achievable values if the device is set to<br>expand the range to the limit. | Same |
| f. | Conformance to a<br>grayscale-to-luminance<br>function | Measurements of the mapping between<br>image values and the luminance output<br>following a target model response for<br>256 or more levels. | Equivalent |
| g. | Luminance at 30° and 45°<br>in diagonal, horizontal,<br>and vertical directions at<br>center and four corners | Measurements of the luminance<br>response at off-normal viewing related<br>to the target model for the luminance<br>response. | Equivalent |
| h. | Luminance uniformity or<br>Mura test | Measurements of the uniformity of the<br>luminance across the display screen. | Equivalent |
| i. | Stability of luminance and<br>chromaticity response<br>with temperature and time<br>of operation (on-time) | Measurements of the change in<br>luminance and chromaticity response<br>with temperature and use time. | Same |
| j. | Spatial noise | Measurements of the spatial noise level<br>as represented by the noise power<br>spectrum using an appropriate ratio of<br>camera and display pixels. Spatial noise<br>and resolution affect the way images<br>are presented to the viewer and can<br>alter features that are relevant to the<br>interpretation process of the physician<br>or radiologist. | Equivalent |
| k. | Reflection coefficient | Measurements of the reflection<br>coefficients of the display device.<br>Specular and diffuse reflection<br>coefficients can be used as surrogates<br>for the full bidirectional reflection<br>distribution function. | Same |
| l. | Veiling glare or small-spot<br>contrast | Measurements of the contrast obtained<br>for small targets. | Equivalent |
| m. | Color tracking | Chromaticity at different luminance<br>levels of primary colors as indicated by<br>the color coordinates in an appropriate<br>units system (e.g., CIE u'v') and the<br>color gamut enveloped by the primary<br>colors. | Equivalent |
| n. | Gray tracking | Chromaticity at different luminance<br>levels of gray shades, including the<br>white point, as indicated by the color<br>coordinates in an appropriate units<br>system | Equivalent |
According to the guidance, we test following performance items:
{8}------------------------------------------------
{9}------------------------------------------------
According to the above test results, there are no significant performance differences between 21HQ613D and the predicate devices that would adversely affect the use of the product. It has substantially equivalent performance compared to the predicate devices.
#### Clinical Test Summary:
No clinical studies were considered necessary and performed.
#### 9. Conclusion [21 CFR 807.92(b)(3)]
In according with the Federal Food & Drug and cosmetic Act, 21 CFR Part 807, and based on the information provided in this premarket notification LG Electronics, concludes that the 21HQ613D is substantially equivalent in safety and effectiveness to the predicate devices as described herein.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.