32HL512D

K192925 · Lg Electronics.Inc · PGY · Mar 13, 2020 · Radiology

Device Facts

Record IDK192925
Device Name32HL512D
ApplicantLg Electronics.Inc
Product CodePGY · Radiology
Decision DateMar 13, 2020
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.2050
Device ClassClass 2

Indications for Use

This Medical Monitor is indicated for use in displaying radiological images for review, analysis, and diagnosis by trained medical practitioners. The display is not intended for mammography.

Device Story

32HL512D is a 31.5-inch TFT LCD medical monitor designed for PACS and radiology systems. It accepts video input via DisplayPort or HDMI interfaces. The device transforms digital signals into high-resolution (3840 x 2160) color and grayscale images. Used in clinical settings by trained medical practitioners, the monitor facilitates the review, analysis, and diagnosis of radiological images. It provides high-resolution visualization to assist clinicians in interpreting medical imaging data, thereby supporting clinical decision-making. The device does not support mammography.

Clinical Evidence

No clinical studies were performed. Evidence is based on bench testing, including electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), and usability (IEC 60601-1-6) testing, alongside software verification and validation.

Technological Characteristics

31.5" TFT LCD module (LM315WR2) with LED backlight; 3840 x 2160 UHD resolution; 10-bit color drivers; 30-pin eDP interface; 4-lane eDP; input ports: 2x DisplayPort, 1x HDMI. Power consumption max 65W. Complies with IEC 60601-1:2005/AMD1:2012, IEC 60601-1-2:2014, and IEC 60601-1-6:2010/A1:2013.

Indications for Use

Indicated for trained medical practitioners to display, review, analyze, and diagnose radiological images. Not for mammography.

Regulatory Classification

Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo for the U.S. Food and Drug Administration (FDA). The logo features a stylized human figure on the left, followed by the acronym "FDA" in a blue square. To the right of the blue square, the words "U.S. FOOD & DRUG" are stacked above the word "ADMINISTRATION", all in blue. The logo is clean and professional, reflecting the organization's role in public health and safety. LG Electronics Inc. % Jinhwan Jun Chief Research Engineer 222, LG-ro, Cheongho-ri, Jinwi-myeon Pyeongtaek-si, Gyeonggi-do 17709 REPUBLIC OF KOREA March 13, 2020 #### Re: K192925 Trade/Device Name: 32HL512D Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: Class II Product Code: PGY Dated: February 6, 2020 Received: February 14, 2020 Dear Jinhwan Jun: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for {1}------------------------------------------------ devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. For Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K192925 Device Name 32HL512D Indications for Use (Describe) This Medical Monitor is indicated for use in displaying radiological images for review, analysis, and diagnosis by trained medical practitioners. The display is not intended for mammography. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|--| |-------------------------------------------------|--| X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary [As Required by 21 CFR 807.92] # 1. Date Prepared [21 CFR 807.92(a)(a)] K192925 February 7, 2020 # 2. Submitter's Information [21 CFR 807.92(a)(1)] | Name of Sponsor: | LG Electronics Inc. | |-----------------------|-----------------------------------------------------------------------------------------------| | Address: | 222, LG-ro, Cheongho-ri, Jinwi-myeon, Pyeongtaek-si,<br>Gyeonggi-do, 17709, Republic of Korea | | Contact Name: | Jinhwan Jun / Chief Research Engineer | | Telephone No.: | +82-31-8066-5641 | | Email Address: | jinhwan.jun@lge.com | | Name of Manufacturer: | LG Electronics Inc. | | Address: | 77, Sanho-daero, Gumi-si, Gyeongsangbuk-do, 39381, | #### 3. Trade Name, Common Name, Classification [21 CFR 807.92(a)(2)] | • Trade Name: | 32HL512D | |---------------|----------| |---------------|----------| - . Common Name: Medical Monitor Republic of Korea - Classification: ● | Classification Name | Picture archiving and communications system | |-----------------------|---------------------------------------------| | Classification Number | 21 CFR 892.2050 | | Product Code | PGY | | Device Class | II | | Review Panel | Radiology | {4}------------------------------------------------ # 4. Identification of Predicate Device(s) [21 CFR 807.92(a)(3)] The identified predicate devices within this submission are shown as follow; # Predicate Device ● ● ● | 510(k) Number: | K131090 | |----------------------|---------------------------------------------| | Applicant: | EIZO Corporation | | Classification Name: | Picture archiving and communications system | - Trade Name: RadiForce MX215 ● # 5. Description of the Device [21 CFR 807.92(a)(4)] The Medical monitor is intended to provide high resolution color and grayscale medical imaging for PACS and Radiology system. This Medical Monitor is intended to be used by trained medical practitioners for displaying, reviewing, and analysis of medical images # 6. Indications for use [21 CFR 807.92(a)(5)] This Medical Monitor is indicated for use in displaying radiological images for review, analysis, and diagnosis by trained medical practitioners. The display is not intended for mammography. {5}------------------------------------------------ # 7. Technological Characteristics (Equivalence to Predicate Device) [21 CFR 807.92(a)(6)] The table below presents comparisons between the subject device (32HL512D) and the legally marketed predicate device (K131090): | | Proposed Device | Predicate Device | |-----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | K Number | Not known | K131090 | | Manufacturer | LG Electronics Inc. | EIZO Corporation | | Model Name | 32HL512D | RadiForce MX215 | | Classification Name | Picture archiving and<br>communications system | Picture archiving and<br>communications system | | Classification Number | 21 CFR 892.2050 | 21 CFR 892.2050 | | Indications for Use | This Medical Monitor is indicated<br>for use in displaying radiological<br>images for review, analysis, and<br>diagnosis by trained medical<br>practitioners. The display is not<br>intended for mammography. | This product is intended to be<br>used in displaying and viewing<br>digital images for review and<br>analysis by trained medical<br>practitioners. It does not support<br>the display of mammography<br>images for diagnosis. | | Display Technology | The 32HL512D has a LM315WR2<br>module for displaying. The<br>LM315WR2 is a 31.5" TFT Liquid<br>Crystal Display module with LED<br>Backlight unit and one 30-pin<br>eDP port has 4 lane eDP<br>interface. This module supports<br>3840 x 2160 UHD mode and can<br>display color driven by 10bit<br>drivers. | RadiForce MX215 is a color LCD<br>monitor for viewing medical<br>images other than those of<br>mammography. The color panel<br>employs in-plane switching (IPS)<br>technology allowing wide<br>viewing angles and the matrix<br>size (or resolution) is 1,200 x<br>1,600 pixels (2MVP). | | Power Consumption | MAX. 65W<br>Sleep Mode ≤ 0.5W<br>Off Mode ≤ 0.3W | Max. 48W<br>Standby < 0.5W | | Screen size | 718.2 x 414.3 mm | 54 cm / 21.3" | | LCD Screen | TFT LCD | TFT LCD | | Pixel Pitch | 0.18159 x 0.18159 mm | 0.270 x 0.270 mm | | Resolution | 3,840 x 2,160 pixels | 1,200 x 1,600 pixels | | Horizontal Frequency | 30 kHz to 135kHz | 24 kHz to 80 kHz | | Vertical Frequency | 56 Hz to 61 Hz | 49 Hz to 76 Hz | | Viewing angle | View angle free (R/L 178(Typ.),<br>U/D 178(Typ.)) | Horizontal: 178°, vertical: 178° | | Input video signals | DisplayPort x 2<br>HDMI x 1 | DisplayPort x 1<br>DVI-I x 1 | [Table 1. Comparison of Proposed Device to Predicate Device] The comparison table shows that the subject device (32HL512D) has the similar indications for use the predicate one. Although the devices have some different technological characteristics (screen size, pixel pitch, resolution, horizontal/vertical frequency), these differences do not make the subject device less safe and reliable, so the subject device fits for diagnostic use as the predicate device does. There are no significant differences in the technological characteristics of the subject device. All the differences between the subject and predicate devise do not raise different questions of safety and {6}------------------------------------------------ effectiveness. It is substantially equivalent to a predicate device in indications for use and technology characteristics. # 8. Non-Clinical Test summary - 1) Electrical Safety and Electromagnetic Compatibility The test results demonstrated that the proposed device complies with the following standards: - · Electrical Basic Safety and Essential Performance requirements in accordance with IEC 60601-1:2005/AMD1:2012 - · Electromagnetic Compatibility Testing in accordance with IEC 60601-1-2 Edition 4.0:2014 - · Medical Electrical Equipment Part 1-6: General Requirements for Basic Safety and Essential Performance - Collateral Standard: Usability in accordance with IEC 60601-1-6:2010/A1:2013 - 2) Software Validation The 32HL512D contains MODERATE level of concern software. The software was designed and developed according to a software development process and was verified and validated. The software information is provided in accordance with FDA quidance: The content of premarket submissions for software contained in medical devices, on May 11, 2005. #### Clinical Test Summary: No clinical studies were considered necessary and performed. # 9. Conclusion [21 CFR 807.92(b)(3)] In according with the Federal Food & Drug and cosmetic Act, 21 CFR Part 807, and based on the information provided in this premarket notification LG Electronics, concludes that the 32HL512D is substantially equivalent in safety and effectiveness to the predicate devices as described herein.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...