OEC 3D

K203346 · Ge Oec Medical Systems, Inc. · OXO · Mar 5, 2021 · Radiology

Device Facts

Record IDK203346
Device NameOEC 3D
ApplicantGe Oec Medical Systems, Inc.
Product CodeOXO · Radiology
Decision DateMar 5, 2021
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1650
Device ClassClass 2
AttributesPediatric

Indications for Use

The OEC 3D mobile fluoroscopy system is designed to provide fluoroscopic and digital spot images of adult and pediativ populations during diagnostic, interventional, and surgical procedures. Examples of a clinical application may include: orthopedic, gastrointestinal, endoscopic, neurologic, vascular, cardiac, citical care and emergency procedures.

Device Story

OEC 3D is a mobile fluoroscopic C-arm system for intraoperative imaging. Inputs: X-ray projections captured via a 31cm x 31cm Thallium-doped Cesium Iodide flat panel detector with CMOS light imager. Transformation: System performs 3D volumetric reconstruction using a dedicated computer; utilizes an isocentric C-arm gantry for rotational acquisition. Outputs: 2D fluoroscopic images and 3D volumetric reconstructions displayed on a high-resolution LCD monitor. Used in OR/clinic settings by surgeons and physicians. Features include a live view camera for alignment, recirculating oil cooler for the X-ray tube, and six 3D imaging modes. Output assists in localizing anatomy and pathology, guiding interventional tools, and verifying post-procedure outcomes. Benefits include real-time visualization of anatomy and surgical progress.

Clinical Evidence

No clinical data. Substantial equivalence established via engineering bench testing, verification, and validation under GE OEC Medical Systems' Quality Management System. Testing included system integration, performance, safety, and simulated use.

Technological Characteristics

Mobile isocentric C-arm fluoroscopy system. X-ray chain: Thallium-doped CsI flat panel detector, CMOS imager, high voltage generator. Connectivity: Wireless capability, DICOM-compatible. Software: Modular architecture, IEC 62304 compliant. Cooling: Recirculating oil cooler. Standards: IEC 60601-1, 60601-1-3, 60601-2-43, 60601-2-54, NEMA XR-27, 21 CFR 1020.30/32, 1040.10.

Indications for Use

Indicated for adult and pediatric populations requiring fluoroscopic and digital spot imaging during diagnostic, interventional, and surgical procedures, including orthopedic, gastrointestinal, endoscopic, neurologic, vascular, cardiac, critical care, and emergency applications.

Regulatory Classification

Identification

An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Special Controls

*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{1}------------------------------------------------ March 5, 2021 Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. GE OEC Medical Systems, Inc. % Rachel Schandel Regulatory Affairs Leader 384 N Wright Brothers Drive SALT LAKE CITY UT 84116 Re: K203346 Trade/Device Name: OEC 3D Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: Class II Product Code: OXO, OWB, JAA Dated: February 5, 2021 Received: February 8, 2021 Dear Rachel Schandel: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see {2}------------------------------------------------ https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, For Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K203346 Device Name OEC 3D Indications for Use (Describe) The OEC 3D mobile fluoroscopy system is designed to provide fluoroscopic and digital spot images of adult and pediativ populations during diagnostic, interventional, and surgical procedures. Examples of a clinical application may include: orthopedic, gastrointestinal, endoscopic, neurologic, vascular, cardiac, citical care and emergency procedures. | Type of Use (Select one or both, as applicable) | | |----------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------| | <div> <span> ☒ Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect. and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ Image /page/3/Picture/2 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized, cursive font, enclosed within a blue circle. There are decorative white swirls around the outer edge of the circle, adding a touch of elegance to the design. The logo is simple, recognizable, and has been used by GE for many years. # 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS This 510(k) Summary of Safety and Effectiveness information is submitted in accordance with the requirement of 21 CFR Part 807.87(h). In accordance with 21 CFR 807.92 the following summary of information is provided: | Date: | February XX , 2020 | |-------------------------|--------------------------------------------------------------------------------------------------------------------------------| | Submitter: | GE OEC Medical Systems, Inc (GE Healthcare)<br>384 Wright Brothers Drive<br>Salt Lake City, Utah 84116 | | Primary Contact: | Rachel Schandel<br>Regulatory Affairs Leader<br>Tel: +1-385-237-7519<br>e-mail: rachel.schandel@ge.com | | Secondary Contacts: | Philip Malca<br>Regulatory Affairs Director<br>General Electric Company<br>Tel: +33 646-373-852<br>e-mail: philip.malca@ge.com | | PRODUCT IDENTIFICATION | | | Device Trade Name: | OEC 3D | | Regulation Name: | Image-intensified Fluoroscopic x-ray system | | Classification Panel: | Radiology | | Regulation: | 21CFR 892.1650 | | Classification: | Class II | | Product Codes: | OXO, OWB, JAA | | Manufacturer: | GE OEC Medical Systems, Inc (GE Healthcare)<br>384 Wright Brothers Drive.<br>Salt Lake City, Utah 84116 | | Manufacturing Location: | GE OEC Medical Systems, Inc (GE Healthcare)<br>384 Wright Brothers Drive. | #### Predicate Device: Device Name: 510(k) number: Manufacturer: Regulation Name: Regulation: Classification: Product Code: OEC Elite K192819 GE OEC Medical Systems, Inc Image-intensified Fluoroscopic x-ray system 21CFR 892.1650 Class II OWB, OXO, JAA {5}------------------------------------------------ #### GE Healthcare 510(k) Premarket Notification Submission - OEC 3D Image /page/4/Picture/2 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized, cursive font, enclosed within a blue circle. The circle is surrounded by four decorative, swirling shapes, one at each cardinal direction (top, bottom, left, and right). ### Primary Reference Device: | Device Name: | Cios Spin (VA30) | |------------------|---------------------------------------------| | 510(k) number: | K181550 | | Manufacturer: | Siemens Medical Solutions USA, Inc. | | Regulation Name: | Image-intensified Fluoroscopic x-ray system | | Regulation: | 21CFR 892.1650 | | Classification: | Class II | | Product Code: | OWB, OXO, JAA | #### Secondary Reference Device: | Device Name: | INNOVA IGS 5 | |------------------|---------------------------------------------| | 510(k) number: | K181403 | | Manufacturer: | GE Medical Systems SCS | | Regulation Name: | Image-intensified Fluoroscopic x-ray system | | Regulation: | 21CFR 892.1650 | | Classification: | Class II | | Product Code: | OWB, JAA, IZI and OXO | ### Device Description: The OEC 3D is a mobile fluoroscopic C-arm imaging system used to assist trained surgeons and other qualified physicians. The system is used to provide fluoroscopic X-ray images and volumetric reconstructions during diagnostic, interventional, and surgical procedures. These images help the physician visualize the patient's anatomy and interventional tools. This visualization helps to localize clinical regions of interest and pathology. The images provide real-time visualization and records of pre-procedure anatomy, in vivo-clinical activity and post-procedure outcomes. The system is composed of two primary physical components. The first is referred to as the "C -Arm" because of its "C" shaped image gantry; the second is referred to as the "Workstation", and this is the primary user interface for the user to interact with the system. The C-arm has an interface tablet allowing a technician to interact with the system. The C-arm is a stable mobile platform capable of performing linear motions (vertical, horizontal) and rotational motions (orbital, lateral) that allow the user to position the X-ray image chain at various angles and distances with respect to the patient anatomy to be imaged. The C-Arm is comprised of the high voltage generator, software, X-ray control, and a "C" shaped image gantry, which supports an X-ray tube and a Flat Panel Detector, The workstation is a stable mobile platform with an articulating arm supporting a color image high resolution LCD display monitor. It also includes image processing equipment/software, recording devices, data input/output devices and power control systems. On the C-Arm, the generator remains unchanged from the OEC Elite. This is also true for the 31 cm x 31 cm image receptor, consisting of a Thallium-doped Cesium Iodide [Cs] (TI)] solid state flat panel X-ray detector with Complementary Metal Oxide Semiconductor (CMOS) light imager. The X-ray tube housing and insert remains the same as on the predicate OEC Elite (K192819). C-Arm functionality is managed by a digital flat tablet control panel mounted on the C-arm base. Motion is controlled by a joystick. {6}------------------------------------------------ Image /page/5/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined in a stylized script, enclosed within a blue circle. There are decorative swirls around the letters, adding to the logo's distinctive design. The logo is simple, yet recognizable. On the workstation, the main hardware includes a computer with integrated wireless capability and a dedicated computer for 3D reconstruction located within the storage bay. The OEC 3D employs the same software architecture and platform design that fully supports the flat panel detector as the OEC Elite and complies with IEC 60601-1. The OEC 3D includes the existing 2D imaging functionalities available on the OEC Elite including imaging and post processing applications. The change that triggered this submission for the subject device compared to the predicate, OEC Elite, is the introduction of the new 3D functionality. However, the new 3D function is substantially equivalent to the Siemens Inc. Cios Spin (K181550) and the 3D algorithm is identical to the Medical Systems INNOVA IGS 5 (K181403). Additionally, software and hardware are comparable to the OEC Elite (K192819). This premarket notification is being submitted as a Traditional 510(k) to request clearance for the Subject Device the OEC 3D device. The subject device OEC 3D is substantially equivalent to the GE Healthcare predicate device OEC Elite (K192819) and the reference devices: Cios Spin (K181550) as well as INNOVA IGS 5(K181403), they are within the same classification and regulation with these devices. ### Proposed Device Modification: The proposed OEC 3D device is a modification of GE's own predicate OEC Elite system (K192819). Its built upon the existing technologies of the predicate device "OEC Elite" (K192819). The primary modifications for the OEC 3D as compared to OEC Elite are summarized below: - 1. Composite isocentric C gantry - 2. Dedicated 3D reconstruction computer - 3. Live view camera embedded in the detector housing - 4. Recirculating oil cooler on the X-ray tube The wheels and wheel brakes, cable wraps, working surface, ergonomic handles, aesthetic open frame, LCD monitor, articulating arm and latch system remain unchanged from the OEC Elite. #### Indications for Use: The OEC 3D mobile fluoroscopy system is designed to provide fluoroscopic and digital spot images of adult and pediatric populations during diagnostic, interventional, and surgical procedures. Examples of a clinical application may include: orthopedic, gastrointestinal, endoscopic, urologic, neurologic, vascular, cardiac, critical care and emergency procedures. ### Technology: The indications for use are identical and technology is similar to the predicate device. X-ray generation and control used with the subject device OEC 3D is identical to the technology used with the predicate device OEC Elite. The subject OEC 3D device employs the same fundamental technology as that of the predicate device. The image chain including the X-ray tube, high voltage generator, collimator, X-ray filters, and detectors, remains unchanged from the predicate, OEC Elite. The 3D functionality is similar to the reference device Cios Spin and INNOVA IGS 5. Table 1 provides predicate and reference device comparison. {7}------------------------------------------------ Image /page/6/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized script, enclosed within a blue circle. There are four white teardrop shapes symmetrically placed around the circle, adding a decorative element to the design. The system continues to meet all applicable IEC 60601-1 series of standards, NEMA XR-27, and applicable parts of 21CFR Subchapter J. The new performance claims did not require clinical data in order to establish safety or efficacy. The OEC 3D device was built upon the existing modular and extensible software architecture, following the same design control process and software development lifecycle process that is compliant to IEC62304 used in the predicate OEC Elite. The changes described above do not change the control mechanism, operating principle, energy type, or the scientific technology of the predicate devices. | Device Name and Manufacturer | 510(k)<br>Number | Clearance<br>Date | Comparable Properties | |------------------------------------------------------------------------------------------------------------------|------------------|-------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Predicate Device: OEC Elite<br>GE OEC Medical Systems, Inc<br>Product Codes: OWB, OXO, JAA | K192819 | 11/08/2019 | - Indications for Use<br>- X-ray tube assembly<br>- X-ray generator<br>- Collimator<br>- Image receptor<br>- X-ray control modes<br>- X-ray initiation and termination<br>- Imaging modes<br>- Dose area product<br>- Imaging features<br>- Image noise reduction<br>- Power requirements<br>- C-arm dimensions<br>- C-arm positioning<br>- Motorization<br>- Workstation dimensions<br>- Display articulation<br>- Primary monitor<br>- Control panel<br>- Software platform<br>- Computing hardware platform<br>- Image management<br>- Image storage<br>- Data transfer<br>- Alignment Aid<br>- Standards compliance | | Primary Reference Device: Cios<br>Spin<br>Siemens Medical Solutions USA,<br>Inc.<br>Product Codes: OWB, OXO, JAA | K181550 | 10/30/2018 | - Isocentric C-arm design<br>- 3D imaging modes<br>- 3D imaging features<br>- Navigation interface | Table 1: Predicate and Reference Device Comparison {8}------------------------------------------------ Image /page/7/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized, cursive font, enclosed within a blue circle. The circle has decorative flourishes or wave-like designs emanating from it, giving it a dynamic and recognizable appearance. | Device Name and Manufacturer | 510(k)<br>Number | Clearance<br>Date | Comparable Properties | |---------------------------------------------|------------------|-------------------|-----------------------| | Secondary Reference Device:<br>INNOVA IGS 5 | K181403 | 11/02/2018 | - 3D Algorithm | | GE Medical Systems SCS | | | | | 283 Rue De La Miniere, 78530<br>BUC, France | | | | | Product Codes: JAA, IZI and OXO | | | | ### Determination of Substantial Equivalence: The main change in the proposed device is the introduction of the 3D functionality. The table below summarizes the substantive feature/technological differences between the predicate device and the proposed device: | Key<br>Difference<br>Category | Subject Device:<br>OEC 3D | Predicate Device:<br>OEC Elite<br>(K192819) | Reference Device:<br>Cios Spin<br>(K181550) | Comparison Results | |-----------------------------------|----------------------------------------------------------------------------------------------------|------------------------------------------------------------|---------------------------------------------|-----------------------------------------------------------------------------------------------------------| | X-ray Tube<br>Assembly | X-ray tube assembly<br>with housing, tube<br>insert, and<br>recirculating oil<br>cooling unit. | X-ray tube<br>assembly with<br>housing and tube<br>insert. | NA | Improved cooling<br>rate and capacity<br>does not raise any<br>new issues of safety<br>and effectiveness. | | Power<br>Requirements | 115-120V 20A<br>power for additional<br>computing<br>hardware. | 100-120V 15A<br>power. | NA | Change in power<br>consumption does<br>not raise any new<br>issues of safety and<br>effectiveness. | | C-arm<br>Dimensions | Isocentric gantry<br>design changes<br>overall dimensions<br>and weight. | Offset gantry<br>design. | NA | C-arm dimensions do<br>not raise any new<br>issues of safety and<br>effectiveness. | | C-arm<br>Positioning | Increased orbital<br>range for 3D<br>scanning. | Orbital range<br>commensurate<br>with 2D imaging. | NA | C-arm positioning<br>does not raise any<br>new issues of safety<br>and effectiveness. | | Workstation<br>Dimensions | Workstation weight<br>increased due to<br>additional computer<br>and support<br>hardware. | Base configuration<br>workstation<br>weight. | NA | Workstation weight<br>does not raise any<br>new issues of safety<br>and effectiveness. | | Computing<br>Hardware<br>Platform | Two computers, one<br>computer for<br>imaging functions<br>and features and<br>second computer for | One computer for<br>imaging functions<br>and features. | NA | Additional computer<br>does not raise any<br>new issues of safety<br>and effectiveness. | {9}------------------------------------------------ Image /page/8/Picture/2 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized, cursive font, enclosed within a blue circle. The circle is surrounded by decorative, swirling lines, also in blue, which give the logo a classic and recognizable appearance. | Key<br>Difference<br>Category | Subject Device:<br>OEC 3D | Predicate Device:<br>OEC Elite<br>(K192819) | Reference Device:<br>Cios Spin<br>(K181550) | Comparison Results | |-------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------|------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------| | | 3D feature. | | | | | Alignment<br>Aid | Laser aimer and<br>preview camera. | Laser aimer. | NA | Addition of preview<br>camera does not raise<br>any new issues of<br>safety and<br>effectiveness. | | C-arm<br>Design | Isocentric gantry<br>capable of rotating<br>200 degrees. | NA | Isocentric gantry<br>capable of rotating<br>196 degrees. | Gantry design<br>change does not raise<br>any new issues of<br>safety and<br>effectiveness. | | 3D Imaging<br>Modes | Six modes with 200-<br>400 images per<br>scan. | NA | Four modes with<br>100-400 images<br>per scan. | Imaging modes are<br>similar and do not<br>raise any new issues<br>of safety and<br>effectiveness. | | 3D Imaging<br>Features | Comprehensive set<br>of adjustments and<br>viewing options for<br>visualizing 3D<br>reconstructions, plus<br>a maximum<br>intensity projection<br>(MIP) view. | NA | Comprehensive set<br>of adjustments and<br>viewing options<br>for visualizing 3D<br>reconstructions. | Imaging features are<br>similar and do not<br>raise any new issues<br>of safety and<br>effectiveness. | ### Non-Clinical Performance Testing The OEC 3D device has successfully completed verification testing per GE Healthcare quality system as well engineering bench testing in support of this submission. The system has been tested and is compliant with the IEC 60601-1, including IEC 60601-1-3, 60601-1-3, 60601-2-43, and 60601-2-54. All applicable 21CFR Subchapter J performance standards are met., 1020.30 Diagnostic X-Ray Systems and their major components, 1020.32 Fluoroscopic equipment, 1040.10 Laser products. The OEC 3D system was developed under the GE OEC Medical Systems Quality Management System, including design controls, risk management and software development life cycle processes. The following quality assurance measures were applied to the development of the system: - Risk Analysis - Required Reviews - · Design Reviews - · Integration testing (System verification) - · Performance testing (Verification) - · Safety testing (Verification) - · Simulated use testing (Validation) The substantial equivalence was also based on software documentation for a "Moderate" level of concern device. {10}------------------------------------------------ #### GE Healthcare 510(k) Premarket Notification Submission - OEC 3D Image /page/9/Picture/2 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in a stylized font in the center. The circle is surrounded by a decorative border. The logo is simple and recognizable, and it is often used to represent the company's brand. ### Clinical Testing The subject OEC 3D device is substantially equivalent to the predicate and reference devices. The indication for use is identical and has equivalent/identical technological characteristics. This type of notification supports using scientific, established, engineering-based performance testing. The system can be fully tested/evaluated using engineering bench testing and clinical data is not required to demonstrate substantial equivalence. ### Summary: OEC 3D uses identical imaging, identical Indications for Use, and has equivalent/identical technological characteristics to the predicates. This type of device supports using scientific, established, engineeringbased performance testing. The system can be fully tested/evaluated using engineering bench testing and clinical data is not required to demonstrate substantial equivalence. Results of all conducted testing assessments were found acceptable and do not raise any new issues of safety or effectiveness. ### Substantial Equivalence Conclusion: The OEC 3D device was built upon the existing modular and extensible predicate OEC Elite device (K192819), following the same design control process and software development lifecycle processes. The differences discussed in this section do not introduce any adverse effects nor raise new questions of safety and effectiveness. Based on the successful verification and validation testing, additional engineering bench testing, conformance to standards, and development under GE OEC Medical System's Quality Management System, we believe that the subject OEC 3D device is of comparable type and substantially equivalent to the predicate device OEC Elite (K192819) with support from the reference devices Siemens Cios Spin device (K181550) and INNOVA IGS 5 (K181403), and therefore is safe and effective for its intended use.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...