Cios Spin
K181550 · Siemens Medical Solutions USA, Inc. · OWB · Oct 30, 2018 · Radiology
Device Facts
| Record ID | K181550 |
| Device Name | Cios Spin |
| Applicant | Siemens Medical Solutions USA, Inc. |
| Product Code | OWB · Radiology |
| Decision Date | Oct 30, 2018 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 892.1650 |
| Device Class | Class 2 |
| Attributes | Pediatric |
Intended Use
The Cios Spin is a mobile X-Ray system designed to provide X-ray imaging of the anatomical structures of patient during clinical applications. Clinical applications may include but are not limited to: interventional fluoroscopic, gastro- intestinal, endoscopic, urologic, pain management, orthopedic, neurologic, vascular, cardiac, critical care and emergency room procedures. The patient population may include pediatric patients.
Device Story
Mobile fluoroscopic C-arm X-ray system for surgical environments; provides image acquisition for orthopedic/vascular procedures. System comprises C-arm (X-ray source/flat panel detector) and mobile trolley (image processing/display). Operator (physician/technician) positions C-arm; system acquires X-ray signals, processes them via VA30 software, and displays images on trolley monitors. Features include CMOS detector, metal artifact reduction, 3D reconstruction (Retina 3D), screw localization (Screw Scout), and trajectory guidance (Target Pointer). Output assists surgeons in real-time visualization of anatomy, implants, and devices; supports image-guided surgery. Benefits include improved image quality, precise 3D visualization, and workflow efficiency. Anti-microbial coating on C-arm/trolley reduces contamination risk.
Clinical Evidence
Bench testing only. No clinical data required. Performance evaluated via non-clinical testing per FDA guidance for Solid State X-ray Imaging Devices (SSXI), including DQE, dynamic range, MTF, and pixel pitch. Software validated through verification and validation testing, including human factors usability assessment.
Technological Characteristics
Mobile C-arm X-ray system; CMOS flat panel detector; X-ray source; mobile trolley with integrated PC. Connectivity: wired/wireless footswitch, data network, optional NaviLink 3D interface. Software: VA30 version. Anti-microbial coating on C-arm/trolley. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-54, IEC 62304, ISO 14971.
Indications for Use
Indicated for X-ray imaging of anatomical structures during clinical procedures including interventional fluoroscopic, GI, endoscopic, urologic, pain management, orthopedic, neurologic, vascular, cardiac, critical care, and ER procedures. Includes pediatric patients.
Regulatory Classification
Identification
An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
- Cios Alpha (K132094)
- ARTIS pheno (K163286)
- Arcadis Orbic (3D) (K042793)
- Arcadis Avantic (K051133)
Reference Devices
- syngo Application Software (VA20) (K170747)
- ZIEHM Vision RFD (K132904)
- ZIEHM Solo FD Mobile Imaging System (K161976)
Related Devices
- K251523 — Cios Spin · Siemens Medical Solutions USA, Inc. · Jul 29, 2025
- K210055 — Cios Alpha · Siemens Medical Solutions USA, Inc. · Feb 5, 2021
- K251520 — Cios Alpha; Cios Flow · Siemens Medical Solutions USA, Inc. · Jul 9, 2025
- K203504 — Cios Flow · Siemens Medical Solutions USA, Inc. · Dec 22, 2020
- K233748 — CIARTIC Move (VB10) · Siemens Medical Solutions USA, Inc. · Mar 15, 2024
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left is the Department of Health & Human Services logo, which features a stylized human figure. To the right of that is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
October 30, 2018
Siemens Medical Solutions USA, Inc. Patricia D. Jones Senior Regulatory Affairs Technical Specialist 40 Liberty Boulevard 65-1A Malvern, Pennsylvania 19355
Re: K181550
Trade/Device Name: Cios Spin Regulation Number: 21 CFR 892.1650 Regulation Name: Image-Intensified Fluoroscopic X-Ray System Regulatory Class: Class II Product Code: OWB, OXO Dated: June 11, 2018 Received: June 13, 2018
Dear Patricia Jones:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice
(https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Image /page/1/Picture/7 description: The image shows the name Michael D. O'hara -S in a large font. Below the name is the text 'DN: c=US, o=U.S. Government, ou=HHS, ou=FDA, ou=People, 0.9.2342.19200300.100.1.1=1300 226759, cn=Michael D. O'hara -S Date: 2018.10.30 09:01:21 -04'00' For'. The text appears to be a digital signature.
Robert A. Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange. To the right of the word "Healthineers" is a graphic of orange dots.
| DEPARTMENT OF HEALTH AND HUMAN SERVICES | Form Approved: OMB No. 0910-0120 |
|--------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Food and Drug Administration | Expiration Date: 06/30/2020 |
| Indications for Use | |
| | See PRA Statement below. |
| 510(k) Number (if known) | K181550 |
| Device Name | Cios Spin |
| Indications for Use (Describe) | The Cios Spin is a mobile X-Ray system designed to provide X-ray imaging of the anatomical structures of patient during clinical applications. Clinical applications may include but are not limited to: interventional fluoroscopic, gastro- intestinal, endoscopic, urologic, pain management, orthopedic, neurologic, vascular, cardiac, critical care and emergency room procedures. The patient population may include pediatric patients. |
| Type of Use (Select one or both, as applicable) | |
| <label><input checked="checked" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> | <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
| FORM FDA 3881 (7/17) | Page 1 of 1 | PSC Publishing Services (301) 443-6740 | EF |
|----------------------|-------------|----------------------------------------|----|
|----------------------|-------------|----------------------------------------|----|
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the Siemens Healthineers logo. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the word "Healthineers" is a cluster of orange dots.
# 510(k) Summary: Cios Spin (VA30)
Siemens Medical Solutions USA, Inc. Company: 40 Liberty Boulevard, 65-1A Malvern, PA 19355
Date Prepared: October 2, 2018
This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR § 807.92.
- 1. General Information: Importer / Distributor: Siemens Medical Systems USA, Inc. 40 Liberty Boulevard, 65-1A Malvern, PA 19355 Establishment Registration Number: 2240869
Location of Manufacturing Site/Manufacturer: SIEMENS AG Sector Healthcare Röntgenstrasse 19 - 21 D-95478 Kemnath, Germany Establishment Registration Number: 3002466018
## 2. Contact Person:
Ms. Patricia D Jones Technical Specialist, Regulatory Submissions Siemens Medical Solutions USA, Inc. 40 Liberty Boulevard, 65-1A Malvern, PA 19355 Phone: (610) 448-6474 Email: patricia.d.jones@siemens-healthineers.com
## 3. Device Name and Classification:
Classification Requlation:
Trade Name: Classification Name:
Classification Panel:
Device Class:
Product Codes:
Cios Spin Image-Intensified Fluoroscopic X-Ray System Radiology 21 CFR §892.1650 Class II OWB, OXO
- 4. Legally Marketed Primary Predicate Device Trade Name: Cios Alpha 510(k) Clearance #: K132094
{4}------------------------------------------------
Classification Name:
Classification Panel: Classification Requlation: Device Class: Product Code: Recall Information:
March 11, 2014 Image-Intensified Fluoroscopic X-Ray System Radiology 21 CFR §892.1650 Class II OWB, OXO All product Recall incidents are considered during the Design Input phase of development to ensure the latest models will not be affected by any of the applicable issues.
# Legally Marketed Secondary Predicate Device
Trade Name: 510(k) Clearance #: Clearance Date: Classification Name:
Classification Panel: Classification Regulation: Device Class: Product Code: Recall Information:
ARTIS pheno (VE10) K163286 March 9, 2017 Image-Intensified Fluoroscopic X-Ray System Radiology 21 CFR §892.1650 Class II OWB. JAA All product Recall incidents are considered the Design Input phase during of development to ensure the latest models
will not be affected by any of the applicable issues.
# Legally Marketed Secondary Predicate Device
Trade Name: 510(k) Clearance #: Clearance Date: Classification Name:
Classification Panel: Classification Regulation: Device Class: Product Code: Recall Information:
Arcadis Orbic (3D) K042793 October 29, 2004 Image-Intensified Fluoroscopic X-Ray System Radiology 21 CFR §892.1650 Class II OWB, OXO, JAA All product Recall incidents are considered Design Input phase during the of development to ensure the latest models will not be affected by any of the applicable issues.
{5}------------------------------------------------
qally Marketed Secondary Predicate Device Trade Name: 510(k) Clearance #: Clearance Date: Classification Name:
Arcadis Avantic K051133 June 1, 2005 Image-Intensified Fluoroscopic X-Rav System Radiology 21 CFR §892.1650 Class II ОХО No Recalls
## 5. Device Description:
Device Class:
Product Code:
Recall Information:
Classification Panel:
Classification Regulation:
The Cios Spin mobile fluoroscopic C-arm X-ray System designed for the surgical environment. The Cios Spin provides comprehensive image acquisition modes to support orthopedic and vascular procedures. The system consists of two major components:
- a) The C-arm with X-ray source on one side and the flat panel detector on the opposite side. The c-arm can be angulated in both planes and be lifted vertically, shifted to the side and move forward/backward by an operator.
- b) The second unit is the image display station with a moveable trolley for the image processing and storage system, image display and documentation. Both units are connected to each other with a cable.
The main unit is connected to the main power outlet and the trolley is connected to a data network.
The following modifications were made to the predicate device the Cios Alpha. Siemens Medical Solutions USA, Inc. submits this Traditional 510(k) to request clearance for the Subject Device the Cios Spin for the following device modification made to the Predicate Device (Cios Alpha (VA10).
# Proposed Device Modifications:
- 1) Updated the Indications for Use Statement to include the Subject Device Name: "Cios Spin"
- 2) New Software VA30 due to new functionality
- New Flat Panel Detector Model (Software & Hardware) a)
- New Algorithm for Metal Artifact Reduction (Software) b)
- New Retina 3D (Software) C)
- Screw Scout (Software) d)
- Target Pointer (Software) e)
- f) Cios Open Apps (Hardware & Software
- High Power 3D (Hardware & Software) g)
- Easy 3D (Software) h)
{6}------------------------------------------------
- ote Control Unit on Cart/Trolley (Hardware)
- New Digital Cine Mode (DCM) 3D (Software)
- 5) New Green Color for Laser Aim (Hardware)
- Optional Navilink 3D (3rd Party Component) ର)
- Anti-microbial Coating on C-Arm and Trollev (Hardware) w/ cleaning instructions 7)
- 8) Optional Wireless or Wired Footswitch (Software & Hardware)
- റി Proposed Product Claims List
The Subject Device Cios Spin (VA30) is within the same classification regulation with the same indication for use as the predicate devices.
### 6. Indications for Use:
The Cios Spin is a mobile X-Ray system designed to provide X-ray imaging of the anatomical structures of patient during clinical applications. Clinical applications may include but are not limited to: interventional fluoroscopic, gastro-intestinal endoscopic, urologic, pain management, orthopedic, neurologic, vascular, cardiac, critical care and emergency room procedures. The patient population may include pediatric patients.
### 7. Substantial Equivalence:
The Cios Spin (VA30) is substantially equivalent to the commercially available primary predicate device Siemens Cios Alpha (VA10), cleared 11/03/2014 with K132094.
Indications for use remain unchanged the technology is similar. The predicate Flat Panel detector (a-Si technology) has been replaced by a CMOS Flat Panel detector. X-ray generation and control used with the Cios Spin is similar to the technology used with the primary predicate device Cios Table 1 provides primary and secondary Predicate Devices Alpha. comparable information.
| Predicate Device(s) Name and<br>Manufacturer | 510(k)<br>Number | Clearance Date | Comparable Properties |
|---------------------------------------------------------------------------|------------------|------------------|--------------------------------------------------------------------------------------------------|
| Siemens<br>Primary Predicate: Cios Alpha<br>Product Codes: OWB, OXO | K132094 | March 11, 2014 | -Indications for Use-<br>-System for Image Acquisition<br>-Post processing Software<br>-Detector |
| Siemens<br>Secondary Predicate: ARTIS<br>pheno<br>Product Codes: OWB, JAA | K163286 | March 09, 2017 | -Anti-Microbial Coating<br>-Ant-Microbial Coating Claims<br>-Cleaning Instructions |
| Siemens<br>Primary Predicate: ARCADIS<br>Avantic<br>Product Codes: OXO | K051133 | June 1, 2005 | -Digital Cine Mode (DCM) |
| Siemens<br>Secondary Predicate: ARCADIS | K042793 | October 29, 2004 | -3D functionality |
# Table 1: Primary and Secondary Comparable Properties
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. To the right of the word "Healthineers" is a graphic of orange dots.
| Predicate Device(s) Name and<br>Manufacturer | 510(k)<br>Number | Clearance Date | Comparable Properties |
|------------------------------------------------------------------------------------------------------------|------------------|-------------------|---------------------------------------|
| Orbic 3D<br>Product Codes: OWB, OXO, JAA | | | -C-arm Range |
| Reference 510(k) Information | | | |
| Siemens<br>Reference 510(k): syngo<br>Application Software (VA20)<br>Product Codes: LLZ | K170747 | June 9, 2017 | -Metal Artifact Reduction<br>Software |
| ZIEHM IMAGING GMBH<br>Reference 510(k): ZIEHM Vision<br>RFD Product Codes:<br>OWB, OXO JAA | K132904 | December 12, 2013 | -Detector Information |
| ZIEHM IMAGING GMBH<br>Reference 510(k):<br>Solo FD Mobile Imaging System<br>Product Codes:<br>OWB, OXO JAA | K161976 | October 6, 2016 | -Detector Information |
## 8. Summary of Technological Characteristics of the Subject Device as Compared to the Predicate Devices:
The Cios Spin (VA30) is substantially equivalent to the commercially available Siemens Cios Alpha (VA10), cleared 03/11/2014 with K132094.
The Indication for use remains unchanged and technology and design of the Cios Spin (VA30) is based on the Cios Alpha (VA10). The predicate Flat Panel detector (a-Si technology) has been replaced by a CMOS Flat Panel detector and an optional wireless footswitch.
An anti-microbial coating is added to the trolley and C-Arm which are components of the Cios Spin (VA30). The anti-microbial coating and associated claims were cleared in the secondary Predicate Device the ARTIS pheno (K163286) on March 9, 2017.
Table 2 provides comparison of the Subject Device modifications to the Predicate Devices.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the words is a cluster of orange dots.
# Table 2: Comparison of Technological Characteristics
| Modifications | Subject Device<br>Cios Spin (VA30) | Primary<br>Predicate<br>Device<br>Cios Alpha<br>(VA10) K132094 | Secondary<br>Predicate<br>Device<br>ARTIS pheno<br>(VE10) K163286 | Comparison Results |
|----------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------|-------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1. IFU<br>Statement | Same IFU Statement<br>– Changed Name to<br>Cios Spin | Same IFU<br>cleared in 510(k) | N/A | Same except for Name<br>Change from Cios Alpha<br>to Cios Spin. The<br>intended use and IFU of<br>the device remains the<br>same. |
| 2. System<br>Software | Software Version<br>VA30 | Software Version<br>VA10 | N/A | VA30 was developed to<br>support identified<br>modifications (a-f).<br>Testing requirements are<br>acceptable per Software<br>and SSIX guidance<br>requirements. |
| | a) CMOS<br>technology<br>Detector<br>(Software &<br>Hardware) | a-Si technology<br>Detector | N/A | Equivalent image quality<br>does not raise any new<br>issues of safety of<br>effectiveness. Non-clinical<br>testing was conducted<br>and is acceptable per<br>SSIX Guidance Document |
| | b) Algorithm for<br>Metal Reduction<br>(Software)<br>important<br>diagnostic<br>information can<br>be obscured by<br>metal artifacts.<br>Metal Artifact<br>Reduction<br>improves the<br>image quality by<br>reducing these<br>artifacts | N/A | N/A | Reference 510(k)<br>syngo Application<br>Software VD20B<br>(K170747) contains a<br>Metal Reduction software<br>feature that is comparable<br>to the Metal Reduction<br>Software feature in the<br>VA30 software for the<br>Cios Spin, the algorithms<br>is unchanged. |
| | c) Retina 3D<br>(Software)<br>provides precise<br>3D visualization<br>of fine structures<br>in exceptional<br>image quality lets<br>you confidently<br>see and evaluate<br>anatomical<br>structures,<br>implants, screws,<br>and devices. 3D<br>images<br>reconstruction<br>from 2D images. | N/A | Yes | Non-clinical Summaries<br>for Retina 3D, Screw<br>Scout, Larger C-Arm<br>range, Target Pointer,<br>Open Apps non-clinical<br>testing and Software<br>Verification /validation<br>testing was conducted<br>and is acceptable per<br>Software Guidance<br>document. The Retina 3D<br>Feature has the same<br>reconstruction algorithm<br>as the predicate ARTIS<br>Pheno. |
| Modifications | Subject Device<br>Cios Spin (VA30) | Primary<br>Predicate<br>Device<br>Cios Alpha<br>(VA10) K132094 | Secondary<br>Predicate<br>Device<br>ARTIS pheno<br>(VE10) K163286 | Comparison Results |
| | The 2D images<br>are acquired<br>during an orbital<br>sweep. | | | |
| d) | Screw Scout<br>(Software) Screw<br>Scout<br>automatically<br>localizes screws<br>and prepares the<br>optimal view of<br>any screw in the<br>3D dataset. | N/A | N/A | |
| e) | Target Pointer<br>(Software)<br>delivers a<br>trajectory that<br>helps surgeons to<br>optimally position<br>k-wires or other<br>devices and<br>ultimately save<br>time. | N/A | N/A | |
| f). | Cios Open Apps –<br>(Hardware &<br>Software) This<br>feature allows the<br>user to use<br>different external<br>applications in a<br>separate<br>integrated PC (in<br>the trolley). | N/A | N/A | |
| g) | High Power 3D<br>(Software) | N/A | N/A | Does not raise any new<br>issues of safety of<br>effectiveness per Software<br>Guidance. |
| h) | Easy 3D<br>(Software) | N/A | N/A | Does not raise any new<br>issues of safety of<br>effectiveness per Software<br>Guidance. |
| 3. Mounted<br>Remote<br>Control Unit<br>on<br>Cart/Trolley | Mounted Remote<br>Control Unit on Cart | N/A | N/A | Does not raise any new<br>issues of safety of<br>effectiveness per<br>verification and validation<br>testing. |
| 4. Digital Cine<br>Mode | Digital Cine Mode<br>(DCM) 3D | N/A | Secondary<br>Predicate<br>Device<br>Arcadis Avantic<br>K163286 | Same except for being<br>able to view in 3D.<br>Feature does not raise<br>any new safety or<br>effectiveness issues |
| Modifications | Subject Device<br>Cios Spin (VA30) | Primary<br>Predicate<br>Device<br>Cios Alpha<br>(VA10) K132094 | Secondary<br>Predicate<br>Device<br>ARTIS pheno<br>(VE10) K163286 | Comparison Results |
| | | | Digital Cine Mode | per verification and<br>validation testing. |
| 5. Laser<br>Aiming<br>Beam | Green Color Laser<br>Aiming Beam | Red color Aiming<br>Beam | N/A | New Green color for<br>laser Aiming Beam<br>does not impact the<br>Safety or the<br>effectiveness of the<br>feature. |
| 6. NaviLink 2D<br>Intergrated<br>2D<br>navigation<br>interface | Optional NaviLink<br>3D integration allows<br>connection to an<br>external Navigation<br>system.<br>NaviLink 3D is a<br>digital interface to<br>connect with certified<br>navigation systems.<br>It automatically<br>transfers 3D datasets<br>to the navigation<br>systems for<br>combined use<br>of image guided and<br>navigated surgery. | NaviLink 2D<br>Intergrated 2D<br>navigation<br>interface | N/A | Similar functionality. 3D<br>integration does not<br>raise any new issues of<br>safety and<br>effectiveness |
| 7. Anti-<br>Microbial<br>Coating<br>w/Cleaning<br>Instructions | Optional Anti-<br>Microbial coating on<br>C-Arm and Trolley | N/A | Optional Anti-<br>Microbial coating<br>on C-Arm and<br>Trolley | Same |
| 8. Footswitch | Wired footswitch<br>Optional Wireless<br>footswitch | Wired Footswitch<br>N/A | N/A<br>N/A | Same<br>Optional wireless<br>footswitch with same<br>functionality like wired<br>footswitch. Wireless<br>version does not raise<br>any new safety or<br>effectiveness issues.<br>Tested per Wireless<br>Guidance. |
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the Siemens Healthineers logo. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. There are also some orange dots to the right of the word "Healthineers."
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. To the right of the words is a graphic of orange dots.
## Non-clinical Performance Testing: 9.
The Siemens Cios Spin has been tested to meet the requirements for conformity to multiple industry standards. Performance testing confirmed, that the Siemens Cios Spin complies with the following 21 CFR Federal Performance Standards:
1020.30 Diagnostic X-Ray Systems and their major components 1020.32 Fluoroscopic equipment
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the logo for Siemens Healthineers. The words "SIEMENS" and "Healthineers" are stacked on top of each other, with "SIEMENS" in teal and "Healthineers" in orange. To the right of the words is an orange graphic of several circles of varying sizes. Below the logo is the text "1040.10 Laser products" in black.
Siemens conforms to the following relevant voluntary FDA Recognized Consensus Standards as listed in the table below: The Cios Spin is certified by Siemens Healthcare GmbH Corporate Testing Laboratory to comply with the following voluntary standards listed in Table 3 below:
# Table 3: Conformance Standards
| Standards Development<br>Organization Reference Number and<br>Date | Standards Development<br>Organization Reference Number and<br>Date | | |
|--------------------------------------------------------------------|--------------------------------------------------------------------|----|------------------------------------------|
| 1 | AAMI ANSI ES60601-1:2005/<br>(R)2012 | 7 | IEC 60601-2-28:2010 |
| 2 | AMI ANSI IEC 60601-1-2:2014 | 8 | IEC 60601-2-43:2017<br>(recognized 2010) |
| 3 | IEC 60601-1-3:2013 | 9 | IEC 60601-2-54:2009/A1:2015 |
| 4 | IEC 60601-1-6:2010/A1:2013 | 10 | ISO 14971:2007 |
| 5 | IEC 60825-1:2014<br>(recognized: 2007) | 11 | IEC 62366-1:2015/Cor1:2016 |
| 6 | IEC 62304:2015 | | |
# Table 4: FDA Guidance Documents
| FDA Guidance Documents and Effective Date | |
|-------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1. | Guidance for Industry and FDA Staff – User Fees and Refunds for Premarket<br>Notification Submissions 510(k)<br>Document issued on October 2, 2017 |
| 2. | Guidance for Industry and Food and Drug Administration Staff: Refuse to Accept<br>Policy for 510(k)s<br>Document issued on January 30, 2018 |
| 3. | Guidance for Industry and FDA Staff: Format for Traditional and Abbreviated<br>510(k)s - Guidance for Industry and FDA Staff<br>Document issued on August 12, 2005 |
| 4. | Guidance for Industry and FDA Staff: Deciding when to submit a 510(k) for a<br>change to an existing device.<br>Document issued on October 25, 2017 |
| 5. | Guidance for Industry and Food and Drug Administration Staff: The 510(k)<br>Program: Evaluating Substantial Equivalence in Premarket Notifications<br>[510(k)]<br>Document Issued on July 28, 2014 |
| 6 | Guidance for Industry and FDA Staff: Guidance for the Submission Of 510(k)'s<br>for Solid State X-ray Imaging Devices<br>Document issued on September 1, 2016 |
| 7. | Guidance for Industry and FDA Staff: Guidance for the Content of Premarket<br>Submission for Software in Medical Devices<br>Document issued on May 11, 2005 |
| 8. | Guidance for Industry and FDA Staff: Guidance for Off-The-Shelf Software Use in<br>Medical Devices |
{12}------------------------------------------------
Image /page/12/Picture/0 description: The image contains the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the words is a graphic of orange dots.
| FDA Guidance Documents and Effective Date | |
|-------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Document issued on September 9, 1999 |
| 9. | Guidance for Industry and FDA Staff: Applying Human Factors and Usability<br>Engineering to Medical Devices. |
| | Document issued February 3, 2016 |
| 10. | Guidance for Industry and FDA Staff: Pediatric Information for X-ray Imaging<br>Device Premarket Notifications. |
| | Document issued on November 28, 2017 |
| 11. | Guidance for Industry and FDA Staff: Content of Premarket Submissions for<br>Management of Cybersecurity in Medical devices. |
| | Document issued on October 2, 2014 |
| 12. | Guidance for Industry and FDA Staff: Radio Frequency Wireless Technology in<br>Medical Device |
| | Document issued on August 14, 2007. |
| 13. | Guidance for Industry and FDA Staff: Information to Support a Claim of<br>Electromagnetic Compatibility (EMC) of Electrically-Powered Medical Devices |
| | Document issued on July 11, 2016 |
# Verification and Validation:
Software Documentation for a Moderate Level of Concern software per FDA's Guidance Document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" issued on May 11, 2005 and "Off-The-Shelf Software Use in Medical Devices" is also included as part of this submission. The performance data demonstrates continued conformance with special controls for medical devices containing software. Non-clinical tests were conducted on the Subject Device Cios Spin software version VA30 during product development.
The Risk analysis was completed and risk control implemented to mitigate identified hazards. The testing results support that all the software specifications have met the acceptance criteria. Testing for verification and validation for the device was found acceptable to support the claims of substantial equivalence.
The Cios Spin software VA30 was tested and found to be safe and effective for intended users, uses and use environments through the design control verification and validation process. All new software functions present in the Subject Device (Metal Artifact Reduction, Retina 3D, Screw Scout, Target Pointer, High power 3D, and Easy 3D) have been validated through detailed software testing and it was founded they worked as intended. The Human Factor Usability Validation showed that Human factors are addressed in the system test according to the operator's manual and in clinical use tests with customer report and feedback form. Customer employees are adequately trained in the use of this equipment.
Siemens conforms to the cybersecurity requirements by implementing a process of preventing unauthorized access, modifications, misuse or denial of use, or the unauthorized use of information that is stored, accessed or transferred from a medical device to an external recipient. Provided in this
{13}------------------------------------------------
mission is a cybersecurity statement that considers IEC 80001-1:2010. The responsibility for compliance with IEC 80001-1-2010 is the hospital. Provided in the Software Section, is the required cybersecurity information.
Additional engineering bench testing was performed including: the nonclinical testing identified in the guidance for submission of 510(k) s for Solid State X-Ray Imaging Devices (SSXI); demonstration of system performance; and an imaging performance evaluation. X-ray Imaging Devices- Laboratory Image Quality and Dose Assessment, Tests and Standards were performed with acceptable results.
With regards to the flat panel detector (SSXI), test documentation provided in this submission demonstrates compliance of the Cios Spin to "Guidance for the Submission of 510(k)'s for Solid State X-ray Imaging Devices". The evaluation includes but is not limited to the following performance metrics identified in the SSXI guidance, showing comparable or better performance of the Cios Spin to its predicate and referenced devices. Provided in Table 5 for detector is equivalent comparative information.
| SSXI Metrics | Cios Spin Performance compared to Predicate Device and Reference<br>Devices | | | |
|---------------------------------------------|-----------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------|----------------------------------------------|
| | Subject Device<br>Cios Spin<br>(VA30) | Pr11edicate<br>Device Cios<br>Alpha (VA10)<br>K132094 | Reference<br>Device<br>Ziehm Vision<br>RFD<br>K132904 | Reference Device<br>Ziehm Solo FD<br>K161976 |
| Imaging<br>Modes | Pulsed<br>fluoroscopy<br><br>Single images | Pulsed<br>fluoroscopy<br><br>Single images | Pulsed<br>Fluoroscopy<br><br>Digital Spot | Pulsed<br>Fluoroscopy<br><br>Digital Spot |
| DQE | 72% | 76% | Information Not<br>Available | 70% |
| Dynamic<br>Range | 96dB | 94dB | Information Not<br>Available | Equivalent |
| Modulation<br>Transfer<br>Function<br>(MTF) | 58% at 1 lpmm<br>large) | 55% at 1 Lp/mm | Information Not<br>Available | 4lp/mm |
| Digitization<br>depth | 16 bit | 16 bit | 16 bit | 16 bit |
| Pixel Pitch | 152 μm | 194μm | 194 μm | 100 μm |
| Field of View | FPD:<br>* 30 cm x 30 cm<br>* 20 cm x 20 cm | Small FPD:<br>* 20 cm x 20 cm<br>* 15 cm x 15 cm<br>* 10 cm x 10 cm<br>Large FPD:<br>* 30 cm x 30 cm<br>* 20 cm x 20 cm | FPD 20 cm:<br>* 20 cm<br>* 15 cm<br>* 10 cm | FPD 20 cm:<br>* 20 cm<br>* 15 cm<br>* 10 cm |
# Table 5: SSXI Metrics
{14}------------------------------------------------
Image /page/14/Picture/0 description: The image contains the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange. To the right of the words is a graphic of orange dots arranged in a circular pattern.
ause Cios Spin' new flat panel detector does not change the system's intended use and represent equivalent technological characteristics, clinical images are not required. Also, Siemens has demonstrated that clinical images are not needed to demonstrate that the device is as safe and as effective as the predicate and reference devices.
# Summarv:
Performance tests were conduct to test the functionality of the Cios Spin (VA30). These tests have been performed to assess the functionality of the Subject Device. Results of all conducted testing and clinical assessments were found acceptable and do not raise any new issues of safety or effectiveness.
## 10. General Safety and Effectiveness Concerns:
Instructions for use are included within the device labeling, and the information provided will enable the user to operate the device in a safe and effective manner. Several safety features including visual and audible warnings are incorporated into the system design. In addition the Cios Spin is continually monitored and if an error occurs, the system functions will be blocked and an error message will be displayed.
Risk management is ensured via a hazard analysis, which is used to identify potential hazards. These potential hazards are controlled via software development, verification and validation testing. To minimize electrical and mechanical hazards. Siemens adheres to recognized and established industry practice, and all equipment is subject to final performance testing. Furthermore the operators are health care professionals familiar with and responsible for the evaluating and post processing of X-ray images.
## 11. Conclusion as to Substantial Equivalence:
The predicate devices were also cleared based on the non-clinical supportive information and clinical images and data per Guidance documents. Similar non-clinical test results demonstrate that the Cios Spin software version VA30 acceptance criteria are adequate for the intended use of the device. The comparison of technological characteristics, non-clinical performance data and software validation data demonstrates that the Subject Devices are as safe and effective when compared to the Predicate Devices that is currently marketed for the same intended use.