The Cios Alpha is a mobile X-Ray system designed to provide X-ray imaging of the anatomical structures of patient during clinical applications. Clinical applications may include but are not limited to: interventional fluoroscopic, gastro-intestinal, endoscopic, urologic, pain management, orthopedic, neurologic, vascular, cardiac, critical care and emergency room procedures. The patient population may include pediatric patients.
Device Story
Cios Alpha is a mobile fluoroscopy C-arm system for surgical environments. It utilizes an X-ray source and a solid-state flat panel detector to acquire images of anatomical structures. The system includes a monitor trolley for image processing, review, and patient data entry. Operated by healthcare professionals, the device provides real-time fluoroscopic imaging to support orthopedic, vascular, and other interventional procedures. The system features pulsed fluoroscopy to reduce radiation dose and supports high-resolution image matrices (up to 1536x1536). Output is displayed on TFT flat-screen monitors, assisting clinicians in visualizing anatomy during procedures, thereby facilitating surgical guidance and clinical decision-making.
Clinical Evidence
No formal clinical trial was performed. A customer use test was conducted over 8 weeks with >70 patients. Image quality was evaluated across procedures including carotids, peripheral vascular, and EVAR (AAA/TAA). Results indicated good to very good image quality, spatial resolution, and contrast, with investigators noting improved detail compared to the predicate ARCADIS Avantic.
Technological Characteristics
Mobile C-arm fluoroscopy system; solid-state detector (20x20cm or 30x30cm); high-frequency generator (40-125 kV); 1536x1536 pixel matrix; 19" TFT flat-screen displays; pulsed fluoroscopy; DICOM compatible; GUI-based operation. Complies with IEC 60601-1, IEC 62304, ISO 14971, and 21 CFR 1020.30-32.
Indications for Use
Indicated for X-ray imaging of anatomical structures in patients, including pediatric patients, undergoing interventional fluoroscopic, gastro-intestinal, endoscopic, urologic, pain management, orthopedic, neurologic, vascular, cardiac, critical care, and emergency room procedures.
Regulatory Classification
Identification
An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
SIEMENS
K132094 Cios Alpha: Al / Hold Response
# 510(k) Summary: Cios Alpha
Company: Siemens Medical Systems, Inc. 51 Valley Stream Parkway Malvern, PA 19355
This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR § 807.92.
#### 1. General Information:
Importer / Distributor:
Siemens Medical Systems, Inc. 51 Valley Stream Parkway Malvern, PA 19355
## Establishment Registration Number:
2240869
Manufacturing Site:
SIEMENS AG Sector Healthcare Röntgenstrasse 19 - 21 D-95478 Kemnath, Germany
# Establishment Registration Number:
3002466018
#### Contact Person: 2.
Mr. Darren Dorman Regulatory Affairs Specialist Siemens Medical Solutions USA, Inc. 51 Valley Stream Pkwy · Mail Code D02 · Malvern, PA 19355, USA Phone: +1 610 219-9066 · Fax: +1 610 448-1787 Email: darren.dorman@siemens.com
#### 3. Device Name and Classification:
Trade Name: Cios Alpha Device: Interventional fluoroscopic x-ray system; image-intensified fluoroscopic x-ray system, mobile Regulation Description: Image-intensified fluoroscopic x-ray system; Regulation: Radiology
{1}------------------------------------------------
# STEMENS
Review Panel: Radiology OWB, OXO Product Code: Submission Type: Traditional 510(k) Regulation Number: 892.1650 Device Class: 2
#### 4. Legally Marketed Predicate Device:
Trade Name: ARCADIS Avantic image-intensified fluoroscopic x-ray system, mobile Device: Image-intensified fluoroscopic x-ray system. Regulation Description: Regulation: Radiology Review Panel: Radiology Product Code: ОХО Submission Type: Traditional 510(k) Regulation Number: 892.1650 Device Class: 2
#### ട്. Device Description:
The Cios Alpha is a mobile fluoroscopy system designed for the surgical environment. The Cios Alpha provides comprehensive image acquisition modes to support orthopedic and vascular procedures. The system consists of two major components:
- a) The C-arm with X-ray source on one side and the flat panel detector on the opposite side. The c-arm can be angulated in both planes and be lifted vertically, shifted to the side and moved forward/backward by an operator.
- b) The monitor trolley providing image processing, review and patient data entry. The monitor trolley may contain an optional hardcopy (paper) printer and navigational equipment as well.
#### 6. Indication for Use:
The Cios Alpha is a mobile X-Ray system designed to provide X-ray imaging of the anatomical structures of patient during clinical applications. Clinical applications may include but are not limited to: interventional fluoroscopic, gastro-intestinal, endoscopic, urologic, pain management, orthopedic, neurologic, vascular, cardiac, critical care and emergency room procedures. The patient population may include pediatric patients.
#### 7. Substantial Equivalence:
The Cios Alpha is substantially equivalent to the commercially available Siemens ARCADIS Avantic, cleared 06/01/2005 with K051133.
The Indication for use statement is similar to the predicate. The image intensifier has been replaced by a solid state detector. X-ray generation and control used with the
{2}------------------------------------------------
SIEMENS
Cios Alpha is similar to the technology used with the predicate ARCADIS Avantic. A new GUI (graphical user interface) supports the user in the operation of the system.
#### 8. Summary of Technological Characteristics of the Subject Device as Compared with the Predicate Device:
The Cios Alpha design is based on the ARCADIS Avantic and the experience Siemens has with more than 30 years with predicate mobile fluoroscopy C-arms (SIREMOBILE series). The advent of solid state detectors created a need to use this technology for mobile fluoroscopic systems too. The detector replaces the image intensifier and the TV camera (1k x 1k pixel matrix). The imaging system was upgraded to support the processing of up to 1536 pixel 16bit image matrix. Full DICOM compatibility and a new graphical user interface provide networking and an intuitive operation. X-ray generation is similar to the predicate with a 25% increase in output power. Pulsed fluoroscopy, similar to the predicate, helps to reduce dose exposure for long procedures.
The following table compares the dominant performance data of the subject device with the predicate device to substantiate equivalence of both devices.
| Feature | Subject Device<br>Cios Alpha | Predicate Device<br>ARCADIS Avantic |
|-----------------------------------------------------------|-------------------------------------------------------|----------------------------------------------------------------|
| Mobile Fluoroscopy C-arm | Yes | Yes |
| Tube housing assembly<br>with high frequency<br>generator | Yes | Yes |
| kV range: 40 kV - 125 kV | Yes | Yes |
| Max power output | 12 kV<br>25 kW (optional) | 20 kV |
| Pulsed Fluoroscopy | 3 mA to 120 mA (12kW)<br>3 mA to 250 mA (25kW) | Up to 70 mA |
| X-ray detector | Solid State Detector<br>20cm x 20cm or<br>30cm x 30cm | Image Intensifier with<br>Optics and TV System 13"<br>diameter |
| Dose measurement device | Yes | Yes |
| Matrix size | 1536 x 1536 | 1024 x 1024 |
| Monitors | 19" TFT Flat Screen Display<br>Panels, B/W or Color | 18" TFT Flat Screen<br>Display Panels, B/W or<br>Color |
| Navigation interface | Yes | Yes |
| Image post-processing | Yes | Yes |
| DICOM Functionality | Yes | Yes |
#### 9. Summary of Non-Clinical Tests:
{3}------------------------------------------------
The Siemens Cios Alpha complies with the following voluntary standards:
| IEC 60601-1 | 60601-2-28 |
|-----------------------------------------------------------------|---------------------------|
| IEC 60601-1-1 | 60601-2-32 |
| IEC 60601-1-3 | 60601-2-54 |
| IEC 62366 | 61910-1 |
| ISO 14971 | 60601-2-43 |
| IEC 62304 | IEC 60825-1 |
| IEC 60601-1-2 | IEC 10993-1 |
| 60601-2-7 | NEMA PS 3.1 - 3.20 (2011) |
| X-Ray performance tests outlined in: 21 CFR 1020.30, 31, and 32 | |
The following quality assurance measures were applied to the development of the system:
- Risk Analysis .
- ◆ Requirement Specification Reviews
- . Design Reviews
- Integration testing (System verification) .
Non-clinical testing was conducted in accordance with Guidance for Submission of 510(k)s for Solid State X-Ray Imaging Devices (issued August 6, 1999). All test results were satisfied.
#### 11. Summary of Clinical Tests:
For the subject of this premarket submission, Siemens did not perform a formal clinical trial but did conduct a customer use test. In this test the Cios Alpha has been evaluated in a clinical environment for 8 weeks. More than 70 patients have been examined. The following table provides a summary of the image quality outcome with different procedures:
| | IQ over<br>all | Spatial<br>resolution | Motion<br>resolution | Contrast | Noise |
|------------------------|----------------|-----------------------|----------------------|------------|------------|
| Carotids sub | Good | Good | n.a. | Acceptable | Acceptable |
| Vascular<br>peripheral | Very good | Very good | n.a. | Good | Low |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows a handwritten note with the text "K132094" on the top line and "Page 5 of 5" on the second line. The handwriting is in black ink and appears to be somewhat cursive. The note seems to be part of a larger document or set of pages, indicating it is page 5 out of 5.
C. HINS & SONS
K132094 Cios Alpha: Al / Hold Response
| EVAR (AAA) | Very good | Very good | Very good | Good | Acceptable |
|------------|-----------|-----------|-----------|------|------------|
| EVAR (TAA) | Very good | Very good | Very good | Good | Acceptable |
It was the investigators opinion that in this test phase the new Cios Alpha provided an even better, more detailed image quality as compared to the ARCADIS Avantic (predicate device). This includes the aortic procedures as well as the peripheral vascular structure representation i.e. the vascular representation of the foot. The complete clinical evaluation is included in the attachment.
#### 12. General Safety and Effectiveness Concerns:
Instructions for use are included within the device labeling, and the information provided will enable the user to operate the device in a safe and effective manner. Several safety features including visual and audible wamings are incorporated into the system design. In addition the Cios Alpha is continually monitored, and if an error occurs, the system functions will be blocked and an error message will be displayed.
Furthermore the operators are health care professionals familiar with and responsible for the X-ray examinations to be performed. To minimize electrical, mechanical and radiation hazards, Siemens adheres to recognized and established industry practice, and all equipment is subject to final performance testing.
#### Conclusion as to Substantial Equivalence: 13.
The Cios Alpha is intended for similar indications for use as the predicate ARCADIS Avantic. The imaging properties (X-ray source, output power and field of view) are similar. The most significant difference is the use of a solid state detector instead of an image intensifier. It is Siemens opinion, that the Cios Alpha is substantially equivalent to the ARCADIS Avantic (K051133, cleared 06/01/2005).
{5}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles an abstract human figure or a flame-like design. The logo is black and white.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 11, 2014
Siemens Medical Solutions USA, Inc. % Mr. Darren Dorman Regulatory Affairs Specialist 51 Valley Stream Parkway MALVERN PA 19355
Re: K132094
Trade/Device Name: Cios Alpha Regulation Number: 21 CFR 892.1650 Regulation Name: Intensified Fluoroscopic X-Ray System Regulatory Class: II Product Code: OWB, OXO Dated: February 7, 2014 Received: February 11, 2014
Dear Mr. Dorman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events} (21 CFR 803); good manufacturing practice requirements as set
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{6}------------------------------------------------
Page 2 - Mr. Darren Dorman
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers. International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.lda.gov/MedicalDevices/ResourcesforYou/Industrv/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Michael D. O'Hara
ਿ၊
Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows a blurry, low-resolution picture with many small, indistinct shapes scattered across the frame. The shapes appear to be concentrated more towards the top and left side of the image. The overall impression is that of a grainy or noisy visual with no clear subject or focus.
## K132094 Clos Alpha: Al / Hold Response
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page.
Indications for Use
510(k) Number (il known) KI 37094
Device Name Cios Alpha
Indicalions for Use (Describe)
The Cios Alpha is a mobile X-Ray system designed to provide of the annonical structures of panent during chuical applications. Clinical applications may included to: interventional final fineroscopic. gastro-intestinal. endoscopic. urologic, pain management, orthopedic, vascular, cardine, chical care and emergency room procedures. The patient population may include pediatric parients.
Type of Use (Select one or both, as applicable)
[8] Prescription Use (Part 21 CFR 80 | Subpan D)
Over The Counter Use (2) CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
Michael D. O'Hara
FORM FDA 3881 (9/13)
Page 1 of 2
*N.; 2012. Nature Reviews ; 16(1): 44-48*
Siemens Medical Solutions, Inc.
Cios Alpha K132094 Confidential
38 of 1564
Siemens Medical Solutions, Inc.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.