The angiographic X-ray systems are indicated for use for patients from newborn to geriatric in generating fluoroscopic and rotational images of human anatomy for cardiovascular, vascular and non-vascular, diagnostic and interventional procedures. Additionally, with the OR table, the angiographic X-ray systems are indicated for use in generating fluoroscopic and rotational images of human anatomy for image-guided surgical procedures. The OR table is suitable for interventional and surgical procedures.
Device Story
Interventional X-ray systems (Innova IGS 5/6, Discovery IGS 7/7 OR) perform monoplane/biplane fluoroscopic imaging for minimally invasive procedures. System components include C-arm positioner, X-ray tube, power unit, digital detector, and image processing unit. Discovery IGS 7 OR integrates with third-party neurosurgical head holders (skull clamps) and OR tables. Device modifications include: 1) Enhanced collision avoidance software allowing operators to define guard volumes around skull clamps/patient head to prevent gantry/table collisions during movement; 2) Optional wireless footswitch (2.4 GHz proprietary protocol) for X-ray control and table brake release. Used in hybrid ORs or interventional suites by physicians/surgeons. Output is real-time fluoroscopic/rotational imagery used for procedural guidance. Benefits include reduced cable clutter and improved safety during complex neurosurgical/trauma procedures via collision mitigation.
Clinical Evidence
No clinical studies were required. Substantial equivalence supported by bench testing, simulated use testing, and usability validation. Verification confirmed design outputs met inputs for collision avoidance and wireless functionality. Safety testing included EMC and coexistence testing per FDA wireless guidance and conformance to IEC 60601-2-43, IEC 62304, and IEC 62366.
Technological Characteristics
System includes C-arm positioner, digital detector (up to 41cm), X-ray tube, and power unit. Connectivity: Wired or optional 2.4 GHz wireless footswitch. Software: Collision avoidance model with operator-defined guard volumes. Standards: 21 CFR 1020.30/32, IEC 60601-2-43:2010, IEC 62304 (2006), IEC 62366 (2007).
Indications for Use
Indicated for patients newborn to geriatric requiring fluoroscopic and rotational imaging for cardiovascular, vascular, non-vascular, diagnostic, interventional, and image-guided surgical procedures.
Regulatory Classification
Identification
An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
GE Innova/Innova IGS/Discovery IGS/Optima angiographic, fluoroscopic X-ray systems (K122457)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. The logos are simple and professional in appearance.
November 2, 2018
GE Healthcare Michel Genuer Senior Regulatory Affairs Program Manager 283 Rue De La Miniere 78530 Buc, France
Re: K181403
Trade/Device Name: Innova IGS 5, Innova IGS 6, Discovery IGS 7, Discovery IGS 7 OR Regulation Number: 21 CFR 892.1650 Regulation Name: Image-Intensified Fluoroscopic X-Ray System Regulatory Class: Class II Product Codes: OWB, JAA, IZI, OXO Dated: May 17, 2018 Received: May 29, 2018
Dear Michel Genuer:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn
(http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Michael D. O'Hara
For
Robert A. Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
#### 510(k) Number (if known) K181403
Device Name
Innova™ IGS 5, Innova™ IGS 6, Discovery™ IGS 7, Discovery™ IGS 7 OR
Indications for Use (Describe)
The angiographic X-ray systems are indicated for use for patients from newborn to geriative in generating fluoroscopic and rotational images of human anatomy for cardiovascular, vascular, diagnostic and interventional procedures.
Additionally, with the OR table, the angiographic X-ray systems are indicated for use in generating fluoroscopic and rotational images of human anatomy for image-guided surgical procedures.
The OR table is suitable for interventional and surgical procedures.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------------|--|
| <div> <span> <b> </b> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | |
| <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in white inside. The letters are stylized and connected, and the circle has a decorative border. The logo is simple and recognizable, and it is associated with a well-known company.
### 510(k) Summary
In accordance with 21 CFR 807.92 the following summary of information is provided:
| Date: | 31-October-2018 |
|----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | GE MEDICAL SYSTEMS SCS<br>283 RUE DE LA MINIERE<br>78530 BUC, FRANCE |
| Primary Contact Person: | Michel GENUER<br>Senior Regulatory Affairs Program Manager<br>GE Healthcare, (GE MEDICAL SYSTEMS SCS)<br>Tel: (+33)-1-3070-4741<br>Email: michel.genuer@ge.com |
| Secondary Contact Persons: | GE Healthcare, (GE MEDICAL SYSTEMS SCS)<br>Philip Malca<br>Regulatory Affairs Director<br>Tel: (+33)-6 4637 3852<br>Email: philip.malca@ge.com<br>Maxime CORNIERE<br>Regulatory Affairs Leader<br>Tel: (+33)-1-3070-4153<br>Email: maxime.corniere@ge.com |
| Device Trade Name: | Innova™ IGS 5, Innova™ IGS 6, Discovery™ IGS 7, Discovery™<br>IGS 7 OR |
| Common/Usual Name: | interventional fluoroscopic x-ray system, angiographic x-ray system |
| Regulation description: | Image-intensified fluoroscopic x-ray system |
| Regulation Number: | 21CFR 892.1650 |
| Product Code: | OWB |
| Subsequent Product Codes | JAA, IZI and OXO for Discovery™ IGS 7 and Discovery™ IGS 7 OR. |
| Classification: | Class II |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for General Electric (GE). The logo consists of the letters 'GE' in a stylized script, enclosed within a blue circle. There are decorative swirls around the perimeter of the circle, adding a touch of elegance to the design.
| Predicate Device(s): | | Trade Name: | GE Innova/Innova IGS/Discovery IGS/Optima angiographic, fluoroscopic X-ray systems with Cathlab Frontiers solutions. |
|----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------|----------------------------------------------------------------------------------------------------------------------|
| | | 510(k) Clearance: | K122457 |
| | | Regulation Name: | Image-Intensified fluoroscopic x-ray system |
| | | Regulation Number: | 21CFR 892.1650 |
| | | Classification: | Class II |
| | | Product Code: | Primary product code: OWB<br>Secondary Product codes: IZI, JAA |
| | | Trade Name: | Discovery IGS TM 740 |
| | | 510(k) Clearance: | K133278 |
| | | Regulation Name: | Image-Intensified fluoroscopic x-ray system |
| | | Regulation Number: | 21CFR 892.1650 |
| | | Classification: | Class II |
| | | Product Code: | Primary product code: OWB<br>Secondary Product codes: JAA, OXO |
| Device Description: | GE Healthcare IGS interventional x-ray systems are designed to perform monoplane, biplane fluoroscopic x-ray examinations to provide the imaging information needed to perform minimally invasive interventional X-Ray imaging procedures (standard configuration). Additionally, in the OR configuration (with an OR Table), these systems allow to perform surgery and X-Ray image guided surgical procedures in a hybrid Operating Room.<br>DiscoveryTM IGS 7 OR, DiscoveryTM IGS 7, InnovaTM IGS 6, InnovaTM IGS 5 are the GE Healthcare IGS interventional X-Ray system product models.<br>Each product model is designed with a set of components that are combined into different configurations for providing specialized interventional x-ray systems. GE Healthcare IGS interventional x-ray system consists of a C-arm positioner (monoplane or biplane), an x-ray imaging table or the interface to the radiologic table, an x-ray tube assembly (per plane), an x-ray power unit with its exposure control unit (per plane), an x-ray imaging chain (including a digital detector and an image processing unit, per plane).<br>DiscoveryTM IGS 7 is a monoplane system (C-arm with mobile AGV gantry) and is proposed in IGS 740 configuration with a square 41cm digital detector or in IGS 730 configuration with a square 31cm digital detector. These product configurations are declined in sub configurations: Standard or OR configuration. The Innova-IQ table in the OR configuration is the GE OR table | | |
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' intertwined in a stylized script, enclosed within a blue circle. The circle is surrounded by a series of white, stylized water droplets or flourishes, giving the logo a distinctive and recognizable appearance.
| Discovery™ IGS 7 OR is a monoplane system (C-arm positioner with mobile AGV gantry) and is proposed with a square 41cm digital detector or with a square 31cm digital detector configuration. This is the product model compatible with a qualified configuration of the Maquet Magnus OR table system. This product model is provided with a table integration kit. The Magnus OR table configuration compatible with Discovery™ IGS 7 OR includes a flat table top configuration for Interventional X-Ray imaging and surgery procedures, an optional universal table top (table top with articulated joints) to enable expansion to surgical procedures requiring advanced patient positioning and with X-Ray imaging capabilities. A set of Magnus OR table accessories is included in the compatible configuration. |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Innova™ IGS 6 is a biplane system (a dual C-arm positioner, LC and LP gantry) and is provided with an Omega V table. Innova™ IGS 6 is proposed in IGS 630 configuration with a square 31cm digital detector per plane or in IGS 620 configuration with a square 20.5cm digital detector per plane. These product configurations are available in Standard configuration. |
| Innova™ IGS 5 is a monoplane system (C-arm positioner with LC gantry) and is proposed with a square digital detector of 41cm or 31cm or 20.5cm. These product configurations are available in Standard configuration with Innova-IQ or Omega V table or Omega IV table (IGS 520 configuration only); or in OR configuration with GE OR table. |
| The purpose of this Premarket Notification is for 2 changes:<br>First, a change of the collision management of Discovery™ IGS 7 OR to enable the use of a Neurosurgical head holder (skull clamp). During trauma neck surgery and stereotactic navigation, a gantry collision with the skull clamp could result in patient harm, delayed or halted surgical procedures. The subject of this submission is to introduce a functionality for adapting the collision avoidance model to the geometrical characteristic of the skull clamp and patient position with a guard volume defined and validated by the operator during the procedure preparation phase. During the procedure, the collision avoidance software will slow down and stop gantry and/or table axis to minimize risk of collision of a motorized moving part with the patient head or the skull clamp. In this configuration, the OR table, the Neurosurgical head holder and its support are from third party Medical device manufacturers. As a consequence, there is no mechanical interfaces between these devices and the Discovery™ 7 OR. Those devices are registered and cleared in the US. The Neurosurgical head holder is from Pro Med Instrument manufacturer (K063494). The OR table is the registered Magnus OR |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' intertwined within a circle. The circle and the letters are blue, and there are white flourishes around the circle.
| | The second change is for a primary technology change, adding an<br>optional wireless footswitch in the examination room to Innova™<br>IGS 5, Innova™ IGS 6, Discovery™ IGS 7 and Discovery™ IGS 7 OR. It<br>provides identical functionalities as the wired footswitch for the<br>control of X-ray on and off, and for the control of the table top brake<br>release.<br>Labelling is updated for the enhanced collision management for the<br>use with Neurosurgical head holder and for the wireless footswitch. |
|----------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use: | The angiographic X-ray systems are indicated for use for patients<br>from newborn to geriatric in generating fluoroscopic and rotational<br>images of human anatomy for cardiovascular, vascular and non-<br>vascular, diagnostic and interventional procedures.<br>Additionally, with the OR table, the angiographic X-ray systems are<br>indicated for use in generating fluoroscopic and rotational images<br>of human anatomy for image-guided surgical procedures. |
| | The OR table is suitable for interventional and surgical procedures. |
| Indication for Use: | The angiographic X-ray systems are indicated for use for patients<br>from newborn to geriatric in generating fluoroscopic and rotational<br>images of human anatomy for cardiovascular, vascular and non-<br>vascular, diagnostic and interventional procedures.<br>Additionally, with the OR table, the angiographic X-ray systems are<br>indicated for use in generating fluoroscopic and rotational images<br>of human anatomy for image-guided surgical procedures.<br>The OR table is suitable for interventional and surgical procedures. |
| Technology: | The modified GE Healthcare IGS Interventional X-Ray systems<br>employ the same fundamental scientific technology as the<br>unmodified predicate devices.<br>The change to enable the dynamic adaptation of the collision<br>avoidance model with a model fitted to the needs of the operator<br>and of the procedure, is a software change. The collision avoidance<br>software is the same as with the predicate devices. The change is<br>the dynamic adaptation of the collision avoidance model. All causes<br>of hazard relative to the use of a Neurosurgical head holder in<br>combination with the Discovery™ IGS 7 OR have been identified and<br>mitigated.<br>The optional wireless footswitch consists of a footswitch containing<br>a transmitter system, a receiver that is installed near the connection<br>plug of the wired footswitch (at the table base of the Innova-IQ<br>table, of the Omega table or at the Discovery Control Center table<br>side cart), and a charger. The introduction of this wireless footswitch<br>does not change existing electrical or software interface of the<br>predicate devices. The new option is equivalent to the wired<br>footswitch on the predicate in that the functionalities are identical.<br>The communication between the footswitch emitter and the |
| | receiver is performed via a proprietary protocol using 2.4 GHz radio<br>frequency technology.<br>The wireless option provides placement flexibility and reduces cable<br>clutter.<br>Both changes have no impact on x-ray imaging performance.<br>The intended use and indications for use remain unchanged from<br>those of the predicate devices. The expansion to x-ray<br>interventional, minimally invasive surgery and conventional open<br>surgery procedures requiring the use of a neurosurgical holder in<br>combination with Discovery™ IGS 7 OR is included in Discovery IGS<br>7 OR's indications for use of interventional, image-guided surgical<br>and surgical procedures. |
| Determination of Substantial<br>Equivalence: | The modifications to Discovery™ IGS 7 OR collision management<br>software and to IGS interventional x-ray systems for the wireless<br>footswitch were developed under the GE Healthcare's design<br>controls processes and overall quality management system. |
| | IGS interventional x-ray systems conform to 21CFR 1020.30 and 32,<br>and with voluntary standards IEC 60601-2-43:2010, IEC 62304<br>(2006) and IEC 62366 (2007). |
| | Risk management activities using risk analysis to identify any<br>potential issues incorporating the wireless footswitch and the use of<br>a skull clamp were performed. These issues were assessed and<br>mitigated. |
| | The following testing was used to assess safety and effectiveness<br>and, thus, to establish the substantial equivalence with the<br>predicate devices. |
| | Summary of Non-Clinical Tests:<br>The following quality management measures were applied to the<br>development of these modifications:<br>Simulated Use Testing ensured the system conforms to user<br>needs and intended uses through simulated clinical<br>workflow using step-by step procedures that would be<br>performed for representative clinical applications. Usability validation testing was conducted to confirm that<br>the product can be used safely and effectively. Participants<br>were representatives of actual users, with knowledge of the<br>customer needs and clinical applications. Product verification ensured the system conforms to its<br>requirements including hazard mitigations risk<br>management requirements. The verification tests<br>confirmed that design output meet design input<br>requirements. Tests were executed at component, software<br>subsystems and system levels. |
| | Safety testing were performed to confirm that the product<br>meets the requirements of the aforementioned standards.<br>It includes the EMC and coexistence testing per the "Radio<br>Frequency Wireless Technology in Medical Devices"<br>Guidance for Industry and Food and Drug Administration<br>Staff issued on August 14, 2013 (Wireless Guidance). |
| | Summary of Clinical Tests:<br>The subject of this premarket submission, change to Discovery™ IGS<br>7 OR collision management to enable the use of a neurosurgical<br>head holder and the change to add an optional wireless footswitch<br>in the examination room to GE Healthcare interventional x-ray<br>systems did not require clinical studies to support substantial<br>equivalence. The expansion to x-ray and surgical procedures<br>requiring the use of a neurosurgical holder in combination with<br>Discovery™ IGS 7 OR is included in Discovery IGS 7 OR's indications<br>for use of image-guided surgical and surgical procedures. It does not<br>introduce new indications for use. Substantial equivalence relies on<br>clinical information that is pre-existing on the cleared predicate<br>devices.<br>Design verification and validation testing were performed to<br>confirm that the safety and effectiveness of the devices has not |
| | been affected. The test plans and results have been executed with<br>acceptable results. |
| Conclusion: | GE Healthcare considers that the Discovery™ IGS 7 OR incorporating<br>the enhanced collision management for the use of a skull clamp or<br>that a GE Healthcare IGS interventional x-ray system incorporating<br>the optional wireless footswitch to be as safe and as effective and<br>substantially equivalent to the predicate devices.<br>This conclusion is based on the fact that Discovery™ IGS 7 OR<br>incorporating the collision management optimization for the use in<br>combination with the Neurosurgical head holder and that GE<br>Healthcare interventional x-ray systems incorporating the optional<br>wireless footswitch have: the same indications for use as their predicate devices. The modified devices have the same technological<br>characteristics as their predicate devices. Verification and Validation testing has demonstrated that<br>the design inputs, user requirements, and risk mitigations<br>have been met. Engineering bench testing per the FDA guidance Radio<br>frequency wireless technology in medical devices as well as<br>conformance to IEC standards and guidance documents<br>were provided. |
| | The successful completion of the above testing was sufficient to assess safety and effectiveness and, thus, to establish the substantial equivalence. |
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in a stylized font inside. The letters are white and have a swirling design around them. The logo is simple and recognizable, and it is often used to represent the company.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for GE Healthcare. The text is black and in a sans-serif font. The words "GE Healthcare" are on one line.
Image /page/8/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" intertwined in a stylized script, enclosed within a blue circle. The letters are also blue, matching the color of the circle.
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined in a stylized, cursive font. The letters are enclosed within a circular border, and the entire logo is presented in a blue color scheme.