K110040 · Modern Module, Inc. · LLZ · Jan 26, 2011 · Radiology
Device Facts
Record ID
K110040
Device Name
OMNIVISION
Applicant
Modern Module, Inc.
Product Code
LLZ · Radiology
Decision Date
Jan 26, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device, 3rd-Party Reviewed
Indications for Use
OmniVision is intended to be used for the acquisition, storage, communication and viewing of medical Images.OmniVision receives images from imaging modalities via DICOM or imported It archives and displays these images for the use of medical specialists who are qualified directly. to operate radiological equipment and to record and diagnose medical images. It provides the user with a range of tools to assist them in viewing the images, such as zoom, filters and measurements, and with facilities to exchange images with other specialists. OmniVision is not intended for the acquisition of mammographic image data.
Device Story
OmniVision is a Picture Archiving and Communications System (PACS) for acquisition, storage, communication, and viewing of medical images; receives inputs from imaging modalities via DICOM or direct import; modular architecture includes viewer, database management, and printing modules; used by radiologists and orthopaedists in clinical settings; provides tools for window leveling, zoom, filters, and measurements; facilitates image exchange between specialists; does not control life-sustaining devices; requires qualified medical personnel for interpretation; benefits patient through efficient image management and diagnostic support.
Clinical Evidence
Bench testing only. Device performance verified against specifications and DICOM conformance standards. No clinical data provided.
Technological Characteristics
Autonomous software PACS; no hardware. Runs on standard PC hardware (e.g., HP, Dell, IBM) with MS Windows XP. Modular architecture; DICOM compliant. Features include image acquisition, storage, communication, and viewing tools (zoom, filters, measurements).
Indications for Use
Indicated for medical specialists qualified to operate radiological equipment to record and diagnose medical images. Not intended for acquisition of mammographic image data.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
510(k) Premarket Notification Modern Module Inc. - OminiVision March 30th, 2010
JAN 2 6 2011
Image /page/0/Picture/2 description: The image shows the text "KI10040" at the top, followed by a logo that says "Modern Module". The logo has a stylized "M" with the rest of the text in a simple font. The text is black and the background is white.
# 510 (K) Summary
Summary of functions of the device and its major components provided as part of the Premarket Notification for OmniVision
ﮧ ..
| Date : | 30 March 2010 |
|-----------------|------------------------------------------------------------------------------------------------|
| Company Name : | Modern Module Inc. |
| Address: | C Bldg., Suite # 809 Woolim Lions Valley,<br>425 CheongChun-dong, Bupyung-ku<br>Incheon, Korea |
| Contact Person: | Samuel Koh / CEO |
| E-Mail : | samuelkoh@e-mmi.com |
| Tel.: | +82-32-623-6721 |
| Fax.: | +82-32-623-6720 |
| Device Trade Name: | OmniVision |
|-------------------------|----------------------------------------------------|
| Device Common Name : | Picture Archiving and Communications System (PACS) |
| Product Code : | LLZ |
| Regulation Number | 892.2050 |
| Device Classification : | Class II |
| Predicate Devices | dicomPACS |
| Predicate Devices | | Items | Predicate Device : | Subject Device | Remark |
|-----------------------|----------------------------------------------------|---------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| Trade Name | dicomPACS | 510(k) No. | K070618 | | |
| Common Name | Picture Archiving and Communications System (PACS) | Trade Name | dicomPACS | OmniVision | |
| Regulation Number | 892.2050 | | dicomPACS® is a software<br>system for the administration,<br>archiving, improvement and<br>compression of medical image<br>data for diagnosis. The images<br>are either acquired from imaging<br>modalities via DICOM or imported<br>directly. All images are archived<br>in a database as DICOM<br>compliant files. The data is<br>displayed on a computer monitor | OmniVision is intended to be<br>used for the acquisition, storage,<br>communication and viewing of<br>medical Images.<br>OmniVision receives images from<br>imaging modalities via DICOM or<br>imported directly. It archives<br>and displays these images for the<br>use of medical specialists who<br>are qualified to operate | |
| Device Classification | Class II | Indications for use | for diagnosis.<br>dicomPACS® also provides<br>services for administering the<br>data.<br>Lossy compressed<br>mammographic images and<br>digitized film screen images must<br>not be reviewed for primary image<br>interpretations. Mammographic<br>images may only be using an FDA<br>approved monitor that offers at<br>least 5 MP resolution and meets<br>other technical specifications<br>reviewed and accepted by FDA.<br>Functions to be carried out using<br>dicomPACS® are, for example,<br>but not limited to, adjustment of<br>window leveling, rotation, zoom, | radiological equipment and to<br>record and diagnose medical<br>images.<br>It provides the user with a range<br>of tools to assist them in viewing<br>the images, such as zoom, filters<br>and measurements, and with<br>facilities to exchange images with<br>other specialists.<br>OmniVision is not intended for the<br>acquisition of mammographic<br>image data. | |
| Product code | LLZ | | | | |
| Manufacturer | Oehm und Rehbein GmbH | | | | |
| 510(k) number | K070618 | | | | |
: : :
{1}------------------------------------------------
, .
Image /page/1/Picture/1 description: The image shows the words "Modern Module" in a stylized font. The word "Modern" is on the top line, and the word "Module" is on the bottom line. The first letter of "Modern" is stylized with two thick vertical lines.
Image /page/1/Picture/2 description: The image shows a grainy black and white picture. There are some numbers in the image, including 1, 315, and 23. The image is blurry and difficult to make out any specific details.
# Comparison between the Predicate Device and the OmniVlsion
:"
{2}------------------------------------------------
### 510(k) Premarket Notification Modern Module Inc. - OminiVision
| | and measurements.<br>dicomPACS® is meant to be<br>used by qualified medical<br>personnel only. All users must be<br>qualified to create and diagnose<br>radiological image data. | | |
|--------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------|------------|
| Product Availability | Available now | Available now | Similarity |
| | System | | |
| Architecture | Server & Client, Web-based<br>versions,<br>local servers & remote<br>archiving | Server & Client, Web-based<br>versions,<br>local servers & remote<br>archiving | Similarity |
| Operating System | Window XP | Window XP | Similarity |
| Hardware | HP, Dell, IBM | HP, Dell, IBM | Similarity |
| | Administration | | |
| System failure Alert | Available | Available | Similarity |
| | DICOM Compatibility | | |
| DICOM Worklist | Yes | Yes | similarity |
| DICOM Storage | Yes | Yes | similarity |
| DICOM Printer | Yes | Yes | Similarity |
| | Image Acquisition | | |
| Image Acquisition | DICOM radiology systems<br>Digital film scanner<br>Other imaging modalities | DICOM radiology systems<br>Digital film scanner<br>Other imaging modalities | Similarity |
| | Image Manipulation | | |
| Rotation * | Yes | Yes | similarity |
| Invert | Yes | Yes | similarity |
| Zoom In/Out | Yes | Yes | similarity |
| W/L Adjustment | Yes | Yes | similarity |
| Pan Image | Yes | Yes | similarity |
| Flipping | Yes | Yes | similarity |
| Image Crop | No | Yes | |
| | Image Post-Processing | | |
| Secondary<br>Processing | Organ specific type | Organ Specific Type | similarity |
| | Measurement | | |
| Length Measurement | Yes | Yes | similarity |
| Angle Measurement | Yes | Yes | similarity |
| Multiple Length<br>Measurement | Yes | Yes | similarity |
| | Backup | | |
| Storage Backup<br>method | CD/DVD, Flash Drive | CD/DVD, Flash Drive, URL | |
| Export Format | DICOM, TIFF, BMP, JPEG,<br>PNG | DICOM, BMP | |
| | Services | | |
| Support | 24/7 customer support for<br>application software. | 24/7 customer support for<br>application software, | similarity |
| Training | On/off site available | On/off site available | similarity |
: .. . :
> ・・ .
{3}------------------------------------------------
#### 510(k) Premarket Notification Modern Module Inc. - OminiVision
.
:
.
000019-3
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the words "Modern Module" stacked on top of each other. The word "Modern" is on the top and the word "Module" is on the bottom. The font is a simple sans-serif font and the text is black.
#### Description:
OmniVision is a Picture Archiving and Communications System (PACS), as is the predicate Device dicomPACS. All of them have been developed to acquire, store, communicate, display and process medical images. They offer features (e.g. window leveling, zoom, measurements, annotations etc.) routinely used by medical professionals, such as radiologists and orthopaedists, and required of all PACS solutions.
OmniVision has a modular system architecture, It consists of the basic application viewer for image viewing and processing, image storage and communication, and a number of other modules for database management, image acquisition, printing etc.
OmniVision conforms to the DICOM (Digital Image and Communications in Medicine) standard.
#### Indications for Use:
OmniVision is intended to be used for the acquisition, storage, communication and viewing of medical images.
OmniVision receives images from modalities via DICOM or imported directly. It archives and displays these images for the use of medical specialists who are qualified to operate radiological equipment and to record and diagnose medical images. It provides the user with a range of tools assist them in viewing the images. such as zoom, filters and measurements, and with facilities to exchange images with other specialists.
OmniVision is not intended for the acquisition of mammographic image data.
#### Technological Characteristics:
OmniVision is intended to be used for the acquisition, storage, communication and viewing of medical images. OmniVision is an autonomous software and involves no hardware. This largely applies to the predicate device dicomPACS.
OmniVision receives images from modalities via DICOM or imported directly. It archives and displays these images for the use of medical specialists who are qualified to operate radiological equipment and to record
000020
{5}------------------------------------------------
## 510(k) Premarket Notification
Modern Module Inc. - OminiVision
Image /page/5/Picture/2 description: The image shows the words "Modern Module" stacked on top of each other. The word "Modern" is on the top line, and the word "Module" is on the bottom line. The first letter of the word "Modern" is stylized with a black rectangle and a white letter M. The rest of the letters are in a simple, sans-serif font.
and diagnose medical images. It provides the user with a range of tools assist them in such as zoom, filters and measurements; and with facilities to exchange images with other specialists.
OmniVision does not control any life-sustaining devices. Specialists with adequate expert knowledge for competent human intervention interpret displayed or printed images and information.
.:
Omni\/ision runs on any hardware platform meeting the minimum system requirements.
OmniVision can be used with the server operating systems MS Windows XP
#### Testing:
Omnivision has been tested according to the specifications documented in this notification. It conforms to the DICOM standard as laid out in the included DICOM Conformance Statement. (Tab. 9 section 2)
#### Conclusion:
Omnivision is a medical device. OmniVision provides functionality comparable to that of its Predicate Device, dicomPACS and is intended for the same user and patient groups as its Predicate Device. Based on the indications for use, Technological Characteristics, performance testing and comparison to predicate device, the proposed OmniVision is substantially equivalent and has been shown to be safe and effective for its intended use
Image /page/5/Picture/12 description: The image is a black and white graphic that appears to be a map or some type of spatial representation. The image is mostly white, with black dots and lines scattered across the surface. The black markings are concentrated in the lower left quadrant of the image, with fewer markings in the upper right quadrant. The overall impression is that of a sparse and abstract representation of a geographical area or network.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the bird-like figure. The logo is black and white.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room – WO66-G609 Silver Spring, MD 20993-0002
Modern Module Inc c/o Marc M. Mouser Engineering Leader Underwriters Laboratories. Inc. 2600 NW Lake RD. CAMAS, WA 98607
JAN 26 2011
Re: K110040
Trade/Device Name: OmniVision Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: December 10, 2010 Received: January 6, 2010
Dear Mr. Mouser:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Insting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of
{7}------------------------------------------------
medical device-related adverse events) (21 CFR 803): and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely Yours.
Mary S. Postel
Mary Pastel, ScD. Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
## Indications for Use
510(k) Number (if known): K110040
Device Name: OmniVision
Indications for Use:
OmniVision is intended to be used for the acquisition, storage, communication and viewing of medical Images.OmniVision receives images from imaging modalities via DICOM or imported It archives and displays these images for the use of medical specialists who are qualified directly. to operate radiological equipment and to record and diagnose medical images.
It provides the user with a range of tools to assist them in viewing the images, such as zoom, filters and measurements, and with facilities to exchange images with other specialists.
OmniVision is not intended for the acquisition of mammographic image data.
AND/OR
: '
14 1.15.
Prescription Use X
(Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) :
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Barl D. Kunn
(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices
510(k) Number
Page 1 of _1_
000018
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.