K103465 · Fujifilm Medical Systems U.S.A, Inc. · LLZ · Jan 31, 2011 · Radiology
Device Facts
Record ID
K103465
Device Name
SYNAPSE 3D CARDIAC TOOLS
Applicant
Fujifilm Medical Systems U.S.A, Inc.
Product Code
LLZ · Radiology
Decision Date
Jan 31, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
Synapse 3D Cardiac Tools is medical imaging software used with Synapse 3D Basic Tools that is intended to provide trained medical imaging professionals, including Physicians and Radiologists, with tools to aid them in reading, interpreting, reporting, and treatment planning. Synapse 3D Cardiac Tools accepts DICOM compliant medical images acquired from a variety of imaging devices including, CT, MR, NM, and XA. This product is not intended for use with or for the primary diagnostic interpretation of Mammography images. Addition to the tools with Synapse 3D Basic Tools, Synapse 3D Cardiac Tools provides the tools for specific clinical applications which provide targeted workflows, custom UI, targeted measurements and reporting functions including: > Functional cardiac analysis for CT left ventriculography images: which is intended to evaluate the functional characteristics of heart > Functional cardiac analysis for non contrasted MR heart images: which is intended to evaluate the functional characteristics of heart > Coronary artery analysis for CT coronary arteriography images: which is intended for the qualitative and quantitative analysis of coronary arteries > Calcium scoring for non contrasted CT heart images: which is intended for non invasive identification and quantification of calcified atherosclerotic plaques in the coronary arteries using tomographic medical image data and clinically accepted calcium scoring algorithms > Cardiac Fusion: which is intended to analyze cardiac anatomy and pathology with a fused image of functional data (e.g. NM image, Bulls eye) and anatomical data.
Device Story
Synapse 3D Cardiac Tools is medical imaging software; integrates with Synapse 3D Basic Tools and Synapse Workstation/Cardiovascular systems. Inputs: DICOM-compliant CT, MR, NM, and XA images. Processing: automated extraction of left ventricular boundaries for functional analysis; path extraction for coronary arteries; Agatston score calculation for calcium scoring; image fusion of anatomical and functional data (e.g., NM, Bulls eye). Output: quantitative measurements (ejection fraction, volumes, calcification scores) and fused images for clinical review. Used in clinical settings by physicians and radiologists to assist in cardiac evaluation, treatment planning, and reporting. Benefits: provides targeted workflows and quantitative analysis to support diagnostic decision-making.
Clinical Evidence
No clinical data provided; substantial equivalence is based on bench testing and comparison to predicate devices.
Technological Characteristics
Software-based medical image processing tool. Operates on DICOM-compliant images (CT, MR, NM, XA). Features include automated boundary detection, vessel path extraction, and image fusion. Implements clinically accepted calcium scoring algorithms (Agatston). Designed for integration with existing Synapse workstation infrastructure.
Indications for Use
Indicated for trained medical imaging professionals, including physicians and radiologists, to aid in reading, interpreting, reporting, and treatment planning for cardiac patients using CT, MR, NM, and XA imaging data. Not for primary diagnostic interpretation of mammography.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
K103465
JAN 3 1 2011
# 5. 510(k) Summary
#### Date Prepared:
November 23, 2010
## Submitter's Information:
FUJIFILM Medical Systems USA, Inc. 419 West Avenue Stamford, Connecticut 06902
Telephone: (203) 602-3774 Facsimile: (203) 363-3813 Contact: Debra A. Peacock
#### Device Trade Name:
Synapse 3D Cardiac Tools
## Device Common Name:
Medical Image Processing and Analysis Software
### Regulation Number:
21 CFR 892.2050
## Device Classification:
Class II
#### Device Classification Name
Picture Archiving Communication System (PACS)
### Panel:
Radiology
#### Product Code:
90-LLZ
#### Date Received:
TBD
{1}------------------------------------------------
## Decision Date:
TBD
## Decision:
TBD
## Predicate Devices:
- · Voxar 3D Enterprise with CardiaMetrix, K061326
- · ZIO Soft MR Cardiac Function Analysis, K091262
- · GE CardIQ Fusion, K061370
# Description of the Device
Synapse 3D Cardiac Tools is with Fujifilm's Synapse 3D Basic Tools (K101662, cleared by CDRH on July 26, 2010) and can be integrated with our cleared Fujifilm's Synapse Workstation, version 3.1.0 and above as a part of a Synapse system. Synapse 3D Cardiac Tools also can be integrated with Fujifilm's Synapse Cardiovascular for cardiology purposes.
Synapse 3D Cardiac Tools offers physicians the following clinical applications in addition to the features available from our cleared Synapse 3D Basic Tools (K101662) to analyze the image data retrieved from various connected devices.
# Functional cardiac analysis (CT)
Functional cardiac analysis (CT) is an application for cardiac function evaluation which obtains the boundary between left ventricle and cardiac wall from CT left ventriculography images retrieved from multiple time phases and calculates ejection fraction, end-diastolic volume, endsystolic volume, stroke volume, etc.
## Functional cardiac analysis (MR)
Functional cardiac analysis (MR) is an application for cardiac function evaluation which obtains the boundary between left ventricle and cardiac wall from non-contrasted MR images retrieved from multiple time phases and calculates ejection fraction, end-diastolic volume, end-systolic volume, output volume per beat, etc.
## Coronary artery analysis (CT)
Coronary artery analysis is an application using CT coronary arteriography images to extract the path of the target blood vessels and to perform coronary artery evaluation.
{2}------------------------------------------------
### Calcium Scoring
The calcium scoring is an application which uses non-contrasted CT images to display the calcification area in the coronary artery with color separation and calculates the calcification quantitative values using the Agatston score method.
#### Cardiac Fusion
Cardiac fusion is an application to create an image having the mutual characteristics of source images of heart. Source images could be original image of CT, MR or NM and the functional image derived from the original image.
#### Indication for Use
Synapse 3D Cardiac Tools is medical imaging software used with Synapse 3D Basic Tools that is intended to provide trained medical imaging professionals, including Physicians and Radiologists, with tools to aid them in reading, interpreting, and treatment planning, Synapse 3D Cardiac Tools accepts DICOM compliant medical images acquired from a variety of imaging devices including, CT, MR, NM, and XA,
This product is not intended for use with or for the primary diagnostic interpretation of Mammography images.
Addition to the tools with Synapse 3D Basic Tools, Synapse 3D Cardiac Tools provides the tools for specific clinical applications which provide targeted workflows, custom UI, targeted measurements and reporting functions including:
- > Functional cardiac analysis for CT left ventriculography images: which is intended to evaluate the functional characteristics of heart
- > Functional cardiac analysis for non contrasted MR heart images: which is intended to evaluate the functional characteristics of heart
- > Coronary artery analysis for CT coronary arteriography images: which is intended for the qualitative and quantitative analysis of coronary arteries
- > Calcium scoring for non contrasted CT heart images: which is intended for non invasive identification and quantification of calcified atherosclerotic plaques in the coronary arteries using tomographic medical image data and clinically accepted calcium scoring algorithms
- > Cardiac Fusion: which is intended to analyze cardiac anatomy and pathology with a fused image of functional data (e.g. NM image, Bulls eye) and anatomical data.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three curved lines representing its body and wings. The logo is surrounded by text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" in a circular arrangement.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
Ms. Debbie Peacock Sr. Regulatory Affairs Specialist FUJIFILM Medical Systems USA, Inc. 419 West Avenue STAMFORD, CT 06092
JAN 3 1 201
and the country of the country of the county of the county of
Re: K103465
Trade/Device Name: Synapse 3D Cardiac Tools Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: November 23, 2010 Received: November 24, 2010
Dear Ms. Peacock:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of
{4}------------------------------------------------
medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely Yours.
Mary S Postel
Mary Pastel. ScD. Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known):
Device Name: Synapse 3D Cardiac Tools
Indications for Use:
Synapse 3D Cardiac Tools is medical imaging software used with Synapse 3D Basic Tools that is intended to provide trained medical imaging professionals, including Physicians and Radiologists, with tools to aid them in reading, interpreting, reporting, and treatment planning. Synapse 3D Cardiac Tools accepts DICOM compliant medical images acquired from a variety of imaging devices including, CT, MR, NM, and XA.
This product is not intended for use with or for the primary diagnostic interpretation of Mammography images.
Addition to the tools with Synapse 3D Basic Tools, Synapse 3D Cardiac Tools provides the tools for specific clinical applications which provide targeted workflows. custom UI, targeted measurements and reporting functions including:
- > Functional cardiac analysis for CT left ventriculography images: which is intended to evaluate the functional characteristics of heart
- > Functional cardiac analysis for non contrasted MR heart images: which is intended to evaluate the functional characteristics of heart
- > Coronary artery analysis for CT coronary arteriography images: which is intended for the qualitative and quantitative analysis of coronary arteries
- > Calcium scoring for non contrasted CT heart images: which is intended for non invasive identification and quantification of calcified atherosclerotic plaques in the coronary arteries using tomographic medical image data and clinically accepted calcium scoring algorithms
- > Cardiac Fusion: which is intended to analyze cardiac anatomy and pathology with a fused image of functional data (e.q. NM imaqe. Bulls eye) and anatomical data.
Prescription Use X Pressmption USD
(Part 21 CFR 801 Subpart D)
Over-The-Counter Use AND/OR (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Device (OIVD)
(Division Sign-Off)
810K K103465
Synanse 3D Cardian Tools
Page 19 of 1101
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.