DC Neuro is an image processing software package designed to run on standard PC hardware (workstation). The required workstation hardware consists of standard, "off-the-shelf" computer components. DC Neuro software receives image data from medical imaging devices such as Magnetic Resonance Imaging systems or from image archives, PACS. DC Neuro can be used to perform image viewing, image manipulation, image quantification, communication, and printing. DC Neuro provides both analysis and viewing capabilities which integrate both anatomical and physiologic/functional imaging data sets, including blood oxygen level dependent (BOLD) fMRI, diffusion, fiber tracking, dynamic analysis, and vessel analysis in both 2D and/or 3D views. BOLD fMRI: The BOLD analysis is useful in quantifying small susceptibility changes in the human brain, created by the execution of a specific task. Diffusion/Fiber tracking: The diffusion of water molecules through brain tissue can be measured with diffusion-weighted MRI scanning. Actual diffusion properties depend on the local tissue. With fiber tracking, the directional dependency can be used to visualize the white matter structure in the brain. Dynamic Analysis: Dynamic analysis is intended for visualization and analysis of MR dynamic studies, showing changes in contrast over time, where such techniques are useful or necessary. Vessel Analysis: Allows the visualization of vascular morphology from contrast-enhanced MR images. 3D Visualization: 3D rendering allows the visualization of bone, vascular structures, and selected anatomy as three dimensional object surfaces projected on a flat screen.
Device Story
DC Neuro is a software-based image processing system for DICOM MRI data; operates on standard off-the-shelf PCs running Windows XP. Inputs include MRI images from scanners or PACS; software automatically sorts headers into exams. Processing includes co-registration, BOLD fMRI analysis, diffusion/fiber tracking, dynamic analysis, vessel analysis, and 3D rendering. Used in clinical settings by healthcare professionals to visualize and quantify anatomical and functional brain data. Output is displayed on a workstation monitor for clinical review; assists in assessing susceptibility changes, white matter structure, vascular morphology, and contrast dynamics. Benefits include enhanced visualization and quantification of complex MR datasets to support clinical decision-making.
Clinical Evidence
No clinical data provided; bench testing only. Substantial equivalence is based on functional comparison to predicate devices.
Technological Characteristics
Software-only image processing system; runs on standard off-the-shelf PC hardware with Windows XP Professional. Inputs: DICOM MRI data. Processing: Automated sorting, co-registration, and analysis algorithms for BOLD, diffusion, dynamic, vessel, and 3D rendering. Connectivity: Networked via PACS or direct MR system interface. Standalone workstation deployment.
Indications for Use
Indicated for clinicians requiring image processing, viewing, manipulation, quantification, and printing of DICOM MRI data. Supports analysis of brain anatomy and physiology, including BOLD fMRI, diffusion, fiber tracking, dynamic contrast studies, and vascular morphology in 2D/3D. No specific age or gender contraindications stated.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Invivo Corporation - DC Neuro Premarka Notification [510(k)] Submission
K081242
pg. 1 of 3
#### 510(K) SUMMARY
This summary is being submitted in accordance with 21 CFR 807.92.
# 1. GENERERAL INFORMATION
### 1.1. Submitter Information
Manufacturer's Name: Address:
Invivo Corporation N27 W23676 Paul Road Pewaukee, WI 53072 2183683.
treiskertajnvivocorp.com
JUN 1 7 2008
Establishment Registration #
#### 1.2. Contact Person Name and Information
| Contact: | Theodore J. Reisker |
|--------------|-------------------------------|
| Title: | Director, Product Development |
| Company: | Invivo Corporation |
| Address: | N27 W23676 Paul Road |
| | Pewaukee, WI 53072 |
| Telephone # | (262) 524-1402 ext. 175 |
| Facsimile #: | (262) 524-1403 |
# 1.3. Trade name and common name of device
| Trade name: | DC Neuro |
|--------------|-------------------------|
| Common name: | Image Processing System |
#### 1.4. Classification of the device
| Classification: | Image Processing System, 21 CFR 892.2050 |
|-----------------|------------------------------------------|
| Class: | Class II |
| Procode: | LLZ |
#### 1.5. Predicate Devices
E-mail:
Table 1 summarizes the correspondence between various features in DC Neuro and predicate devices, along with the pertinent 510(k) submission references.
| Feature | Predicate Device Name | Predicate<br>510(k)<br>Submission<br>Reference |
|---------------------------------------|-------------------------------------------------------------------------|------------------------------------------------|
| Diffusion / FiberTrak | Philips Achieva, Intera, & Panorama 1.0T | K052078 |
| BOLD / fMRI | Philips EasyVision Family Workstation<br>Option - BOLD Analysis Package | K990329 |
| Dynamic Analysis,<br>Vessels Analysis | Philips Quantitative Analysis Package | K971965 |
| 3D Rendering | Philips EasyVision Family Workstation<br>Option - 3D Volume Rendering | K953095 |
| Image processing<br>workstation | Philips View Forum | K032096 |
#### Summary of Predicate Devices and 510(k) references Table 1:
{1}------------------------------------------------
### Invivo Corporation - DC Neuro Premarket Notification [510(k)] Submission
K081262
pg. 2 of 3
# 2. BASIS FOR SUBSTANTIAL EQUIVALENCE DETERMINATION 2.1. Device Description
# 2.1.1. Summary of device functions and the devices major components
DC Neuro is a software package available for viewing, printing, storing, transferring, and quantifying DICOM magnetic resonance imaging (MRI) studies. The DC Neuro software is able to run on "off the shelf" standard PC components using the Microsoft Windows XP Professional Operating Software.
DC Neuro provides both analysis and viewing capabilities that integrate both anatomical and physiologic/functional imaging data sets, including blood oxygen level dependent (BOLD) fMRI, diffusion, fiber tracking, dynamic analysis, and vessel analysis in both 2D and/or 3D views.
DC Neuro software provides automated processing of DICOM images. Imaging cxams are transferred to the DC Neuro system in DICOM format either directly from the MR system or from a picture archiving and communications system (PACS) or via supported digital media. The DC Neuro system reads the incoming image headers and automatically sorts the images into exams consisting of multiple data sets. Pre-set processing algorithms may be performed on the incoming data as defined by the user such as image co-registration, fMRI analysis, diffusion analysis, dynamic analysis, vessel analysis, and 3D displays.
# 2.1.2. Description of device (major components)
DC Neuro is a software package that uses standard PC hardware, with a standard uscr/operator interface (keyboard and mouse), and a standard Operating System: Microsoft Windows XP Professional.
# 2.2. Intended use
DC Neuro is an image processing software package designed to run on standard PC hardware (workstation). The required workstation hardware consists of standard, "off-the-shelf" computer components. DC Neuro software receives image data from medical imaging devices such as Magnetic Resonance Imaging systems or from image archives, PACS. DC Neuro can be used to perform image viewing, image manipulation, image quantification, communication, and printing,
DC Neuro provides both analysis and viewing capabilities which integrate both anatomical and physiologic/functional imaging data sets, including blood oxygen level dependent (BOLD) fMRI, diffusion, fiber tracking, dynamic analysis, and vessel analysis in both 2D and/or 3D views.
{2}------------------------------------------------
K081262
pg. 3 of 3
BOLD fMRI: The BOLD analysis is useful in quantifying small susceptibility changes in the human brain, created by the execution of a specific task.
Diffusion/Fiber tracking: The diffusion of water molecules through brain tissue can be measured with diffusion-weighted MRI scanning. Actual diffusion properties depend on the local tissue. With fiber tracking, the directional dependency can be used to visualize the white matter structure in the brain.
Dynamic Analysis: Dynamic analysis is intended for visualization and analysis of MR dynamic studics, showing changes in contrast over time, where such techniques are useful or necessary.
Vessel Analysis: Allows the visualization of vascular morphology from contrast-enhanced MR images.
3D Visualization: 3D rendering allows the visualization of bone, vascular structures, and selected anatomy as three dimensional object surfaces projected on a flat screen.
# 2.3. Safety Information
No new hazards are introduced by Invivo Corporation's DC Neuro, as compared to predicated devices and image processing functionality.
# 2.4. Conclusion
Invivo Corporation's DC Neuro is substantially equivalent to the Philips ViewForum 2003 workstation, and the Philips BOLD fMRI. Diffusion/Fiber Tracking, Dynamic and Vessel Analyses, and 3D visualization image processing functionality.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird in flight, composed of three curved lines. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol. The logo is simple and uses a monochrome color scheme.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# JUN 1 7 2008
Invivo Corporation % Mr. Mark Job Responsible Third Party Official Regulatory Technology Services LLC 1394 25th Street NW BUFFALO MN 55313
Re: K081262
Trade/Device Name: DC Neuro Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: June 5, 2008 Received: June 6, 2008
#### Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter.
| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrb/industry/support/index.html.
Sincerely yours,
Nancy Cbrogdon
Nancy C. Brogdon Director, Division of Reproductive. Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
| | INDICATIONS FOR USE |
|-------|---------------------|
| own): | K081262 |
510(k) Number (if knov
Device Name:
DC Neuro
#### Indications For Use:
DC Neuro is an image processing software package designed to run on standard PC hardware (workstation). The required workstation hardware consists of standard, "off-the-shelf" computer components. DC Neuro software receives image data from medical imaging devices such as Magnetic Resonance Imaging systems or from image archives, PACS. DC Neuro can be used to perform image viewing, image manipulation, image quantification, communication, and printing.
DC Neuro provides both analysis and viewing capabilities which integrate both anatomical and physiologic/functional imaging data sets, including blood oxygen level dependent (BOLD) fMRI, diffusion, fiber tracking, dynamic analysis, and vessel analysis in both 2D and/or 3D views.
BOLD fMRI: The BOLD analysis is useful in quantifying small susceptibility changes in the human brain, created by the execution of a specific task.
Diffusion/Fiber tracking: The diffusion of water molecules through brain tissue can be measured with diffusion-weighted MRI scanning. Actual diffusion properties depend on the local tissue. With fiber tracking, the directional dependency can be used to visualize the white matter structure in the brain.
Dynamic Analysis: Dynamic analysis is intended for visualization and analysis of MR dynamic studies, showing changes in contrast over time, where such techniques are useful or necessary.
Vessel Analysis: Allows the visualization of vascular morphology from MR images.
AND/OR
3D Visualization: 3D rendering allows the visualization of bone, vascular structures, and selected anatomy as three dimensional object surfaces projected on a flat screen.
Prescription Use J (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Arjun Mr. Whang
ductive Abdominal, and 510(k) Numbe
Page 1 of 1
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.