The KA6 is a portable X-ray system for diagnostic imaging of body extremities.
Device Story
Handheld, battery-operated X-ray system; generates radiographic images of body extremities. Input: operator-selected exposure parameters (kVp, mA, time). Transformation: Canon X-ray tube (0.4 mm focal spot) generates X-rays; device controls exposure intensity (40-70 kV), dosage (2-6 mA), and time (0.06-2.00 sec). Output: X-ray beam for capture by external Flat Panel X-ray Detector (FXRD). Used in clinical settings by physicians or technicians. Operator uses built-in software (RPG-F-0706) to manage exposure settings via pushbuttons and LED indicators. Device includes SSD cage to maintain 300 mm distance. Output images are displayed/stored on external computer system. Benefits: portable diagnostic imaging for extremities; lightweight design (2.4 kg) allows mobility compared to traditional stationary or stand-mounted systems.
Clinical Evidence
Clinical evaluation consisted of a qualitative assessment of diagnostic images produced by the Remex KA6. Images were reviewed by a board-certified radiologist and determined to be of acceptable diagnostic quality. No comparative clinical trial data provided.
Technological Characteristics
Handheld mobile X-ray system; Canon D-041SB X-ray tube; 0.4 mm focal spot; 40-70 kVp range; 2-6 mA range; 1.6 mm Al equivalent filtration. Powered by 22.2 VDC Lithium Polymer battery. Connectivity: interfaces with external FXRD. Software: RPG-F-0706 v1.00. Standards: IEC 60601-1, 60601-1-3, 60601-1-6, 60601-2-28, 60601-2-54, 60601-1-2.
Indications for Use
Indicated for diagnostic imaging of body extremities in adult and pediatric patients. Intended for use by qualified/trained physicians or technicians.
Regulatory Classification
Identification
A mobile x-ray system is a transportable device system intended to be used to generate and control x-ray for diagnostic procedures. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
K221417 — CLAROX plus (Model: VX-100) · Vsi Co., Ltd. · Sep 13, 2022
K200976 — SR-8230, SR-8230S Portable X-ray Unit · Shantou Institute of Ultrasonic Instruments Co., Ltd. (Siui) · Jun 10, 2020
Submission Summary (Full Text)
{0}------------------------------------------------
August 3, 2021
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Remedi Co. Ltd. % Mr. W. Lee Strong Quality Systems Manager 510K FDA, Inc. 100 E Granada Blvd. Suite 219 ORMOND BEACH FL 32176
### Re: K212144
Trade/Device Name: Remex KA6 Regulation Number: 21 CFR 892.1720 Regulation Name: Mobile x-ray system Regulatory Class: Class II Product Code: IZL Dated: July 6, 2021 Received: July 9, 2021
Dear Mr. Strong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K212144
Device Name REMEX KA6
Indications for Use (Describe)
The KA6 is a portable X-ray system for diagnostic imaging of body extremities.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows a stylized black and white drawing of an eagle's head and wing. The eagle's head is facing to the right, and its beak is slightly open. The wing is spread out above the head, and it has several feathers. The drawing is simple and elegant, and it captures the power and grace of the eagle.
# 510k FDA Consulting
## Medical Device Clearance
100 East Granada Blvd., Suite 219 Ormond Beach, FL 32176
386-506-8711
# 510(k) Summary
# K212144
## Submitter
Remedi Co., Ltd. #1409, IS BIZ Tower, 26 Yangpyeong-ro 21 Seoul, Republic of Korea 072NN
+82-(0)2-2135-5879 Tel: +82-(0)2-2135-5889 Fax: Gyuun Choi, Regulatory Manager(gyuun.choi@remedihc.com ) Contact: Date: July 6, 2021
## Consultant
510K FDA, Inc. 100 East Granada Blvd., Suite 219 Ormond Beach, FL 32176
| Phone: | 386-506-8711 |
|--------------------|-------------------------------------------------|
| eFax: | 855-235-7902 |
| Primary Contact: | W. Lee Strong, Quality Systems Manager () |
| Secondary Contact: | Claude Berthoin, President (Claude@510kfda.com) |
### Device Classification
| Trade Names: | Remex KA6 |
|--------------------|---------------------|
| Common Name: | Mobile X-Ray System |
| Regulation Name: | Mobile X-Ray System |
| Regulation Number: | 21 CFR 892.1720 |
| Medical Specialty: | Radiology |
| Regulatory Class: | II |
| Product Code: | IZL |
| Submission Type: | 510(k) |
| Regulatory Class: | 2 |
{4}------------------------------------------------
#### Predicate Devices
The following predicates are legally marketed, post-amendment devices:
| 510(k) Number: | K182207 |
|--------------------|------------------------------------------------------------------------------------|
| Clearance Date: | December 14, 2018 |
| 510(k) Trade Name: | MinXray TR90BH |
| Submitters: | Mikasa X-Ray Co., Ltd. (Tokyo, Japan)<br>MinXray, Inc. (Northbrook, Illinois, USA) |
| Regulatory Class: | II |
| Regulation Number: | 21 CFR 892.1720 |
| Regulation Name: | Mobile X-Ray System |
| Product Code: | IZL |
#### Device Description
The Remex KA6 is a handheld battery-operated x-ray system intended for use by a qualified/trained physician or technician on both adult and pediatric subjects for taking diagnostic x-rays.
The subject device is designed as an x-ray source to take diagnostic x-rays as radiographic images. The x-ray tube by Canon (tube voltage 40-70kV with 0.4 mm focal spot) is located inside the handheld device. The device is used with a Flat Panel X-ray Detector(FXRD), not part of the system, which is attached to a computer with imaging software to enable image capture, display, manipulation, storage, and transmission. The device is not designed to be used with a radiographic grid.
The operator controls three key variables to obtain the best image at minimal exposure:
- 1. X-Ray Exposure Intensity can be changed from 40 kV to 70 kV in 1 kV increments.
- 2. X-Ray Exposure Dosage can be changed from 2mA 6mA in 1mA increments.
- 3. X-Ray Exposure Time can be changed from 0.06 sec 2.00 sec in 0.01 increments.
#### Indications for Use
The KA6 is a handheld portable X-ray system for diagnostic imaging of body extremities.
The device was previously submitted to the FDA in submission number K202559 and found not-substantially equivalent (NSE). The Remex KA6 is being re-submitted with an adjusted IFU statement.
{5}------------------------------------------------
## Comparison of Technological Characteristics with Predicates
The following table compares technological and other characteristics of the subject and predicate devices.
| Characteristic | Remex KA6<br>(K212144) | MinXray TR90BH<br>(K182207) |
|----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | Intended for use by a<br>qualified/trained physician or<br>technician on both adult and pediatric<br>subjects for taking diagnostic x-rays. | Intended for use by a<br>qualified/trained physician or<br>technician on both adult and pediatric<br>subjects for taking diagnostic x-rays. |
| Indications for<br>Use | The KA6 is a portable X-ray system<br>for diagnostic imaging of body<br>extremities. | The TR90BH is a portable X-ray<br>system with following limitations of<br>use:<br>The device may be used for handheld<br>diagnostic imaging of body<br>extremities.<br>The device may be used for stand<br>mounted diagnostic imaging of head,<br>abdomen, or extremities.<br>The device may be used for stand<br>mounted imaging of the chest when<br>used without a grid.<br>- Not to be used on bariatric patients,<br>unless imaging body extremities.<br>- Not for mammography use<br>- The TR90BH is not intended to<br>replace a stationary radiographic<br>system, which may be required for<br>full optimization of image quality and<br>radiation exposure for different exam<br>types. |
| Comment on<br>Indications | Certain limitations apply because of<br>the lower maximum kVp output as<br>compared to the predicate. See<br>indications above. | Original Indications |
| Characteristic | Remex KA6<br>(K212144) | MinXray TR90BH<br>(K182207) |
| Size: Body | 165 x 176 x 255 mm | 219 x 442 x 190 mm |
| Weight | 2.4 kg (including cone 190g) | 7.5kg |
| Energy Source | Lithium Polymer Rechargeable<br>Battery (22.2 VDC), 300 exposures<br>per charge | Lithium-ion Rechargeable Battery<br>(57.6 VDC), 300 exposures per<br>charge. |
| Mounting<br>method | Unit is usually handheld, or mounted<br>on optional arms or tripod | Unit is usually mounted to a MinXray<br>XGS MKIII Portable Stand |
| User Interface | Up-Down pushbuttons for selection<br>of exposure time, kVp, and mA, with<br>LED indicators | Up-Down pushbuttons for kVp<br>selections and exposure time<br>selections with LED indicators |
| Exposure switch | Dual-stage, deadman type | Dual-stage, deadman type |
| Controls | Built-in Software<br>S/W name: RPG-F-0706<br>S/W version: 1.00 | Analog/digital, no software |
| Exposure times | 0.06 -2.0 sec (0.01 sec. steps) | 0.03-0.2 sec (0.01 sec. Steps)<br>0.2-04 sec ( 0.02 sec. Steps)<br>04-1.0 sec (0.05 sec. Steps)<br>1.0-4.0 sec (0.1 sec. Steps) |
| Tube potential<br>(kVp) | 40 - 70 kVp (1 kVp steps) | 40 - 90 kVp |
| Characteristic | Remex KA6<br>(K212144) | MinXray TR90BH<br>(K182207) |
| Tube current<br>(mA) | 2 mA - 6 mA (1 mA steps) | 20 mA @ 40 kVDC – 60 kVDC (2 kVP steps)<br><br>15 mA @ 62 kVDC – 80 kVDC (2 kVP steps)<br><br>10 mA @ 82 kVDC – 90 kVDC (2 kVP steps)<br><br>High Power Mode<br><br>15 mA @ 82 kVDC – 90 kVDC (2 kVP steps) |
| Alerts and<br>Alarms | At least three specific alerts and alarms indicate equipment state of readiness to use. | At least two specific alerts and alarms indicate equipment state of readiness to use. |
| X-ray tube | Canon D-041SB | Toshiba D-0814 |
| Focal Spot Size | 0.4 mm | 1.2 mm |
| Total filtration | 1.6 mm Al equivalent | 3.2mm AL equivalent |
| Collimator | Square collimator with LED Light Field Center Indicator | Four manually and steplessly adjustable shutters with light beam type central x-ray indicator (Advantech R72) |
| Triggering<br>Mechanism | Two stage triggering | Two stage triggering |
| Source to Skin<br>Distance (SSD) | 300 mm | 300 mm |
| Performance<br>Standard | IEC 60601-1-3:2013<br>IEC 60601-2-28:2017<br>IEC 60601-2-54:2009 | IEC 60601-1-3:2008<br>IEC 60601-1-2-28:2010<br>IEC 60601-2-54:2009 |
| Characteristic | Remex KA6<br>(K212144) | MinXray TR90BH<br>(K182207) |
| Electrical safety | IEC 60601-1:2012 | IEC 60601-1:2012 |
| | IEC 60601-1-2:2014 | IEC 60601-1-2:2007 |
| | IEC 60601-1-6:2013 | IEC 60601-1-6:2010 |
| | IEC 62304:2006 | IEC 62304:2006 |
| | IEC 62366:2007 | IEC 62366:2007 |
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
Both the subject Remex KA6 and predicate MinXray TR90BH device are intended for mobile x-ray examination of adult and pediatric populations.
The subject device is lightweight and designed for handheld use and transport in a carry case, while the predicates are heavy enough to generally require mounting and transport on a portable stand.
The subject is DC-powered by battery, while the predicate is AC-powered via wall outlet. Both the subject and predicates are operated by push buttons and/or touch screen on the device itself.
The above comparison of technological characteristics shows the Remex KA6 to be smaller, lighter, and less powerful than the predicate, with lower tube voltage and current being the more obvious examples. These variations raise no new issues of safety or effectiveness.
#### Performance Data
The following performance data were provided in support of the substantial equivalence determination...
- 1. Software Remex KA6 completed software validation and was determined to be a moderate level of concern software. The risk management test was performed in accordance with ISO 14971 and met all requirements of the standard. These tests were performed by the manufacturer and all requirements were met.
- 2. Electrical Safety -Accredited Testing Laboratory KCTL Inc. performed tests for 60601-1, 60601-1-3, 60601-1-6, and 60601-2-28, and 60601-2-54 and met all requirements for electrical safety.
- 3. Electromagnetic Compatibility (EMC)—KCTL performed tests for 60601-1-2and met all requirements for safety.
{9}------------------------------------------------
- 4. Bench Testing Performance tests for radiation protection in x-ray equipment was completed by an outside laborabory, KCTL, and met the test requirements.
- 5. Clinical Evaluation Images from the Remex KA6 were evaluated for diagnostic quality by a board certified radiologist and were approved for diagnostic quality.
#### Risk Analysis Information
Risks associated with the handheld design include increased operator exposure due to leakage radiation and backscatter radiation. Methods to reduce exposure include proper lead lining around the x-ray source assembly, measurement of a typical exposure near and around the unit, and recommended safety precautions such as wearing personnel monitoring and protective equipment.
NOTE: The use of a tripod stand is recommended if you need the device to be more stable while taking the images to prevent blurry imaging. This is an accessory to the Remex KA6.
The Remex KA6 has a Source-Skin Distance (SSD) cage that is attached to the device to maintain a minimum SSD of 300 mm. Operators are not to remove this cage to bring the device closer to the patient. Images may be taken at a distance greater than the cage to the source, but cannot be taken closer than the cage allows.
The petition was prepared in compliances with the following FDA guidance instructions and documents: "FDA Guidance on Radiation Safety Considerations for X-Ray Equipment Designed for Hand-Held Use. "
#### Conclusions
The comparison of intended use and technological characteristics shows the subject device to be at least as safe and effective as the predicate, and, furthermore, warrants a finding of substantial equivalence between the Remex KA6 and MinXray TR90BH predicate.
The non-clinical data support the safety of the device and demonstrate that Remex KA6 should perform as intended in the specified use conditions.
Clinical images from the Remex KA6 were evaluated for diagnostic quality by a board certified radiologist and were approved for diagnostic quality.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.