K182207 · Minxray, Inc. · IZL · Dec 14, 2018 · Radiology
Device Facts
Record ID
K182207
Device Name
TR90BH
Applicant
Minxray, Inc.
Product Code
IZL · Radiology
Decision Date
Dec 14, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1720
Device Class
Class 2
Indications for Use
The TR90BH is a portable X-ray system with following limitations of use: The device may be used for handheld diagnostic imaging of body extremities The device may be used for stand mounted diagnostic imaging of head, abdomen, or extremities. The device may be used for stand mounted imaging of the chest when used without a grid. - Not to be used on bariatric patients, unless imaging body extremities - Not for mammography use - The TR90BH is not intended to replace a stationary radiographic system, which may be required for full optimization of image quality and radiation exposure for different exam types.
Device Story
Portable X-ray system; battery-powered; used for diagnostic imaging of extremities, head, abdomen, and chest. Operates via high-frequency generator; tubehead/control unit; light beam collimator; 2-stage exposure switch. Operated by physician or technician; handheld or stand-mounted. Input: user-selected technique factors (kVp, mA, exposure time). Output: X-ray emission for image capture. Safety features: dead-man switch; power-on password for cybersecurity; radiation shielding for handheld use. Benefits: portable imaging at point-of-care or disaster sites; battery-operated for remote use.
Clinical Evidence
Limited clinical testing performed to demonstrate imaging capability for anatomy beyond extremities. Review conducted by a Board Certified Radiologist; results confirmed successful imaging performance. Bench testing verified compliance with IEC 60601 series standards and battery safety per UN transport criteria.
Technological Characteristics
Portable X-ray system; high-frequency generator; Toshiba D-0814 X-ray tube (0.8mm focal spot); Lithium-ion battery powered; Mikasa BLD34L collimator. Dimensions: 219 x 442 x 190 mm; Weight: 7.5 kg. Connectivity: standalone. Software: moderate level of concern; includes power-on password for cybersecurity. Standards: IEC 60601-1, 60601-1-2, 60601-1-3, 60601-1-6, 60601-2-28, 60601-2-54, IEC 62304, IEC 62366.
Indications for Use
Indicated for diagnostic X-ray imaging of body extremities (handheld or stand-mounted) and head, abdomen, or chest (stand-mounted only). Contraindicated for bariatric patients (except extremities) and mammography.
Regulatory Classification
Identification
A mobile x-ray system is a transportable device system intended to be used to generate and control x-ray for diagnostic procedures. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the U.S. Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
December 14, 2018
MinXray, Inc % Daniel Kamm Principal Engineer Kamm & Associates 8870 Ravello Ct NAPLES, FLORIDA 34114
Re: K182207
Trade/Device Name: TR90BH Regulation Number: 21 CFR 892.1720 Regulation Name: Mobile x-ray system Regulatory Class: Class II Product Code: IZL Dated: October 9, 2018 Received: October 22, 2018
Dear Daniel Kamm:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Hole 2. Nils
for Robert A. Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K182207
Device Name TR90BH (Portable Diagnostic X-Ray System)
Indications for Use (Describe)
The TR90BH is a portable X-ray system with following limitations of use:
The device may be used for handheld diagnostic imaging of body extremities
The device may be used for stand mounted diagnostic imaging of head, abdomen, or extremities.
The device may be used for stand mounted imaging of the chest when used without a grid.
- Not to be used on bariatric patients, unless imaging body extremities
- Not for mammography use
- The TR90BH is not intended to replace a stationary radiographic system, which may be required for full optimization of image quality and radiation exposure for different exam types.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) Summary: 510(k) Number K182207 MinXray, Inc. 3611 Commercial Avenue Northbrook, Illinois 60062, USA Toll Free 1-800-221-2245 (USA & Canada) Date Prepared: December 11, 2018 Contact: Keith Kretchmer, President
- 1. Identification of the Device: Trade/Device Names: TR90BH Regulation Number: 21CFR892.1720 Regulation Name: Mobile X-Ray System Regulatory Class: II Product Code: IZL Common/Usual Name: Mobile Diagnostic X-Ray System
2. Equivalent legally marketed device: K153059 Trade/Device Name: HF1202H PowerPlus™ Portable X-Ray Equipment Manufacturer: MinXray, Inc. Regulation Number: 21CFR892.1720 Regulation Name: Mobile X-Ray System Regulatory Class: II Product Code: IZL Common/Usual Name: Mobile Diagnostic X-Ray System
- 3. Indications for Use (intended use): The TR90BH is a portable X-ray system with following limitations of use: The device may be used for handheld diagnostic imaging of body extremities The device may be used for stand mounted diagnostic imaging of head, abdomen, or extremities. The device may be used for stand mounted imaging of the chest when used without a grid.
- Not to be used on bariatric patients, unless imaging body extremities
- Not for mammography use
- The TR90BH is not intended to replace a stationary radiographic system, which may be required for full optimization of image quality and radiation exposure for different exam types.
- 4. Reference Device: MinXray Model HF80H cleared for general purpose radiography in K945707. This device provides a range of technique factors closely comparable to the new TR90BH but using older technology. The HF80H was a 10 mA model. It was soon followed by a 15 mA model known as the HF80H+. FDA advised us that it was not necessary to submit a new 510(k) when we released HF80H+ with 15 mA and the same kV range. The concern is that higher mAs factors needed for properly imaging thicker body parts (e.g. chest, abdomen) would need to be achieved via longer exposure times leading to patient motion artifact, but this reference device has a fixed ma capability of 15 ma and a maximum kVp of 80, whereas the new TR90BH has the same ma capability of 15 mA at a maximum kVp of 90. Therefore longer exposure times would not be needed with the new device as compared to this reference device. The same Technique Chart we provided for the HF80H+ could be used for the TR90BH. The new device does have higher mA capability, 20 mA at up to 60 kVp. A comparison chart follows.
{4}------------------------------------------------
| | HF80H+ K945707 | TR90BH K182207 |
|-----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| mA & kVp | 15 mA, fixed | 20 mA @ 40 kVDC – 60 kVDC (2 kVP steps) |
| | 50-80 kVp | 15 mA @ 62 kVDC – 80 kVDC (2 kVP steps) |
| | | 10 mA @ 82 kVDC – 90 kVDC (2 kVP steps) |
| | | |
| | | High Power Mode |
| | | 15 mA @ 82 kVDC – 90 kVDC (2 kVP steps) |
| Exposure | 0.08-3.98 sec, 0.02 steps. | 0.01-1.0 sec in 0.01 sec steps. |
| times | | High Power Mode 82-90 kV @ 15 mA the |
| | | maximum exposure time is 0.30 sec. |
| | HF1202H PowerPlus™ K153059 | TR90BH |
| Intended Use: | This radiographic system is intended for use<br>by a qualified/trained physician or technician<br>on both adult and pediatric subjects for<br>taking diagnostic x-rays. Not for<br>mammographic use. | The TR90BH is a portable X-ray system with<br>following limitations of use:<br>The device may be used for handheld diagnostic<br>imaging of body extremities. The device may be<br>used for stand mounted diagnostic imaging of<br>head, abdomen, or extremities. The device may<br>be used for stand mounted imaging of the chest<br>when used without a grid. - Not to be used on<br>bariatric patients, unless imaging body<br>extremities - Not for mammography use<br>- The TR90BH is not intended to replace a<br>stationary radiographic system, which may be<br>required for full optimization of image quality<br>and radiation exposure for different exam types. |
| Comment on<br>Indications | Original Indications | Certain limitations apply because of the lower<br>maximum kVp output as compared to the<br>predicate and because of the possibility of<br>handheld use on extremities. See indications<br>above. |
| Weight: | 19.5kgs | 7.5kgs |
| Size / | 305 x 460 x 225 mm | 219 x 442 x 190 mm |
| Energy Source: | 120/240 V 50 - 60AC | Lithium-ion Rechargeable Battery<br>(57.6DC), 300 exposures per charge. |
| Use Interface: | Soft touch push buttons | SAME |
| Exposure times: | (0.02–0.2sec) in 0.01sec. Step<br>(0.2-0.4sec) in 0.02sec. Step<br>(0.4-1.0sec) in 0.05sec Step<br>(1.0-5.0sec) in 0.1sec Step | 0.01 sec – 1.0 sec : 0.01 sec Step<br>High Power Mode<br>0.01 sec - 0.3 sec : 0.01 sec Step |
| mA: | 60 mA(0.02-0.05sec), 45mA (0.06 –0.2sec),<br>33 mA(0.22-5.0sec), @ 40 – 60 kVDC<br>55 mA(0.02-0.05sec), 41.3mA (0.06 –0.2sec),<br>30.3 mA(0.22-5.0sec), @ 62 – 70 kVDC<br>50 mA(0.02-0.05sec), 37.5mA (0.06 –0.2sec),<br>27.5 mA(0.22-5.0sec), @ 72 – 80 kVDC<br>45 mA(0.02-0.05sec), 33.8mA (0.06 – 0.2sec),<br>24.8 mA(0.22-5.0sec), @ 82 – 90 kVDC<br>40 mA(0.02-0.05sec), 30.0mA (0.06 – 0.2sec),<br>22.0 mA(0.22-5.0sec), @ 92 – 100 kVDC<br>30 mA(0.02-0.05sec), 22.5mA (0.06 – 0.2sec),<br>16.5 mA(0.22-5.0sec), @ 102 – 110 kVDC<br>25 mA(0.02-0.05sec), 18.8mA (0.06 –0.2sec),<br>13.8 mA(0.22-5.0sec), @ 112 – 120 kVDC | 20 mA @ 40 kVDC - 60 kVDC (2 kVP steps)<br>15 mA @ 62 kVDC - 80 kVDC (2 kVP steps)<br>10 mA @ 82 kVDC - 90 kVDC (2 kVP steps)<br>High Power Mode<br>15 mA @ 82 kVDC – 90 kVDC (2 kVP steps)<br>Comment: 90 kVP maximum instead of 120 kVP<br>maximum. |
| | HF1202H PowerPlus™ K153059 | TR90BH |
| Memory<br>settings<br>(technique) | 10 memories via pushbutton | 5 memories via pushbutton |
| HF Generator | High Frequency | SAME |
| kW | 1.5 kW | 1.35 kW |
| kVp: | 40 – 120kVp | 40 – 90kVp |
| X-ray Tube | SXR 130-15-1.2 | D-0814 |
| FDA<br>Performance<br>Standard | Complies | Complies |
| Collimator | Collimare® | Mikasa BLD34L |
| Photo | | |
- 5. Description of the Device: We have developed compact, lightweight, battery powered easy-carry TR90BH based on the technology of HF1202H. This device can be used with a stand or hand-held. A detailed comparison table with an equivalent device is provided below. The MinXray TR90BH consists of a battery powered X-ray generator (tubehead/control), continuously adjustable light beam collimator, mounting trunnion, exposure cord with 2-stage exposure switch, battery charger, AC adaptor and AC power cord, and those are provided in a portable carry case with wheels. If a stand is purchased with the TR90BH, such as the MinXray XGS series of gas spring portable mobile stands, instructions for assembly of the stand and the attachment of the TR90BH are included with the stand. The device works with a dedicated secondary Lithiumion battery, therefore it can be operated at no power supply places such as disaster sites. A few number of shots can be taken in short chanks to the newly designed and superior patent licensed battery. The x-ray tube is Toshiba Electron Tubes & Devices D-0814 0.8mm, 9.86kHU which is designed for diagnostic x-ray applications. 2-stage exposure switch: The exposure switch has 2-stage button. The first stage preheats the filament of the x-ray tube and lights collimator enabling operator to confirm exposure area. After keeping the button position at the first stage end for approximately 1 second, exposure preparation will be done and the STAND BY indicator will light. The second stage initiates the x-ray exposure for the sec display in advance. When the second stage is fully pressed, x-ray will be emitted, the X-RAY indicator will light, and an audible signal will be heard. One must PRESS AND HOLD THE SECOND STAGE END POSITION UNTIL THE EXPOSURE HAS TERMINATED. The switch operates in a "dead man" mode in that x-ray exposure will terminate immediately as a safety feature when the button is released.
Technology: The basic technology has not changed, however the TR90BH uses a new designed secondary battery unit. Our labeling shows that it is possible to operate the unit in a "hand held" mode. We took the following FDA guidance into account: Guidance for Industry and FDA Staff Radiation Safety Considerations for X-Ray Equipment Designed for Hand-Held Use; Document issued on December 24, 2008. The software and microprocessors are the same as those used in our predicate design, with a moderate level of concern applicable to this kind of software. A power-on password routine has been added to address cybersecurity concerns. This would prevent unauthorized use of the device.
- 6. Safety and Effectiveness, comparison to predicate device. The results of bench testing indicates that the new device is as safe and effective as the predicate device. Proper system operation is fully verified upon installation.
{5}------------------------------------------------
{6}------------------------------------------------
- 8. Summary of non-clinical testing: The software is unchanged with a moderate level of concern. Risk analysis was performed. Laboratory testing was performed according to the following standards:
IEC 60601-1: 2005 + CORR. 1 (2006) + CORR. 2 (2007) + AM1 (2012) Medical Electrical Equipment, Part 1: General Requirements for Safety
IEC 60601-1-2:2007 Medical Electrical Equipment-Part 1-2: General Requirements for Safety – 2. Collateral Standard-Electromagnetic compatibility - Requirements and tests
IEC 60601-1-3:2008 (Second Edition) for use with IEC 60601-1: 2005 (Third Edition) Medical electrical equipment Part 1-3: General Requirements for Radiation Protection in Diagnostic X-Ray Equipment
IEC 60601-1-6:2010 (Third Edition) for use in conjunction with IEC 60601-1: 2005 (Third Edition)Medical electrical equipment Part 1-6: General requirements for safety – Collateral Standard: Usability
IEC 60601-2-28:2010 (Second Edition) for use in conjunction with IEC 60601-1: 2005 (Third edition) Medical electrical equipment Part 2: Particular requirements for the safety of X-ray source assemblies and X-ray tube assemblies for medical diagnosis
IEC 60601-2-54 (First Edition): 2009 Medical electrical equipment Part 2: Particular requirements for
{7}------------------------------------------------
the basic safety and essential performance of X-ray equipment for radiography and radioscopy
IEC 62304:2006 (First Edition) Medical device software: Software life-cycle processes
IEC 62366: 2007 (First Edition) + A1: 2014 Medical devices - Application of usability engineering to medical devices.
The test results showed compliance with the above standards. We also confirmed overall operation by taking and reviewing test images.
Battery performance testing: A 3rd Party Testing Lab was engaged to do safety performance testing on the rechargeable lithium-ion battery pack, done according to the United Nations "Recommendations on the Transport of Dangerous Goods Manual of Tests and Criteria." The following testing was successfully performed: Altitude simulation, Thermal test, Vibration, Shock, External short circuit, Impact, Overcharge, and Forced discharge. In addition, battery life testing was performed to establish how many x-ray exposures can be made on one battery charge. The AC/DC adapter used with the battery charger is medical grade UL Listed (E356265).
During the design process, we took the following FDA guidance into account: Guidance for Industry and FDA Staff Radiation Safety Considerations for X-Ray Equipment Designed for Hand-Held Use; Document issued on December 24, 2008. This involved employing extra shielding and precautionary information in the labeling. Meets the US Performance standard on leakage.
- 9. Summary of clinical testing: A limited clinical test was performed to demonstrate that the device could image anatomy other than extremities. A Board Certified Radiologist performed a successful review.
- 10. Conclusion: After analyzing bench and clinical testing, it is the conclusion of MinXray Inc. that the TR90BH Portable X-Ray Equipment is as safe and effective as the predicate device, has few technological differences, and has similar indications for use, thus rendering it substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.