Digital Diagnostic Mobile X-ray System

K181626 · Samsung Electronics Co., Ltd. · IZL · Jul 20, 2018 · Radiology

Device Facts

Record IDK181626
Device NameDigital Diagnostic Mobile X-ray System
ApplicantSamsung Electronics Co., Ltd.
Product CodeIZL · Radiology
Decision DateJul 20, 2018
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1720
Device ClassClass 2

Indications for Use

The GM85 Digital Mobile X-ray imaging System is intended for use in generating radiographic images of human anatomy by a qualified/trained doctor or technician. This device is not intended for mammographic applications.

Device Story

Mobile digital X-ray system; captures radiographic images of human anatomy. X-rays pass through patient to detector; converted to electrical signals; processed by S-Station (OS) for amplification and digital conversion; saved as DICOM files. Features include S-Enhance (image enhancement for tubes, lines, needles, urinary stones), Manual Stitching (captures large body parts via multiple images), and Remote View (web-based remote image access). Used in clinical settings by physicians or technicians. Output viewed on PACS for clinical decision-making. Benefits include improved visualization of foreign bodies and anatomical structures, and efficient imaging of large areas.

Clinical Evidence

Bench-only and phantom-based evaluation. No human clinical trials. Anthropomorphic phantom images evaluated by a professional radiologist confirmed image quality equivalence to predicate. S-Enhance performance verified for visibility of foreign bodies (tubes, lines, needles) and urinary stones in chest, abdomen, and L-spine protocols. Detectors verified via MTF/DQE measurements per IEC 62220-1.

Technological Characteristics

Mobile X-ray system; high-frequency generator (32/40kW); 40-150kVp range. Detectors: CsI indirect flat-panel, 14"x17" or 17"x17", IP54 rated. Connectivity: DICOM, wireless/wired. Software: S-Station OS. Standards: ES 60601-1, IEC 60601-1-2, IEC 60601-1-3, IEC 60601-2-28, IEC 60601-2-54, ISO 14971.

Indications for Use

Indicated for generating radiographic images of human anatomy for qualified/trained doctors or technicians. Not for mammographic applications.

Regulatory Classification

Identification

A mobile x-ray system is a transportable device system intended to be used to generate and control x-ray for diagnostic procedures. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. July 20, 2018 Samsung Electronics Co., Ltd. % Soeun Chang Senior Professional, Regulatory Affairs 129, Samsung-ro, Yeongtong-gu Suwon-si, Gyeonggi-do 16677 REPUBLIC OF KOREA Re: K181626 Trade/Device Name: GM85 Regulation Number: 21 CFR 892.1720 Regulation Name: Mobile x-ray system Regulatory Class: II Product Code: IZL Dated: June 18, 2018 Received: June 20, 2018 Dear Soeun Chang: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); {1}------------------------------------------------ and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Hsl 2. Nils Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ### Indications for Use 510(k) Number (if known) K181626 Device Name GM85 Indications for Use (Describe) The GM85 Digital Mobile X-ray imaging System is intended for use in generating radiographic images of human anatomy by a qualified/trained doctor or technician. This device is not intended for mammographic applications. Type of Use (Select one or both, as applicable) | <div> <span style="font-size: 10pt;">☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> </div> | |---------------------------------------------------------------------------------------------------| | <div> <span style="font-size: 10pt;">☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> | CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ 510(k) Premarket Notification - Traditional ### Section 5: 510(k) Summary This summary of 510(k) safety and effectiveness information is being submitted accordance with requirements of 21 CFR 807.92 - 1. Date: June 18, 2018 #### 2. Submitter - Company Name: SAMSUNG ELECTRONICS Co., Ltd. A. - Address: 129, Samsung-ro, Yeongtong-gu, Suwon-si, Gyeonggi-do, 16677, B. Republic of Korea #### 3. Primary Contact Person - A. Name: SOEUN CHANG & CHUNSIN KIM - B. Title: Regulatory Affairs, Senior Professional - Phone Number: +82-2-2193-2436 & +82-2-2193-2437 ﻥ - FAX Number: +82-31-200-1601 D. E-Mail: soeun.chang@samsung.com & chulsin.kim@samsung.com #### 4. Secondary Contact Person - A. Name: Genci Omari - B. Title: Manager, Regulatory Affairs - C. Phone Number: 978-564-8602 - D. FAX Number: 978-560-0602 E-Mail: gomari@samsungneurologica.com #### 5. Proposed Device - A. Trade Name: GM85 - B. Device Name: GM85 - C. Common Name: Digital Diagnostic Mobile X-ray System - D. Classification Name: Mobile X-ray System - E. Product Code: IZL - ட் Regulation: 21 CFR 892.1720 #### 6. Predicate Devices | | Predicate Device | |----------------------|---------------------| | Device Name | GM85 | | Classification Name | MobileX-ray system. | | Product Code | IZL | | Regulation | 21 CFR 892.1720 | | 510(K)# | K180543 | | 510(K) Decision Date | May 24, 2018 | {4}------------------------------------------------ 510(k) Premarket Notification - Traditional #### 7. Device Description The GM85 Digital Mobile X-ray imaging System is used to capture images by transmitting X-ray to a patient's body. The X-ray passing through a patient's body is sent to the detector and then converted into electrical signals. These signals go through the process of amplification and digital data conversion in the signal process on the S-Station, which is the Operation Software (OS) of Samsung Digital Diagnostic X-ray System, and save in DICOM file, a standard for medical imaging. The captured images are tuned up by an Image Post-processing Engine (IPE) which is exclusively installed in S-Station, and send to the Picture Archiving & Communication System (PACS) sever for reading images. The GM85 Digital Mobile X-ray imaging System was previously cleared with K180543, and through this premarket notification, we would like to add more configurations in the previously cleared GM85 as a fixed column type and two detectors are newly added, and three software features are newly added as stated below. S-Enhance is renamed from Tube & Line Enhancement (TLE), which was cleared before with GM85 at K171119, to enhance visibility of tubes and lines and provide enhanced images separately from original images. In this submission, the scope of S-Enhance is expanded from tubes and lines on chest images to foreign body (e.g. tubes, lines and needles) and urinary stones on chest, abdomen, and L-spine. And Manual Stitching to capture a body part that is larger than the detector's by capturing multiple images and Remote View function to remote access to view the current image on the workstation through a web browser. It was determined that the level of concern for the software contained in the GM85 Digital Mobile X-ray imaging System was Moderate in accordance with the FDA guidance, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Device". #### 8. Intended Use The GM85 Digital Mobile X-ray imaging System is intended for use in generating radiographic images of human anatomy by a qualified/trained doctor or technician. This device is not intended for mammographic applications. #### 9. Summary of Technological characteristic of the proposed device compared with the predicate devices The proposed device, GM85, has the same technological characteristics and hardware as its original predicate device. GM85 (K180543), and added a new fixed column type, new types of detectors and image processing technology(S-Enhance, Manual stitching, and Remote View). It does not have significant changes in materials, energy source or technological characteristics compared to the predicated device. Comparisons of technological characteristics were executed and demonstrate the substantial equivalence to the predicates. #### A. Comparing with Predicate Device The proposed device is shown as its parts are identical or equivalent with predicate device while some differences are made as below, which does not show significant difference in safety and effectiveness. {5}------------------------------------------------ | Specification | Predicate Device | Proposed Device | Discussion | | |---------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------|------| | Device Name | GM85 | GM85 | - | | | Manufacturer | SAMSUNG ELECTRONICS | SAMSUNG ELECTRONICS | - | | | 510(k) Number | K180543 | - | - | | | Appearances | Image: Predicate Device Appearance | Image: Proposed Device Appearance<br>[C-Type*]<br>[F-Type**]<br>*Collapsible column type (C-Type)<br>**Fixed column type (F-Type) | Difference(1) | | | Intended Use | The GM85 Digital Mobile X-ray<br>imaging System is intended for<br>use in generating radiographic<br>images of human anatomy by a<br>qualified/trained doctor or<br>technician. This device is not<br>intended for mammographic<br>applications. | The GM85 Digital Mobile X-ray<br>imaging System is intended for<br>use in generating radiographic<br>images of human anatomy by a<br>qualified/trained doctor or<br>technician. This device is not<br>intended for mammographic<br>applications. | Same | | | Manufacturer Contents | Predicate Device | GM85 | Discussion | | | (1)High Voltage Generator | | | | | | Type | High Frequency | High Frequency | - | | | Max. Power | 32kW / 40kW | 32kW / 40kW | Same | | | Output<br>RANGE | kVp Range | 40 to 150kVp | 40 to 150kVp | Same | | | mA Range | 10 - 500mA | 10 - 500mA | Same | | | Exposure Time | 1msec-10sec | 1msec-10sec | Same | 510(k) Premarket Notification - Traditional | Manufacturer Contents | | Predicate Device | GM85 | | Discussion | |-----------------------|------------------|-----------------------|-----------------------|-----------------------|---------------| | | | | C-type | F-type | | | (2)Tube assembly | | | | | | | Moving<br>Range | Horizontal | 793~1355mm | 793~1355mm | | Same | | | Vertical | 550~2030mm | 550~2030mm | 550~2030 or<br>1850mm | Difference(2) | | Rotation<br>Range | Column | $\pm$ 315° | $\pm$ 315° | | Same | | | Tube (Arm axis) | $\pm$ 180° | $\pm$ 180° | | Same | | | Tube (Tube axis) | -30°~90° | -30°~90° | | Same | | Tube | Model | LUC-13L,<br>XRR-3332X | LUC-13L,<br>XRR-3332X | | Same | | | Focal spot | 0.6/1.2mm | 0.6/1.2mm | | Same | | | Target Angle | 14° | 14° | | Same | {6}------------------------------------------------ 510(k) Premarket Notification - Traditional | Manufacturer Contents | | Predicate Device | GM85 | | Discussion | |-----------------------|-------------------------|-----------------------------------------------|------------------------------------------------------|---------------------------------------------------|---------------| | | | | C-type | F-type | | | | Target Material | Rhenium-tungsten faced<br>molybdenum | Rhenium-tungsten faced<br>molybdenum | | Same | | | Nominal Tube<br>Voltage | 150kVp | 150kVp | | Same | | | Max.Anode HU | 300kHU | 300kHU | | Same | | Collimator | | Automatic<br>SDR-OGCL40U<br>212 X 306 X 179mm | Automatic<br>SDR-<br>OGCL40U<br>212 X 306 X<br>179mm | Manual<br>SDR-<br>OGCL41U<br>222 X 271 X<br>140mm | Difference(3) | | Manufacturer Contents | Predicate Device | GM85 | Discussion | | | |----------------------------------------------|------------------------------------------------------------------------------------------------|------------------------------|------------------------------------------------------------------------------------------------|------------------------------|-----------------| | (3) Detector | | | | | | | Name | S4335-W<br>S4343-W<br>S3025-W | | S4335-W<br>S4343-W<br>S3025-W<br>S4335-AW<br>S4343-AW | | Difference(4) | | | S4335-W | S4343-W | S4335-AW | S4343-AW | | | Detector Type | Csl | Csl | Csl | Csl | Same | | | Indirect | Indirect | Indirect | Indirect | Same | | Detector Area | 14"X17"<br>(345mmX42<br>5mm) | 17"X17"<br>(425mmX42<br>5mm) | 14"X17"<br>(345mmX4<br>25mm) | 17"X17"<br>(425mmX4<br>25mm) | Same | | Number of pixels | 2466X3040 | 3036X3040 | 2466X3040 | 3036X3040 | Same | | Pixel Pitch(um) | 140 | 140 | 140 | 140 | Same | | High Contrast Limiting<br>Resolution (LP/mm) | 3.57 | 3.57 | 3.57 | 3.57 | Same | | Communication | Wired /<br>Wireless | Wired /<br>Wireless | Wired /<br>Wireless | Wired /<br>Wireless | Same | | Dust/Water-resistance | IPx1 | | IP54 | | Difference(4)-1 | | Max.load capacity | 150 kg for uniform load,<br>100 kg for local load (40<br>mm in diameter disk at the<br>center) | | 400 kg for uniform load,<br>200 kg for local load (40<br>mm in diameter disk at the<br>center) | | Difference(4)-2 | | Manufacturer Contents | Predicate Device | GM85 | Discussion | |-----------------------------|------------------|--------------|------------| | (4) Grid | | | | | Lines/cm | 84.6 | 84.6 | Same | | Grid mechanism | Stationary | Stationary | Same | | Removability | Removable | Removable | Same | | (5) Weight Distribution Cap | | | | | Model Name | SDR-OGWD80U | SDR-OGWD80U | Same | | Size(mm) | 505x553x37.4 | 505x553x37.4 | Same | | (6) Software Features | | | | | SimGrid | SimGrid | SimGrid | Same | {7}------------------------------------------------ | Manufacturer Contents | Predicate Device | GM85 | Discussion | |------------------------------------------------|-------------------------------|------------------|---------------| | Tube & Line<br>Enhancement(TLE)<br>/ S-Enhance | TLE*<br>*FDA-cleared(K171119) | S-Enhance | Difference(5) | | Pediatric Exposure<br>Management(PEM) | PEM | PEM | Same | | S-DAP<br>(Dose Area Product) | S-DAP | S-DAP | Same | | S-Align | S-Align | S-Align | Same | | S-Share | S-Share | S-Share | Same | | Bone Suppression<br>Image(BSI) | BSI | BSI | Same | | Remote View | | Remote View | Difference(6) | | Manual Stitching | | Manual Stitching | Difference(7) | 510(k) Premarket Notification - Traditional | No | Differences | Explanation | |-------|---------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | (1) | Appearances | The proposed device GM85 adds newly a fixed column type in comparison with the predicated device's one that has only collapsible column type. The fixed column type has two options about the column's height, 1,800mm and 1,980mm. It does not contribute any adverse impacts to the device's image quality. The variety of column's types can be more helpful to choose the right equipment for fitting user's needs. | | (2) | Moving Range<br>(Vertical) | The vertical moving range of proposed device GM85 varies depending on the column type. | | (3) | Collimator | The proposed device GM85 has different types of collimator depending on column type. The collapsible column type has the same collimator(automatic, SDR-OGCL40U) applied to the predicate device and the fixed column type has the new type of collimator(manual, SDR-OGCL41U) which does not contribute any adverse impacts to the device's image quality. | | (4)-1 | Detector<br>Dust/Water-resistance | The new detectors which are added to the GM85 device have better dust/water-resistance than the predicate device's dust/water-resistance and this change does not contribute any adverse impact to the device's safety and effectiveness. | | (4)-2 | Detector<br>Max.load capacity | The new detectors which are added to the GM85 device have higher max load capacity than the predicate device's max load capacity and this change does not contribute any adverse impact to the device's safety and effectiveness. | | (5) | Tube & Line<br>Enhancement(TLE) / S-Enhance | Tube & Line Enhancement (TLE), which was cleared with K171119, is to enhance visibility of tubes and lines on chest images and provide enhanced images separately from original images. TLE is renamed as S-Enhance, expanding its coverage and scope in this proposed device. S-Enhance is expanded from tubes and lines on chest images to foreign body (e.g. tubes, lines and needles) and urinary stones on chest, abdomen, and L-spine. This change is considered low risk and does not | {8}------------------------------------------------ | | | contribute any adverse impact to the device's safety and<br>effectiveness. | |-----|------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | (6) | Remote View | The function of Remote View, which allows remote access to<br>view the current image on the workstation through a web<br>browser, is applied to the GM85 and this change does not<br>contribute any adverse impact to the device's safety and<br>effectiveness. | | (7) | Manual Stitching | Manual Stitching can capture a body part that is larger than the<br>detector's area by capturing multiple images. This function is<br>considered low risk and does not contribute any adverse impact<br>to the device's safety and effectiveness. | 510(k) Premarket Notification - Traditional #### B. Safety, EMC and Performance Data Electrical, mechanical, environmental safety and performance testing according to standard ES 60601-1,IEC 60601-1-2, IEC 60601-1-3, IEC 60601-2-28, IEC 60601-2-54, ISO14971, 21CFR1020.30 and 21CFR1020.31 were performed, and EMC testing was conducted in accordance with standard IEC 60601-1-2. Wireless function was tested and verified followed by guidance, Radio frequency Wireless Technology in Medical Devices. All test results were satisfying the standards. #### C. Non-clinical data Non-clinical data was provided in conformance to the FDA "Guidance for the Submission of 510(k)'s for Solid-State X-ray Imaging Devices", which includes MTF and DQE measurements as tested by IEC 62220-1. The proposed device has new types of detectors that have equivalent image characteristics as the existing ones. Specific description is added to make it clear with the non-clinical data and phantom image evaluation report. And those detectors evaluated by Software System Test Case for verification and validation. The S-Enhance is same image processing technology as TLE of predicate device GM85 (K171119) and extends the applicable body parts to chest, abdomen, and L-spine so it shows no difference in non-clinical testing data such as TLC and CER measurements from TLE of the predicate device GM85 (K171119). Manual Stitching is optional software to capture a body part that is larger than the detector's area and evaluated Software System Test Case for verification and validation. #### D. Clinical data In clinical data, phantom image evaluations of the new types of detectors and S-Enhance were performed in accordance with FDA guidance for the submission of 510(k)'s for Solid State X-ray Imaging Devices. Anthropomorphic phantom images were provided; these images were not necessary to establish substantial equivalence based on the modifications to the device (note X-ray flat-panel detector similar to the predicate detector) but they provide further evidence in addition to the performance data to show {9}------------------------------------------------ #### 510(k) Premarket Notification - Traditional that the complete system works as intended. They were evaluated by a professional radiologist and found to be equivalent to the predicate devices. There is no significant difference in the average score considering the standard deviation of image quality evaluation between the proposed device and the predicate device and it is confirmed that S-Enhance is able to generate a companion image which provide clear visibility for foreign bodies such as lines, tubes, and needles, and urinary stones in chest, abdomen, and L-spine protocol in addition to the original images. Therefore, these changes do not affect either the safety or the effectiveness, compared to the predicated device. #### E. Conclusions The non-clinical and clinical demonstrates that the proposed device is as safe, as effective, and performs as well as the legally marketed predicate devices.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...