The GC85A Digital X-ray Imaging System is intended for use in generating radiographic images of human anatomy by a qualified/trained doctor or technician. This device is not intended for mammographic applications.
Device Story
Stationary digital X-ray system; captures radiographic images of human anatomy. X-rays pass through patient to detector; converted to electrical signals; processed by S-station (Operation Software) for amplification and digital conversion; saved as DICOM files. Image Post-processing Engine (IPE) tunes images; sent to PACS for radiologist review. Operated by qualified doctors or technicians in clinical settings. Includes S-Enhance (image enhancement for tubes, lines, needles, urinary stones), Pediatric Exposure Management (PEM) for dose optimization, and Remote View for workstation access. Enhancements aid clinical decision-making by improving visibility of foreign bodies and anatomical structures. Benefits include optimized pediatric imaging and improved diagnostic clarity.
Clinical Evidence
Bench testing only; no clinical trials on human subjects. Performance validated via MTF and DQE measurements per IEC 62220-1. Phantom image evaluations performed for new detectors and S-Enhance feature. Radiologist assessment of anthropomorphic phantom images confirmed image quality equivalence to predicates, with S-Enhance providing clear visibility for foreign bodies and urinary stones.
Technological Characteristics
Stationary X-ray system; CsI indirect flat-panel detectors; high voltage generator. Connectivity: DICOM standard, PACS integration, Remote View web-based access. Software: S-station OS with IPE. Standards: ES 60601-1, IEC 60601-1-2, IEC 60601-1-3, IEC 60601-2-28, IEC 60601-2-54, ISO 14971, 21 CFR 1020.30/31. Detectors feature IP54 dust/water resistance and up to 400kg load capacity.
Indications for Use
Indicated for generating radiographic images of human anatomy for patients requiring diagnostic X-ray imaging. Not intended for mammographic applications.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
July 20, 2018
Samsung Electronics Co., Ltd. % Jaesang Noh Senior Professional, Regulatory Affairs 129, Samsung-ro, Yeongtong-gu Suwon-si, Gyeonggi-do 16677 REPUBLIC OF KOREA
Re: K181629
Trade/Device Name: GC85A Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: KPR Dated: June 15, 2018 Received: June 20, 2018
Dear Jaesang Noh:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice
(https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Hole 2. Nils
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
510(k) Number (if known)
K181629
Device Name GC85A
Indications for Use (Describe)
The GC85A Digital X-ray Imaging System is intended for use in generating radiographic images of human anatomy by a qualified/trained doctor or technician. This device is not intended for mammographic applications.
> Prescription Use (Part 21 CFR 801 Subpart D) ] Over-The-Counter Use (21 CFR 801 Subpart C) CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services
Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Type of Use (Select one or both, as applicable)
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the Samsung logo. The word "SAMSUNG" is written in white, bold letters. The word is set against a blue, oval-shaped background.
510(k) Premarket Notification - Traditional
### Section 5: 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted accordance with requirements of 21 CFR 807.92
- 1. Date: June 15, 2018
#### 2. Submitter
- A. Company Name: SAMSUNG ELECTRONICS Co., Ltd.
- B. Address: 129, Samsung-ro, Yeongtong-gu, Suwon-si, Gyeonggi-do, 16677, Republic of Korea
#### 3. Primary Contact Person
- A. Name: JAESANG NOH & CHUNSIN KIM
- B. Title: Regulatory Affairs, Senior Professional
- C. Phone Number: +82-2-2193-2444 & +82-2-2193-2437
- D. FAX Number: +82-31-200-6401
- E-Mail: jaesang.noh@samsung.com & chulsin.kim@samsung.com ய்
#### 4. Secondary Contact Person
- A. Name: Genci Omari
- B. Title: Manager, Regulatory Affairs
- C. Phone Number: 978-564-8602
- D. FAX Number: 978-560-0602
- E. E-Mail: gomari@samsungneurologica.com
### 5. Proposed Device
- A. Trade Name: GC85A
- B. Device Name: GC85A
- C. Common Name: Digital Diagnostic X-ray System
- D. Regulation Name: Stationary X-ray system
- E. Product Code: KPR
- ட் Regulation: 21 CFR 892.1680
#### 6. Predicate Devices
| | Predicate Device #1 | Predicate Device #2 |
|----------------------|----------------------------|---------------------|
| Device Name | GC85A | GM85 |
| Classification Name | Stationary X-ray<br>System | Mobile X-ray System |
| Product Code | KPR | IZL |
| Regulation | 21 CFR 892.1680 | 21 CFR 892.1720 |
| 510(K)# | K172229 | K171119 |
| 510(K) Decision Date | Nov., 22, 2017 | May, 12, 2017 |
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the Samsung logo. The word "SAMSUNG" is written in white, bold letters. The word is placed on a blue, rounded, rectangular background.
510(k) Premarket Notification - Traditional
#### 7. Device Description
The GC85A digital X-ray imaging system is a stationary x-ray system designed for general radiography and used to capture images by transmitting X-ray to a patient's body. The Xray passing through a patient's body is sent to the detector and then converted into electrical signals. These signals go through the process of amplification and digital data conversion in the signal process on the S-station, which is the Operation Software (OS) of Samsung Digital Diagnostic X-ray System, and save in DICOM file, a standard for medical imaging. The captured images are tuned up by an Image Post-processing Engine (IPE) which is exclusively installed in S-station, and sent to the Picture Archiving & Communication System (PACS) sever for reading images.
The GC85A digital X-ray imaging system was previously cleared with K172229, and through this premarket notification, we would like to add more configurations in the previously cleared GC85A as three High Voltage Generators and two detectors are newly added, and three software features are newly added as stated below.
S-Enhance is renamed from Tube & Line Enhancement (TLE), which was cleared before with the predicate device GM85 at K171119, to enhance visibility of tubes and lines and provide enhanced images separately from original images. In this submission, the scope of S-Enhance is expanded from tubes and lines on chest images to foreign body (e.g. tubes, lines and needles) and urinary stones on chest, abdomen, and L-spine. And Pediatric Exposure Management (PEM), which was cleared before with the predicate device GM85 at K171119, is subdivided patient size and exposure conditions especially for pediatric patients based on weight and protocols, and Remote View to enable the images on the device is being displayed on the remote monitor. It was determined that the level of concern for the software contained in the GC85A digital X-ray imaging system was Moderate in accordance with the FDA guidance, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Device".
#### 8. Intended Use
The GC85A Digital X-ray Imaging System is intended for use in generating radiographic images of human anatomy by a qualified/trained doctor or technician. This device is not intended for mammographic applications.
#### 9. Summary of Technological characteristic of the proposed device compared with the predicate device
The proposed device GC85A apply software features as S-Enhance and Pediatric Exposure, which were cleared with K171119, and Remote View to GC85A product which was cleared with K172229 without changes in technical characteristics, materials, energy sources and biocompatibility such as X-ray Tube Assembly except of high voltage generator and detector. Optional high voltage generators and detectors which were improved durability and dust/water-resistance are added to GC85A.
Comparisons of technological characteristics were executed and demonstrate the substantial equivalence to the predicate.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the Samsung logo. The word "SAMSUNG" is written in white, bold letters. The word is placed inside of a blue, oval shape.
510(k) Premarket Notification - Traditional
| Specification | Predicate<br>Device #1 | Proposed<br>Device | Predicate<br>Device #2 | Discussion |
|---------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|
| Device Name | GC85A | GC85A | GM85 | |
| Manufacturer | SAMSUNG<br>ELECTRONICS<br>Co., Ltd. | SAMSUNG<br>ELECTRONICS<br>Co., Ltd. | SAMSUNG<br>ELECTRONICS<br>Co., Ltd. | |
| 510(k) Number | K172229 | None | K171119 | |
| Appearances | Image: GC85A X-ray system | Image: GC85A X-ray system | Image: GM85 Mobile X-ray system | Same as PD#1 |
| Intended Use | The GC85A<br>digital X-ray<br>imaging system<br>is intended for<br>use in<br>generating<br>radiographic<br>images of<br>human anatomy<br>by a<br>qualified/trained<br>doctor or<br>technician. This<br>device is not<br>intended for<br>mammographic<br>applications. | The GC85A<br>Digital X-ray<br>Imaging System<br>is intended for<br>use in<br>generating<br>radiographic<br>images of<br>human anatomy<br>by a<br>qualified/trained<br>doctor or<br>technician. This<br>device is not<br>intended for<br>mammographic<br>applications. | The GM85<br>Digital Mobile X-<br>ray imaging<br>System is<br>intended for use<br>in generating<br>radiographic<br>images of<br>human anatomy<br>by a<br>qualified/trained<br>doctor or<br>technician. This<br>device is not<br>intended for<br>mammographic<br>applications. | Same as PD#1 |
| t | Comparing with Predicate Device | | | |
|---|---------------------------------|--|--|--|
|---|---------------------------------|--|--|--|
| Manufacturer<br>Contents | GC85A<br>(K172229) | GC85A | GM85<br>(K171119) | Discussion |
|---------------------------|--------------------|------------|-------------------|---------------|
| (1)High Voltage Generator | | | | |
| Max. Power | 82kW | 82kW<br> | 32kW / 40kW | Difference(1) |
| Output<br>RANGE | 40-150kV | 40 - 150kV | 40-150kV | Same |
| | Tube<br>Voltage | <br> | | |
| | Tube<br>Curr | 10-1000mA | 10 - 500mA | 10 - 500mA |
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the Samsung logo. The word "SAMSUNG" is written in white, bold letters. The background is a blue, rounded rectangle.
510(k) Premarket Notification - Traditional
| Manufacturer<br>Contents | | GC85A<br>(K172229) | GC85A | | | GM85<br>(K171119) | Discussion |
|----------------------------------------------------|--------------------------|--------------------|-------------------------------------|-------------------------------------|-------------------------------------|-------------------|--------------|
| | ent | | m<br>A | m<br>A | A<br>A | | |
| | Exp<br>osur<br>e<br>Time | 1msec-10sec | 1<br>ms<br>ec<br>-<br>10<br>se<br>c | 1<br>ms<br>ec<br>-<br>10<br>se<br>c | 1<br>ms<br>ec<br>-<br>10<br>se<br>c | 1msec-10sec | Same |
| AEC<br>(Automatic<br>Exposure<br>Control) | | Yes | Yes | | | - | Same as PD#1 |
| APR<br>(Anatomically<br>Programmed<br>Radiography) | | Yes | Yes | | | - | Same as PD#1 |
| Manufacturer<br>Contents | GC85A (K172229) | GC85A | GM85<br>(K171119) | Discussion |
|----------------------------------------------------|---------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------|---------------------|
| (2) Detector | | | | |
| Name | S4335-W<br>S4343-W<br>S3025-W | S4335-W<br>S4343-W<br>S3025-W<br>S4335-AW<br>S4343-AW | S4335-W<br>S4343-W<br>S3025-W | Difference(3) |
| | S4335-W | S4335-AW | S4335-W | |
| | S4343-W | S4343-AW | S4343-W | |
| Detector Type | Csl | Csl | Csl | Same |
| | Indirect | Indirect | Indirect | Same |
| Detector Area | 14"X17"<br>(345mm<br>X425m<br>m) | 17"X17"<br>(425mm<br>X425m<br>m) | 14"X17"<br>(345mm<br>X425m<br>m) | Same |
| | 17"X17"<br>(425mm<br>X425m<br>m) | 14"X17"<br>(345mm<br>X425m<br>m) | 17"X17"<br>(425mm<br>X425m<br>m) | Same |
| Number of pixels | 2466X3<br>040 | 3036X3<br>040 | 2466X3<br>040 | Same |
| | 3036X3<br>040 | 2466X3<br>040 | 3036X3<br>040 | Same |
| Pixel Pitch(um) | 140 | 140 | 140 | Same |
| High Contrast<br>Limiting<br>Resolution<br>(LP/mm) | 3.57 | 3.57 | 3.57 | Same |
| Communication | Wired /<br>Wireless | Wired /<br>Wireless | Wired /<br>Wireless | Same |
| Dust/Water-<br>resistance | IPx1 | IP54 | IPx1 | Difference(3)<br>-1 |
| Max.load<br>capacity | 150 kg for uniform<br>load,<br>100 kg for local<br>load (40 mm in<br>diameter disk at | 400 kg for uniform<br>load,<br>200 kg for local<br>load (40 mm in<br>diameter disk at | 150 kg for uniform<br>load,<br>100 kg for local<br>load (40 mm in<br>diameter disk at | Difference(3)<br>-2 |
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the SAMSUNG logo. The logo is white text on a blue background. The text is in a bold, sans-serif font.
510(k) Premarket Notification - Traditional
| Manufacturer<br>Contents | GC85A (K172229) | GC85A | GM85<br>(K171119) | Discussion |
|--------------------------|-----------------|-------------|-------------------|------------|
| | the center) | the center) | the center) | |
| Manufacturer<br>Contents | GC85A (K172229) | GC85A | GM85<br>(K171119) | Discussion |
|--------------------------|-----------------|-------------|-------------------|-----------------|
| (3) Software Features | | | | |
| Feature<br>Names | SimGrid | SimGrid | SimGrid | Same |
| | S-Guide | S-Guide | - | Same as<br>PD#1 |
| | - | S-Enhance | TLE | Difference(4) |
| | - | PEM | PEM | Same as<br>PD#2 |
| | S-DAP | S-DAP | S-DAP | Same |
| | S-Align | S-Align | S-Align | Same |
| | S-Share | S-Share | S-Share | Same |
| | BSI | BSI | BSI | Same |
| | - | Remote View | - | Difference(5) |
| No | Differences | Explanation |
|-------|-----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| (1) | HVG Power rating | Optional HVGs which are added to the GC85A device have lower max power than the predicate device's max power and these changes do not contribute any adverse impact to the device's safety and effectiveness. |
| (2) | HVG mA range | Optional HVGs which are added to the GC85A device have lower max current than the predicate's max current and these changes do not contribute any adverse impact to the device's safety and effectiveness. |
| (3)-1 | Dust/Water-resistance | The new detectors which are added to the GC85A device have better dust/water-resistance than the predicate device's dust/water-resistance and this change does not contribute any adverse impact to the device's safety and effectiveness. |
| (3)-2 | Max.load capacity | The new detectors which are added to the GC85A device have higher max load capacity than the predicate device's |
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the Samsung logo. The word "SAMSUNG" is written in white, bold, sans-serif font. The word is placed inside a blue, rounded, oval shape. The logo is simple and modern.
510(k) Premarket Notification - Traditional
| | | max load capacity and this change does not contribute any<br>adverse impact to the device's safety and effectiveness. |
|-----|-------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| (4) | S-Enhance | S-Enhance is renamed from Tube & Line Enhancement<br>(TLE), which was cleared with K171119, is to enhance<br>visibility of tubes and lines and provide enhanced images<br>separately from original images. This software feature is<br>applied to the GC85A with the scope of application for S-<br>Enhance, which is expanded from tubes and lines on chest<br>images to foreign body (e.g. tubes, lines and needles) and<br>urinary stones on chest, abdomen and L-spine images.<br>This change is considered low risk and does not contribute<br>any adverse impact to the device's safety and<br>effectiveness. |
| (5) | Remote View | The function of Remote View, which allows remote access<br>to view the current image on the workstation through a web<br>browser, is applied to the GC85A and this change does not<br>contribute any adverse impact to the device's safety and<br>effectiveness. |
#### B. Safety, EMC and Performance Data
Electrical, mechanical, environmental safety and performance testing according to standard ES 60601-1,IEC 60601-1-2,IEC 60601-1-3,IEC 60601-2-28, IEC 60601-2-54, ISO14971, 21CFR1020.30 and 21CFR1020.31 were performed, and EMC testing was conducted in accordance with standard IEC 60601-1-2. Wireless function was tested and verified followed by guidance, Radio frequency Wireless Technology in Medical Devices. All test results were satisfying the standards.
#### C. Non-clinical data
Non-clinical testing data was provided in conformance to the FDA "Guidance for the Submission of 510(k)'s for Solid-State X-ray Imaging Devices", which includes MTF and DQE measurements as tested by IEC 62220-1. The proposed device shows no difference in non-clinical testing data such as MTF and DQE measurements form the predicate device.
#### D. Clinical data
Phantom image evaluations for the new detector and S-Enhance were performed in accordance with FDA guidance for the submission of 510(k)'s for Solid State X-ray Imaging Devices. Anthropomorphic phantom images were provided; these images were not necessary to establish substantial equivalence based on the modifications to the device (note X-ray flat-panel detector similar to the predicate detector) but they provide further evidence in addition to the performance data to show that the complete system works as intended. They were evaluated by professional radiologists and found to be equivalent to the predicate devices. There is no significant difference in the average score considering the standard deviation of image quality evaluation between the proposed device and the predicate device and it is confirmed that S-Enhance is able to generate a companion image which provide clear visibility for foreign bodies such as lines, tubes, and
{9}------------------------------------------------
Image /page/9/Picture/1 description: The image shows the Samsung logo. The logo consists of the word "SAMSUNG" in white, bold, sans-serif font. The word is set against a blue, oval-shaped background. The oval is tilted slightly upward from left to right.
510(k) Premarket Notification - Traditional
needles, and urinary stones in chest, abdomen, and L-spine protocol in addition to the original images. Therefore, these changes do not affect either the safety or the effectiveness, compared to the predicated device.
#### E. Conclusions
The non-clinical and clinical data demonstrate that the proposed device is as safe, as effective, and performs as well as the legally marketed device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.