DRX-Revolution Nano Mobile X-ray System

K173924 · Carestream Health, Inc. · IZL · Feb 5, 2018 · Radiology

Device Facts

Record IDK173924
Device NameDRX-Revolution Nano Mobile X-ray System
ApplicantCarestream Health, Inc.
Product CodeIZL · Radiology
Decision DateFeb 5, 2018
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1720
Device ClassClass 2
AttributesPediatric

Indications for Use

The device is designed to perform radiographic x-ray examinations on pediatric and adult patients, in all patient treatment areas. The Nano system used to generate and control x-rays for diagnostic procedures.

Device Story

Mobile digital radiography (DR) system; generates/controls x-rays for diagnostic imaging outside stationary rooms. Input: x-ray exposure parameters; Output: digital radiographic images. Used in clinical settings (e.g., patient treatment areas) by healthcare professionals. Modifications to x-ray tube (tungsten thickness, focal spot size) and generator power (reduced to 5.3 kW) improve reliability and lifespan. System includes x-ray generator, image receptor, display, and software. Images reviewed on PACS workstation; radiologists use window/level/magnification tools for diagnostic assessment. Facilitates rapid diagnostic decision-making at point-of-care; benefits patients by providing mobile imaging capability.

Clinical Evidence

Phantom-based imaging study using anthropomorphic phantoms (adult chest/pelvis, pediatric upper/lower limb, neonatal chest). Three board-certified radiologists evaluated 60 images using the RadLex rating scale (1.0-4.0). Mean RadLex rating was 3.595 (SD 0.220). 100% of images rated as 'Diagnostic' (3) or better. One-sample t-test confirmed mean rating significantly greater than 3.5 (p < 0.05). No significant difference between readers (ANOVA p > 0.05).

Technological Characteristics

Mobile x-ray system; digital radiography (DR) technology. Components: self-contained x-ray generator, image receptor, display, software. Modifications: adjusted tungsten anode thickness, focal spot size, 5.3 kW generator power. Connectivity: PACS workstation integration. Software: embedded, modified to accommodate tube design changes.

Indications for Use

Indicated for radiographic x-ray examinations on pediatric and adult patients in all patient treatment areas.

Regulatory Classification

Identification

A mobile x-ray system is a transportable device system intended to be used to generate and control x-ray for diagnostic procedures. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration. The Department of Health & Human Services logo is on the left, and the U.S. Food & Drug Administration logo is on the right. The FDA logo is blue and white and includes the words "FDA U.S. FOOD & DRUG ADMINISTRATION". February 5, 2018 Carestream Health, Inc. % Ms. Victoria Wheeler Sr. Regulatory Affairs Manager 150 Verona Street ROCHESTER NY 14608 Re: K173924 Trade/Device Name: DRX-Revolution Nano Mobile X-ray System Regulation Number: 21 CFR 892.1720 Regulation Name: Mobile x-ray system Regulatory Class: II Product Code: IZL Dated: December 27, 2017 Received: December 28, 2017 Dear Ms. Wheeler: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Michael D. O'Hara For Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ### Indications for Use 510(k) Number (if known) K173924 Device Name DRX-Revolution Nano Mobile X-ray System Indications for Use (Describe) The device is designed to perform radiographic x-ray examinations on pediatric and adult patients treatment areas. Type of Use (Select one or both, as applicable) X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # Carestream ## "510(k) Summary" | 510(k) Owner Name:<br>510(k) Owner Address: | Carestream Health, Inc.<br>150 Verona Street<br>Rochester, New York 14608 | |-------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | 510(k) Owner Phone:<br>510(k) Owner Fax: | 585-627-8706<br>585-627-8802 | | Contact Person & Info: | Victoria Wheeler<br>Senior Regulatory Affairs Manager<br>victoria.wheeler@carestream.com<br>585-627-8706 | | Date Summary Prepared: | January 24, 2018 | | Device Trade Name:<br>Device Common Name:<br>Classification Name: | DRX-Revolution Nano Mobile X-ray System<br>Mobile x-ray system<br>Mobile x-ray system | | Device Class:<br>Device Code:<br>Regulation Number: | Class II<br>IZL<br>21 CFR 892.1720 | | Predicate Device: | DRX- Revolution Nano Mobile X-ray System<br>Manufactured by: Carestream Health, Inc.<br>510(k) No.: K170755 (June 21, 2012)<br>Classification Regulation: 21 CFR 892.1720<br>Classification Name: Mobile x-ray system<br>Primary Product Code: IZL | #### Device Description: Carestream Health, Inc. is submitting this Traditional 510(k) premarket notification for a modification to the cleared DRX-Revolution Nano Mobile X-ray System (K170755). The product will continue to be marketed as the DRX-Revolution Nano Mobile X-ray System. Consistent with the original system, the modified DRX-Revolution Nano Mobile X-ray System (also referred to throughout this document as the Nano system) is a diagnostic mobile x-ray system utilizing digital radiography (DR) technology. The system consists {4}------------------------------------------------ of a self contained x-ray generator, image receptor(s), imaging display and software for acquiring medical diagnostic images outside of a standard stationary x-ray room. The modified Nano system can also be used to expose Computed Radiography (CR) storage phosphor or film cassettes. Modifications have been made to the design of the system's x-ray tube and its operational characteristics for the purpose of improving the x-ray tube reliability and life span. This required modifications to the embedded software to accommodate these tube design changes. #### Indications for Use The Indications for Use for the device, as described in its labeling, are: "The device is designed to perform radiographic x-ray examinations on pediatric and adult patients, in all patient treatment areas." #### Intended Use The intended use for this device, as determined by descriptions and the proposed labeling contained in this submission, is similar to the Indications for Use statement provided above. The Nano system used to generate and control x-rays for diagnostic procedures. We believe that the Nano system and the predicate device have the same intended use. #### Comparison of Technological Characteristics: Modifications have been made to the design of the x-ray tube and its operational characteristics for the purpose of improving the x-ray tube reliability and life span. The technical characteristics of the modified device are not different from the predicate device except for minor adjustments to tungsten thickness (on the anode), focal spot size, and maximum exposure current. These adjustments led to a reduction in generator power from 6.5 kW to 5.3 kW. Based upon information provided within this submission, we believe that the modified DRX-Revolution Nano Mobile X-ray System is substantially equivalent to the legally marketed DRX-Revolution Nano X-ray System (K170755). #### Discussion of Non-clinical Testing: The performance characteristics and operation / usability of the DRX-Revolution Mobile Nano System were evaluated in non-clinical (bench) testing. These studies have demonstrated the intended workflow, related performance, overall function, shipping performance, verification and validation of requirements for intended use, and reliability of the system including both software and hardware requirements. Non-clinical test results have demonstrated that the device conforms to its specifications. Predefined acceptance criteria were met and demonstrated that the device is as safe, as effective, and performs as well as or better than the predicate device. {5}------------------------------------------------ #### Discussion of Clinical Testing (using Phantoms): The diagnostic imaging performance of the modified DRX-Revolution Nano system was evaluated by executing a phantom imaging study. #### Anthropomorphic Phantoms The phantoms used for the imaging study were purchased commercial products intended for x-ray imaging tests. The phantoms represent different sections of the human anatomy and are anthropomorphic both in appearance, size, and x-ray attenuation characteristics. For example, the adult chest phantom ("Lungman") was manufactured by Kyoto Kagaku and it is widely accepted in the scientific community for x-ray imaging quality assessment. The particular phantom used contains actual human bone and simulated lung tissues. Similarly, the adult pelvis phantom, also used in this study, contains actual human bones. As both the chest and the pelvis phantom only mimic a medium sized adult patient, larger sized patients were simulated by additional PMMA (acrylic glass) plates each of 2.5 cm in thickness to the two phantoms. For example, images were taken with no additional PMMA on the phantoms, one PMMA plate in the front of the phantoms, and one PMMA plate in both the front and the back of the phantoms (two PMMA plates in total). In summary, these two phantoms cover the anatomical regions of the lung, mediastinum, diaphragm, ribs, clavicles, shoulder, inferior c-spine, t-spine, l-spine, pelvis, and superior femur. The other two anthropomorphic phantoms used in the clinical study were an upper limb and a lower limb representing anatomical size and structure of a typical five year old child. These phantoms were recently manufactured by CIRS Inc (Model number 715-A-L-FX and 715-L-R). They contain simulated human bone skeletons and also simulated soft tissue/muscle/fat around the skeletons. These phantoms are intended for imaging at both diagnostic and therapy x-ray energy ranges. These two phantoms cover the anatomical regions of the interior femur, knee, tibia/fibula (lower leg), ankle, foot, inferior humerus, elbow, ulna/radius (forearm), ankle, and hand. The standard Gammex 610 neonatal chest phantom manufactured by Supertech, was used for the neonate imaging. It is designed for routine quality assurance monitoring of computed and digital radiography systems. This phantom replicates both the anatomic structure and the tissue attenuation characteristics of an actual human neonate. In summary, we believe the anthropomorphic phantoms used in this study adequately represent the device's broad range of intended uses. The study objective was to evaluate the imaging performance (diagnostic image quality) of the modified DRX-Revolution Nano Mobile X-ray System using a radiologist reader rating scale (RadLex Rating) that is applicable to projection radiography. Three (3) board certified radiologists with extensive reading experience (Table 1) rated the images using the RadLex scale for diagnostic quality. The radiologists provided image quality ratings from 1.0 to 4.0 using a graduated RadLex rating scale (to one decimal place) to account for ratings between the absolute values (e.g. 3.1, 3.5, or 3.9). {6}------------------------------------------------ For each reading session, the radiologists were given work lists of image cases to review on the CARESTREAM PACS Workstation. Images were randomized and displayed following a single stimulus format. In addition, the image work lists were randomized for the radiologists to avoid bias due to study learning curve or reader fatigue. The radiologists were allowed to pan, adjust window/level, and magnify on the PACS workstation while evaluating image quality. #### Summary of Statistical Analysis Methodology The statistical analysis is performed using Minitab® statistical software and Microsoft Excel for generating charts. Diagnostic quality ratings (RadLex Scale) are treated as a continuous variable for the descriptive statistics and hypothesis tests. The inferential tests are performed at the 5% level of significance. Descriptive statistics including mean, standard error of the mean, median, standard deviation, and sample size are presented in Table 1 for all RadLex rating data. | Statistic | Investigational RadLex<br>Rating | |--------------------|----------------------------------| | Mean | 3.595 | | Standard Error | 0.028 | | Median | 3.6 | | Standard Deviation | 0.220 | | Observation Count | 60 | Table 1 - RadLex Rating Descriptive Statistics Applying the one-sample t-test to the mean RadLex rating we find statistical significance. At a 95% significance level the mean of the DRX-Revolution Nano RadLex rating data is greater than the mid-value (3.5) of the Diagnostic (3) quality image rating scale. There is no significant difference in ratings across Readers. The ANOVA vields a p-value greater than alpha (0.05) which fails to reject the null hypothesis "All means are equal" (referring to the RadLex rating means of each Reader). In summary, the three different Readers are similar in their evaluation of the images. #### Conclusion The mean RadLex rating for the 60 responses is 3.6. 100% of the image ratings were Diagnostic (3) quality. Approximately 78 % of the total number of rating responses falls in the mid to upper end of the Diagnostic (3) RadLex Rating scale between Diagnostic and Exemplary. The one- sample t-test yields statistical significance at the 95% significance level. This result implies quality ratings equal to or better than a Diagnostic (3) quality rating. There is no significant difference in ratings across Readers. #### Summary The statistical test results and graphical summaries demonstrate the DRX-Revolution Nano Mobile X-ray System Study using Phantoms delivers quality imaging performance that is rated Diagnostic. Results of the Phantom Reader Study indicated that the diagnostic capability of the modified DRX-Revolution Nano Mobile X-ray System is statistically equivalent to or better than that of the predicate device.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...