K170755 · Carestream Health, Inc. · IZL · Jun 21, 2017 · Radiology
Device Facts
Record ID
K170755
Device Name
DRX-Revolution Nano Mobile X-ray System
Applicant
Carestream Health, Inc.
Product Code
IZL · Radiology
Decision Date
Jun 21, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1720
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The device is designed to perform radiographic x-ray examinations on pediatric and adult patients, in all patient treatment areas.
Device Story
Mobile digital radiography (DR) system for diagnostic imaging outside stationary x-ray rooms. System comprises self-contained X-ray generator, X-ray tube, collimator, touchscreen operator console, and battery-powered mobile cart. Captures images using wireless or tethered flat-panel detectors; compatible with CR storage phosphor screens or film cassettes. Operated by healthcare professionals in clinical settings. System software manages hardware, controls X-ray generation, and processes images. Wireless components include badge reader for sign-in, optional barcode reader for patient ID, and hospital network connectivity. Provides diagnostic images to clinicians to support clinical decision-making and patient diagnosis. Benefits include improved mobility and workflow efficiency compared to stationary systems.
Clinical Evidence
Comparative reader study evaluated diagnostic capability against predicate (K120062). Study used adult cadavers for 2D imaging (chest, abdomen, skull, spine, pelvis, extremities) and pediatric phantoms. Images reviewed in single-stimulus random format. Results demonstrated diagnostic capability of DRX-Revolution Nano is statistically equivalent to or better than the predicate device.
Technological Characteristics
Mobile X-ray system; stationary anode X-ray tube; battery-powered cart; articulating arm. Digital flat panel detectors (wireless/tethered). Connectivity: DICOM, hospital network, wireless badge/barcode readers. Software: DirectView platform. Standards: ISO 14971:2012, IEC 60601-1-2 Ed 3, IEC 60601-2-54.
Indications for Use
Indicated for radiographic x-ray examinations on pediatric and adult patients in all patient treatment areas.
Regulatory Classification
Identification
A mobile x-ray system is a transportable device system intended to be used to generate and control x-ray for diagnostic procedures. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Carestream Health, Inc. Victoria Wheeler Seinor Regulatory Affairs Manager, Regulatory & Quality Systems 150 Verona Street Rochester, New York 14608
June 21, 2017
Re: K170755
Trade/Device Name: DRX-Revolution Nano Mobile X-ray System Regulation Number: 21 CFR 892.1720 Regulation Name: Mobile X-Ray System Regulatory Class: Class II Product Code: IZL Dated: May 26, 2017 Received: May 30, 2017
Dear Ms. Victoria Wheeler:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Michael D. O'Hara
For
Robert A. Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K170755
#### Device Name
DRX-Revolution Nano Mobile X-ray System
Indications for Use (Describe)
The device is designed to perform radiographic x-ray examinations on adult patients, in all patient treatment areas.
Type of Use (Select one or both, as applicable)
| > Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Carestream
## "510(k) Summary"
| 510(k) Owner Name:<br>510(k) Owner Address: | Carestream Health, Inc.<br>150 Verona Street<br>Rochester, New York 14608 |
|-------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Owner Phone:<br>510(k) Owner Fax: | 585-627-8706<br>585-627-8802 |
| Contact Person & Info: | Victoria Wheeler<br>Senior Regulatory Affairs Manager<br>victoria.wheeler@carestream.com<br>585-627-8706 |
| Date Summary Prepared: | May 26, 2017 |
| Device Trade Name:<br>Device Common Name:<br>Classification Name: | DRX-Revolution Nano Mobile X-ray System<br>Mobile x-ray system<br>Mobile x-ray system |
| Device Class:<br>Device Code:<br>Regulation Number: | Class II<br>IZL<br>21 CFR 892.1720 |
| Predicate Device: | DRX- Revolution Mobile X-ray System<br>Manufactured by: Carestream Health, Inc.<br>510(k) No.: K120062 (April 13, 2012)<br>Classification Regulation: 21 CFR 892.1720<br>Classification Name: Mobile x-ray system<br>Primary Product Code: IZL |
### Device Description:
#### Device Description
The DRX-Revolution Nano Mobile X-ray System (also referred to as the DRX-Revolution Nano) is a diagnostic mobile x-ray system utilizing digital radiography (DR) technology. The system consists of a self contained x-ray generator, image receptor(s), imaging display and software for acquiring medical diagnostic images outside of a standard stationary x-ray room. The DRX-Revolution Nano system can also be used to expose CR storage phosphor screen or film cassettes.
{4}------------------------------------------------
The DRX-Revolution Nano system is designed to be used with the following legally marketed solid state imaging devices (referred to throughout this document as "detectors"):
- Carestream DRX-1 Detector (K090318); o
- Carestream DRX-1C Detector (K120062)
- DRX Plus 3543 Detector and DRX Plus 3543C Detector (K150766) ●
- DRX Plus 4343 Detector and DRX Plus 4343C Detector (K153142) ●
- DRX Core 3543 Detector (Relabeled version of the DRX Plus 3543 Detector) ●
- DRX Core 3543C Detector (Relabeled version of DRX Plus 3543C Detector) ●
- . DRX Core 4343 Detector (Relabeled version of the DRX Plus 4343 Detector)
- DRX Core 4343C Detector (Relabeled version of the DRX Plus 4343C Detector ●
- DRX 2530C Detector (K130464) ●
These detectors are commercially available for use with other Carestream products or as standalone devices. Image acquisition software for the detectors is integrated with the mobile x-ray system's user interface.
The DRX-Revolution Nano contains several wireless components that facilitate its use. Specifically, the Carestream detectors communicate to the console wirelessly, eliminating the need for cables that can hinder efficient workflow. The detector is also capable of communicating to the console via a wired ("tethered") connection if desired. Other peripheral wireless components include a badge reader for employee sign-in, an optional 2D barcode reader (for patient ID), and wireless communication capability with the hospital network.
#### Indications for Use
The Indications for Use for the device, as described in its labeling, are: «The device is designed to perform radiographic x-ray examinations on pediatric and adult patients, in all patient treatment areas."
#### Intended Use
The intended use for this device, as determined by descriptions and the proposed labeling contained in this submission, is similar to the Indications for Use statement provided above. The DRX-Revolution Nano is a mobile system used to generate and control x-rays for diagnostic procedures. We believe that the DRX-Revolution Nano and the predicate device have the same intended use.
{5}------------------------------------------------
#### Comparison of Technological Characteristics:
Based upon information provided within this submission, we believe that the DRX-Revolution Nano Mobile X-ray System is substantially equivalent to the legally marketed DRX-Revolution Mobile X-ray System. At a high level, the subject device and predicate devices are based on the same technological elements as listed below:
- X-ray generator ●
- X-ray tube ●
- Collimator
- Graphical user interface ●
- Operator console with Touchscreen Display ●
- Battery powered mobile cart ●
- Flat panel detector and o
- System software with image processing capability ●
- Optional IR remote o
The following technological characteristics in the subject device are different from those of the predicate device:
- Stationary anode X-ray tube
- Lower powered generator
- Collimator has lower projection radiography range ●
- Elimination of auxiliary Touchscreen Display on the tube head ●
- Use of different battery technology
- o Smaller sized cart
- Manually driven ●
- Use of articulating arm with increased rotation o
The DRX-Revolution Nano Mobile X-ray System (Nano) and the DRX-Revolution Mobile X-ray System both consist of an x-ray generator, x-ray tube, collimator, optional Infrared remote and graphical user interface on an operator console with touch screen monitor. These components are mounted on a mobile cart to enable the device to be moved from location to location. Both systems utilize a digital flat panel detector for image capture and can also be used to expose CR storage phosphor screen or film cassettes. Both systems have an operator console with the same image processing software. The Intended Use is the same for both devices.
Both systems have operator console software that is based on the DirectView platform, but specific user interface screens and software components are specific to the DRX-Revolution Nano device. These screens were developed to control and manage the DRX-Revolution Nano's hardware subsystems including its X-ray generator. All of the embedded software and subsystem firmware of the DRX-Revolution Nano was developed specifically for the new (Nano) hardware system.
{6}------------------------------------------------
The DRX-Revolution Nano offers hardware features (such as an articulating arm) that are different from the predicate device. These features are designed to aid the user in achieving an efficient workflow but do not result in additional indications for use (i.e., there is no change to the intended diagnostic effect) for the new device as compared to the predicate.
#### Performance Testing:
The DRX-Revolution Nano conforms to the following standards:
- ISO 14971:2012 Medical device application of risk management to medical device ●
- Digital Imaging and Communication in Medicine (DICOM)
- 0 IEC 60601-1-2 Edition 3: Medical electrical equipment, Part 1-2: General requirements for basic safety and essential performance
- IEC 60601-2-54: Medical electrical equipment Part 2-54: Particular requirements ● for the basic safety and essential performance of X-ray equipment for radiography and radioscopy.
#### Discussion of Testing
The performance characteristics and operation / usability of the DRX-Revolution Nano Mobile X-ray System were evaluated in non-clinical (bench) testing. These studies have demonstrated the intended workflow, related performance, overall function, shipping performance, verification and validation of requirements for intended use, and reliability of the system including both software and hardware requirements. Non-clinical test results have demonstrated that the device conforms to its specifications. Predefined acceptance criteria were met and test results havedemonstrated that the device is as safe, as effective, and performs as well as or better than the predicate device.
The following performance data were provided in support of the substantial equivalence determination:
- Electrical safety and electromagnetic compatibility (EMC)
Electrical safety and EMC testing were conducted on the DRX-Revolution Nano Mobile X-ray System in compliance to IEC 60601-1-2:2007 – Medical electrical equipment Part 1-2, Electromagnetic Compatibility.
#### ● Software Verification and Validation Testing
Software verification and validation testing was conducted and documentation has been provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in
{7}------------------------------------------------
Medical Devices." The software for this device is considered to present a "moderate" level of concern. This is based on the fact that a malfunction of or latent design flaw in the software component could lead to an erroneous diagnosis, or to a delay in delivery of appropriate medical care that may lead to a minor injury.
#### Clinical Studies
A Clinical Study was performed to investigate the imaging performance of the DRX-Revolution Nano Mobile X-ray System ((investigational device) as compared to the currently marketed DRX-Revolution Mobile X-ray System (K120062), (predicate device). Both devices were paired with the DRX Plus 3543 detector (K150766) for execution of the clinical study.
As compared to the predicate device, the DRX-Revolution Nano is a smaller, manually driven and operated mobile x-ray system. The Nano has a collapsible arm assembly that pivots rather than the predicate's telescoping column. The image acquisition module and image processing software are unchanged from the predicate. The DRX-Revolution Nano can interface with various Carestream digital detectors including the DRX 2530C, DRX-1, DRX Plus, and DRX Core detectors, as can the predicate.
Adult cadavers were used to perform pair-wise investigational vs. predicate acquisitions using similar radiographic techniques and body positioning for 2D imaging. The images acquired included targeting various exam types such as the chest, abdomen, skull, C-Spine, L-Spine, pelvis, hip, nasal bones, shoulder, and extremity radiographs. In addition, eight pediatric phantom images were acquired on both the investigational and predicate devices in the laboratory at Carestream Health, Inc. The images were assembled and reviewed in a single stimulus random format in order to perform a Comparative Evaluation to statistically confirm the findings of this evaluation.
Results of the Reader Study demonstrated that the diagnostic capability of the DRX-Revolution Nano Mobile X-ray System is statistically equivalent to or better than that of the predicate device.
#### Conclusions
Since the predicate device was cleared based in part on the results of clinical studies, clinical testing was required to support substantial equivalence in terms of diagnostic capability. Non-clinical test results support the safety and effectiveness of the device and demonstrate that the DRX-Revolution Nano Mobile X-ray System performs as intended in the specified conditions. Results of both non-clinical and clinical tests demonstrate that the DRX-Revolution Nano Mobile X-ray System is as safe and as effective as the predicate device that is currently marketed for the same intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.