Paragon 2

K162687 · Meridian Medical Systems, LLC · IZL · Oct 21, 2016 · Radiology

Device Facts

Record IDK162687
Device NameParagon 2
ApplicantMeridian Medical Systems, LLC
Product CodeIZL · Radiology
Decision DateOct 21, 2016
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1720
Device ClassClass 2
AttributesPediatric

Indications for Use

Intended for use by a qualified/trained doctor or technician on both adult and pediatric subjects for taking diagnostic radiographic exposures of the skull, spinal column, extremities, and other body parts. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position. Not for mammography.

Device Story

Paragon 2 is a mobile digital X-ray system for bedside, emergency, orthopedic, and operating theater use. Operated by qualified doctors or technicians, the device captures radiographic images using integrated digital X-ray panels (Toshiba FDX series). The system features a telescoping tubestand for transport visibility and is battery or line-powered. It utilizes a high-frequency generator to produce X-ray exposures; images are acquired via digital panels and processed by ECOM DROC software. Output is displayed for clinical review to assist in diagnostic decision-making. The device benefits patients by providing portable, high-quality diagnostic imaging without requiring transport to a stationary radiology suite.

Clinical Evidence

No clinical data. Bench testing only. Testing included performance to DHHS Radiation Safety standards, electrical safety (IEC-60601), EMC (IEC-60601-1-2), and radiation protection (IEC 60601-1-3, IEC 60601-2-54). Software integration and risk analysis were performed.

Technological Characteristics

Mobile X-ray system; high-frequency generator (20, 32, 40, 50 kW); telescoping tubestand; digital X-ray panels (Toshiba FDX series, 140μm pixel pitch); AC line or battery power; Ethernet/Wi-Fi connectivity; DICOM compliant. Safety standards: IEC-60601-1, IEC-60601-1-2, IEC-60601-1-3, IEC-60601-2-54. Software: ECOM DROC.

Indications for Use

Indicated for adult and pediatric patients requiring diagnostic radiographic imaging of the skull, spine, extremities, chest, and abdomen. Contraindicated for mammography.

Regulatory Classification

Identification

A mobile x-ray system is a transportable device system intended to be used to generate and control x-ray for diagnostic procedures. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles three human profiles facing to the right, with flowing lines extending from the bottom of the profiles. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 October 21, 2016 Meridian Medical Systems, LLC % Daniel Kamm, P.E. Submission Correspondent Kamm & Associates 8870 Ravello Ct NAPLES FL 34114 Re: K162687 Trade/Device Name: Paragon 2 Regulation Number: 21 CFR 892.1720 Regulation Name: Mobile x-ray system Regulatory Class: II Product Code: IZL, MQB Dated: September 24, 2016 Received: September 27, 2016 Dear Mr. Kamm: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Robert Oals Robert Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K162687 Device Name Paragon 2 Indications for Use (Describe) Intended for use by a qualified/trained doctor or technician on both adult and pediatric subjects for taking diagnostic radiographic exposures of the skull, spinal column, extremities, and other body parts. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position. Not for mammography. Type of Use (Select one or both, as applicable) 2 Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED. #### FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary 510(k) Number K162687 MERIDIAN MEDICAL SYSTEMS LLC 352 Harris Drive Aurora , OH 44202 Date Prepared: October 17, 2016 Contact: Larry Cornell, <Icornell@meridianmed.net> ### 1. Identification of the Device: Trade/Device Name: Paragon 2 Regulation Number: 21CFR892.1720 Regulation Name: Mobile X-Ray System Regulatory Class: II Product Code: IZL and MQB Common/Usual Name: Digital Mobile Diagnostic X-Ray System ### 2. Equivalent legally marketed device: K161345 Trade/Device Name: RadPRO® Mobile 40kW; RadPRO® Mobile 40kW FLEXPLUS, Model SM-40HF-B-D-VIR. Manufacturer: Sedecal SA (Spain) Regulation Number: 21CFR892.1720 Regulation Name: Mobile X-Ray System Regulatory Class: II Product Code: IZL and MQB Common/Usual Name: Digital Mobile Diagnostic X-Ray System - 3. Reference device: K143232 (The mobile device incorporates this device) Trade/Device Name: Universal Digital Interface 2W, UDI 2W. Manufacturer: MERIDIAN MEDICAL SYSTEMS LLC Regulation Number: 21CFR892.1680 Regulation Name: Stationary X-ray System Regulatory Class: II Product Code: MQB Common/Usual Name: Digital X-ray panel and software - 4. Indications for Use (intended use): Intended for use by a qualified/trained doctor or technician on both adult and pediatric subjects for taking diagnostic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position. Not for mammography. - 5. Description of the Device: The Paragon 2 is a mobile x-ray unit that covers all the specific needs of any radiographic examination at the patient's bed, first aid, and emergency, orthopedics, pediatric, and operating theater. These battery or line operated units combine stand-alone feature for exposures with battery assisted motor drive. Any of the following digital x-ray acquisition panels are available: Toshiba's FDX4343RPW, FDX 3543RPW, or FDX2520RPW. Integration of the panels and the software with the mobile system was straightforward since the panels and software have already been cleared. (K130883) The device complies with the US Federal Safety Performance Standard and is UL listed. The Paragon 2 has a telescoping tubestand which enhances visibility while transporting the unit. The Paragon 2 employs the digital x-ray panels and software employed in our premarket notification K143232. {4}------------------------------------------------ - 6. Safety and Effectiveness, comparison to predicate device. Bench, test laboratory results, rick analysis and software validation and integration testing results indicate that the new device is as safe and effective as the predicate devices. All digital panels have previous FDA clearance and are provided unmodified. | Characteristic | K161345 Trade/Device Name: RadPRO®<br>Mobile 40kW; RadPRO® Mobile 40kW<br>FLEXPLUS, | Meridian Medical Paragon 2 | |------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Intended Use: | Intended for use by a qualified/trained<br>doctor or technician on both adult and<br>pediatric subjects for taking diagnostic<br>radiographic exposures of the skull,<br>spinal column, chest, abdomen,<br>extremities, and other body parts.<br>Applications can be performed with the<br>patient sitting, standing, or lying in the<br>prone or supine position. Not for<br>mammography | SAME | | Configuration | Battery or line operated mobile | SAME | | Performance<br>Standard | 21 CFR 1020.30 | SAME | | Generator | High frequency made by Sedecal | SAME | | Generator<br>power level | One available power level: 40 kW | Four available power levels:<br>20 KW, 32 KW, 40 KW, 50 KW. These<br>four power levels were cleared in<br>Sedecal's K101517 which was the<br>predicate for K161345. Those same<br>generators are used in the Paragon 2. | | Collimator | Ralco R221 DHHS Manual Collimator | SAME | | Characteristic | K161345 Trade/Device Name: RadPRO®<br>Mobile 40kW; RadPRO® Mobile 40kW<br>FLEXPLUS, | Meridian Medical Paragon 2 | | Image<br>acquisition | Canon CXDI 401C Wireless (CSI) K133693<br>Pixel size: 125 μm 3320 × 3408 pixels<br>Canon CXDI 701C Wireless (CSI) K131106<br>Pixel size: 125 μm 2800 × 3408 pixels<br>Canon CXDI 701G Wireless (GOS) K131106<br>Pixel size: 125 μm 2800 × 3408 pixels<br>Canon CXDI 801C Wireless (CSI) K131106<br>Pixel size: 125 μm 2800 × 2192 pixels<br>Canon CXDI 801G Wireless (GOS) K131106<br>Pixel size: 125 μm 2800 × 2192 pixels | Toshiba panels are employed:<br>FDX4343R/RPW,<br>Active Area 43 (H)×43 (V) cm<br>Pixel Matrix 3036 (H)×3040 (V)<br>Pixel Pitch 140μ m<br>FDX3543RP/RPW<br>Active Area 345(H)×423(V)mm<br>Pixel Matrix 2466(H)×3040(V)<br>Pixel Pitch 140μ m<br>FDX2530RPW.<br>Active Area 245 (H)×295 (V) mm<br>Pixel Matrix 1750 (H)×2108 (V)<br>Pixel Pitch 140μ m<br>The Paragon 2 employs the digital x-ray<br>panels and software employed in our<br>premarket notification K143232. | | Software | Canon control software CXDI-NE | ECOM DROC, cleared with Toshiba<br>panels (Sedecal) in K130883 AND<br>K143232. | | Connection | Ethernet or Wireless Wi-Fi | SAME | | DICOM | YES | YES | | Power Source | AC Line or Rechargeable Battery | SAME | | Electrical<br>safety and EMC | Electrical Safety per IEC-60601. UL listed;<br>EMC per IEC-60601-1-2;<br>IEC 60601-1-3 Radiation protection in<br>diagnostic X-ray equipment<br>IEC 60601-2-54 Particular Requirements<br>For The Basic Safety And Essential<br>Performance Of X-Ray Equipment for<br>Radiography and Radioscopy | SAME | # 7. Substantial Equivalence Chart {5}------------------------------------------------ {6}------------------------------------------------ Image /page/6/Picture/0 description: This image shows the characteristics of the RadPRO® Mobile 40kW and RadPRO® Mobile 40kW FLEXPLUS, as well as the Meridian Medical Paragon 2. There are photos of the RadPRO® Mobile, RadPRO® Mobile Flex Plus, Standard Column, and Telescopic Column. The trade/device name is K161345. - 8. Summary of Laboratory Testing and Bench Testing: Laboratory testing included performance testing to the DHHS Radiation Safety Performance Standard, Electrical Safety per IEC-60601. EMC per IEC-60601-1-2; IEC 60601-1-3 Radiation protection in diagnostic X-ray equipment, IEC 60601-2-54 Particular Requirements For The Basic Safety And Essential Performance Of X-Ray Equipment for Radiography and Radioscopy. Software integration testing and risk analysis was performed to confirm proper implementation of the (already cleared) software. Every unit manufactured undergoes thorough performance testing which includes visual and configuration testing, leakage testing, hi-pot testing, power supply testing, (batteries, etc.), generator controls (voltage, current, timing), and image acquisition. - 9. Summary of Clinical Testing: Not required because all of the proposed digital panels have prior FDA clearance. - 10. Conclusion After analyzing bench and laboratory testing to applicable standards, it is the conclusion of Meridian Medical that the Paragon 2 Mobile X-Ray Systems are as safe and effective as the predicate devices, have few technological differences, and has no new indications for use, thus rendering them substantially equivalent to the predicate devices.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...