This product is intended primarily for on-line or remote viewing of cardiac catheterization angiography studies communicated by network, or transported by CD-R interchange media. The CD-R media may also be used as a full resolution digital image archive of complete cath lab studies, to replace cine film. The ARS2 can optionally be used with other digital imaging modalities, such as R/F, nuclear and ultrasound. The ARS2 provides the capability to write and read CD-R optical discs which are produced in accordance with the ACC/NEMA DICOM standard.
Device Story
Digital image recording/display system; inputs include cardiac catheterization, R/F, nuclear, and ultrasound images via network or CD-R media. Operates as workstation for viewing, archiving, and quantitative analysis of studies. Features include edge enhancement, zoom, window/level adjustment, gamma correction, and subtraction. Quantitative analysis includes ventricular wall motion/volume and arteriogram analysis. Used in clinical settings by medical professionals to replace cine film; facilitates diagnostic review and clinical decision-making through high-resolution digital display and standardized DICOM data interchange.
Clinical Evidence
No clinical data provided; performance data not required for determination of substantial equivalence for this class of device.
Technological Characteristics
Digital imaging workstation; supports 8-bit grayscale, 1024x1024 display matrix, 1280x1024 video output. Storage via hard disk and CD-R. Connectivity via network and DICOM-compliant optical media. Image processing includes edge enhancement, zoom, window/level, gamma correction, and subtraction. Quantitative analysis modules for ventricular and arteriogram assessment.
Indications for Use
Indicated for clinicians requiring on-line or remote viewing and archiving of digital medical images, including cardiac catheterization angiography, radiography/fluoroscopy, nuclear medicine, and ultrasound studies.
Regulatory Classification
Identification
An angiographic x-ray system is a device intended for radiologic visualization of the heart, blood vessels, or lymphatic system during or after injection of a contrast medium. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
K023467 — HCP DICOM NET · Softlink International, Inc. · Jan 7, 2003
Submission Summary (Full Text)
{0}
Camtronics MEDICAL SYSTEMS
5519
K955519
MAY - 9 1996
Date prepared: November 22, 1995
Name of contact person: Robert Kriedermann
Device trade name: Analytical Review Station (ARS2, subject to change)
Common name: Digital Imaging Workstation
Classification name: Medical Image Picture Archiving and Communications Device
Predicate substantially equivalent devices: Camtronics Ltd. "Video Plus" (K941979), Camtronics Ltd. "Archium" (K934496)
Device description: Digital image recording and display system for use in cardiac catheterization, radiography/fluoroscopy, ultrasound, or other digital imaging modalities.
Intended use: This product is intended primarily for on-line or remote viewing of cardiac catheterization angiography studies communicated by network, or transported by CD-R interchange media. The CD-R media may also be used as a full resolution digital image archive of complete cath lab studies, to replace cine film. The ARS2 can optionally be used with other digital imaging modalities, such as R/F, nuclear and ultrasound. The ARS2 provides the capability to write and read CD-R optical discs which are produced in accordance with the ACC/NEMA DICOM standard.
Predicate device specifications comparison:
| | Camtronics Analytical Review Station | Camtronics Analytical Workstation K941979 |
| --- | --- | --- |
| Image Storage and Display | | |
| Gray scale resolution | 8 bit, 256 levels | 8 bit, 256 levels |
| Stored image matrix size | 512 x 512, 1024 x 512, or 1024 x 1024 | 512 x 512, 1024 x 512, or 1024 x 1024 |
| Displayed image matrix size | 1024 x 1024 | 1024 x 1024 |
| Display video output | 1280 x 1024, 72 hz refresh, color or monochrome; progressive | 1280 x 1024, 72 hz refresh, color, progressive |
| Screen size | 16 in., 19 in., or 20 in. | 16 in., 19 in., or 20 in. |
| Image formats | DICOM, Siemens, GE, Philips, Picker, others; 1, 4, or 16 on 1 | Siemens, GE, Philips, Picker, others; 1, 4, or 16 on 1 |
| Media | Hard disk, CD-R | Hard disk, CD-R |
| Compression | JPEG Processes 1 and 14, dual-track lossy or lossless | JPEG Process 14, lossless |
| Image Processing Functions | Edge enhancement | Edge enhancement |
| | Zoom | Zoom |
| | Window and level adjustment | Window and level adjustment |
| | Gamma correction | Gamma correction |
| | Subtraction | Subtraction |
| Quantitative Analysis | | |
| Ventricular Analysis | Sheehan/Dodge centerline algorithm for regional wall motion analysis, and LV volume computation; others | Sheehan/Dodge centerline algorithm for regional wall motion analysis, and LV volume computation. |
| Arteriogram Analysis | Saunders Data Systems, others | Artrek by QCS, Inc. |
Performance data: Not required for determination of substantial equivalence for this class of device.
Conclusions drawn from clinical and nonclinical test data: Not required for determination of substantial equivalence for this class of device.
Camtronics Ltd., A Subsidiary of ANALOGIC Corporation
900 Walnut Ridge Drive • P.O. Box 950 • Hartland, WI 53029 • 414-367-0700 • Fax: 414-367-0717
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.