DIGITAL IMAGING SYSTEM
Device Facts
| Record ID | K954159 |
|---|---|
| Device Name | DIGITAL IMAGING SYSTEM |
| Applicant | Camtronics, Ltd. |
| Product Code | LLZ · Radiology |
| Decision Date | Mar 19, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 892.2050 |
| Device Class | Class 2 |
Intended Use
Image Acquisition Display Control/Image Processing Input/Output Control Patient Data Management
Device Story
Digital imaging system for cardiac catheterization or radiography/fluoroscopy suites; records and displays medical images. Inputs include fluoroscopic and radiographic signals; system performs image processing (window/level, edge enhancement, zoom, roam, roadmapping, peak opacification, remask, pixel shift, annotation, qualitative analysis). Outputs include processed images and patient data displayed for clinical review. Operated by physicians or technicians in clinical settings. Supports various acquisition modes including DSA, peripheral angiography, digital tomography, rotational angiography, and ECG-gated acquisition. Facilitates clinical decision-making by providing enhanced visualization and management of diagnostic images.
Clinical Evidence
No clinical data provided; performance data not required for determination of substantial equivalence for this class of device.
Technological Characteristics
Digital image recording and display system. Matrix size: 512² to 2068². Bit depth: 8, 10, or 12. Storage: 1 gigabyte minimum. Features include networking, cine, filmless cine, and synchronized biplane acquisition. Supports various acquisition modes including DSA, peripheral angiography, digital tomography, rotational angiography, and ECG-gated acquisition.
Indications for Use
Indicated for use in cardiac catheterization laboratories or radiography/fluoroscopy suites for the acquisition, display, processing, and management of medical images.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
Predicate Devices
- Camtronics Video Plus (K941979)
- Siemens Medical Systems Fluorospot H version 2 (K914525)
Related Devices
- K955519 — ANALYTICAL REVIEW STATION · Camtronics, Ltd. · May 9, 1996
- K030544 — RFX FLUROSCOPIC DIGITAL IMAGING SYSTEM · Sigma Vision, Inc. · May 21, 2003
- K024200 — INNOVA 2000 AND INNOVA 2000S (MOBILE VERSION) · Ge Medical Systems, Inc. · Jan 16, 2003
- K983088 — INSTACINE D-2000 · Fidelity Medical, Inc. · Sep 22, 1998
- K041027 — DIGITAL RADIOGRAPHY SYSTEM WITH FLAT PANEL DETECTOR, MODEL DFP8000D · Toshibamedical Systems Corporation · May 14, 2004