K023467 · Softlink International, Inc. · LLZ · Jan 7, 2003 · Radiology
Device Facts
Record ID
K023467
Device Name
HCP DICOM NET
Applicant
Softlink International, Inc.
Product Code
LLZ · Radiology
Decision Date
Jan 7, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
HCP DICOM Net® is a software device intended to acquire and store medical images studies from catheterization laboratories, Echo cardiograms, CT & MRI and other such modalities. In addition, HCP DICOM Net® provides image enhancement and filtering tools for viewing of selected images. It also imports image studies from DICOM CD's/DVD's, makes copies of image studies and converts and exports image studies into DICOM3 and other commercial file formats. HCP DICOM Net® also integrates industry standard third party approved software within its application to allow for analysis of the acquired image.
Device Story
HCP DICOM Net® is a software-only system for medical image management; operates on standard Windows-based PCs. Inputs include medical images from catheterization labs, echocardiograms, CT, and MRI via video source or DICOM3 interface. Non-DICOM3 video inputs are converted to DICOM3 format. Images are stored on hard drives, magnetic tape, CD-R, or DVD. Device provides image enhancement and filtering tools for viewing; integrates third-party software for image analysis. Used in clinical settings to replace film-based workflows; enables retrieval, viewing, and analysis of studies. Output is displayed on workstations for clinician review, facilitating clinical decision-making through improved image access and analysis capabilities.
Clinical Evidence
No clinical data. Safety and effectiveness supported by bench testing, including software validation and verification testing conducted in accordance with EN-1441 (Risk Analysis) and adherence to DICOM voluntary standards.
Technological Characteristics
Software-only device; runs on industry-standard PCs with Windows operating systems. Supports DICOM3 image format; provides image enhancement and filtering tools. Connectivity via DICOM3 interface or video source. Complies with DICOM voluntary standards. Risk analysis performed per EN-1441.
Indications for Use
Indicated for the acquisition, storage, enhancement, and viewing of medical images from catheterization laboratories, echocardiograms, CT, and MRI modalities. No specific patient population or contraindications are defined.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{1}------------------------------------------------
K02 3467
# EXHIBIT #1
# 510(K) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
The assigned 510(k) number is:
#### 1. Submitter's Identification:
Softlink International Inc. 199 Main Street Suite 314 White Plains, NY 10601
Contact: Mr. Praveen Lobo Softlink International Tel: 914-582-1047 Fax: 914-686-4666
Date Summary Prepared: October 14, 2002
October 8, 2002
#### 2. Name of the Device:
HCP DICOM Net®
#### 3. Predicate Device Information:
- Digital Imaging Systems, K#954159, Camtronics, Ltd., Hartland, WI 1. (892.2050, Product Code LLZ).
- 2. The TCS-Telecardiology System, Medcon Ltd., Efrat, Israel (892.1600, Product Code IZI).
- 3. WriteStar and ViewStar, K#964274, ComView Corp., Montrose, Ny (892.2010, Product Code LMB).
#### 4. Device Description:
HCP DICOM Net® is a computerized system that acquires images which are captured by cardiac catherterization laboratories and other modalities like Echocardiograms. MRI & CT. These images are acquired through a video source or a DICOM3 interface of the
{2}------------------------------------------------
modality. These images are then stored in DICOM3 format on Hard Disk Drives, Magnetic Tape. CD-R or DVD. If images are acquired through a video source, in a non-DICOM3 format then these images are converted into a DICOM3 format before they are stored. After storage these images are retrieved and displayed on workstations for convenient review and analysis. Analysis of images is provided via approved industry standard third party software. These workstations are also capable of reading DICOM3 images from CD and DVD media and uploading them to the image store for retrieval, viewing and analysis at a later time.
# 5. Intended Use:
HCP DICOM Net® is a software device intended to acquire and store medical images studies from catherterization laboratories. Echo cardiograms. CT & MRI and other such modalities. In addition, HCP its DICOM Net provides image enhancement and filtering tools for viewing of selected images. It also imports image studies from DICOM CD's/DVD's, makes copies of image studies and converts and exports image studies into DICOM3 and other commercial file formats. HCP DICOM Net® also integrates industry standard third party approved software within its application to allow for analysis of the acquired image.
# 6. Comparison to Predicate Devices:
HCP DICOM Net® provides convenient acquisition, storage and review of DICOM3 images like the ComView WriteStar and Medcon TCS-Cardiology System. HCP DICOM Net® unlike the Camtronics Digital Imaging System and ComView ViewStar and WriteStar, is a software only device that runs on industry standard PCs running windows based operating systems. HCP DICOM Net®, like the Camtronics Digital Imaging System and the ComView ViewStar/WriteStar System, can acquire analog images and convert them to a digital DICOM 3 image format and this functionality is intended to replace film.
# 7. Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence are as follows:
Testing information demonstrating safety and effectiveness of the HCP DICOM Net® device in the intended environment of use is supported by testing that was conducted in accordance with EN-1441 (Risk Analysis) and software validation and verifications testing. In addition, a DICOM Conformance Statement was provided with this submission to DICOM Voluntary Standards.
### 8. Discussion of Clinical Tests Performed:
- Not applicable
{3}------------------------------------------------
## 9. Conclusions:
The HCP DICOM Net® device has similar intended uses and similar characteristics as the predicate devices. Moreover, software testing contained in this submission demonstrates that any differences in their technological characteristics do not raise any new questions of safety or effectiveness. Thus, the HCP DICOM Net® device is substantially equivalent to the predicate devices.
{4}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle with three stripes forming its body and wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JAN 0 7 2003
SoftLink International, Inc. % Ms. Susan D. Goldstein-Falk mdi Consultants, Inc. 55 Northern Blvd. Suite 200 GREAT NECK NY 11021
Re: K023467 Trade/Device Name: HPC DICOM Net Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: 90 LLZ Dated: October 14, 2002 Received: October 16, 2002
Dear Ms. Goldstein-Falk
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 8xx.1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely vours.
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive. Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Exhibit B
Page ________ of ___1___
510(k) Number (if known): _ Ko 2 3467
Device Name Softlink International Inc. HCP DICOM Net®
Indications For Use:
HCP DICOM Net® is a software device intended to acquire and store medical images studies from catheterization laboratories, Echo cardiograms, CT & MRI and other such modalities. In addition, HCP DICOM Net® provides image enhancement and filtering tools for viewing of selected images. It also imports image studies from DICOM CD's/DVD's, makes copies of image studies and converts and exports image studies into DICOM3 and other commercial file formats. HCP DICOM Net® also integrates industry standard third party approved software within its application to allow for analysis of the acquired image.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
| Concurrence of CDRH, Office of Device Evaluation (ODE) |
|--------------------------------------------------------|
|--------------------------------------------------------|
| Prescription Use |
|------------------|
|------------------|
| (Division Sign-Off) | |
|---------------------------------------------------------------|---------|
| Division of Reproductive, Abdominal, and Radiological Devices | |
| 510(k) Number | K023467 |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.