Browse hierarchy Radiology (RA) Subpart B — Diagnostic Devices 21 CFR 892.1550 Product Code IYN K020754 — NICOLET PIONEER TC8080
NICOLET PIONEER TC8080
K020754 · Nicolet Biomedical · IYN · Apr 21, 2003 · Radiology
Device Facts
Record ID K020754
Device Name NICOLET PIONEER TC8080
Applicant Nicolet Biomedical
Product Code IYN · Radiology
Decision Date Apr 21, 2003
Decision SESE
Submission Type Abbreviated
Regulation 21 CFR 892.1550
Device Class Class 2
Intended Use
Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: The Pioneer TC8080 applies ultrasound probes of 1.6 MHz Pulsed Wave (PW), 2 MHz PW, 4 MHz PW and Continuous Wave (CW), 8 MHz,PW/CW, 16 MHz PW and 20 MHz PW to support Transcranial Doppler (TCD) Adult Cephalic and Peripheral Vascular blood flow evaluation.
Device Story
Pioneer TC8080 is a diagnostic ultrasound system for blood flow evaluation. It utilizes various ultrasound probes (1.6 MHz to 20 MHz) operating in Pulsed Wave (PW) and Continuous Wave (CW) modes. The system processes ultrasonic signals to analyze blood flow in the adult cranium and peripheral vasculature. It is intended for use by healthcare professionals in clinical settings. The device provides real-time blood flow data, assisting clinicians in assessing vascular conditions. The output aids in clinical decision-making regarding patient vascular health and potential interventions.
Clinical Evidence
No clinical data provided; substantial equivalence is based on bench testing and technological comparison to previously cleared devices.
Technological Characteristics
Diagnostic ultrasound system; operates with multiple probes ranging from 1.6 MHz to 20 MHz; supports Pulsed Wave (PW) and Continuous Wave (CW) modes; crystal diameters include 1.5 mm, 2.0 mm, 10 mm, and 15 mm; intended for adult cephalic and peripheral vascular applications.
Indications for Use
Indicated for diagnostic ultrasound imaging or fluid flow analysis for Transcranial Doppler (TCD) Adult Cephalic and Peripheral Vascular blood flow evaluation using 1.6 MHz to 20 MHz probes.
Regulatory Classification
Identification An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Reference Devices
Related Devices
K993939 — NICOLET EME LEGEND TC22 · Nicolet Biomedical · Jun 29, 2000
K053648 — PIONEER TC8080, COMPANION III · Viasys Neuro Care · Jan 24, 2006
K213009 — Transcranial Doppler Ultrasound System · Shenzhen Delica Medical Equipment Co., Ltd. · Jul 8, 2022
K990517 — EXPLORER CVS TRANSCRANIAL DOPPLER · Diagnostic Medical Systems · Aug 1, 2000
K060421 — SONARA; SONARA/TEK · Viasys Healthcare, Inc. · Nov 3, 2006
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three lines representing its wings and head. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.
APR 2 1 2003
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Nicolet Vascular Division of VIASYS Healthcare % Mr. Gary Syring Regulatory Consultant Quality & Regulatory Associates, LLC 800 Levanger Lane STOUGHTON WI 53589
Re: K020754
Trade Name: Pioneer TC8080 Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulation Number: 21 CFR 892.1570 Regulation Name: Diagnostic ultrasonic transducer Regulatory Class: II Product Code: 90 IYN and ITX Dated: February 3, 2003 Received: February 5, 2003
Dear Mr. Syring:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
This determination of substantial equivalence applies to the following transducers intended for use with the Pioneer TC8080, as described in your premarket notification:
Transducer Model Number
1.6 MHz Probe 2 MHz Probe 4 MHz Probe 8 MHz Probe 16 MHz Probe
{1}------------------------------------------------
# 20 MHz Probe
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801, please contact the Office of Compliance at (301) 594-4591. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or at (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html".
If you have any questions regarding the content of this letter, please contact Rodrigo C. Perez at (301) 594-1212.
Sincerely vours.
David A. Lyman
for
Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure(s)
{2}------------------------------------------------
# System: Pioneer TC8080 Transcranial (Adult Cephalic) and Peripheral Vascular Diagnostic Device
### Diagnostic Ultrasound Indications for Use Form
#### Fill out one form for each ultrasound system and each transducer.
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | |
|----------------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | | | | | | | |
| Abdominal | | | | | | | | | | |
| Intraoperative (specify) | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ (specify) | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | N | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral Vascular | | | | N | N | | | | | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | |
| Other (specify) | | | | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
Additional Comments: The Pioneer TC8080 applies ultrasound probes of 1.6 MHz Pulsed Wave (PW), 2 MHz PW, 4 MHz PW and Continuous Wave (CW), 8 MHz,PW/CW,
16 MHz PW and 20 MHz PW to support Transcranial Doppler (TCD) Adult Cephalic and
Peripheral Vascular blood flow evaluation.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
David R. Lyman
(Division Sign-Off) Division of Reproductive, Abdominal, and Radiglogical Devices 510(k) Number _
{3}------------------------------------------------
### Fill out one form for each ultrasound system and each transducer.
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | | Mode of Operation | | | | | | | | | | | |
|----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------|---|--------------------------|-----|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------|--------------------|--|--|--|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) | | | |
| Ophthalmic | | | | | | | | | | | | | |
| Fetal | | | | | | | | | | | | | |
| Abdominal | | | | | | | | | | | | | |
| Intraoperative (specify) | | | | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | | | | |
| Pediatric | | | | | | | | | | | | | |
| Small Organ (specify) | | | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | | | |
| Adult Cephalic | | | | E | | | | | | | | | |
| Cardiac | | | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | | | |
| Transurethral | | | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | | | | |
| Other (specify)<br>1<br>C<br>A 1 | ****************************************************************************************************************************************************************************** | | | -1 ------- -1 L. Pro A . | ﺎ | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | A | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>L | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
Additional Comments: E - 1.6 MHz Pulsed Wave
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
Division Sign-Off
Division of Reproductive, Abdomina ge B-3 cal Devices
{4}------------------------------------------------
#### Appendix F
### Diagnostic Ultrasound Indications for Use Form
## Fill out one form for each ultrasound system and each transducer.
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | |
|----------------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | | | | | | | |
| Abdominal | | | | | | | | | | |
| Intraoperative (specify) | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ (specify) | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | P | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | |
| Musculo-skeletal Superficiai | | | | | | | | | | |
| Other (specify) | | | | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
Additional Comments: P - Same probe (Transducer) cleared to market by K993939
Two (2) crystal diameters are indicated: 10 mm diameter and 15 mm diameter.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Eyalfuation
Daniel R. Logemann
Division Sign Off
of Reproductive, Abdomina and Radiological Devices 510(k) Number
Prescription Use (Per 21 CFR 801.109)
F-3
{5}------------------------------------------------
# Fill out one form for each ultrasound system and each transducer.
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) |
|----------------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Clinical Application | Mode of Operation | | | | | | | | | |
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | | | | | | | |
| Abdominal | | | | | | | | | | |
| Intraoperative (specify) | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ (specify) | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral Vascular | | | | P | P | | | | | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | |
| Other (specify) | | | | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
Additional Comments: P - Same probe (Transducer) cleared to market by K993939
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Reproductive, Abdominal, and Radiological Devices
510(k) Number ________________________________________________________________________________________________________________________________________________________________
Prescription Use (Per 21 CFR 801.109)
Page B-5
{6}------------------------------------------------
### Fill out one form for each ultrasound system and each transducer.
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | |
|----------------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | | | | | | | |
| Abdominal | | | | | | | | | | |
| Intraoperative (specify) | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ (specify) | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral Vascular | | | | P | P | | | | | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | |
| Other (specify) | | | | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
Additional Comments: P - Same probe (Transducer) cleared to market by K993939
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)Prescription Use (Per 21 CFR 801.109)
(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number _
{7}------------------------------------------------
#### Fill out one form for each ultrasound system and each transducer.
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | |
|----------------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | | | | | | | |
| Abdominal | | | | | | | | | | |
| Intraoperative (specify) | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ (specify) | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral Vascular | | | | N | | | | | | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | |
| Other (specify) | | | | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
Additional Comments: N - 16 MHz PW microvascular transducer for peripheral
vascular blood flow evaluation.
Two (2) crystal diameters are indicated: 1.5 mm and 2.0 mm.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NERL
Concurrence of CDRH, Office of Device Evaluation (ODM)
(Division Sign-Off Division of Reproductive, and Radiological Devices 510(k) Number _
Prescription Use (Per 21 CFR 801.109)
Page B-7
{8}------------------------------------------------
Appendix F
### Diagnostic Ultrasound Indications for Use Form
#### Fill out one form for each ultrasound system and each transducer.
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | |
|----------------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | | | | | | | |
| Abdominal | | | | | | | | | | |
| Intraoperative (specify) | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ (specify) | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral Vascular | | | | N | | | | | | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | |
| Other (specify) | | | | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
Additional Comments: N - 20 MHz PW microvascular transducer for peripheral vascular blood flow evaluation.
Two (2) crystal diameters are indicated: 1.5 mm and 2.0 mm.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Daniel A. Lyman
(Division Sign-Off)
Division of Reproductive, Abdomina and Radiological Devices Page B 510(k) Number _
Prescription Use (Per 21 CFR 801.109)
F-3