K990517 · Diagnostic Medical Systems · IYN · Aug 1, 2000 · Radiology
Device Facts
Record ID
K990517
Device Name
EXPLORER CVS TRANSCRANIAL DOPPLER
Applicant
Diagnostic Medical Systems
Product Code
IYN · Radiology
Decision Date
Aug 1, 2000
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1550
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Explorer CVS Doppler is intended to be used to produce images of peripheral vasculature and intracranial vasculature to be analysed by trained medical personnel for the assessment and diagnosis of vascular status and disorders.
Device Story
Explorer CVS is a diagnostic ultrasonic Doppler system; utilizes multiple frequency transducers (2 MHz, 4 MHz, 8 MHz) to acquire ultrasound signals from peripheral and intracranial vasculature; processes signals to provide high-resolution images and fluid flow analysis; operated by trained medical personnel in clinical settings; output used for assessment and diagnosis of vascular status and disorders; aids clinical decision-making by visualizing blood flow and vascular anatomy.
Clinical Evidence
Bench testing only. No clinical data provided. Substantial equivalence is supported by technical specifications and acoustic output measurements.
Technological Characteristics
Diagnostic ultrasonic Doppler system; utilizes multiple frequency transducers (2 MHz, 4 MHz, 8 MHz); supports PW-Doppler and CW-Doppler modes; intended for peripheral, intracranial, and intravascular applications; complies with 21 CFR 892.1550; requires post-clearance submission of acoustic output measurements.
Indications for Use
Indicated for diagnostic ultrasound imaging or fluid flow analysis of the human body, specifically peripheral vascular, intracranial (adult/neonatal cephalic), and intravascular applications using pulsed-wave (PW) and continuous-wave (CW) Doppler modes.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{1}------------------------------------------------
K990517
1 2000 AUG
510 (k) Summary
Manufacturer : A.
DIAGNOSTIC MEDICAL SYSTEMS DHS -Parc de la Mediterranee District de Montpellier 34470 Perols France
Submitted By:
Ferguson Medical Consultant to DMS
+33.467.50.49.00 Phone : B. Contact Information: PAX: +33.467.50.49.09
- Classification Name: System, Imaging, Pulsed Doppler, ਂ Ultrasonic
Transcranial/Vascular Doppler Common/usual Name:
Explorer CVS Transcranial and Proprietary Name: Peripheral Vascular Doppler
- Classification Number: 901 YN D.
- Rimed Limited Intra-View E. Substantial Equivalence : Transcranial Doppler Device (K974588), Bilateral Transcranial Doppler Inc. Neuroguard Medasonics, Medasonics, (K962796), Inc. System Ultrasound System; CDS (K914862), Cerebrovascular Diagnostic 2 Transcranial Probe Biosound, Phase Inc. Acuson Corp. Acuson S228 Transducer Fox (K894163), (K894163), Medasonics, Inc. Doppler Transcranial (K872292) , and Doppler Transpect Transcranial others.
The Explorer CVS device is a Device Description: r.
{2}------------------------------------------------
#### diagnostic ultrasonic Doppler.
- device 18 Explorer CV8 Doppler Intended Use : The G. produce images of be used to intended to peripheral vasculature to be and intracranial the medical personnel 17 analysed trained by vascular status and assessment and diagnosis of disorders.
- The Explorer CVS Doppler ਜ -Technological Characteristics: device utilizes multiple frequencies of transducers to provide high resolution images.
{3}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or a stylized human figure with outstretched arms. The logo is rendered in black and white.
1 2008 AUG
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Diagnostic Medical Systems, Inc. c/o Frank Ferquson Official Correspondent Ferquson Medical P.O. Box 12038 LaJolla, CA 92039
Re: K990517 Explorer CVS Transcranial and Peripheral Vascular Doppler Regulatory Class: II/21 CFR 892.1550 Product Code: 90 IYN Dated: May 15, 2000 Received: May 30, 2000
Dear Mr. Ferguson:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Ameraments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. "The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
This determination of substantial equivalence applies to the following transducers intended for use with the Explorer CVS Transcranial and Peripheral Vascular Doppler, as described in your premarket notification:
Transducer Model Number
2 MHz PW-Doppler 4 MHz CW/PW-Doppler 8 MHz CW/PW-Doppler
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval) it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. ನ್ substantially equivalent determination assumes compliance with the Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General (GMP) requlation (21 CFR Part 820) and that, through periodic QS inspections, the FDA will verify such Failure to comply with the GMP regulation may result in assumptions. In addition, the Food and Drug Administration (FDA) may regulatory action. publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification does not affect any obligation you may have under sections 531 and 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{4}------------------------------------------------
#### Page 2 - Frank Ferguson
This determination of substantial equivalence is granted on the condition that Inis decermination of Subbeancear oquon submit a postclearance special report. prior co shipping the rise amplete information, including acoustic output measurements based on production line devices, requested in Appendix G,
measurements based on production 1ine derices , requested in Appartuages (enclosed) of the Center 3 September 30) III, III, III, International Systems and Transducers." Seeking Marketing Crearance of Disacted or contains unacceptable values (e.g., II the Special report 15 Incompressed overs), then the 510(k) clearance may acoustic oucpar groduction units which as a result may be considered adulterated or misbranded.
The special report should reference the manufacturer's 510(k) number. It The special report should forceines the manufacturing and should be submitted in duplicate to:
> Food and Drug Administration Center for Devices and Radiological Health Document Mail Center (HFZ-401) 9200 Corporate Boulevard Rockville, Maryland 20850
This letter will allow you to begin marketing your device as described in your The FDA finding of substantial equivalence of your premarket notification. premarket notification. Inc predicate device results in a classification for device to a logally master your device to proceed to market.
If you desire specific advice for your device on our labeling requlation (21 II you desire specific advice ac- 10 for in vitro diagnostic devices), please critic of the dualemaniance at (301) 594-4591. Additionally, for contact the office or compand advertising of your device, please contact the questions on the promote (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Other general information on your responsibilities under the Act may 807.97). be obtained from the Division of Small Manufacturers Assistance at its tollfree number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html" .
If you have any questions regarding the content of this letter, please contact Rodrigo C. Perez at (301) 594-1212.
Sincerely yours,
David C. Lippman
Schultz, M.D. Daniel G. Captain, USPHS Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosures
{5}------------------------------------------------
### EXPLORER CVS Transcranial Doppler System 2,4 and 8MHz Transducers
### Diagnostic Ultrasound Indications for Use Form
## Fill out one form for each ultrasound system and each transducer.
# Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | |
|-------------------------------|-------------------|---|---|-----|-----|---------------|-------------------|------------------------|--------------------|-----------------|
| Clinical Application | A | B | M | PWD | CWD | Color Doppler | Amplitude Doppler | Color Velocity Imaging | Combined (specify) | Other (specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | | | | | | | |
| Abdominal | | | | | | | | | | |
| Intraoperative (specify) | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ (specify) | | | | | | | | | | |
| Neonatal Cephalic | | | | X | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral Vascular | | | | X | X | | | | | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal Conventional | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | |
| Other (specify) | | | | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
Additional Comments:
(PLEASE DO NOT WRITE BELOW THUS LINE - CONTINUE ON ANOTHER PAGE IF NEFDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
።
... ..........................................................................................................................................................................
Daniel C. Lynn
Bivision Sign=And દ-૩ Division of Reproduetive, Abdominal, ENT, and Radiological Devices
> 310(k) Number • ﺑﺎﻟﻤﺴﺎﻋﺔ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻘﺮﻥ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻘﺮﻥ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻘﺮﻥ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻘﺮﻥ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻘﺮﻥ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻘﺮﻥ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻘﺮﻥ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻘﺮﻥ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻘﺮﻥ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻘﺮﻥ ﺍﻟﻤﺘﻮﻗﻒ ﺍﻟﻤﺘﻮﻗﻒ ﺍﻟﻤﺘﻮﻗ
{6}------------------------------------------------
## 2 MHz PW-Doppler Transducer (Serial Number 98-311)
Appendix F
### Diagnostic Ultrasound Indications for Use Form
## Fill out one form for each ultrasound system and each transducer.
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | |
|-------------------------------|-------------------|---|---|-----|-----|---------------|-------------------|------------------------|--------------------|-----------------|
| Clinical Application | A | B | M | PWD | CWD | Color Doppler | Amplitude Doppler | Color Velocity Imaging | Combined (specify) | Other (specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | | | | | | | |
| Abdominal | | | | | | | | | | |
| Intraoperative (specify) | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ (specify) | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | X | | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal Conventional | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | |
| Other (specify) | | | | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
Additional Comments:
f
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTIMUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Toril A. Lynn
---
(Division Sign-Off)
Prescription Use (Per 21 CFR 801.109)
Division of Reproductive, Abdominal, ENT, and Radiological Devices
510(k) Number K990517
{7}------------------------------------------------
# 4 MHz CW/PW-Doppler (Serial Number 115)
#### Diagnostic Ultrasound Indications for Use Form
### Fill out one form for each ultrasound system and each transducer.
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | |
|----------------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | | | | | | | |
| Abdominal | | | | | | | | | | |
| Intraoperative (specify) | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ (specify) | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | X | X | | | | | |
| Peripheral Vascular | | | | X | X | | | | | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | |
| Other (specify) | | | | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Apper
**Additional Comments:**
Prescription Use (Per 21 CFR 801.109)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTRIVE ON ANOTHER PAGE IF NEEDED)
Concurrence of CORH, Office of Device Evaluation (ODE)
Yoriel A. Sgenon
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT, and Radiological Devices
510(k) Number K990517
{8}------------------------------------------------
### Diagnostic Ultrasound Indications for Use Form
## Fill out one form for each ultrasound system and each transducer.
Intended Use: Diagnostic uttrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | |
|----------------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | | | | | | | |
| Abdominal | | | | | | | | | | |
| Intraoperative (specify) | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ (specify) | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | X | | | | | |
| Peripheral Vascular | | | | X | X | | | | | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | |
| Other (specify) | | | | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
Additional Comments:
*Rrescription Use (Per 21 CFR 801.109)
PLEASE DO NOT WRITE BELOW THIS UNE - CONTINUE ON ANOTHER PAGE IF NEEDEDI Concurrence of CDRH, Office of Device Evaluation (ODE)
Daniel G. Regner
(Division Sign-Off) Division of Repreductive, Abdominal, ENT,
and Radiological Devices
510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.