K060421 · Viasys Healthcare, Inc. · IYN · Nov 3, 2006 · Radiology
Device Facts
Record ID
K060421
Device Name
SONARA; SONARA/TEK
Applicant
Viasys Healthcare, Inc.
Product Code
IYN · Radiology
Decision Date
Nov 3, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1550
Device Class
Class 2
Attributes
Pediatric
Indications for Use
Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: [Table of clinical applications including Ophthalmic, Intraoperative, Pediatric, Neonatal Cephalic, Adult Cephalic, and Peripheral Vascular using PWD and CWD modes]. The Sonara and Sonara/tek Transcranial Doppler are medical ultrasound devices for measuring the blood flow velocities in arteries and in veins noninvasively, consistent with the FDA Ultrasound System and Transducer Indication for Use forms cleared with the 510(k).
Device Story
Sonara/Sonara/tek TCD systems measure blood flow velocities and hemodynamic parameters in intracranial, extracranial, and peripheral vessels. System utilizes 2MHz, 4MHz, and 8MHz ultrasound probes in unilateral or bilateral configurations; includes 2MHz monitoring probe for headset use. Inputs processed via integrated PC/touchscreen display; provides M-mode imaging and spectral Doppler displays. Operated by clinicians in clinical or surgical settings; supports online/offline modes. Output allows real-time monitoring of blood flow and emboli detection (HITS) to assist clinical decision-making during vascular procedures or monitoring. Sonara/tek variant connects to external PC via USB.
Clinical Evidence
Bench testing only. Device subjected to safety and performance validation, including functional specification testing and compliance with IEC/EN 60601-1, 60601-1-2, and 60601-2-37 standards. Acoustic output measured per FDA 1997 ultrasound guidance.
Technological Characteristics
Pulsed Doppler ultrasound system; 2MHz, 4MHz, 8MHz transducers. Integrated 15" touch screen LCD, PC board, hard disk. Connectivity: USB (Sonara/tek). Standards: IEC/EN 60601-1, 60601-1-2, 60601-2-37. Modes: PWD, CWD, M-mode. Power levels displayed during scanning.
Indications for Use
Indicated for noninvasive measurement of blood flow velocities in intracranial, extracranial, and peripheral arteries and veins. Used for transcranial Doppler monitoring, carotid monitoring, and emboli detection (HITS) in adult, pediatric, and neonatal populations. Applicable for intraoperative use during surgical procedures. No invasive transducer application.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
## NOV - 3 2006
### 510(K) SUMMARY
# SONARA TCD SYSTEM (INCLUDING SONARA AND SONARA/TEK)
## 510(k) Number: K060421
| Applicant's Name: | VIASYS Healthcare, Inc.<br>227 Washington St., Suite 200<br>Conshohocken, PA 19428 |
|--------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Sponsor Contact: | Mr. Glen Hermanson<br>Global Manager, Quality Engineering<br>VIASYS NeuroCare, Inc.<br>5225 Verona Road<br>Madison, WI 53711<br>Phone: 608 441-2065<br>Fax: 608 441-2007 |
| Date Prepared: | September 29, 2006 |
| Trade Name: | Sonara and Sonara/tek Transcranial Doppler (TCD) System |
| Classification Name: | CFR Classification section 892.1550 (Product code IYN); CFR<br>Classification section 892.1570 (Product code ITX) |
| Classification: | Class II medical Device |
| Predicate Device: | The Sonara and Sonar/tek TCD devices are comparable to the following<br>predicate devices: |
| | - Doppler-Box (K051085) manufactured by Compumedics Gmbh<br>(DWL). Doppler-Box is a Transcranial Doppler Ultrasound device,<br>providing similar data to the Sonara/tek device. |
| | - TCD 100M (K002533) manufactured by Spencer Technologies. TCD<br>100M is a Transcranial Doppler Ultrasound device, providing similar<br>data to the Sonara device. |
| | - TC8080 Pioneer /Companion III (K053648) manufactured by VIASYS<br>HealthCare, Inc., Neurocare Group.<br>The TC8080 Pioneer /<br>Companion III are TCD devices with equivalent indications for use. |
| Device Description: | The Sonara device is a Transcranial and peripheral vascular Doppler<br>system, which is designed to measure blood flow velocities and other<br>hemodynamic parameters in a non-invasive manner, in intracranial,<br>extracranial and peripheral blood vessels.<br>The Sonara system includes an integrated 15" touch screen color LCD<br>display, integrated PC board and hard disk for data management and |
| Performance Standards: | None.<br>The design of the Sonara and Sonara/tek Transcranial Doppler (TCD)<br>devices conforms to the following voluntary standards:<br>IEC/EN-60601-1: Medical Electrical Equipment; Part 1: General<br>Requirements for Safety. Second edition (1990), including amendments<br>#1(1993), #2(1995), #13(1996).<br>IEC/EN 60601-1-2: Medical Electrical Equipment; Part 1-2: Collateral<br>Standard: Electromagnetic Compatibility- Requirements and Tests<br>(2001)<br>IEC/EN 60601-2-37: Medical Electrical Equipment; Part 2: Particular<br>Requirements for the Safety of Ultrasonic Medical Diagnostic and<br>Monitoring Equipment (2001) |
| Acoustic Output: | Acoustic output reporting for the Sonara devices was carried in<br>compliance with "Information for manufacturers seeking marketing<br>clearance of diagnostic ultrasound systems and transducers", September<br>30, 1997 (FDA Ultrasound Guidance (1997), Table 6-3, Page 6-9. |
| Test Data: | The Sonara and Sonara/tek Transcranial and vascular Doppler devices<br>have been subjected to extensive safety, performance testing, and<br>validation before release. Final testing of the Sonara device included<br>various performance tests designed to ensure that the device met all its<br>functional specifications. Tests have been performed to ensure the<br>device complies with industry and safety standards. |
| Substantial Equivalence: | The Sonara and Sonara/tek TCD are similar to currently distributed<br>pulsed Doppler Ultrasound systems with 2MHz, 4MHz and 8MHz<br>transducers intended for transcranial and peripheral Doppler<br>applications. Maximum acoustic output levels are below pre-amendment<br>levels for acoustic intensity for this application, and for Mechanical<br>Index for all applications. Power levels are displayed at all times during<br>scanning. A standard spectrum display is shown in both viewing |
{1}------------------------------------------------
display. The Sonara/tek is the same device, utilizing a personal computer for its opcration (via USB connection). The TCD system supports 2MHz, 4MHz and 8MHz ultrasound probe frequencies, either in a unilateral or bilateral configuration. For monitoring purpose, a special 2MHz monitoring probe (which fits into a headset) is provided. Online and Offline modes of operation are available. The system also includes a remote control and footswitch as accessories.
Intended Use / Indication for Use: The Sonara and Sonara/tek Transcranial Doppler are medical ultrasound devices for measuring the blood flow velocities in arteries and in veins noninvasively, consistent with the FDA Ultrasound System and Transducer Indication for Use forms cleared with the 510(k).
### Performance Standards: None
{2}------------------------------------------------
| | formats. The device includes an M-mode image. All of the above features are similar to these features in the predicate devices. |
|--------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Conclusions: | The conclusions drawn from the above Performance Testing and comparison to Predicate devices is that the Sonara and Sonara/tek Transcranial and peripheral vascular Doppler device and transducers are substantially equivalent in safety and efficacy to the predicate devices listed above. |
:
:
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird symbol, with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
NOV - 3 2006
VIASYS Healthcare, Inc., NeuroCare Group % Mr. Gary Syring Principal Consultant Quality & Regulatory Associates, LLC 800 Levanger Lane MADISON WI 53711
Re: K060421
Trade Name: Sonara and Sonara/tek Systems Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulation Number: 21 CFR 892.1570 Regulation Name: Diagnostic ultrasonic transducer Regulatory Class: II Product Code: IYN and ITX Dated: October 6, 2006 Received: October 10, 2006
Dear Mr. Syring:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
This determination of substantial equivalence applies to the following transducers intended for use with the Sonara and Sonara/tek Systems, as described in your premarket notification:
Transducer Model Number
2 MHz PW Hand-Held Transducer 2 MHz PW Monitoring Transducer 4 MHz Transducer 8 MHz Transducer
Image /page/3/Picture/14 description: The image is a circular logo or emblem. The text "1906-2006" is at the top, and the letters "FDA" are in the center. The word "Centennial" is below the letters. There are three stars below the word "Centennial". The text "Dedicated to Protecting and Promoting Your Health" surrounds the circle.
*Protecting and Promoting Public Health*
{4}------------------------------------------------
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This determination of substantial equivalence is granted on the condition that prior to shipping the first device, you submit a postclearance special report. This report should contain complete information, including acoustic output measurements based on production line devices, requested in Appendix G, (enclosed) of the Center's September 30, 1997 "Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers." If the special report is incomplete or contains unacceptable values (e.g., acoustic output greater than approved levels), then the 510(k) clearance may not apply to the production units which as a result may be considered adulterated or misbranded.
The special report should reference the manufacturer's 510(k) number. It should be clearly and prominently marked "ADD-TO-FILE" and should be submitted in duplicate to:
> Food and Drug Administration Center for Devices and Radiological Health Document Mail Center (HFZ-401) 9200 Corporate Boulevard Rockville, Maryland 20850
This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html
{5}------------------------------------------------
If you have any questions regarding the content of this letter, please contact Dr. Ewa Czerska at (240) 276-3666 .
Sincerely yours,
Nancy C Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure(s)
{6}------------------------------------------------
## Sonara System
#### Diagnostic Ultrasound Indications for Use Form
#### Fill out one form for each ultrasound system and each transducer.
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | |
|----------------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) |
| Ophthalmic | | | | X | | | | | | |
| Fetal | | | | | | | | | | |
| Abdominal | | | | | | | | | | |
| Intraoperative (specify) | | | | X | | | | | | |
| Intraoperative Neurological | | | | | | | | | | |
| Pediatric | | | | X | | | | | | |
| Small Organ (specify) | | | | | | | | | | |
| Neonatal Cephalic | | | | X | | | | | | |
| Adult Cephalic | | | | X | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral Vascular | | | | X | X | | | | | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | |
| Other (specify) | | | | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
an her marcanent of the Sonara device can be used during surgery to support Transcranial Doppler
Additional Comments:
Monitoring, Carotid Monitoring, Emboli Detection (HITS).
No associated transducer probes are applied invasively.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
Nancy Brogdon
(Division Sign-Off) Division of Reproductive, Abd and Radiological Devices 510k) Number _
{7}------------------------------------------------
## Sonara/tek System
Appendix F
#### Diagnostic Ultrasound Indications for Use Form
#### Fill out one form for each ultrasound system and each transducer.
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | |
|------------------------------------------------------------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) |
| Ophthalmic | | | | X | | | | | | |
| Fetal | | | | | | | | | | |
| Abdominal | | | | | | | | | | |
| Intraoperative (specify) | | | | X | | | | | | |
| Intraoperative Neurological | | | | | | | | | | |
| Pediatric | | | | X | | | | | | |
| Small Organ (specify) | | | | | | | | | | |
| Neonatal Cephalic | | | | X | | | | | | |
| Adult Cephalic | | | | X | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral Vascular | | | | X | X | | | | | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | |
| Other (specify) | | | | | | | | | | |
| N= new indication; P= previously cleared by FDA; E= added under Appendix E | | | | | | | | | | |
Additional Comments: The Sonara/tek device can be used during surgery to support Transcranial Doppler
Monitoring, Carotid Monitoring, Emboli Detection (HITS).
No associated transducer probes are applied invasively.
Prescription Use (Per 21 CFR 801.109)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Nancy C. hogdon
(Division Sign-Off) Division of Reproductive, Abdo and Radiological Devices 510(k) Number __
$$\varepsilon \text{-3}$$
Page 210
{8}------------------------------------------------
#### Appendix F
## 2 MHz PW Hand-Held Transducer
#### Diagnostic Ultrasound Indications for Use Form
#### Fill out one form for each ultrasound system and each transducer.
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | |
|----------------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) |
| Ophthalmic | | | X | | | | | | | |
| Fetal | | | | | | | | | | |
| Abdominal | | | | | | | | | | |
| Intraoperative (specify) | | | X | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | |
| Pediatric | | | X | | | | | | | |
| Small Organ (specify) | | | | | | | | | | |
| Neonatal Cephalic | | | X | | | | | | | |
| Adult Cephalic | | | X | | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral Vascular | | | X | | | | | | | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | |
| Other (specify) | | | | | | | | | | |
Intraoperative: Noninvasive use during surgery to support Transcranial Doppler Additional Comments:_
Monitoring, Carotid Monitoring, Emboli Detection (HITS).
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
Nancy Brogdon
(Division S Division of Reproductive, Abd and Radiological Devices F-3 10(k) Number
Page 211
{9}------------------------------------------------
#### Appendix F
# 2 MHz PW Monitoring Transducer
### Diagnostic Ultrasound Indications for Use Form
#### Fill out one form for each ultrasound system and each transducer.
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | |
|------------------------------------------------------------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) |
| Ophthalmic | | | X | | | | | | | |
| Fetal | | | | | | | | | | |
| Abdominal | | | | | | | | | | |
| Intraoperative (specify) | | | X | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | |
| Pedialric | | | X | | | | | | | |
| Small Organ (specify) | | | | | | | | | | |
| Neonatal Cephalic | | | X | | | | | | | |
| Adult Cephalic | | | X | | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral Vascular | | | X | | | | | | | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | |
| Other (specify) | | | | | | | | | | |
| N= new indication; P= previously cleared by FDA; E= added under Appendix E | | | | | | | | | | |
Additional Comments: Intraoperative: Noninvasive use during surgery to support Transcranial Doppler
Monitoring, Carotid Monitoring, Emboli Detection (HITS)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
Danay Shrodon
Division Slar of Reproductive. liological Devices 510(k) Numbe --3
{10}------------------------------------------------
## 4 MHz Transducer
Appendix F
#### Diagnostic Ultrasound Indications for Use Form
#### Fill out one form for each ultrasound system and each transducer.
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | |
|----------------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | | | | | | | |
| Abdominal | | | | | | | | | | |
| Intraoperative (specify) | | | X | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | |
| Pediatric | | | X | | | | | | | |
| Small Organ (specify) | | | | | | | | | | |
| Neonatal Cephalic | | | X | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral Vascular | | | | X | X | | | | | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | |
| Other (specify) | | | | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
#### Additional Comments:_
Prescription Use (Per 21 CFR 801.109)
Intraoperative: Use during surgical procedures to support noninvasive extracranial monitoring.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Nancy Broadon
Division S Division of Active, Abdominal, and Radiological Devices 510/k) Number
F-3
Page 213
{11}------------------------------------------------
## 8 MHz Transducer
Appendix F
#### Diagnostic Ultrasound Indications for Use Form
### Fill out one form for each ultrasound system and each transducer.
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | | | | | | | | | |
|----------------------------------|-------------------|---|---|-----|-----|---------------|-------------------|------------------------|--------------------|-----------------|
| | A | B | M | PWD | CWD | Color Doppler | Amplitude Doppler | Color Velocity Imaging | Combined (specify) | Other (specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | | | | | | | |
| Abdominal | | | | | | | | | | |
| Intraoperative (specify) | | | X | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | |
| Pediatric | | | X | | | | | | | |
| Small Organ (specify) | | | | | | | | | | |
| Neonatal Cephalic | | | X | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral Vascular | | | X | X | | | | | | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | |
| Other (specify) | | | | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
Additional Comments:
Intraoperative: Use during surgical procedures to support noninvasive extracranial monitoring.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
Nancyc brogdon
(Division Sign-Off) Division of Reproductive, Al ominal and Radiological Device 510(k) Number ....
-- Page 214
E-3
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.