Slim UP ULTRA

K151722 · Aime Medical, Inc. · NGX · Mar 18, 2016 · Physical Medicine

Device Facts

Record IDK151722
Device NameSlim UP ULTRA
ApplicantAime Medical, Inc.
Product CodeNGX · Physical Medicine
Decision DateMar 18, 2016
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 890.5850
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Slim UP® ULTRA is indicated for the use by healthy persons to apply transcutaneous electrical muscle stimulation (EMS) through skin contact electrode for the purpose of improvement of muscle tone of buttocks muscles.

Device Story

Slim UP® ULTRA is a powered muscle stimulator used in health spas and beauty salons. It applies transcutaneous electrical muscle stimulation (EMS) to the buttocks via 16 skin-contact electrodes. The device features 8 channels delivering a compensated two-phase rectangular waveform with 5-7 selectable multi-phase treatment programs. It includes an infrared (IR) unit with 6-7 lamps on a mobile arm for patient comfort; the IR component has no therapeutic effect. The device is operated by staff in a clinic setting. The system uses a microcontroller to manage stimulation parameters, intensity, and safety features, including an automatic shut-off and a safety circuit that resets the machine if current exceeds 100mA. The output is controlled by the operator via a control panel. By stimulating muscle contraction, the device aims to improve muscle tone, providing a non-invasive aesthetic benefit to the user.

Clinical Evidence

Bench testing only. Evidence includes biocompatibility testing per ISO 10993-1, electrical safety and EMC testing per IEC 60601-1, IEC 60601-1-2, and IEC 60601-2-10, and software verification and validation testing.

Technological Characteristics

Powered muscle stimulator; 8-channel EMS; 6-7 IR lamps (non-therapeutic). Materials: Polycarbonate arc, nylon screws. Dimensions: 95x129x187 cm; Weight: 161 kg. Power: 230VAC. Waveform: Biphasic symmetric rectangular. Connectivity: Standalone. Sterilization: Non-sterile. Software: Microcontroller-based firmware.

Indications for Use

Indicated for healthy persons to improve muscle tone of buttocks via transcutaneous electrical muscle stimulation (EMS).

Regulatory Classification

Identification

A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized graphic of three human profiles facing right, layered on top of each other. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the graphic. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 March 18, 2016 Aime Medical, Inc % Mr. John O'Brien Regulatory Consultant AJW Technology Consultants, Inc. 445 Apollo Beach Blvd. Apollo Beach, Florida 33572 Re: K151722 Trade/Device Name: Slim UP ULTRA Regulation Number: 21 CFR 890.5850 Regulation Name: Powered muscle stimulator Regulatory Class: Class II Product Code: NGX Dated: February 18, 2016 Received: February 19, 2016 Dear Mr. O'Brien: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in {1}------------------------------------------------ the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Jennifer R. Stevenson -A For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K151722 Device Name SLIM UP® ULTRA Indications for Use (Describe) Slim UP® ULTRA is indicated for the use by healthy persons to apply transcutaneous electrical muscle stimulation (EMS) through skin contact electrode for the purpose of improvement of muscle tone of buttocks muscles. Type of Use (Select one or both, as applicable) | <input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) | |----------------------------------------------------------------------------------| | <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) | ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) SUMMARY (As required by 807.92) #### I. SUBMITTER Company Name: Aime Medical Inc. Address: 1258 West Bay Drive, Suite F. Largo, FL 33770 USA Phone: +1-727-286-6227 Contact Person: Paul Guilbaud Email: paulg@AimeMedicalinc.com Date Prepared: 16 Feb 2016 ## REGULATORY CORRESPONDENT AJW Technology Consultants, Inc 445 Apollo Beach, Blvd Apollo Beach, FL 33572 Phone: 813-645-2855 x101 Fax: 813-645-2856 Contact Person: John O'Brien Email: jobrien@ajwtech.com #### II. DEVICE Name of Device: Common or Usual Name: Classification Name: Regulation Number: Device Panel: Regulatory Class: Main Product Code: Slim UP® ULTRA Power muscle Stimulator. Power Muscle Stimulator 890.5850 Physical Medicine II NGX {4}------------------------------------------------ #### III. PREDICATE DEVICE | 510(k) Number: | K123075 | |----------------|---------------------------------------------------------------------------------------| | Product Code: | NGX | | Manufacturer: | Well Life Healthcare Limited | | Trade Name: | Buttock Muscle Stimulator Model WL 2413B; Buttock Muscle<br>Stimulator Model WL-2413E | The predicate has not been subject to a design-related recall. #### IV. DEVICE DESCRIPTION Slim UP® ULTRA is a new beauty appliance that works on the principles of Electrostimulation. Slim UP® is used in environments such as Health Spa's and Beauty Salons. Slim UP® ULTRA has following major components: - Main power switch ■ - . Control panel - Infrared unit - 트 Distribution of electro-stimulation channels. Slim UP® ULTRA electrostimulation section has 8 channels for body electrostimulation with compensated two-phase rectangular waveform, with individual intensity adjustment and 5 different multi-phase treatment programs diversified to meet the needs of the single case. Slim UP® ULTRA infrared section has 6 infrared ray sources housed in a height-adjustable, mobile arm that adapts to any couch, ensuring the ideal working distance is maintained at all times. IR source is used for patient comfort only. It does not have any therapeutic effect. Slim UP® ULTRA is a latest-generation machine occupying only half a square meter. It can be located in any booth, can be adapted to any bed and can be moved anywhere. {5}------------------------------------------------ The table below describes the list of associated accessories included with device. | Description | Quantity | |------------------------------------------------------------|----------| | Polycarbonate arc | 1 | | 100W IR lamp | 7 | | Nylon screw+ washer for fixing the polycarbonate arc | 5 | | Protective goggles | 3 | | 4 mm hexagonal wrench | 1 | | Threaded pivot for fixing IR apparatus | 2 | | Temperature probe | 1 | | Disc for electrostimulation | 100 | | 510k cleared electrode with conductive gel under (K091163) | 54 | #### V. INDICATIONS FOR USE Slim UP® ULTRA is indicated for the use by healthy persons to apply transcutaneous electrical muscle stimulation (EMS) through skin contact electrode for the purpose of improvement of muscle tone of buttocks muscles. {6}------------------------------------------------ | And Sun Intelle | | |--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--| | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | A has and a manufacturer and a manufacturer and the may be and | | | | | | | | | | | | | | | CONSTITUTE CLOUIS CLO WE THE WATERS WAS WITH A STATE TO PARTY OF PARTY OF PARTY OF PARTY OF PARTY OF PARTY OF PARTY OF PARTY OF PARTY OF PARTY OF PARTY OF PARTY OF PARTY OF P | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | 1 | | | | | | | | | | | | 1<br>1 | | | | | | Clear Income Income | | | | | | | | | I | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | 1 | | | | | | | | | is vaaa vaa vaavaama vaamaalis amaali aasta<br>1 | | | | | | 1 | | | 2 | | | | | | A C A T M M A W L L A | | | | | | | | | | | | | | | | | | | | | | | | Concellined of Concellented | | | | | | | | | | | | | | | | | I Subject and predicate devices are intended for similar treatment condition based on the principle of operation they uses | Parameter | Name of Device | Product Code | Regulation Number | 510 K Number | Intended use | Predicate Device | Predicate Difference | |----------------|----------------|--------------|--------------------|--------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------| | Slim UP® ULTRA | Slim UP® ULTRA | NGX, NUV | 878.4810, 890.5850 | K151722 | Slim UP® ULTRA is indicated for the<br>use by healthy persons to apply<br>transcutaneous electrical muscle<br>stimulation (EMS) through skin contact<br>electrode for the purpose of<br>improvement of muscle tone of<br>buttocks muscles. | Buttock Muscle Stimulator<br>(WL 2413 E Model) | - | | | | | | | | NGX | - | | | | | | | | 890.5850 | - | | | | | | | | K123075 | - | | | | | | | | Buttock Muscle Stimulator, model WL-<br>2413E, is -intended for use by healthy<br>persons to apply transcutaneous<br>electrical muscle stimulation (EMS)<br>through skin contact electrode for the<br>purpose of improvement of muscle tone<br>of the buttocks muscles | Same | {7}------------------------------------------------ | Parameter | Slim UP® ULTRA | Predicate Device | Predicate<br>Difference | |-----------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------| | Name of Device | Slim UP® ULTRA | Buttock Muscle Stimulator<br>(WL 2413 E Model) | | | Product Component | 1. Main power switch<br>2. Control panel<br>3. Infrared generation unit<br>4. Electro-stimulation channels<br>5. Skin contact electrodes | 1. Electronic Stimulator Module<br>2. Skin contact electrodes<br>3. Conductive gel,<br>4. 3AAA batteries,<br>5. 2 read wires<br>6. Storage bag | - | | Treatment Energy | Electrostimulation | Electrostimulation | Same | | Energy source | 1) 16 Electrodes (K091163)<br>2) 6 IR Lamps | K082065, Silicon pad electrodes (7.0 cm diameter) | Different | | Pigmentation Effect | No | No | Similar | | Dimensions | 95 cm x 129 cm x 187 cm (W*D*H) | 64 X20 X90 (W*D*H) | Different | | Weight | 161 Kg | 0.08 Kg (80 Gm) (including batteries) | Different | | Electrical Requirement | 230VAC, Single-phase | 1.5 V x 3 AAA size battery | Different | | Safety features and<br>compliance with safety<br>standard | Compliant | Compliant | Same | | Patient body contact<br>material | Biocompatible | Biocompatible | Same | | Parameter | Slim UP® ULTRA | Predicate Device | Predicate Difference | | Name of Device | Slim UP® ULTRA | Buttock Muscle Stimulator (WL 2413 E Model) | Different | | Treatment Time | Between 15 and 60 minutes depend on treatment program selected | 20-40 Minutes | Different | | Compliance with Voluntary Standards? | Compliant | Compliant | Same | | Housing Materials and Construction | Polycarbonate arc<br>Nylon screw and washer for fixing the polycarbonate arc | ABS | Different | | Principle of Operation | Slim UP® ULTRA works on the basic principles of Electrostimulation. Skin contact electrodes are applied on treatment area. | The Buttock Muscle Stimulator, model WL-241 3E, comprises an electronic stimulator module with silicon pad electrodes which connect the signals from the stimulator to the skin. The electrodes are placed on skin | Same | | Anatomical Site of Application | Buttocks | Buttocks | Same | | Number of Electrode Used | 16 Electrostimulation electrode | K082065, Silicon Pad Electrode (7.0 cm diameter) | Different | | Mode OR Program name | 2 TONE (F2 work phase) | WL-2413E | Different | | Wave form | Biphasic | Biphasic | Same | | Shape | Rectangular | Rectangular | Same | {8}------------------------------------------------ AIME MEDICAL - SLIM UP ULTRA -AI RESPONSE 7 of 16 {9}------------------------------------------------ | Parameter | Name of Device | Slim UP® ULTRA | Slim UP® ULTRA | Predicate Device<br>Buttock Muscle Stimulator<br>(WL 2413 E Model) | Predicate Difference | |----------------------------------|----------------|----------------------------------------------------------------------------------------------------------|-----------------------|--------------------------------------------------------------------|----------------------| | Maximum output voltage | | 100Vpp @ 500Ω<br>184Vpp @ 1 KΩ<br>234Vpp @ 2 KΩ<br>240Vpp @ 10 KΩ<br>(Single Phase) | | 40.8 V @500Ω<br>70.0 V @2KΩ<br>106.0 V @10KΩ | Different | | Maximum output current | | 100mA @ 500Ω<br>92 mA @ 1 KΩ<br>58.5 mA @ 2 KΩ<br>12 mA @ 10 KΩ<br>(calculated on single positive pulse) | | 81.6mA@500 Ω<br>35.0mAr@2KΩ<br>10.6mA@10KΩ | Different | | Duration of primary phase (µsec) | | 400 µsec | | 300 µsec | Different | | Pulse duration | | 400 µsec | | 720 µsec | Different | | Frequency (Hz) | | 40 Hz | | 70 Hz | Different | | For multiphase | | Symmetrical phases<br>Yes | | Yes | Same | | Waveform Phase duration ms only | | | 800 µsec | Not Applicable | Same | | Method of Line current Isolation | | | Type BF | Type BF | Same | | Number of output mode | | | | 04 | Same | | Pulse per burst | | | Same for each program | Same for each program | Same | | Burst per second | | | Same for each program | Same for each program | Same | {10}------------------------------------------------ | Parameter | Slim UP® ULTRA | Predicate Device<br>Buttock Muscle Stimulator<br>(WL 2413 E Model) | Predicate<br>Difference | |-----------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------|-------------------------| | Name of Device | Slim UP® ULTRA | | | | Burst duration | Same for each program | Same for each program | Same | | Duty cycle | Same for each program | Same for each program | Same | | Method of achieving<br>zero net charge/pulse | Biphasic symmetric wave for each<br>pulse | Biphasic symmetric wave for each<br>pulse | Same | | Regulated Current or<br>Regulated Voltage? | Voltage | Voltage | Same | | Software/Firmware/Mic<br>roprocessor control? | Yes | Yes | Same | | Automatic Shut Off? | Yes | Yes | Same | | RMS (Root Mean<br>Square) Current | 100mA during the active phase of the<br>current pulse, square wave with null<br>medium value and period of 600μs | 38.644mA (500Ω)<br>16.907 mA (2KΩ)<br>5.120 mA (10 Ω) | Different | | Prescription or OTC | Prescription | OTC | Different | | Number of Treatment<br>Programs | 7 different programs that should be<br>chosen based on the individual<br>requirements of each case | N/A | Different | | Method for Channel<br>Isolation | Sole transformer on all channels from<br>12V and the tension that goes to the<br>electrostimulation channels (it is not an<br>insulating transformer).<br>If current goes over 100mA (short<br>output transistor) a safety circuit<br>intervenes by resetting the machine | Output coil | Different | | Sterilization | Provided Non Sterile | Provided Non sterile | Same | 9 of 16 {11}------------------------------------------------ | Comparison Feature | Slim UP Ultra | Predicate Device | Remarks | |------------------------------------|-----------------------------------------------------|-------------------------------------------|-----------| | Net Charge | 0 @ 500Ω (symmetrical positive and negative pulses) | 0 (symmetrical stimulation biphasic wave) | Same | | Max Phase Charge | 40 μC @ 500Ω | 24μc | Different | | Max Average Current Density (jmax) | 0.089 mA/cm² @ 500Ω | 0.0997 mA/cm2 | Different | | Max Power Density (W/cm²) | 0.00014 W/cm² @ 500Ω with electrodes 4x9 cm | 0.00399 watts/cm2 | Different | | Burst Mode | Yes | Yes | Same | Significant output characteristics {12}------------------------------------------------ ## Similarities in Technological characteristics: - o Intended use - Product components ● - Treatment energy . - Pigmentation effect . - Safety feature and compliance with safety standard 0 - Patient body contact material o - Compliance with voluntary standards o - Principle of operation o - Anatomical site of application o - o Waveform - Shape o - Symmetrical phases for multiphase waveform only ● - Method of line current isolation ● - Number of output mode ● - Pulse per Burst ● - Burst per Second ● - Bust duration ● - Duty cycle - Method for achieving zero net charge/pulse ● - Current/ Voltage regulation ● - Software/firmware control ● - Automatic shut off ● - Sterilization ● - Net Charge ● - Max Phase Charge ● - o Burst Mode # Differences in Technological characteristics - Energy Source ● - Physical characteristics (weight, dimensions) ● - Electrical Requirements ● - Treatment Time o - Housing material & construction ● - Number of Electrode ● - Mode or program name ● - Maximum output voltage ● - o Maximum output current - Duration of primary phase o {13}------------------------------------------------ - Pulse duration ● - Frequency ● - Root Mean Square (RMS) Current ● - Prescription or OTC ● - Number of Treatment programs - 0 Method for channel isolation - 0 Maximum Phase Charge - O Max. current density - Max, power density ## Justification of predicate differences: Differernce-1: Energy Source - IR source Predicate device: Buttock muscle stimulator does not contain IR modality. New Device: The Slim UP® ULTRA contains IR modality in addition to Electrostimulation modality. Intended Use: Same ## Different questions of Safety and effectiveness? No ## Justification of Difference: IR source does not have any therapeutic effect it is just used for patient comfort only. Also IR modality used in subject device is not able to raise and maintain body temperature above 38-41 degree Celsius for 10 minutes. IR source used in Slim UP® ULTRA is similar to reference device i.e. Syneron Medical's Velashape (K122579). So this minor technological difference does not raise any new questions of safety and efficacy. ## Difference-2: Physical Characteristics (Dimension and weight) Predicate device: Buttock muscle stimulator is 0.08 Kg in weight and has physical dimensions of 64cm Width X20cm Diameter X90 cm Height. New Device: The Slim UP® ULTRA has physical dimension of 95cm Width X 129cm Diameter X 187cm Height and 161 Kg in weight. Intended Use: Same # Different questions of Safety and effectiveness? No ## Justification of Difference: The difference in physical dimensions and weight doesn't introduce any additional safety issues. Due to its compact size Slim UP® ULTRA can be located in any booth, can be adapted to any bed and can be moved anywhere. {14}------------------------------------------------ ## Difference-3: Housing material and construction ### Predicate device: Buttock muscle stimulator has asbestos as primary housing material. ### New Device: Different parts of Slim UP® ULTRA were made from different housing material. It has polycarbonate arc, nylon screw and washer for fixing the polycarbonate arc. - Intended Use: Same ## Different questions of Safety and effectiveness? No Justification of difference: This predicate difference will not raise any questions of safety and efficacy as patient body contact parts of Slim UP® ULTRA are biocompatible and Slim UP® ULTRA complies with international safety standard EN ISO 14121-1:2007- Safety of the equipment. Risk evaluation part-1: Principles. ## Difference-4: Number of electrodes Predicate device: The information regarding number of electrodes in predicate device is not available. New device: Slim UP® ULTRA has 16 skin contact electrodes. Intended Use: Same # Different questions of Safety and effectiveness? No ### Justification of difference: The difference in number of electrodes do not raise new question of safety or effectiveness. Slim UP® ULTRA use 510k cleared electrode FIAB PG474W (K091163). Apart from this electromagnetic safety of subject device has been established by conducting EMC testing in complies with international EMC standard 60601-1-2 and 60601-2-10. So this difference will not raise any additional questions of safety and efficacy. Difference-5: Functional parameters like Mode or Program name, Pulse duration, Number of output mode, pulse per burst, burst per second, Burst duration, Duty cycle, RMS current, Number of treatment program Method of channel isolation, Maximum current and power density. ### Predicate device: Buttock muscle stimulator has following different functional parameters: Mode or Program name: WL-2413E Pulse Duration: 720 usec Number of output mode: 04 Pulse per burst: Same for each program Burst per second: Same for each program {15}------------------------------------------------ Burst duration: Same for each program Duty cycle: Same for each program RMS current: 38.644mA (500Ω), 16.907mA (2KΩ), 5.120 mA (10 Ώ) Number of treatment program: N/A Method of Channel isolation: Output coil Maximum Current density: 0.0997 mA/cm2 Maximum Power density: 0.00399 watts/cm2 ### New Device: Slim UP® ULTRA has following different functional parameters: Mode or Program name: 2 TONE (F2 work phase) Pulse Duration: 400 usec Number of output mode: N/A Pulse per burst: 160 Burst per second: 1/8 Burst duration: 4 Sec Duty cycle: 1/2 RMS current: 100mA during the active phase of the current pulse, square wave with null medium value and period of 600us Number of treatment program: 7 different programs that should be chosen based on the individual requirements of each case. Method of channel isolation: Sole transformer on all channels from 12V and the tension that goes to the electrostimulation channels (it is not an insulating transformer). If current goes over 100mA (short output transistor) a safety circuit intervenes by resetting the machine Maximum current density: 2, 8 mA/cm2 @ 500Ω Maximum power density: 8, 96 mW/cm² @ 500Ω with electrodes 4x9 cm ## Intended Use: Same #### Different questions of safety and effectiveness: No Justification of difference: The minor difference in functional parameter between predicate device and subject device do not raise new questions of safety or efficacy as Electromagnetic safety of Slim UP® ULTRA has been established by conducting EMC testing in complies with international EMC standard 60601-1. 60601-1-2 and 60601-2-10. ### Difference-6: Electrical Requirement ## Predicate device: Runs on 3 AAA batteries at 1.5v. New Device: 230VAC, Single-Phase Intended Use: Same Different questions of Safety and effectiveness? No Justification of difference: {16}------------------------------------------------ The difference in Electrical Requirements between predicate device and subject device do not raise new questions of safety or efficacy as Electromagnetic safety of Slim UP® ULTRA has been established by conducting EMC testing in complies with international EMC standard 60601-1, 60601-1-2 and 60601-2-10. The conclusion of this technical comparison is that Slim UP® ULTRA is substantially equivalent to the Buttock Muscle Stimulator and have been verified by subsequent testing. Any difference between Predicate device and Slim UP® ULTRA do not raise any new questions for safety and efficacy. ## PERFORMANCE DATA The following performance testing was provided in support of the substantial equivalence determination. ## Biocompatibility Testing Patient applied parts of Slim UP® ULTRA are biocompatible as per the International standard ISO 10993-1:2009- Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process. ## Electrical safety and electromagnetic compatibility (EMC) The following standards have been complied within designing, manufacturing and testing of Slim UP® ULTRA: - IEC 60601-1:2012, Medical Electrical Equipment - Part 1: General Requirements for Safety. - IEC 60601-1-2:2007, Medical Electrical Equipment Part 1-2: General ● Requirements for Safety - Collateral Standard: Electromagnetic Compatibility -Requirements and Tests. - IEC 60601-2-10:2000: Medical electrical equipment Part 2-10: Particular . requirements for the basic safety and essential performance of nerve and muscle stimulators - EN ISO 14121-1:2007- Safety of machinery -- Risk assessment -- Part 1: Principles. o ### Software Verification and Validation Testing Software validation was completed to check the safety requirements implemented by the firmware resident on the Microcontroller used on the Slim Up® Ultra, in order to provide objective evidence that the firmware itself operates as indicated in the user manual and does not cause any risks connected to its use. {17}------------------------------------------------ #### CONCLUSION I. The performance data demonstrates that the Slim UP® ULTRA performs as intended and similar to that of predicate devices and exhibits comparable mechanical, functional and technological characteristics to that of the predicate device. Any technological or engineering differences existing between subject device and predicate devices is insignificant in terms with safety and effectiveness of subject device. Based on these factors, the Slim UP® ULTRA can be considered substantially equivalent to the predicate device and safe and effective for its intended use.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...