BUTTOCK MUSCLE STIMULATOR MODEL WL-2413B; BUTTOCK MUSCLE STIMULATOR MODE WL-2413E

K123075 · Well-Life Healthcare Limited · NGX · Jul 11, 2013 · Physical Medicine

Device Facts

Record IDK123075
Device NameBUTTOCK MUSCLE STIMULATOR MODEL WL-2413B; BUTTOCK MUSCLE STIMULATOR MODE WL-2413E
ApplicantWell-Life Healthcare Limited
Product CodeNGX · Physical Medicine
Decision DateJul 11, 2013
DecisionSESE
Submission TypeAbbreviated
Regulation21 CFR 890.5850
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Buttock Muscle Stimulator, model WL-2413E, is intended for use by healthy persons to apply transcutaneous electrical muscle stimulation(EMS) through skin contact electrode for the purpose of improvement of muscle tone of the buttocks muscles

Device Story

Buttock Muscle Stimulator (model WL-2413E) is a two-channel, battery-operated EMS device. System includes electronic stimulator module, shorts with integrated silicon pad electrodes, conductive gel, and lead wires. Device applies transcutaneous electrical muscle stimulation to buttocks muscles to improve muscle tone. Operated by healthy individuals in home setting. User wears shorts with electrodes in contact with skin; stimulator generates biphasic, rectangular stimulation signals. Device features four output modes, timer (20-40 minutes), and automatic shut-off. Output parameters (voltage, current, pulse duration) are controlled by internal microprocessor. Device provides non-invasive muscle stimulation; intended for OTC use.

Clinical Evidence

Bench testing only. Compliance with IEC 60601-1, IEC 60601-1-2, and IEC 60601-2-10 standards. Software verification performed per FDA guidance.

Technological Characteristics

Powered muscle stimulator; 3x AAA battery source. ABS housing. Silicon pad electrodes (7.0 cm diameter). Biphasic, rectangular waveform. 4 output modes. Type BF patient isolation. Microprocessor-controlled. Connectivity: None. Sterilization: Not applicable (non-sterile). Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-10.

Indications for Use

Indicated for healthy persons seeking improvement of buttock muscle tone via transcutaneous electrical muscle stimulation (EMS).

Regulatory Classification

Identification

A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## 510(K) Summary The assigned 510(k) number is: K123075 #### Submitter's Identifications: 1. Well Life Healthcare Limited 1F., No. 16, Lane 454, Jungjeng Road, Yunghe City, Taipei County 234, Taiwan, ROC Contact: Jenny Hsieh Telephone: + 886 2 2928 2112 Date of Summary Preparation: May 25, 2013 #### 2. Name of the Device: Trade/Device Name: Buttock Muscle Stimulator, model: WL-2413E. Regulation Number: 21 CFR 890.5850 Requlation Name: Power muscle stimulator. Regulatory Class: II Product Code: NGX #### 3. Information of the 510(k) Cleared Device (Predicate Device): SPORT-ELEC models Body Control System "BS"( K092142) #### Device Description: 4. The Buttock Muscle Stimulator, model WL-2413E, is a two channel battery operated muscle stimulation system specifically designed to exercise the buttocks' muscle. The device comprises mainly the electronic stimulator module which generates the required stimulation signals. The Buttock Muscle Stimulator, model WL-2413E, comprises a short with silicon pad electrodes which connect the signals from the stimulator to the skin. The electrodes are placed on the inner surface of the accessories. The product is supplied with a 510(K) cleared conductive gel, a user's manual, 3AAA batteries, 2 lead wires and a storage bag. The required power of the device is derived from 3 batteries located in a compartment protected by a removable battery cover. The electrodes are in the inner surface of the shorts which is worn as shown on the picture of user manual. There is no current passed from side to side. The user cannot access the wiring or connectors within the shorts. The Buttock Muscle Stimulator, model WL-2413E, is intended for use by healthy persons to apply transcutaneous electrical muscle stimulation(EMS) through skin contact electrode for the purpose of improvement of muscle tone of the buttocks muscles {1}------------------------------------------------ #### 5. Intended Use: The Buttock Muscle Stimulator, model WL-2413E, is intended for use by healthy persons to apply transcutaneous electrical muscle stimulation(EMS) through skin contact electrode for the purpose of improvement of muscle tone of the buttocks muscles #### Discussion of Non-Clinical Tests Performed Determination of Substantial Equivalence are റ. as follows: Compliance to applicable voluntary standards includes IEC 60601-2-10, as well as IEC 60601-1, and IEC 60601-1-2 requirement. In addition to the compliance of voluntary standards, the software verification has been carried out according to the FDA software guidance. Discussion: The Compliance to applicable voluntary standards as above mentioned indicates that the new device in this submission used the same standards as that of predicate device in addition to the software validation standard. For our new device, the software validation was prepared according to FDA guidance, however the predicate device used IEC 60601-1-4 to prepare their software validation report. For this difference we believe that we used more adequate reference tool to prepare software validation report than the predicate device. Therefore, we consider that the compliance of standards included in our submission is adequate for the determination of substantial equivalence. ## 7. Comparison of Significant device features | Comparison<br>Features | Models to be Compared | | |------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------| | Model | SPORT-ELEC Body Control System<br>"BS" | WL-2413E | | 510(K) No. | K092142 | K123075 | | Prescription or OTC | OTC | Same | | Indication for use | intended for use by healthy persons to<br>apply transcutaneous electrical muscle<br>stimulation(EMS) through skin contact<br>electrode for the purpose of improvement<br>of muscle tone of the buttocks muscles | Same | | FDA product code | NGX | Same | | Electrode Used | Self Adhesive Electrode( 5 X 5 cm) | K082065, Silicon Pad Electrode<br>( 7.0 cm diameter) | | Manufacturer | SPORT-ELECT | Well-Life | {2}------------------------------------------------ ## 8. Comparison of Unit Characteristics & Output Specification | Mode or Program Name | Predicate Device | New Device | | | | | |--------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------|---------------------------------------------|----------------------------|--------------------|----------------------------------------|----------------------------------------| | Waveform (e.g., pulsed monophasic, biphasic) | Biphasic | Biphasic | | | | | | Shape (e.g., rectangular, spike, rectified<br>sinusoidal) | Rectangular | Rectangular | | | | | | Maximum Output Voltage (volts)-<br>(+/- 20%) | 31.2V @500Ω<br>66.0V @2KΩ<br>66.0V @10KΩ | 40.8V @500Ω<br>70.0V @2KΩ<br>106.0V @10KΩ | | | | | | Maximum Output Current (mA)-<br>(+/- 20%) | 62.4A @500Ω<br>33.0A @2KΩ<br>6.6A @10KΩ | 81.6mA @500Ω<br>35.0mA @2KΩ<br>10.6mA @10KΩ | | | | | | Duration of primary phase (usec) | 200 Max | 300 Max | | | | | | Pulse Duration (usec) | 540 Max. | 720 Max. | | | | | | Frequency (Hz) [or Rate (pps)] | 70 Max | 70 Max | | | | | | For<br>multiphasic<br>waveforms<br>only: | Symmetrical phases?<br>Phase Duration (include units),<br>(Stage range, if applicable),<br>(both phases, if asymmetrical) | Yes<br>Not applicable | Yes<br>Not applicable | | | | | 510(K) Number | K092142 | K123075 | | | | | | Device Name and Model | BS | WL-2413E | | | | | | Manufacturer | SPORT-ELECT | Well-Life | | | | | | Power Source(s) | 1.5Vx3<br>(AAA Size) | 1.5Vx3<br>(AAA Size) | | | | | | - Method of Line current Isolation<br>- Patient Leakage Current | Type BF<br>--- | Type BF<br>--- | | | | | | - Normal condition (uA) | Under 0.1 | Under 0.1 | | | | | | - single Fault condition (uA) | Under 0.5 | Under 0.5 | | | | | | Average DC current through electrodes when<br>device is on but no pulses are being applied ( uA) | Not applicable | Not applicable | | | | | | Number of Output Modes | 4 | 4 | | | | | | Number of Output<br>Channels: | Synchronous or<br>Alternating?<br>Method of Channel<br>Isolation | Synchronous<br>Output Coil | Synchronous<br>Output Coil | | | | | Regulated Current or Regulated Voltage? | Voltage | Voltage | | | | | | Software/Firmware/Microprocessor control? | Yes | Yes | | | | | | Automatic Overload Trip? | No | No | | | | | | Automatic No-Load Trip? | Yes | Yes | | | | | | Automatic Shut Off? | Yes | Yes | | | | | | User Override control? | No | No | | | | | | Indicator Display: | On/Off Status?<br>Low Battery?<br>Voltage/Current<br>Level? | Yes<br>Yes<br>Yes | Yes<br>Yes<br>Yes | | | | | Timer Range (Minutes) | 20-40 | 20-40 | | | | | | Compliance with Voluntary Standards? | IEC 60601-2-10 | IEC 60601-2-10 | | | | | | Compliance with 21 CFR 898? | Yes | Yes | | | | | | Weight (g) including battery | 171.6 | 80 | | | | | | Dimensions (mm.) [W x H x D] | 87×43×69 | 64×90×20 | | | | | | Housing Materials and construction | ABS | ABS | | | | | | Pulse per burst | Same for each program | Same for each program | | | | | | Burst per second | Same for each program | Same for each program | | | | | | Bust duration | Same for each program | Same for each program | | | | | | Duty Cycle | Same for each program | Same for each program | | | | | | Method of achieving zero net charge for net | Method of achieving zero net charge/pulse | Biphasic symmetric | | Biphasic symmetric | Biphasic symmetric wave for each pulse | Biphasic symmetric wave for each pulse | · {3}------------------------------------------------ | Comparison feature | Model | | Remarks | |-------------------------|--------------------|--------------------|---------| | | SPORT-ELECT BS | WL-2413E | | | Net charge | 0 | 0 | Note 1 | | Max. phase charge | 14 uc | 24 uc | | | Max. current Density | 0.0784 mA/cm² | 0.0997 mA/cm² | | | Max. Average<br>current | 38.43mA | 38.644mA | | | 500 Ω | | | | | 2K Ω | 14.892mA | 16.907 mA | | | (RMSA) | | | | | 10 Ω | 4.996mA | 5.120 mA | | | Max. Power Density | 0.00274 Watts/ cm² | 0.00399 Watts/ cm² | | | Burst Mode | Yes | Yes | | ## 9. Significant output characteristics comparison table: Note 1: Net Charge is Zero for the symmetrical stimulation biphasic wave ### 10. Summary for the technology comparison. Basically the WL-2413E has the similar technological characteristics with the predicate device in the product design, material, energy source type, main program mode and the main output waveform... etc. There exist some difference in the detailed output parameters( mainly in the pulse duration and electrode sizes). Through the detailed calculation comparison of stimulation output energy for each operation mode(in particular the output current density), we found the output level in each operation mode for our WL-2413E and predicate device are very close and within the acceptable range as specified in FDA guidance. So we believe the difference in detailed output parameters does not affect the determination of substantial equivalence. ## 11. Conclusions The Buttock Muscle Stimulator, model: WL-2413E have the same intended use and the similar technological characteristics as the cleared device- SPORT-ELEC models Body Control System "BS"( K092142). Moreover, verification and validation tests contained in this submission demonstrate that the difference in the submitted models could maintain the same safety and effectiveness as that of cleared device. In the other words, those engineering difference do not: (1) affect the intended use or (2) alter the fundamental scientific technology of the device. {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. Inside the circle is a stylized symbol that resembles a person embracing another person, or a bird in flight, composed of three curved lines. #### DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 ### July 11, 2013 Well-Life Healthcare Ltd C/O Ms. Jennv Hsieh 1F. No. 16, Lane 454, Jungjeng Road Yunghe City, Taipei County, Taiwan R.O.C. Rc: K123075 Trade/Device Name: Well-Life Buttock Muscle Stimulator. model W1.-2413E Regulation Number: 21 CFR 890.5850 Regulation Name: Powered muscle stimulator Regulatory Class: Class II Product Code: NGX Dated: June 2. 2013 Received: June 5, 2013 Dear Ms. Hsieh: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH docs not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (sec above) into cither class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 {5}------------------------------------------------ Page 2 - Ms. Jenny Hsieh CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, # Joyce M. Whang -S Victor Krauthamer, Ph.D. for Acting Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ ## Indications for Use 510(k) Number (if known): K123075 Device Name: Well-Life Buttock Muscle Stimulator, model: WL 2413E Indications For Use: The Buttock Muscle Stimulator, model WL-2413E, is intended for use by healthy persons to apply transcutaneous electrical muscle stimulation(EMS) through skin contact electrode for the purpose of improvement of muscle tone of the buttocks muscles. . Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use × (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) # Joyce M. Whang -S (Division Sign Off) Division of Neurological and Physical Medicine Devices (DNPMD) 510(k) Number _ K123075_ Page 1 of 1
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...