Body Contouring Machine can be used to: - Improvement of abdominal tone, strengthening of the abdominal muscles, development of firmer abdomen. - Strengthening, Toning and Firming of buttocks, thighs and calves. - Improvement of muscle tone and firmness, for strengthening muscles in arms.
Device Story
Non-invasive powered muscle stimulator; generates electromagnetic field to induce muscle contraction via nerve action potentials. Device features 15.6-inch touch screen; four treatment handles (models ZH-01, ZH-03); two operating modes (intelligent, professional). Used in clinical settings by trained operators. Operator selects parameters (intensity, time) via interface; device delivers magnetic pulses (1-150Hz) to target muscle groups. Output facilitates muscle conditioning, toning, and strengthening. Benefits include improved muscle firmness and tone in abdomen, buttocks, thighs, calves, and arms.
Clinical Evidence
No clinical data. Bench testing only, including electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), nerve/muscle stimulator performance (IEC 60601-2-10), biocompatibility (ISO 10993-1, -5, -10), and magnetic density testing.
Technological Characteristics
Powered muscle stimulator; electromagnetic field energy source. Class II, Type BF electrical protection. 15.6-inch touch screen interface. Four-handle configuration. Magnetic field intensity 0.1-2.5T; pulse repetition 1-150Hz; pulse duration 190-300µs. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-10, ISO 14971, ISO 10993. Software lifecycle per IEC 62304.
Indications for Use
Indicated for individuals requiring improvement of abdominal tone, strengthening of abdominal muscles, development of firmer abdomen, and strengthening, toning, and firming of buttocks, thighs, calves, and arms.
Regulatory Classification
Identification
A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters FDA in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
April 24, 2024
Hebei JT Medical Co., Ltd. % Gina Lian General Manager Beijing STYH Medical Technology Service Co. Ltd. No.18, Jianshe Road, Kaixuan Street, Liangxiang Town, Fangshan District. Beijing, Beijing 102488 China
Re: K232181
Trade/Device Name: Body Contouring Machine (Models: CS2, CS3, CS5, CS7, CS8, CS11, CS13) Regulation Number: 21 CFR 890.5850 Regulation Name: Powered Muscle Stimulator Regulatory Class: Class II Product Code: NGX Dated: August 7, 2023 Received: August 9, 2023
## Dear Gina Lian:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of
{1}------------------------------------------------
Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Tushar Bansal -S
{2}------------------------------------------------
for Heather Dean, PhD Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Rehabilitation Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K232181
Device Name
Body Contouring Machine (Models: CS2, CS3, CS5, CS7, CS8, CS11, CS13)
Indications for Use (Describe)
Body Contouring Machine can be used to:
- Improvement of abdominal tone, strengthening of the abdominal muscles, development of firmer abdomen.
- Strengthening, Toning and Firming of buttocks, thighs and calves.
- Improvement of muscle tone and firmness, for strengthening muscles in arms.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <span style="font-size:16px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|------------------------------------------------------------------------------------|
| <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510k Summary K232181
#### 510(k) Summary
This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of Title 21, CFR Section 807.92.
The assigned 510(k) Number:_ K232181
- 1. Date of Preparation
8/7/2023
- 2. Sponsor
Hebei JT Medical Co., Ltd.
C12 Hegu Industrial Park, Chaoyang Road, Development Zone ZhuoZhou city Establishment Registration Number: Not yet registered or the Number Contact Person: Karen Liu Position: Sales manager Tel: +86-13911459627 Fax:86-10-5721-2057 Email: 919309354@qq.com
- 3. Submission Correspondent
Ms. Gina Lian
#### Beijing STYH Medical Technology Service Co. Ltd.
No.18, Jianshe Road, Kaixuan Street, Liangxiang Town, Fangshan District,102488 Beijing,China. Tel: +86-13311543099 Email: 409308721@qq.com
{5}------------------------------------------------
- Identification of Subject Device 4.
Trade Name: Body Contouring Machine
Common Name: Stimulator, muscle, powered, for muscle conditioning Model(s): CS2,CS3,CS5,CS7,CS8,CS11,CS13
Regulatory Information: Classification Name: Stimulator, muscle, powered, for muscle conditioning Classification: II; Product Code: NGX: Regulation Number: 21 CFR 890.5850 ; Review Panel: Physical Medicine ;
Indication for Use: Body Contouring Machine can be used to:
- ✔ Improvement of abdominal tone, strengthening of the abdominal muscles, development of firmer abdomen.
- イ Strengthening, Toning and Firming of buttocks, thighs and calves.
- V Improvement of muscle tone and firmness, for strengthening muscles in arms.
- 5. Device Description
Body Contouring Machine is a non-invasive treatment device. The device generates an electromagnetic field that interacts with human tissue. Body Contouring Machine is equipped with four handles, two each for ZH-01 and ZH-03, which can treat different parts of the same time. The device is equipped with a 15.6-inch true-color touch screen and provides two operating modes, intelligent and professional, to meet the different needs of users.
The Body Contouring Machine includes seven models in this submission, CS2,CS3,CS5,CS7, CS8,CS11, and CS13, all seven models have same principle, software, operation etc., only differences are appearance
The subject device includes the following components:
| Components | Function Description | Applied Model(s) |
|--------------|--------------------------------------------------------------------------------------------------------------------------------|-------------------------------|
| 1.Handle | The handle is the working component of the treatment. Perform treatment according to the treatment parameters set by the host. | CS2、CS3、CS5、CS7、CS8、CS11、CS13 |
| 2.LCD screen | The device is equipped with a 15.6-inch true color touch screen. | CS2、CS3、CS5、CS7、CS8、CS11、CS13 |
| 3.Emergency | In an emergency, press this button to disconnect the | CS2、CS3、CS5、CS7、 |
Table 1 Main Components of Subject Device
{6}------------------------------------------------
| stop switch | control system and power supply, rotate it to the right and eject the button to return to normal. | CS8、CS11、CS13 |
|----------------------|----------------------------------------------------------------------------------------------------|-------------------------------|
| 4.Casters | Push the equipment horizontally to the designated position | CS2、CS3、CS5、CS7、CS8、CS11、CS13 |
| 5.Push button switch | press this button under normal circumstances to open and close the control system and power supply | CS2、CS3、CS5、CS7、CS8、CS11、CS13 |
| 6.Handle bracket | fixed treatment handle | CS2、CS3、CS5、CS7、CS8、CS11、CS13 |
- 6. Identification of Predicate Device
Predicate device 510(k) number: K190456 Trading name: BTL 799-2L Common name: Stimulator, muscle, powered, for muscle conditioning Regulation No.: 21 CFR 890.5850 Classification name: Stimulator, muscle, powered, for muscle conditioning Classification: Class II Product code: NGX
- 7. Non-Clinical Test Conclusion
Non-clinical tests were conducted to verify that the subject device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the subject device complies with the following standards:
- A IEC 60601-1:2012, Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance:
- IEC 60601-1-2:2014+A1:2020, Medical electrical equipment- Part 1-2: General A requirements for basic safety and essential performance- Collateral standard: Electromagnetic compatibility- Requirements and tests.
- A IEC 60601-2-10:2012/AMD2:2023 Medical electrical equipment Part 2:Particular requirements for the basic safety and essential performance of nerve and muscle stimulators.
- > IEC62304:2015, Medical device software – Software life cycle processes
- ISO 14971:2019. Medical devices Application of risk management to medical devices A
- A ISO 10993-1:2018, Biological evaluation of medical devices - Part 1: Evaluation and
{7}------------------------------------------------
testing within a risk management process
- ISO 10993-5:2009, Biological Evaluation of Medical Device, Part 5-Tests for Vitro > cytotoxicity
- ISO 10993-10:2010, Biological Evaluation of Medical Device, Part 10-Test for irritation > and skin sensitization
- > Magnetic density test.
- Clinical Test Conclusion 8.
-
No clinical study is included in this submission.
{8}------------------------------------------------
## 9. Substantially Equivalent (SE) Comparison
## Table 2 General Comparison
| ITEM | Subject Device | Predicate Device |
|-----------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Code | NGX - Stimulator, Muscle, Powered,<br>Muscle Conditioning | NGX - Stimulator, Muscle, Powered,<br>Muscle Conditioning |
| Regulation<br>Number | 21 CFR 890.5850 | 21 CFR 890.5850 |
| Regulatory<br>Class | Class II | Class II |
| Clinical Use | Prescription use | Prescription use |
| Indication<br>for Use | Body Contouring Machine can be used to:<br>● Improvement of abdominal tone,<br>strengthening of the abdominal<br>muscles, development of firmer<br>abdomen.<br>● Strengthening, Toning and Firming<br>of buttocks, thighs and calves.<br>● Improvement of muscle tone and<br>firmness, for strengthening muscles<br>in arms. | BTL 799-2L is indicated to be used<br>for:<br>● Improvement of abdominal tone,<br>strengthening of the abdominal<br>muscles, development of firmer<br>abdomen.<br>● Strengthening, Toning and<br>Firming of buttocks, thighs and<br>calves.<br>● Improvement of muscle tone and<br>firmness, for strengthening<br>muscles in arms. |
| ITEM | Subject Device | Predicate Device |
| Principle of<br>Action | Initiating action potential of nerves results<br>in muscle contraction | Initiating action potential of nerves<br>results in muscle contraction. |
| Electrical<br>Protection | Class II, BF | Class II, BF |
| User Interface | Touch screen | Touch screen |
| Firmware<br>Controlled | Yes | Yes |
| Type of Energy | Magnetic field | Magnetic field |
| Number of<br>outputs | 4 | 2 |
| Number of<br>Magnetic Coils<br>in the<br>Applicator | 1 | 1 |
| Magnetic Field<br>Intensity | ZH-01 applicator: 0.1-1.8T±20% | BTL 299-6 applicator: 0.5 - 1.8 T<br>±20% |
| | ZH-03 applicator: 0.1-2.5T±20% | BTL 299-7 applicator: 0.7 - 2.0 T<br>±20% |
| Pulse<br>Repetition<br>Rate | 1-150Hz | 1 - 150 Hz |
| Pulse Duration | $300\pm20 \mu s$<br><br>$190\pm20 \mu s$ | BTL 299-6 applicator: $280 \pm 20\% \mu s$<br><br>BTL 299-7 applicator: $190 \pm 20\% \mu s$ |
| Selection of<br>parameters<br>(Intensity,<br>Time) | Yes | Yes |
| Therapy Time | Up to 60 min | Up to 60 min |
| Energy Source | 220/110V ,50 - 60 Hz | 100 - 240 V AC, 50 - 60 Hz |
| System<br>Dimensions<br>(W×H×D) | 450*530*1220mm | 580×1380×580 mm<br>(23×55×23 in) |
| Environmental<br>Specifications | For indoor use only | For indoor use only |
| Patient<br>Contacted Part | Skin | Skin |
-
{9}------------------------------------------------
#### Table 4 Safety Comparison
| Item | Subject Device | Predicate Device |
|-------------------|----------------------------|-----------------------------------------|
| Electrical Safety | Comply with IEC 60601-1 | Comply with IEC 60601-1, IEC 60601-2-10 |
| | Comply with IEC 60601-2-10 | Comply with IEC 60601-2-10 |
| EMC | Comply with IEC 60601-1-2 | Comply with IEC 60601-1-2 |
#### Analysis:
These difference between the predicate devices and subject device have no significant influence on safety or effectiveness of the Body Contouring Machine.
The subject device is substantially equivalent to the predicate device. Based on the nonclinical tests performed, the subject device is as safe, as effective, and performs as well as the legally marketed predicate device.
## 10. Substantially Equivalent (SE) Conclusion
Based on the comparison and analysis above, the subject device is determined to be Substantially Equivalent (SE) to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.