K191146 · Fx Shoulder USA, Inc. · PHX · Jul 15, 2020 · Orthopedic
Device Facts
Record ID
K191146
Device Name
Humelock TiN Coated Glenosphere
Applicant
Fx Shoulder USA, Inc.
Product Code
PHX · Orthopedic
Decision Date
Jul 15, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3660
Device Class
Class 2
Attributes
Therapeutic
Intended Use
When used in the Humelock II Reversible Shoulder System: The Humelock II Reversible Shoulder is indicated for primary, fracture or revision total shoulder arthroplasty for the relief of pain and to improve function in patients with a massive and non-repairable rotator cuff tear. The patient's joint must be anatomically and structurally suited to receive the selected implants and a functional deltoid muscle is necessary to use the device. The humeral stems are intended for cementless use. The metaglene baseplate is intended for cementless use with the addition of screws for fixation. When used in the Humelock Reversed Shoulder System: The Humelock Reversed Shoulder System is indicated for primary, fracture or revision total shoulder arthroplasty for the relief of pain and to improve function in patients with a massive and non-repairable cuff tear. The patient's joint must be anatomically and structurally suited to receive the selected implants and a functional deltoid muscle is necessary to use the device. During primary or revision surgery, if the glenoid bone stock appears to be insufficient to bear the reversed glenoid components or the glenoid bone fractures during the procedure, a taper adapter can be used to convert the Humelock Reversed Shoulder to an anatomic hemishoulder prosthesis. The humeral stem of the Humelock Reversed Cemented Shoulder Prothesis is intended for cemented use only. The humeral stem of the Humelock Reversed Cementless Shoulder Prothesis is lockable with two cortical bone screws and is intended for cementless use only. An optional anti-rotational spoiler can be used with either the cementless or the cemented stems. The glenoid baseplate and post extension are intended for cementless use with the addition of screws for fixation. When used in the Humeris Shoulder System: In an anatomic shoulder configuration, the Humeris Shoulder System is indicated for use in total and hemi-shoulder replacement to treat: - A severely painful and/or disabled joint resulting from osteoarthritis or rheumatoid arthritis; - Other difficult clinical problems where shoulder arthrodesis or resection arthroplasty are not acceptable (e.q. revision of a previously implanted primary component, a humeral plate or a humeral nail). In a reverse shoulder configuration, the Humeris Shoulder is indicated for primary or revision total shoulder arthroplasty for the relief of pain and to improve function in patients with a massive and non-repairable rotator cuff tear. The patient's joint must be anatomically and structurally suited to receive the selected implants and a functional deltoid muscle is necessary to use the device. The humeral stem of the Humeris Cementless Shoulder is intended for cementless use only. The humeral stem of the Humeris Cemented Shoulder is intended for cemented use only. The glenoid components of the Humeris Shoulder System are intended for cemented use only. The glenoid baseplate component is intended for cementless use with the addition of screws for fixation.
Device Story
Humelock TiN Coated Glenosphere is a component for reverse total shoulder replacement systems (Humelock II, Humelock Reversed, Humeris). Device consists of a cobalt chromium molybdenum (CoCr) glenosphere with a titanium nitride (TiN) surface coating. Implanted by orthopedic surgeons during shoulder arthroplasty to replace the glenoid surface. TiN coating provides surface modification to the existing CoCr substrate. Device integrates with previously cleared humeral stems and baseplates. Benefits include potential wear reduction and biocompatibility consistent with reference devices. Used in clinical settings for patients requiring shoulder joint reconstruction.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including biocompatibility (ISO 10993-1), wear testing under worst-case loading, surface roughness analysis, and metal/UHMWPE particle analysis.
Technological Characteristics
Glenosphere component made of CoCr with TiN coating. Designed for reverse shoulder arthroplasty. Compatible with existing Humelock/Humeris system components. Biocompatibility per ISO 10993-1. Sterilization and shelf-life specifications identical to predicate.
Indications for Use
Indicated for primary, fracture, or revision total shoulder arthroplasty for pain relief and functional improvement in patients with massive, non-repairable rotator cuff tears; also indicated for osteoarthritis, rheumatoid arthritis, or revision of prior components in anatomic configurations. Requires functional deltoid muscle and anatomically/structurally suitable joint.
Regulatory Classification
Identification
A shoulder joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a shoulder joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a humeral resurfacing component made of alloys, such as cobalt-chromium-molybdenum, and a glenoid resurfacing component made of ultra-high molecular weight polyethylene, and is limited to those prostheses intended for use with bone cement (§ 888.3027).
Special Controls
*Classification.* Class II. The special controls for this device are:(1) FDA's:
(i) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ”
(ii) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),”
(iii) “Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement,”
(iv) “Guidance Document for the Preparation of Premarket Notification (510(k)) Application for Orthopedic Devices,” and
(v) “Guidance Document for Testing Non-articulating, ‘Mechanically Locked’ Modular Implant Components,”
(2) International Organization for Standardization's (ISO):
(i) ISO 5832-3:1996 “Implants for Surgery—Metallic Materials—Part 3: Wrought Titanium 6-aluminum 4-vandium Alloy,”
(ii) ISO 5832-4:1996 “Implants for Surgery—Metallic Materials—Part 4: Cobalt-chromium-molybdenum casting alloy,”
(iii) ISO 5832-12:1996 “Implants for Surgery—Metallic Materials—Part 12: Wrought Cobalt-chromium-molybdenum alloy,”
(iv) ISO 5833:1992 “Implants for Surgery—Acrylic Resin Cements,”
(v) ISO 5834-2:1998 “Implants for Surgery—Ultra-high Molecular Weight Polyethylene—Part 2: Moulded Forms,”
(vi) ISO 6018:1987 “Orthopaedic Implants—General Requirements for Marking, Packaging, and Labeling,” and
(vii) ISO 9001:1994 “Quality Systems—Model for Quality Assurance in Design/Development, Production, Installation, and Servicing,” and
(3) American Society for Testing and Materials':
(i) F 75-92 “Specification for Cast Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implant Material,”
(ii) F 648-98 “Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants,”
(iii) F 799-96 “Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Forgings for Surgical Implants,”
(iv) F 1044-95 “Test Method for Shear Testing of Porous Metal Coatings,”
(v) F 1108-97 “Specification for Titanium-6 Aluminum-4 Vanadium Alloy Castings for Surgical Implants,”
(vi) F 1147-95 “Test Method for Tension Testing of Porous Metal,”
(vii) F 1378-97 “Standard Specification for Shoulder Prosthesis,” and
(viii) F 1537-94 “Specification for Wrought Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implants.”
K191698 — TiN Coated Humeral Head · Fx Shoulder USA, Inc. · Jul 16, 2020
K213124 — Tomier Perform Reversed Glenoid Cannulated Glenospheres · Tornier, Inc. · Oct 26, 2021
K252567 — AltiVate Reverse® ADLC Glenosphere · Encore Medical L.P. · Jan 15, 2026
K203755 — Shoulder System - TiNbN Coating · Medacta International S.A. · Mar 2, 2021
K193122 — Biocore9 Humeral Resurfacing System · Biocore9, LLC · May 15, 2020
Submission Summary (Full Text)
{0}------------------------------------------------
July 22, 2020
Image /page/0/Picture/1 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
FX Shoulder USA, Inc. Kathy Trier VP Regulatory, Quality, Clinical, Compliance 13465 Midway Road, Suite 101 DALLAS, TX 75244
Re: K191146
Trade/Device Name: Humelock TiN Coated Glenosphere Regulation Number: 21 CFR 888.3660 Regulation Name: Shoulder joint metal/polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: PHX, KWT, HSD
Dear Kathy Trier:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated July 15, 2020. Specifically, FDA is updating this SE Letter as an administrative correction.
Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Michael Owens, OHT6: Office of Orthopedic Devices, 301-796-5650, Michaelc.Owens@fda.hhs.gov.
Sincerely.
Image /page/0/Picture/9 description: The image shows the name "Michael C. Owens -S" on the left side of the image. On the right side of the image, it says "Digitally signed by Michael C. Owens -S" followed by the date "2020.07.22" and the time "14:24:54 -04'00'".
Michael Owens Acting Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
FX Shoulder USA, Inc. Kathy Trier VP Regulatory, Quality, Clinical, Compliance 13465 Midway Road, Suite 101 DALLAS, TX 75244
Re: K191146
Trade/Device Name: Humelock TiN Coated Glenosphere Regulation Number: 21 CFR 888.3660 Regulation Name: Shoulder joint metal/polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: PHX Dated: June 29, 2020 Received: July 2, 2020
Dear Kathy Trier:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
July 15, 2020
{2}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-
542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Michael C. Owens -S
Digitally signed by Michael C. Owens -S Date: 2020.07.15 11:13:55 -04'00'
Michael Owens Acting Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K191146
Device Name Humelock TiN Coated Glenosphere
#### Indications for Use (Describe)
When used in the Humelock II Reversible Shoulder System:
The Humelock II Reversible Shoulder is indicated for primary, fracture or revision total shoulder arthroplasty for the relief of pain and to improve function in patients with a massive and non-repairable rotator cuff tear.
The patient's joint must be anatomically and structurally suited to receive the selected implants and a functional deltoid muscle is necessary to use the device.
The humeral stems are intended for cementless use. The metaglene baseplate is intended for cementless use with the addition of screws for fixation.
When used in the Humelock Reversed Shoulder System:
The Humelock Reversed Shoulder System is indicated for primary, fracture or revision total shoulder arthroplasty for the relief of pain and to improve function in patients with a massive and non-repairable cuff tear.
The patient's joint must be anatomically suited to receive the selected implants and a functional deltoid muscle is necessary to use the device.
During primary or revision surgery, if the glenoid bone stock appears to bear the reversed glenoid components or the glenoid bone fractures during the procedure, a taper adapter can be used to convert the Humelock Reversed Shoulder to an anatomic hemi-shoulder prosthesis.
The humeral stem of the Humelock Reversed Cemented Shoulder Prothesis is intended for cemented use only. The humeral stem of the Humelock Reversed Cementless Shoulder Prothesis is lockable with two cortical bone screws and is intended for cementless use only. An optional anti-rotational spoiler can be used with either the cementless or the cemented stems.
The glenoid baseplate and post extension are intended for cementless use with the addition of screws for fixation.
When used in the Humeris Shoulder System:
In an anatomic shoulder configuration, the Humeris Shoulder System is indicated for use in total and hemi-shoulder replacement to treat:
- A severely painful and/or disabled joint resulting from osteoarthritis or rheumatoid arthritis;
- Other difficult clinical problems where shoulder arthrodesis or resection arthroplasty are not acceptable (e.g. revision of a previously implanted primary component, a humeral plate or a humeral nail).
In a reverse shoulder configuration, the Humeris Shoulder is indicated for primary or revision total shoulder athroplasty for the relief of pain and to improve function in patients with a massive and non-repairable rotator cuff tear.
The patient's joint must be anatomically suited to receive the selected implants and a functional deltoid muscle is necessary to use the device.
{4}------------------------------------------------
The humeral stem of the Humeris Cementless Shoulder is intended for cementless use only. The humeral stem of the Humeris Cemented Shoulder is intended for cemented use only. The Humeris Shoulder System are intended for cemented use only. The glenoid baseplate component is intended for cementless use with the addition of screws for fixation.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
#### 7. 510(k) Summary
| Applicant/Sponsor: | FX Shoulder USA, Inc.<br>13465 Midway Road, Suite 101<br>Dallas, Texas 75244<br>Establishment Registration No: 3014128390 |
|-----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer: | FX Solutions<br>1663 Rue de Majornas<br>Viriat, France 01440<br>Establishment Registration No: 3009532798 |
| Contact Person: | Kathy Trier, Ph.D.<br>VP Regulatory, Clinical, Quality, Compliance<br>ktrier@fxshoulder.com<br>800.280.0775 |
| Date: | July 13, 2020 |
| Proprietary Name: | Humelock TiN Coated Glenosphere |
| Common Name: | Shoulder Prosthesis, Reverse Configuration |
| Product Code(s): | PHX, KWT, HSD |
| Classification Name: | 21 CFR 888.3660: shoulder joint metal/polymer<br>semi-constrained cemented prosthesis – Class II |
| Substantially Equivalent Devices: | Primary Predicate:<br>Humelock II Reversible Shoulder System (K150488)<br>Predicates:<br>Humelock Reversed Shoulder System (K162455)<br>Humeris Shoulder System (K163669)<br>Reference Device:<br>LINK® Endo-Model® Knee System with PorEx®; LINK®<br>Sled Knee System with PorEx® (K152431) |
# Device Description
The Humelock TiN Coated Glenosphere is a new component for Humelock II, Humelock Reversed, and Humeris reversed total shoulder replacement systems. The Humelock TiN Coated Glenosphere has a titanium nitride (TiN) coating, which is applied to the predicate glenosphere made of cobalt chromium molybdenum (CoCr). Compatible components for use
{6}------------------------------------------------
with the Humelock TiN Coated Glenosphere to complete the reversed total shoulder replacement construct are the same as those previously cleared compatible components for use with the CoCr glenospheres in the primary predicate device, K150488 and are also components in K162455 for the Humelock Reversed Shoulder System and K163669 for the Humeris Shoulder, when used for a reverse shoulder construct.
#### Intended Use / Indications
When used in the Humelock II Reversible Shoulder System: The Humelock II Reversible Shoulder is indicated for primary, fracture or revision total shoulder arthroplasty for the relief of pain and to improve function in patients with a massive and non-repairable rotator cuff tear.
The patient's joint must be anatomically and structurally suited to receive the selected implants and a functional deltoid muscle is necessary to use the device.
The humeral stems are intended for cementless use. The metaglene baseplate is intended for cementless use with the addition of screws for fixation.
When used in the Humelock Reversed Shoulder System:
The Humelock Reversed Shoulder System is indicated for primary, fracture or revision total shoulder arthroplasty for the relief of pain and to improve function in patients with a massive and non-repairable cuff tear.
The patient's joint must be anatomically and structurally suited to receive the selected implants and a functional deltoid muscle is necessary to use the device.
During primary or revision surgery, if the glenoid bone stock appears to be insufficient to bear the reversed glenoid components or the glenoid bone fractures during the procedure, a taper adapter can be used to convert the Humelock Reversed Shoulder to an anatomic hemishoulder prosthesis.
The humeral stem of the Humelock Reversed Cemented Shoulder Prothesis is intended for cemented use only. The humeral stem of the Humelock Reversed Cementless Shoulder Prothesis is lockable with two cortical bone screws and is intended for cementless use only. An optional anti-rotational spoiler can be used with either the cementless or the cemented stems.
The glenoid baseplate and post extension are intended for cementless use with the addition of screws for fixation.
When used in the Humeris Shoulder System:In an anatomic shoulder configuration, the Humeris Shoulder System is indicated for use in total and hemi-shoulder replacement to treat:
- A severely painful and/or disabled joint resulting from osteoarthritis or rheumatoid arthritis; - Other difficult clinical problems where shoulder arthrodesis or resection arthroplasty are not acceptable (e.q. revision of a previously implanted primary component, a humeral plate or a humeral nail).
In a reverse shoulder configuration, the Humeris Shoulder is indicated for primary or revision total shoulder arthroplasty for the relief of pain and to improve function in patients with a massive and non-repairable rotator cuff tear.
{7}------------------------------------------------
The patient's joint must be anatomically and structurally suited to receive the selected implants and a functional deltoid muscle is necessary to use the device.
The humeral stem of the Humeris Cementless Shoulder is intended for cementless use only. The humeral stem of the Humeris Cemented Shoulder is intended for cemented use only. The glenoid components of the Humeris Shoulder System are intended for cemented use only. The glenoid baseplate component is intended for cementless use with the addition of screws for fixation.
# Summary of Technologies / Substantial Equivalence
The Humelock TiN Coated Glenosphere is substantially equivalent to the primary predicate in that it is identical to the primary predicate on indications, design, dimensions, packaging, single use, sterilization, shelf life, pyrogen testing, biocompatibility, compatible components, instrumentation and surgical technique. It is identical to the primary predicate material CoCr. The change, subject of this 510(k), is to add the titanium nitride (TiN) coating to the surface of the glenosphere, which is substantially equivalent to the technology of coating on the reference device based upon the characterization and testing in accordance with the same scientific methods and standards. Differences between the subject device and the predicate devices do not raise new questions of safety and effectiveness.
## Non-Clinical Testing
Biocompatibility testing has been completed in accordance with FDA Guidance titled, "Use of International Standard ISO 10993-1, 'Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk manaqement process" (September 2016).Non-clinical biocompatibility testing is the same as the reference device and includes Acute System Toxicity Study: Cytotoxicity Study: GC/MS Fingerprint Study: Irritation Study: Sensitization Study. Characterization of the TiN coating demonstrates substantial equivalence of the subject device to the coating of the reference device - Bone Implantation Study, 28 Day Muscle Implantation Study: 90Day Muscle Implantation Study are referenced for the subject device. Wear testing under worst case loading and worst case environment with analysis of surface roughness and both UHMWPE and metal particle analysis have been completed.
Mechanical testing for the complete system was previously submitted with the cleared predicate device.
# Clinical Testing
Clinical testing was not necessary to determine substantial equivalence of the Humelock TiN Coated Glenosphere to the predicate devices.
#### Conclusion
The subject device is identical to the primary predicate with the only modification of added TiN coating. The evidence reviewed to demonstrate substantial equivalence includes risk analysis, design controls, biocompatibility, wear properties of the TIN coating, and verification and validation activities to demonstrate that the TiN coating does not increase risk and does not raise new questions of safety and effectiveness of the Humelock TiN Coated Glenosphere.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.