K162455 · Fx Solutions · PHX · Jan 17, 2017 · Orthopedic
Device Facts
Record ID
K162455
Device Name
Humelock Reversed Shoulder
Applicant
Fx Solutions
Product Code
PHX · Orthopedic
Decision Date
Jan 17, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3660
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Humelock Reversed Shoulder is indicated for primary, fracture or revision total shoulder arthroplasty for the relief of pain and to improve function in patients with a massive and non-repairable rotator cuff tear. The patient's joint must be anatomically and structurally suited to receive the selected implants and a functional deltoid muscle is necessary to use the device. During primary or revision surgery, if the glenoid bone stock appears to be insufficient to bear the reversed glenoid components or the glenoid bone fractures during the procedure, a taper adapter can be used to convert the Humelock Reversed Shoulder to an anatomic hemishoulder prosthesis. The humeral stem of the Humelock Reversed Cemented Shoulder Prosthesis is intended for cemented use only. The humeral stem of the Humelock Reversed Cementless Shoulder Prosthesis is lockable with two cortical bone screws and is intended for cementless use only. An optional anti-rotation spoiler can be used with either the cementless or the cemented stems. The glenoid baseplate and post extension are intended for cementless use with the addition of screws for fixation.
Device Story
Total shoulder prosthesis; inverted articulation design (ball on glenoid, cup on humerus). Components: glenoid baseplate, glenospheres (centered/eccentric), humeral cups (Ti-6Al-4V shell/UHMWPE insert), humeral stems (cemented/cementless), bone screws, optional anti-rotation spoiler, taper adapter for hemi-shoulder conversion. Used by orthopedic surgeons in clinical/OR settings for shoulder arthroplasty. Provides mechanical joint replacement to restore function and relieve pain in patients with non-functional rotator cuffs. Surgeon selects components based on patient anatomy; glenoid baseplate fixed with screws; humeral stem implanted via cement or cementless fixation; glenosphere and humeral cup articulate to replace joint function.
Clinical Evidence
No clinical data. Substantial equivalence determined via non-clinical bench testing, including range of motion analysis, fatigue testing of stem/spacer/cup constructs, post-fatigue disassembly testing, torsional fatigue testing of the spoiler, and pyrogenicity testing (limit <20 EU/device).
Technological Characteristics
Materials: Ti-6Al-4V (ISO 5832-3), Co-Cr-Mo (ISO 5832-12), UHMWPE (ISO 5834-1/2), CP Titanium/Hydroxyapatite coating. Design: Reverse shoulder geometry; modular components (baseplate, glenosphere, humeral cup, stem). Fixation: Cemented or cementless (screws/coating). Connectivity: None. Sterilization: Not specified.
Indications for Use
Indicated for primary, fracture, or revision total shoulder arthroplasty in patients with massive, non-repairable rotator cuff tears requiring pain relief and functional improvement. Requires anatomically/structurally suitable joint and functional deltoid muscle. Contraindicated if glenoid bone stock is insufficient or fractures during procedure (unless converted to hemi-shoulder).
Regulatory Classification
Identification
A shoulder joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a shoulder joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a humeral resurfacing component made of alloys, such as cobalt-chromium-molybdenum, and a glenoid resurfacing component made of ultra-high molecular weight polyethylene, and is limited to those prostheses intended for use with bone cement (§ 888.3027).
Special Controls
*Classification.* Class II. The special controls for this device are:(1) FDA's:
(i) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ”
(ii) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),”
(iii) “Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement,”
(iv) “Guidance Document for the Preparation of Premarket Notification (510(k)) Application for Orthopedic Devices,” and
(v) “Guidance Document for Testing Non-articulating, ‘Mechanically Locked’ Modular Implant Components,”
(2) International Organization for Standardization's (ISO):
(i) ISO 5832-3:1996 “Implants for Surgery—Metallic Materials—Part 3: Wrought Titanium 6-aluminum 4-vandium Alloy,”
(ii) ISO 5832-4:1996 “Implants for Surgery—Metallic Materials—Part 4: Cobalt-chromium-molybdenum casting alloy,”
(iii) ISO 5832-12:1996 “Implants for Surgery—Metallic Materials—Part 12: Wrought Cobalt-chromium-molybdenum alloy,”
(iv) ISO 5833:1992 “Implants for Surgery—Acrylic Resin Cements,”
(v) ISO 5834-2:1998 “Implants for Surgery—Ultra-high Molecular Weight Polyethylene—Part 2: Moulded Forms,”
(vi) ISO 6018:1987 “Orthopaedic Implants—General Requirements for Marking, Packaging, and Labeling,” and
(vii) ISO 9001:1994 “Quality Systems—Model for Quality Assurance in Design/Development, Production, Installation, and Servicing,” and
(3) American Society for Testing and Materials':
(i) F 75-92 “Specification for Cast Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implant Material,”
(ii) F 648-98 “Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants,”
(iii) F 799-96 “Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Forgings for Surgical Implants,”
(iv) F 1044-95 “Test Method for Shear Testing of Porous Metal Coatings,”
(v) F 1108-97 “Specification for Titanium-6 Aluminum-4 Vanadium Alloy Castings for Surgical Implants,”
(vi) F 1147-95 “Test Method for Tension Testing of Porous Metal,”
(vii) F 1378-97 “Standard Specification for Shoulder Prosthesis,” and
(viii) F 1537-94 “Specification for Wrought Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implants.”
Predicate Devices
Fx Solutions Humelock II Reversible Shoulder (K150488)
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 17, 2017
Fx Solutions Cheryl Hastings Official Correspondent 1663 Rue De Majornas Viriat. 01440 FR
Re: K162455
Trade/Device Name: Humelock Reversed Shoulder Regulation Number: 21 CFR 888.3660 Regulation Name: Shoulder joint metal/polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: PHX, HSD Dated: December 15, 2016 Received: December 16, 2016
Dear Cheryl Hastings:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{2}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known): K162455
Device Name: Humelock Reversed Shoulder
### Indications for Use:
The Humelock Reversed Shoulder is indicated for primary, fracture or revision total shoulder arthroplasty for the relief of pain and to improve function in patients with a massive and non-repairable rotator cuff tear.
The patient's joint must be anatomically and structurally suited to receive the selected implants and a functional deltoid muscle is necessary to use the device.
During primary or revision surgery, if the glenoid bone stock appears to be insufficient to bear the reversed glenoid components or the glenoid bone fractures during the procedure, a taper adapter can be used to convert the Humelock Reversed Shoulder to an anatomic hemishoulder prosthesis.
The humeral stem of the Humelock Reversed Cemented Shoulder Prosthesis is intended for cemented use only. The humeral stem of the Humelock Reversed Cementless Shoulder Prosthesis is lockable with two cortical bone screws and is intended for cementless use only. An optional anti-rotation spoiler can be used with either the cementless or the cemented stems.
The glenoid baseplate and post extension are intended for cementless use with the addition of screws for fixation.
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
{4}------------------------------------------------
| 510(k) Summary | |
|--------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Prepared: | December 15, 2016 |
| Submitter: | Fx Solutions<br>1663 Rue de Majornas<br>Viriat, France 01440 |
| Contact: | Jean-Jacques Martin<br>+33 4 74 55 35 55<br>www.fxsolutions.fr |
| Proprietary Name: | Humelock Reversed Shoulder |
| Common Name: | Reverse Shoulder Prosthesis |
| Classification Names: | 21 CFR 888.3660: Shoulder joint metal/polymer semi-constrained<br>cemented prosthesis; Class II<br><br>21 CFR 888.3680 Shoulder joint glenoid (hemi-shoulder) metallic<br>cemented prosthesis. |
| Product Codes: | 87/PHX, 87/HSD |
| Substantially<br>Equivalent Devices: | Fx Solutions Humelock II Reversible Shoulder, K150488<br><br>Tornier Aequalis Reversed Shoulder Prosthesis,<br>K100142, K082120, K030941 |
#### Device Description:
The Humelock Reversed Shoulder is a total shoulder prosthesis designed for use in patients with a non-functional rotator cuff. The articulation of this design is inverted compared to a traditional total shoulder prosthesis. The reverse shoulder is designed so that the ball of the articulation is on the glenoid side and the mating cup fits into the humeral stem. The components of the system include a glenoid baseplate, standard and locking bone screws, optional baseplate post extensions, centered and eccentric glenospheres with and without central screws, humeral cups, cementless and cemented humeral stems, optional cortical bone screws, an optional humeral spacer, an optional anti-rotation spoiler and an optional taper adapter for use in hemi-shoulder replacement.
The Humelock glenoid baseplate has a round base with a central, cannulated post and four peripheral, threaded screw holes. The outer edges of the baseplate are tapered to lock with the glenosphere component.
{5}------------------------------------------------
The glenoid baseplate is used with 4.5mm standard or locking bone screws for added stability. The bone screws are available in lengths from 20 - 50mm in 5mm increments.
Optional post extensions are available to extend the central post of the baseplate and provide additional anchoring in cases with poor bone quality. The post extensions are available in 6mm and 10mm lengths. When used, the post extensions screw into the baseplate post.
The glenoid baseplate, standard and locking screws, and post extensions are manufactured from Ti-6Al-4V alloy conforming to ISO 5832-3. The backside of the baseplate and the post extensions are coated with a plasma sprayed CP Titanium and Hydroxyapatite coating.
The Humelock Reversed Glenosphere is available in 36 and 40 mm diameter sizes in centered and eccentric styles. The eccentric glenospheres are designed to be offset from the center of the glenoid baseplate. All glenospheres have a 10° tilt. Although not physically tilted, the curvature of the glenosphere extends 10° beyond the equator of a hemisphere. This additional articular surface lateralizes the center of rotation to help reduce the potential for scapular notching by the humeral cup. All glenospheres mate with the glenoid baseplate via a taper lock; the glenosphere incorporates a female taper while the edges of the baseplate form a male taper. The glenospheres are also available with an optional central, cannulated screw. This screw can be threaded through the central post of the baseplate for additional security.
The glenospheres are manufactured from Co-Cr-Mo conforming to ISO 5832-12. The glenosphere screw is manufactured from Ti-6Al-4V conforming to ISO 5832-3.
The humeral cups are one-piece constructs consisting of a pre-assembled Ti-6Al-4V alloy shell and a UHMWPE insert. A 24mm diameter tapered post on the inferior surface of the shell locks into the female taper on the superior surface of the humeral stem. The humeral cups are available in 36mm and 40mm diameters and in standard and mobility styles. The standard cups offer a slightly deeper articular surface to provide additional constraint while the mobility cups are not as deep to provide slightly less constraint. The humeral cups are available in heights of +3mm, +6mm.
If additional height of the humeral articulation is needed, a +9mm humeral spacer can be used between the humeral stem and the humeral cup. The +9mm humeral spacer adds 9mm of height resulting in construct heights of +12mm, +15mm. The spacer has a 24mm male taper that mates with the humeral stem and a 24mm female taper that mates with the humeral cup.
{6}------------------------------------------------
The humeral cups are manufactured from Ti-6Al-4V alloy conforming to ISO 5832-3 and UHMWPE conforming to ISO 5834-1 and ISO 5834-2. The +9mm humeral spacers are manufactured from Ti-6Al-4V alloy conforming to ISO 5832-3.
The Humelock Reversed Shoulder includes both cemented humeral stems. The cementless humeral stems are available in diameters 8 to 16 mm. The distal end is cylindrical with a grit blasted surface and two unthreaded screw holes oriented in the medial/lateral direction. Bone screws can be used to provide additional early fixation and stability of the stem. Cortical bone screws are available in lengths from 18 - 40mm in 2mm increments. The proximal portion of the cementless humeral stem has a plasma sprayed CP Titanium and Hydroxyapatite coating.
Cemented stems are available in diameters 6 to 14 mm. The distal end of the cemented humeral stem is trapezoidal with a polished surface. The cemented humeral stems incorporate a 24mm diameter female taper for attachment of compatible components.
Both the cementless and cemented stems are compatible with an optional spoiler, which can be attached to the lateral side of the stem to provided additional resistance to rotation. The spoiler is fixed to either stem using an M6 hex screw.
The spoiler and hex screw are manufactured from Ti-6Al-4V alloy conforming to ISO 5832-3.
During primary or revision surgery, if the glenoid bone stock appears to be insufficient to bear the reversed glenoid components or the glenoid bone fractures during the procedure, the Humelock Reversed cementless and cemented stems can be used with an eccentric taper adapter and 50, 52 or 54mm eccentric humeral heads for conversion to an anatomic shoulder hemi-arthroplasty. The taper adapter has a 24mm male taper to mate with the humeral stem and a 10mm male taper to mate with the humeral head. The taper adapter is manufactured from Ti-6Al-4V alloy conforming to ISO 5832-3. The eccentric humeral heads are manufactured from wrought Co-Cr-Mo conforming to ISO 5832-12.
## Intended Use / Indications:
The Humelock Reversed Shoulder is indicated for primary, fracture or revision total shoulder arthroplasty for the relief of pain and to improve function in patients with a massive and nonrepairable rotator cuff tear.
{7}------------------------------------------------
The patient's joint must be anatomically and structurally suited to receive the selected implants and a functional deltoid muscle is necessary to use the device.
During primary or revision surgery, if the glenoid bone stock appears to be insufficient to bear the reversed glenoid components or the glenoid bone fractures during the procedure, a taper adapter can be used to convert the Humelock Reversed Shoulder to an anatomic hemi-shoulder prosthesis.
The humeral stem of the Humelock Reversed Cemented Shoulder Prosthesis is intended for cemented use only. The humeral stem of the Humelock Reversed Cementless Shoulder Prosthesis is lockable with two cortical bone screws and is intended for cementless use only. An optional anti-rotation spoiler can be used with either the cementless or the cemented stems.
The glenoid baseplate and post extension are intended for cementless use with the addition of screws for fixation.
## Summary of Technologies/Substantial Equivalence:
The Humelock Reversed Shoulder is substantially equivalent to the predicate device in regards to its intended use and indications, materials, design and sizes. Differences between the subject device systems and the predicate device systems do not raise new types of safety and effectiveness questions.
#### Non-Clinical Testing:
Range of motion analysis, stem /spacer / humeral cup construct fatigue testing, post-fatigue disassembly testing and torsional fatigue testing of the spoiler were conducted. The results of these tests indicate that the performance of the Humelock Reversed Shoulder is adequate for its intended use. Pyrogenicity testing was conducted and the Humelock Reversed Shoulder met a recommended limit of <20 EU/device construct.
#### Clinical Testing:
Clinical testing was not necessary to determine substantial equivalence of the Humelock Reversed Shoulder to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.