Cemented Acqualis Reversed prosthesis: It is indicated for patients with a functional deltoid muscle as a total shoulder replacement for the relief of pain and significant disability following arthropathy associated with the massive and non repairable rotator cuff-tear. This device is also indicated for the prosthetic revisions with massive and non repairable rotator cuff-tear. Only the humeral components are for cemented use. The glenoid implant is anchored to the bone with 4 screws and is for non-cemented fixation. When during the primary surgery the glenoid bone stock appears to be insufficient to bear the reversed glenoid components or when glenoid bone fracture occurs during the surgical procedures, the hemi-prosthesis adaptor and the union screw can be adapted to the humeral components in order to transform the Aequalis Reversed prosthesis into a non reversed hemi-prosthesis. When, in case of revision of a Aequalis Reversed prosthesis, the glenoid bone stock appears to be insufficient to again implant a base plate and a sphere of Aequalis Reversed range, the use of the hemi-prosthesis adaptor and the union screw allows for the transformation of the Aequalis Reversed prosthesis in to a non reversed hemi-prosthesis in order to avoid the revision of the humeral components. Uncemented Acqualis Reversed prosthesis: It is indicated for patients with a functional deltoid muscle as a total shoulder replacement for the relief of pain and significant disability following arthropathy associated to massive and non repairable rotator cuff-tear. This device is also indicated for the prosthetic revisions with massive and non repairable rotator cuff-tear. The humeral components are for non-cemented use. The glenoid implant is anchored to the bone with 4 screws and is for non-cemented fixation. When during the primary surgery the glenoid stock appears to be insufficient to bear the reversed glenoid components or when glenoid bone fracture occurs during the surgical procedures, the hemiprosthesis adaptor and the union screw can be adapted to the humeral components in order to transform the Aequalis Reversed prosthesis into a non reversed hemi-prosthesis. When, in case of revision of an Aequalis Reversed prosthesis, the glenoid bone stock appears to be insufficient to again implant a base plate and a sphere of Aequalis Reversed range, the use of the hemi-prosthesis adaptor and the union screw allows for the transformation of the Aequalis Reversed prosthesis in to a non reversed hemi-prosthesis in order to avoid the revision of the humeral components.
Device Story
Aequalis Reversed Shoulder Prosthesis is a semi-constrained total shoulder system; designed to treat arthropathy with massive, non-repairable rotator cuff tears. Device features reversed geometry to medialize center of rotation; increases deltoid muscle lever arm. System comprises humeral and glenoid components; humeral stems available in cemented or uncemented (hydroxylapatite-coated) configurations. Glenoid component anchored via 4 screws. Intraoperative adaptability allows conversion to non-reversed hemi-prosthesis using adaptor and union screw if glenoid bone stock is insufficient or fractured. Used by orthopedic surgeons in clinical settings. Restores joint function; reduces pain; improves patient mobility.
Clinical Evidence
No clinical data; bench testing only. Equivalence established through comparative analysis of design, materials, and fixation methods against predicate devices.
Technological Characteristics
Semi-constrained shoulder prosthesis. Materials: Titanium alloy (stems), Cobalt-chromium alloy and UHMWPE (metaphyses). Hydroxylapatite coating per ASTM F 1185. Fixation: Cemented or uncemented (press-fit with HA coating). Sterilization: Gamma radiation. Modular design.
Indications for Use
Indicated for patients with functional deltoid muscle requiring total shoulder replacement for pain relief and disability reduction due to arthropathy associated with massive, non-repairable rotator cuff tears; also indicated for prosthetic revisions in cases of massive, non-repairable rotator cuff tears.
Regulatory Classification
Identification
A shoulder joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a shoulder joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a humeral resurfacing component made of alloys, such as cobalt-chromium-molybdenum, and a glenoid resurfacing component made of ultra-high molecular weight polyethylene, and is limited to those prostheses intended for use with bone cement (§ 888.3027).
Special Controls
*Classification.* Class II. The special controls for this device are:(1) FDA's:
(i) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ”
(ii) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),”
(iii) “Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement,”
(iv) “Guidance Document for the Preparation of Premarket Notification (510(k)) Application for Orthopedic Devices,” and
(v) “Guidance Document for Testing Non-articulating, ‘Mechanically Locked’ Modular Implant Components,”
(2) International Organization for Standardization's (ISO):
(i) ISO 5832-3:1996 “Implants for Surgery—Metallic Materials—Part 3: Wrought Titanium 6-aluminum 4-vandium Alloy,”
(ii) ISO 5832-4:1996 “Implants for Surgery—Metallic Materials—Part 4: Cobalt-chromium-molybdenum casting alloy,”
(iii) ISO 5832-12:1996 “Implants for Surgery—Metallic Materials—Part 12: Wrought Cobalt-chromium-molybdenum alloy,”
(iv) ISO 5833:1992 “Implants for Surgery—Acrylic Resin Cements,”
(v) ISO 5834-2:1998 “Implants for Surgery—Ultra-high Molecular Weight Polyethylene—Part 2: Moulded Forms,”
(vi) ISO 6018:1987 “Orthopaedic Implants—General Requirements for Marking, Packaging, and Labeling,” and
(vii) ISO 9001:1994 “Quality Systems—Model for Quality Assurance in Design/Development, Production, Installation, and Servicing,” and
(3) American Society for Testing and Materials':
(i) F 75-92 “Specification for Cast Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implant Material,”
(ii) F 648-98 “Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants,”
(iii) F 799-96 “Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Forgings for Surgical Implants,”
(iv) F 1044-95 “Test Method for Shear Testing of Porous Metal Coatings,”
(v) F 1108-97 “Specification for Titanium-6 Aluminum-4 Vanadium Alloy Castings for Surgical Implants,”
(vi) F 1147-95 “Test Method for Tension Testing of Porous Metal,”
(vii) F 1378-97 “Standard Specification for Shoulder Prosthesis,” and
(viii) F 1537-94 “Specification for Wrought Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implants.”
{0}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/0/Picture/1 description: The image is a logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol consists of three stylized human profiles facing to the right, stacked on top of each other. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".
### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 2, 2016
Tornier Mrs. Severine Bonneton Regulatory Affairs Specialist 161, rue Lavoisier - Montbonnot 38334 Saint-Ismier Cedex FRANCE
Re: K100142
Trade/Device Name: Aequalis Reversed Shoulder Prosthesis Regulation Number: 21 CFR 888.3660 Regulation Name: Shoulder joint metal/polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: PHX, KWS, HSD Dated: May 3, 2010 Received: May 5, 2010
Dear Mrs. Bonneton:
This letter corrects our substantially equivalent letter of May 6, 2010.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing
{1}------------------------------------------------
### Page 2 - Mrs. Severine Bonneton
(21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Lori A. Wiggins -S
for
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
### 510(k) Number (if known): K100142
Device Name: Aequalis Reversed Shoulder Prosthesis
Indications For Use:
### Cemented Acqualis Reversed prosthesis:
It is indicated for patients with a functional deltoid muscle as a total shoulder replacement for the relief of pain and significant disability following arthropathy associated with the massive and non repairable rotator cuff-tear. This device is also indicated for the prosthetic revisions with massive and non repairable rotator cuff-tear. Only the humeral components are for cemented use. The glenoid implant is anchored to the bone with 4 screws and is for non-cemented fixation.
When during the primary surgery the glenoid bone stock appears to be insufficient to bear the reversed glenoid components or when glenoid bone fracture occurs during the surgical procedures, the hemi-prosthesis adaptor and the union screw can be adapted to the humeral components in order to transform the Aequalis Reversed prosthesis into a non reversed hemi-prosthesis.
When, in case of revision of a Aequalis Reversed prosthesis, the glenoid bone stock appears to be insufficient to again implant a base plate and a sphere of Aequalis Reversed range, the use of the hemi-prosthesis adaptor and the union screw allows for the transformation of the Aequalis Reversed prosthesis in to a non reversed hemi-prosthesis in order to avoid the revision of the humeral components.
{3}------------------------------------------------
### Uncemented Acqualis Reversed prosthesis:
It is indicated for patients with a functional deltoid muscle as a total shoulder replacement for the relief of pain and significant disability following arthropathy associated to massive and non repairable rotator cuff-tear. This device is also indicated for the prosthetic revisions with massive and non repairable rotator cuff-tear. The humeral components are for non-cemented use. The glenoid implant is anchored to the bone with 4 screws and is for non-cemented fixation.
When during the primary surgery the glenoid stock appears to be insufficient to bear the reversed glenoid components or when glenoid bone fracture occurs during the surgical procedures, the hemiprosthesis adaptor and the union screw can be adapted to the humeral components in order to transform the Aequalis Reversed prosthesis into a non reversed hemi-prosthesis.
When, in case of revision of an Aequalis Reversed prosthesis, the glenoid bone stock appears to be insufficient to again implant a base plate and a sphere of Aequalis Reversed range, the use of the hemi-prosthesis adaptor and the union screw allows for the transformation of the Aequalis Reversed prosthesis in to a non reversed hemi-prosthesis in order to avoid the revision of the humeral components.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| (Division Sign-Off) | |
|-----------------------------------------------------------|---------|
| Division of Surgical, Orthopedic, and Restorative Devices | |
| | K100142 |
510(k) Number
{4}------------------------------------------------
## Implants Chirurgicaux
MAY - 5 2010
### Summary of Safety and Effectiveness information 510(k) Premarket Notification - Aequalis Reversed Shoulder Prosthesis
Date: April 29th, 2010
Regulatory authority: Safe Medical Devices Act of 1990, 21 CRF 807.92
### 1) Device name
Aequalis Reversed Shoulder Prosthesis Trade name: Common name: Total-Shoulder System and Hemi-Shoulder System Classification name: Shoulder joint metal/polymer semi-constrained cemented prosthesis Shoulder, Hemi-, Humeral, Metallic uncemented
### 2) Submitter
Tornier Rue Doyen Gosse 38330 Saint Ismier - France
### 3) Company contact
Tornier Mrs Séverine Bonneton Regulatory affairs Specialist 161, rue Lavoisier - Montbonnot 38334 Saint Ismier Cedex - France Tel: 00 33 4 76 61 35 00 Fax: 00 33 4 76 61 35 59 e-mail: severine.bonneton@tornier.fr
### 4) Classification
| Device class: | Class II |
|-----------------------|------------|
| Classification panel: | Orthopedic |
| Product code: | KWS, HSD |
### 5) Equivalent / Predicate device
Aequalis Reversed Shoulder Prosthesis, TORNIER SA, K030941, K041873, K050316, K061439, K081059.
Aequalis Reversed Fracture Shoulder System, TORNIER SA, K082120. Delta Xtend Reverse Shoulder System, DEPUY, K062250, K071379, K091751.
Image /page/4/Picture/16 description: The image shows a logo with a stylized letter 'T' inside a hexagon. The 'T' is bold and has a unique design, with a horizontal bar at the top and a vertical stem that extends downwards. The hexagon provides a clear border around the central 'T' symbol, making it stand out.
TORNIER S.A.S. 161, rue Lavoisier - Montbonnot 38334 SAINT-ISMIER CEDEX FRANCE
Tહા. : 33 (0)4 76 61 35 00 Fax : 33 (0)4 76 61 35 33 Section 5 - Page 1/ page 4
S.A.S. au capital de 288 000 € SIRET : 070 501 275 000 13 R.C.S. : 070 501 275 CODE APE : 331 B
SIEGIE SOCIAL : rue du Doyen Gosse - 38330 SAINT-ISMIER - FRANCE
{5}------------------------------------------------
# Implants Chirurgicaux
### 6) Device description
The Aequalis Reversed Shoulder Prosthesis is intended to relieve pain and significant disability . following massive and non repairable cuff-tear associated to arthropathy and following massive cuff-tear arthropathy. In this case, the rotator muscles of the shoulder (supraspinatus, teres minor and long head of the biceps) are no more useful for mobility, and only the deltoid (for abduction and external rotation) and the subscapularis (for internal rotation) are functional.
Therefore, the usual goal of such surgery is to restore the shoulder joint to facilitate its working condition and to reduce or eliminate pain. The Aequalis Reversed Shoulder Prosthesis is intended to accomplish these goals. Its reversed design allows to medialize the center of rotation of the shoulder, lengthening the deltoid muscle lever arm.
The Aequalis Reversed Shoulder Prosthesis is a semi-constrained system composed of a humeral and a glenoid parts.
The present device modification submission consists of:
- the addition of humeral stems with hydroxylapatite coating,
- the addition of 2 sizes of metaphyses with hydroxylapatite coating, •
- the update of the Aequalis Reversed Shoulder Prosthesis indications for use. .
### 7) Materials
The stems are manufactured from titanium alloy. The metaphyses are manufactured from cobalt-chromium alloy and ultra high molecular weight polyethylene (UHMWPE).
The hydroxylapatite coating conforms to ASTM standard F 1185. The coating is performed by BioCoat, Inc. according to their Master File MAF-339.
### 8) Indications for use
Cemented Aequalis Reversed prosthesis:
It is indicated for patients with a functional deltoid muscle as a total shoulder replacement for the relief of pain and significant disability following arthropathy associated with the massive and non repairable rotator cuff-tear. This device is also indicated for the prosthetic revisions with massive and non repairable rotator cuff-tear. Only the humeral components are for cemented use. The glenoid implant is anchored to the bone with 4 screws and is for non-cemented fixation.
When during the primary surgery the glenoid bone stock appears to be insufficient to bear the reversed glenoid components or when glenoid bone fracture occurs during the surgical procedures, the hemiprosthesis adaptor and the union screw can be adapted to the humeral components in order to transform the Aequalis Reversed prosthesis into a non reversed hemi-prosthesis.
When, in case of revision of a Aequalis Reversed prosthesis, the glenoid bone stock appears to be insufficient to again implant a base plate and a sphere of Aequalis Reversed range, the use of the hemiprosthesis adaptor and the union screw allows for the transformation of the Aequalis Reversed prosthesis in to a non reversed hemi-prosthesis in order to avoid the revision of the humeral components.
Section 5 . Page 2/ page 4
Image /page/5/Picture/18 description: The image shows a logo with a stylized letter 'T' inside a hexagon. The 'T' is designed with a thick vertical line and a horizontal bar at the top, with additional decorative elements on either side of the vertical line. The logo appears to be a simple, possibly vintage design, rendered in black and white.
TORNIER S.A.S. 161. rue Lavoisier - Montbonnot 38334 SAINT-ISMIER CEDEX FRANCE
Ticl. : 33 (0)4 76 61 35 00 Fax : 33 (0)4 76 61 35 33
S.A.S. au capital de 288 000 € SIRET : 070 501 275 000 13 R.C.S. : 070 501 275 CODE APE : 331 B
S1EGE SOCIAL : rue du Doyen Gosse - 38330 SAINT-ISMIER - FRANCE
{6}------------------------------------------------
### Uncemented Aequalis Reversed prosthesis:
It is indicated for patients with a functional deltoid muscle as a total shoulder replacement for the relief of pain and significant disability following arthropathy associated to massive and non repairable rotator cufftear. This device is also indicated for the prosthetic revisions with massive and non repairable rotator cufftear. The humeral components are for non-cemented use. The glenoid implant is anchored to the bone with 4 screws and is for non-cemented fixation.
Implants Chirurgicaux
KIDOIYA * >14
When during the primary surgery the glenoid stock appears to be insufficient to bear the reversed glenoid components or when glenoid bone fracture occurs during the surgical procedures, the hemi-prosthesis adaptor and the union screw can be adapted to the humeral components in order to transform the Aequalis Reversed prosthesis into a non reversed hemi-prosthesis.
When, in case of revision of an Aequalis Reversed prosthesis, the glenoid bone stock appears to be insufficient to again implant a base plate and a sphere of Aequalis Reversed range, the use of the hemiprosthesis adaptor and the union screw allows for the transformation of the Aequalis Reversed prosthesis in to a non reversed hemi-prosthesis in order to avoid the revision of the humeral components.
| Main features or system<br>characteristics | | Aequalis Reversed<br>(new components) | Aequalis Reversed | Aequalis Reversed<br>Fracture | Delta Xtend<br>Reverse Shoulder |
|--------------------------------------------|-----------------|---------------------------------------|---------------------------------------------------------|-------------------------------|---------------------------------|
| Materials | Humeral<br>stem | Ti6Al4V | CoCr | Ti6Al4V | CoCr + UHMWPE |
| | Metaphysis | CoCr + UHMWPE | CoCr + UHMWPE | - | CoCr + UHMWPE |
| Coating<br>material | Humeral part | Hydroxylapatite | uncoated | Hydroxylapatite | Hydroxylapatite or<br>uncoated |
| Size | Humeral<br>stem | 4 diameters<br>4 lengths | 4 diameters<br>4 lengths | 5 diameters<br>4 lengths | 5 diameters<br>9 lengths |
| | Metaphysis | 2 sizes:36 or 42 mm | 2 sizes:36 or 42 mm | 40 mm | 2 sizes |
| Stem / metaphysis type | | Modular | Modular | Monobloc | Monobloc or<br>modular |
| Method of fixation<br>Humeral component | | uncemented | cemented | partially cemented | cemented or<br>uncemented |
| Terminal sterilization | | Gamma | Gamma | Gamma | Gamma |
| Manufacturer | | Tornier | Tornier | Tornier | DePuy |
| K-number | | K100142 | K030941,<br>K041873,<br>K050316,<br>K061439,<br>K081059 | K082120 | K062250,<br>K071379,<br>K091751 |
### 9) Summary of technological characteristics
The technical characteristics (materials, design and sizing) of the new components of the Aequalis Reversed Shoulder Prosthesis are similar or identical to the predicate devices.
Image /page/6/Picture/7 description: The image shows a logo with a stylized letter 'T' inside a hexagon. The 'T' is designed with thick lines and has a slightly abstract, geometric appearance. The logo is black and white, with the 'T' and the hexagon both being black against a white background.
TORNIER S.A.S. 161, rue Lavoisier - Monthonnot 38334 SAINT-ISMIER CEDEX FRANCE
Tél. : 33 (0)4 76 61 35 00 Fax : 33 (0)4 76 61 35 33
Section 5 - Page 3/ page 4
S.A.S. nu capital de 288 000 € SIRET : 070 501 275 000 13 R.C.S. : 070 501 275 CODE APE : 331 B
SIEGI! SOCIAL : rue du Dayen Gosse - 38330 SAINT-ISMIER - FRANCE
{7}------------------------------------------------
K10014d #4/4
### mplants Chirurgicaux
### 10) Non-clinical testing
Non-clinical testing was not necessary to determinate substantial equivalence between the new uncemented stems and new uncemented metaphysis of the Aequalis Reversed Shoulder Prosthesis and the cited predicate devices:
- The design of the new components of the Aequalis Reversed Shoulder Prosthesis (uncemented . stems and uncemented metaphyses) is identical to the design of predicate Aequalis Reversed Shoulder Prosthesis (K030941, K041873, K050316, K061439, and K081059) except for the addition of an HA coating.
- The raw material of the new components of the Aequalis Reversed Shoulder Prosthesis -(uncemented stems and uncemented metaphyses) is identical to the raw material of predicates Aequalis Reversed Fracture Shoulder Prosthesis (K082120) and Delta Xtend (K062250, K071379, and K091751).
- The coating material of the new components of the Aequalis Reversed Shoulder Prosthesis -(uncemented stems and uncemented metaphyses) is identical to the coating material of the predicate Aequalis Reversed Fracture Shoulder Prosthesis (K082120).
- The fixation method of the new humeral components of the Aequalis Reversed Shoulder Prosthesis is identical to the fixation method of the predicates Aequalis Reversed Fracture Shoulder Prosthesis (K082120) and Delta Xtend (K062250, K071379, and K091751).
### 11) Substantial equivalence conclusion
Based upon this comparative study, substantial equivalence of the new components of the Aequalis Reversed Shoulder Prosthesis to the cleared components of the predicates can be demonstrated on the following grounds, according to the FDA's Guidelines for Substantial Equivalence Decision Making Process:
- The new components of the Aequalis Reversed Shoulder Prosthesis are compared to the predicate devices.
- The new components of the Aequalis Reversed Shoulder Prosthesis have the same intended use as 1 the cleared predicate.
- Major technological characteristics are equivalent between the new components of the Aequalis • Reversed Shoulder Prosthesis and the predicate devices:
- Equivalence of general features l
- Equivalent means of fixation
- Equivalent materials -
.
- Equivalent surgical technique.
Therefore, in the light of the above information, the new components of the Aequalis Reversed Shoulder Prosthesis are found to be equivalent to the predicate devices.
Image /page/7/Picture/18 description: The image shows a logo with a stylized letter 'T' inside a hexagon. The 'T' is designed with a unique, almost tree-like structure, featuring a thick vertical line and two curved shapes on either side near the top. The logo is presented in black and white, with the 'T' appearing solid and the hexagon outlining it.
TORNIER S.A.S. 161, rue Lavoisier - Montbonnot 38334 SAINT-ISMIER CEDEX FRANCE
Tél. : 33 (0)4 76 61 35 00 Fax : 33 (0)4 76 61 35 33
Section 5 - Page 4/ page 4
S.A.S. au capital de 288 000 € SIRET : 070 501 275 000 13 R.C.S. : 070 501 275 CODE APE : 331 B
SIEGE SOCIAL : rue du Doyen Gosse - 38330 SAIN'T-ISMIER - FRANCE
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.