K163669 · Fx Solutions · PHX · May 17, 2017 · Orthopedic
Device Facts
Record ID
K163669
Device Name
Humeris Shoulder
Applicant
Fx Solutions
Product Code
PHX · Orthopedic
Decision Date
May 17, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3660
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
In an anatomic shoulder configuration, the Humeris Shoulder System is indicated for use in total and hemi-shoulder replacement to treat: - A severely painful and/or disabled joint resulting from osteoarthritis or rheumatoid arthritis; - Other difficult clinical problems where shoulder arthrodesis or resection arthroplasty are not acceptable (e.g. revision of a previously implanted primary component, a humeral plate or a humeral nail). In a reverse shoulder configuration, the Humeris Shoulder is indicated for primary or revision total shoulder arthroplasty for the relief of pain and to improve function in patients with a massive and non-repairable rotator cuff tear. The patient's joint must be anatomically suited to receive the selected implants and a functional deltoid muscle is necessary to use the device. The humeral stem of the Humeris Cementless Shoulder is intended for cementless use only. The humeral stem of the Humeris Cemented Shoulder is intended for cemented use only. The glenoid components of the Humeris Shoulder System are intended for cemented use only. The glenoid baseplate component is intended for cementless use with the addition of screws for fixation.
Device Story
Humeris Shoulder System is a modular orthopedic prosthesis for anatomic or reverse shoulder arthroplasty. System components include humeral stems (cemented/cementless), humeral heads, glenoid components, and reverse configuration parts (glenosphere, baseplate, adapters). Surgeons implant the device in a clinical/OR setting to replace damaged shoulder joints. Anatomic configuration treats arthritis or revision cases; reverse configuration treats massive rotator cuff tears. Modular design allows adjustment of head angle and offset to optimize bone coverage and joint function. Reverse configuration lateralizes center of rotation to reduce scapular notching. Device provides pain relief and improved joint function for patients with severe shoulder pathology.
Clinical Evidence
No clinical data. Substantial equivalence determined via non-clinical bench testing, including range of motion analysis, fatigue testing of anatomic constructs, stress analysis of reverse configurations, pyrogenicity, and bacterial endotoxin testing. Coating characterization and glenoid loosening testing (ASTM F2028) were referenced from previous submissions.
Technological Characteristics
Materials: Ti-6Al-4V alloy (ISO 5832-3), Co-Cr-Mo alloy (ISO 5832-12), UHMWPE (ISO 5834-2), Tantalum (ASTM F560). Surface: Plasma sprayed CP Titanium and hydroxyapatite (HA) coating. Modular design with taper lock connections. Reverse configuration includes glenosphere with 10-degree tilt. Fixation: Cemented or cementless humeral stems; cemented glenoids; cementless glenoid baseplate with bone screws.
Indications for Use
Indicated for patients requiring total or hemi-shoulder replacement due to severe pain/disability from osteoarthritis or rheumatoid arthritis, or revision of prior implants. Reverse configuration indicated for primary or revision total shoulder arthroplasty in patients with massive, non-repairable rotator cuff tears. Requires functional deltoid muscle and anatomically suitable joint.
Regulatory Classification
Identification
A shoulder joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a shoulder joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a humeral resurfacing component made of alloys, such as cobalt-chromium-molybdenum, and a glenoid resurfacing component made of ultra-high molecular weight polyethylene, and is limited to those prostheses intended for use with bone cement (§ 888.3027).
Special Controls
*Classification.* Class II. The special controls for this device are:(1) FDA's:
(i) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ”
(ii) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),”
(iii) “Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement,”
(iv) “Guidance Document for the Preparation of Premarket Notification (510(k)) Application for Orthopedic Devices,” and
(v) “Guidance Document for Testing Non-articulating, ‘Mechanically Locked’ Modular Implant Components,”
(2) International Organization for Standardization's (ISO):
(i) ISO 5832-3:1996 “Implants for Surgery—Metallic Materials—Part 3: Wrought Titanium 6-aluminum 4-vandium Alloy,”
(ii) ISO 5832-4:1996 “Implants for Surgery—Metallic Materials—Part 4: Cobalt-chromium-molybdenum casting alloy,”
(iii) ISO 5832-12:1996 “Implants for Surgery—Metallic Materials—Part 12: Wrought Cobalt-chromium-molybdenum alloy,”
(iv) ISO 5833:1992 “Implants for Surgery—Acrylic Resin Cements,”
(v) ISO 5834-2:1998 “Implants for Surgery—Ultra-high Molecular Weight Polyethylene—Part 2: Moulded Forms,”
(vi) ISO 6018:1987 “Orthopaedic Implants—General Requirements for Marking, Packaging, and Labeling,” and
(vii) ISO 9001:1994 “Quality Systems—Model for Quality Assurance in Design/Development, Production, Installation, and Servicing,” and
(3) American Society for Testing and Materials':
(i) F 75-92 “Specification for Cast Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implant Material,”
(ii) F 648-98 “Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants,”
(iii) F 799-96 “Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Forgings for Surgical Implants,”
(iv) F 1044-95 “Test Method for Shear Testing of Porous Metal Coatings,”
(v) F 1108-97 “Specification for Titanium-6 Aluminum-4 Vanadium Alloy Castings for Surgical Implants,”
(vi) F 1147-95 “Test Method for Tension Testing of Porous Metal,”
(vii) F 1378-97 “Standard Specification for Shoulder Prosthesis,” and
(viii) F 1537-94 “Specification for Wrought Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implants.”
Predicate Devices
Fx Solutions Humelock II Reversible Shoulder (K150488)
Tornier Aequalis Ascend Flex Shoulder System (K122698)
Submission Summary (Full Text)
{0}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Fx Solutions Ratsamy Thevenin Regulatory Affairs Ouality Manager 1663 Rue de Maiornas Viriat. 01440 FRANCE
May 17, 2017
Re: K163669
Trade/Device Name: Humeris Shoulder Regulation Number: 21 CFR 888.3660 Regulation Name: Shoulder Joint Metal/Polymer Semi-Constrained Cemented Prosthesis Regulatory Class: Class II Product Code: PHX, KWT, HSD Dated: April 14, 2017 Received: April 17, 2017
Dear Ratsamy Thevenin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
Image /page/0/Picture/11 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other. The profiles are simple and abstract, and they are all the same color.
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
# Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K163669
Device Name Humeris Shoulder
#### Indications for Use (Describe)
In an anatomic shoulder configuration, the Humeris Shoulder System is indicated for use in total and hemi-shoulder replacement to treat:
- A severely painful and/or disabled joint resulting from osteoarthritis or rheumatoid arthritis;
- Other difficult clinical problems where shoulder arthrodesis or resection arthroplasty are not acceptable (e.g. revision of a previously implanted primary component, a humeral plate or a humeral nail).
In a reverse shoulder configuration, the Humeris Shoulder is indicated for primary or revision total shoulder athroplasty for the relief of pain and to improve function in patients with a massive and non-repairable rotator cuff tear.
The patient's joint must be anatomically suited to receive the selected implants and a functional deltoid muscle is necessary to use the device.
The humeral stem of the Humeris Cementless Shoulder is intended for cementless use only. The humeral stem of the Humeris Cemented Shoulder is intended for cemented use only. The glenoid components of the Humeris Shoulder System are intended for cemented use only. The glenoid baseplate component is intended for cementless use with the addition of screws for fixation.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------|---------------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
| 510(k) Summary | |
|-------------------------------|----------------------------------------------------------------------------------------------------|
| Prepared: | May 12, 2017 |
| Submitter: | Fx Solutions<br>1663 Rue de Majornas<br>Viriat, France 01440 |
| Contact: | Ratsamy Thevenin<br>Regulatory Affairs Quality Manager<br>+33 4 74 55 35 55<br>www.fxsolutions.fr |
| Proprietary Name: | Humeris Shoulder |
| Common Name: | Anatomic and Reverse Shoulder Prosthesis |
| Classification Names: | 21 CFR 888.3660: Shoulder joint metal/polymer semi-constrained<br>cemented prosthesis; Class II |
| | 21 CFR 888.3650: Shoulder joint metal/polymer non-constrained<br>cemented prosthesis; Class II |
| | 21 CFR 888.3690 Shoulder joint humeral (hemi-shoulder) metallic<br>uncemented prosthesis; Class II |
| Product Codes: | 87/PHX, 87/KWT, 87/HSD |
| Primary Predicate Device: | Fx Solutions Humelock II Reversible Shoulder, K150488 |
| Additional Predicate Devices: | Tornier Aequalis Ascend Flex Shoulder System, K122698 |
## Device Description:
The Humeris Shoulder is a system of shoulder components that can be used in either an anatomic or a reverse configuration.
The Humeris Cementless Humeral Stem is manufactured from Ti-6Al-4V alloy conforming to ISO 5832-3 and is available in diameters of 8-16mm and a length of 100mm. The distal end of the humeral stem is trapezoidal with a polished surface. The proximal portion of the humeral stem has a plasma sprayed commercially pure Titanium (CP Ti) and hydroxyapatite (HA) coating.
{4}------------------------------------------------
The Humeris Cemented Humeral Stem is also manufactured from Ti-6Al-4V alloy conforming to ISO 5832-3 and is available in diameters of 6-14mm and a length of 100mm. The distal end of the humeral stem is trapezoidal with a polished surface.
Both humeral stems incorporate a female taper for attachment of compatible components. The Humeris humeral stems can be used with a straight or angled double taper connector or a centered spacer, a centered or offset humeral head and a 2 peg or 3 or 4 peg cemented glenoid for use in an anatomical shoulder configuration.
The straight and angled double taper connectors and the centered spacers have a male taper on each end and are used to connect the humeral stem to the humeral head. The straight double taper connector and three options of angled double taper connectors allow the angle of the head to be changed from 135° to 132° or 138° and the offset of the humeral head to be increased from +0 to +3 or +5 mm. The centered spacers also allow the offset of the humeral head to be increased from +0 to +3 or +5 mm. All of the double taper connectors and centered spacers are compatible with all sizes of humeral stems and humeral heads. The double taper connectors and the centered spacers are manufactured from Ti-6Al-4V alloy conforming to ISO 5832-3.
The humeral heads are available in diameters of 39 – 54 mm in centered and offset styles. The offset of the taper allows the head to be rotated relative to the cut surface of the humerus to provide optimal coverage of the bone. A female taper allows attachment to the double taper connector, which connects to the humeral stem. The humeral heads are manufactured from wrought Co-Cr-Mo alloy conforming to ISO 5832-12.
The 2 peg cemented glenoid component is available in sizes extra small, small, medium and large and features two pegs for cemented fixation to the glenoid bone. It is manufactured from ultra-high molecular weight polyethylene (UHMWPE) conforming to ISO 5834-2. Each peg contains a radiopaque marker manufactured from tantalum conforming to ASTM F560.
The 3 or 4 peg cemented glenoid component is also available in sizes extra small, small, medium and large. Sizes extra small and small have three fixation pegs. Sizes medium and large have four fixation pegs. The 3 or 4 peg cemented glenoid component is manufactured from ultra-high molecular weight polyethylene (UHMWPE) conforming to ISO 5834-2. The central peg contains a radiopaque marker manufactured from tantalum conforming to ASTM F560.
The Humeris humeral stems can also be used with a humeral cup and 135/145 adapter or a 135/145 humeral cup, a centered or eccentric glenosphere with or without a central screw, a glenoid baseplate, post extensions and standard (compression) or locking bone screws for use in a reverse shoulder configuration.
{5}------------------------------------------------
The humeral cup and the 135/145 humeral cup are available in two sizes, Ø36 and Ø40 mm. Each size is available in two versions, standard and mobility. Each version is available in three heights: +3mm, +6mm, +9mm; and is compatible with all sizes of humeral stems. A 24mm male taper allows attachment of the 135/145 cup to the humeral stem. If the humeral cup is used, it must be used with a 135/145 adaptor. Male tapers allow attachment of the humeral cup to the 135/145 adaptor and of the 135/145 adaptor to the humeral stem. The humeral cup and 135/145 humeral cup are preassembled, one-piece components manufactured from UHMWPE conforming to ISO 5834-2 and Ti-6Al-4V alloy conforming to ISO 5832-3.
The glenosphere is also available in Ø36 and Ø40 mm sizes and in centered and eccentric styles. The eccentric glenospheres are designed to be offset from the center of the glenoid baseplate. All glenospheres have a 10° tilt, meaning that the curvature of the glenosphere extends 10° beyond the equator of a hemisphere. This additional articular surface lateralizes the center of rotation to help reduce the potential for scapular notching by the humeral cup. All glenospheres mate with the glenoid baseplate via a taper lock. The glenospheres are also available with an optional central, cannulated screw. This screw can be threaded through the central post of the baseplate for additional security. The glenosphere components are manufactured from wrought Co-Cr-Mo alloy conforming to ISO 5832-12. The glenosphere screw is manufactured from Ti-6Al-4V allov conforming to ISO 5832-3.
The glenoid baseplate has a round base with a central, cannulated post and four peripheral, threaded screw holes. The outer edges of the baseplate are tapered to lock with the glenosphere component.
The glenoid baseplate is used with 4.5mm standard or locking bone screws for added stability. The bone screws are available in lengths from 20 - 50mm in 5mm increments. Each screw allows an angulation of +/- 12° around the axial axis of the screw hole.
Optional post extensions are available to extend the central post of the baseplate and provide additional anchoring in cases with poor bone quality. The post extensions are available in 6mm and 10mm lengths. When used, the post extensions screw into the baseplate post.
The glenoid baseplate, standard and locking screws, and post extensions are manufactured from Ti-6Al-4V alloy conforming to ISO 5832-3. The backside of the baseplate and the post extensions are coated with a plasma sprayed CP Titanium and Hydroxyapatite coating.
{6}------------------------------------------------
## Intended Use / Indications:
In an anatomic shoulder configuration, the Humeris Shoulder System is indicated for use in total and hemi-shoulder replacement to treat:
- A severely painful and/or disabled joint resulting from osteoarthritis or rheumatoid arthritis;
- Other difficult clinical problems where shoulder arthrodesis or resection arthroplasty are not acceptable (e.g. revision of a previously implanted primary component, a humeral plate or a humeral nail).
In a reverse shoulder configuration, the Humeris Shoulder is indicated for primary or revision total shoulder arthroplasty for the relief of pain and to improve function in patients with a massive and non-repairable rotator cuff tear.
The patient's joint must be anatomically and structurally suited to receive the selected implants and a functional deltoid muscle is necessary to use the device.
The humeral stem of the Humeris Cementless Shoulder is intended for cementless use only. The humeral stem of the Humeris Cemented Shoulder is intended for cemented use only. The glenoid components of the Humeris Shoulder System are intended for cemented use only. The glenoid baseplate component is intended for cementless use with the addition of screws for fixation.
## Summary of Technologies/Substantial Equivalence:
The Humeris Shoulder is substantially equivalent to the predicate device in regards to its intended use and indications, materials, design and sizes. Differences between the subject device systems and the predicate device systems do not raise new types of safety and effectiveness questions.
# Non-Clinical Testing:
Range of motion analysis, stem / double taper connector / humeral head construct fatigue testing in an anatomic configuration, post-fatigue disassembly testing, stress analysis of the worst case stem in a reverse configuration and pyrogenicity testing were conducted. Characterization of the Ti/HA plasma spray coating and glenoid loosening testing per ASTM F2028 were submitted in previous submissions and were referenced to support the current submission. The results of these tests indicate that the performance of the Humeris Shoulder is adequate for its intended use. Bacterial endotoxin testing was conducted.
# Clinical Testing:
Clinical testing was not necessary to determine substantial equivalence of the Humeris Shoulder to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.