ISSYS LP SPINAL FIXATION SYSTEM

K070281 · Custom Spine, Inc. · NKB · Jun 11, 2007 · Orthopedic

Device Facts

Record IDK070281
Device NameISSYS LP SPINAL FIXATION SYSTEM
ApplicantCustom Spine, Inc.
Product CodeNKB · Orthopedic
Decision DateJun 11, 2007
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3070
Device ClassClass 2
AttributesTherapeutic

Intended Use

The ISSYS LP Spinal Fixation System is intended to help provide immobilization of spinal segments as an adjunct to fusion of the lumbar and/or sacral spine, specifically as follows: For pedicular use: When used as pedicle screw fixation system of the non cervical posterior spine in skeletally mature patients, these systems are indicated for one or more of the following: degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). In addition, this system is indicated for pedicle screw fixation in skeletally mature patients with severe spondylolisthesis (grade 3 & 4) at the L5-S1 joint having fusion with autogenous bone graft, having the device fixed or attached to the lumbar and sacral spine (with pedicle placement at L3 and below) with removal of the implants after the development of a solid fusion mass. When used as non pedicular fixation system: The ISSYS LP Spinal Fixation Systems, when used as an anterior screw fixation system and posterior sacral/iliac screw fixation system are indicated for the following: Degenerative disc disease of the thoracic and lumbar spine (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies). Spondylolisthesis. Fracture. Spinal deformities such as scoliosis, kyphosis, lordosis. Tumor. Revision of failed fusion attempts. Pseudarthrosis. Spinal stenosis. When used in the anterior indication the ISSYS LP Spinal Fixation Systems are indicated for use in the thoracic and lumbar spine.

Device Story

ISSYS LP Spinal Fixation System consists of pedicle screws, rods, cross connectors, and staples; used for spinal immobilization and stabilization as an adjunct to fusion. Implants are custom-fitted to individual patient anatomy. Used by surgeons in clinical settings to treat degenerative disc disease, deformities, and fractures. Provides mechanical stabilization to facilitate bone fusion; implants may be removed after solid fusion mass develops.

Clinical Evidence

Bench testing only.

Technological Characteristics

Constructed from medical grade titanium alloy (Ti6Al4V-ELI). System includes pedicle screws, rods, cross connectors, and staples. Mechanical fixation device.

Indications for Use

Indicated for skeletally mature patients requiring spinal immobilization/stabilization as an adjunct to fusion. Conditions include degenerative spondylolisthesis with neurological impairment, severe spondylolisthesis (grade 3-4 at L5-S1), degenerative disc disease, fracture, dislocation, scoliosis, kyphosis, lordosis, spinal tumor, pseudarthrosis, and spinal stenosis. Used in thoracic, lumbar, and sacral spine.

Regulatory Classification

Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards; (ii) Compliance with mechanical testing standards; (iii) Compliance with biocompatibility standards; and (iv) Labeling that contains these two statements in addition to other appropriate labeling information: “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.” “Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.” (2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls: (i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use. (ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant. (iii) Device components must be demonstrated to be biocompatible. (iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments. (v) Labeling must include the following: (A) A clear description of the technological features of the device including identification of device materials and the principles of device operation; (B) Intended use and indications for use, including levels of fixation; (C) Identification of magnetic resonance (MR) compatibility status; (D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and (E) Detailed instructions of each surgical step, including device removal. (3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls: (i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate. (ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Custom Spine, Inc. #### K070281 # 510(k) Summary according to 807.92(c) JUN 1 1 2007 | Company Contact: | Aaron Markworth<br>Custom Spine<br>322 US Highway 46W<br>Suite 270E<br>Parsippany, NJ 07054<br>Phone: 973-808-0019<br>Fax: 973-808-0707 | |------------------|-----------------------------------------------------------------------------------------------------------------------------------------| | Trade Name: | ISSYS LP Spinal Fixation System | | Classification: | 21 CFR §888.3070 - Pedicle Screw Spinal System<br>And, §888.3050 Spinal Interlaminal Fixation Orthosis | | Class: | III | | Product Codes: | MNI, MNH, KWP, NKB | #### Indications For Use: For pedicular use: When used as pedicle screw fixation system of the non cervical posterior spine in skeletally mature patients, these systems are indicated for one or more of the following: degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis), In addition, this system is indicated for pedicle screw fixation in skeletally mature patients with severe spondylolisthesis (grade 3 & 4) at the L5-S1 joint having fusion with autogenous bone graft, having the device fixed or attached to the lumbar and sacral spine (with pedicle placement at L3 and below) with removal of the implants after the development of a solid fusion mass. The ISSYS LP Spinal Fixation System is also intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either sever spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. #### When used as non pedicular fixation system: The ISSYS LP Spinal Fixation Systems, when used as an anterior screw fixation system and posterior sacral/iliac screw fixation system are indicated for the following: - · Degenerative disc disease of the thoracic and lumbar spine (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies). - o Spondylolisthesis Confidential ~~85~~ pg 1 of 2 {1}------------------------------------------------ Custom Spine, Inc. ### K070281 દિવે - Fracture o - Spinal deformities such as scoliosis, kyphosis, lordosis o - o Tumor - Revision of failed fusion attempts o - o Pseudarthrosis - Spinal stenosis o When used in the anterior indication the ISSYS LP Spinal Fixation Systems are indicated for use in the thoracic and lumbar spine. # Device Description: The ISSYS LP Spinal Fixation System is comprised of a variety of pedicle screws, rods, cross connectors and staples sizes that may uniquely fitted for each individual case. All implants are manufactured from medical grade titanium alloy (Ti6A14V-ELI). # Predicate Device(s): ISSYS LP Spinal Fixation System was shown to be substantially equivalent to previously cleared devices and has the same indications for use, design, function, and materials used. # Performance Testing: Pre-clinical testing performed indicated that the ISSYS LP Spinal Fixation System is substantially equivalent to predicate devices. Confidential {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" arranged around the top half of the circle. Inside the circle is an abstract symbol resembling an eagle or bird in flight, rendered in a stylized, minimalist design. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Custom Spine % Richard Jansen, Pharm. D. Silver Pine Consulting 13540 Guild Avenue Apple Valley, Minnesota 55124 Re: K070281 Trade/Device Name: ISSYS LP Spinal Fixation System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class III Product Code: NKB, KWP, MNI, MNH Dated: May 22, 2007 Received: May 23, 2007 JUN 1 1 2007 Dear Dr. Jansen: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally {3}------------------------------------------------ Page 2 - Richard Jansen, Pharm. D. marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or on the Internet address http://www.fda.gov/cdrh/industry/support/index.html Sincerely yours, for Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Indications for Use 510(k) Number (if known): K670281 Indications for Use: The ISSYS LP Spinal Fixation System is intended to help provide immobilization of spinal segments as an adjunct to fusion of the lumbar and/or sacral spine, specifically as follows: For pedicular use: When used as pedicle screw fixation system of the non cervical posterior spine in skeletally mature patients, these systems are indicated for one or more of the following: degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). In addition, this system is indicated for pedicle screw fixation in skeletally mature patients with severe spondylolisthesis (grade 3 & 4) at the L5-S1 joint having fusion with autogenous bone graft, having the device fixed or attached to the lumbar and sacral spine (with pedicle placement at L3 and below) with removal of the implants after the development of a solid fusion mass. # When used as non pedicular fixation system: The ISSYS LP Spinal Fixation Systems, when used as an anterior screw fixation system and posterior sacral/iliac screw fixation system are indicated for the following: - Degenerative disc disease of the thoracic and lumbar spine (defined as back pain of discogenic origin with 0 degeneration of the disc confirmed by history and radiographic studies). - Spondylolisthesis O - Fracture 0 - Spinal deformities such as scoliosis, kyphosis, lordosis 0 - Tumor 0 - Revision of failed fusion attempts 0 - Pseudarthrosis ට - Spinal stenosis 0 When used in the anterior indication the ISSYS LP Spinal Fixation Systems are indicated for use in the thoracic and lumbar spine. Prescription Use V (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Nol Division of and Neurological De 510(k) Number 4070281
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...