K072866 · Custom Spine, Inc. · KWP · Dec 13, 2007 · Orthopedic
Device Facts
Record ID
K072866
Device Name
ISSYS LP SPINAL FIXATION SYSTEM
Applicant
Custom Spine, Inc.
Product Code
KWP · Orthopedic
Decision Date
Dec 13, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3050
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The ISSYS LP Spinal Fixation System is intended to help provide immobilization and stabilization of the spinal segments as an adjunct to fusion of the lumbar and/or sacral spine, specifically as follows: For pedicular use: When used as pedicle screw fixation system of the non cervical posterior spine in skeletally mature patients, these systems are indicated for one or more of the following: degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). In addition, this system is indicated for pedicle screw fixation in skeletally mature patients with severe spondylolisthesis (grade 3 & 4) at the L5-S1 joint having fusion with autogenous bone graft, having the device fixed or attached to the lumbar and sacral spine (with pedicle placement at L3 and below) with removal of the implants after the development of a solid fusion mass. When used as non pedicular fixation system: The ISSYS LP Spinal Fixation System, when used as an anterior screw fixation system and posterior sacral/iliac screw fixation system are indicated for the following: Degenerative disc disease of the thoracic and lumbar spine (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies). Spondylolisthesis. Fracture. Spinal deformities such as scoliosis, kyphosis, lordosis. Tumor. Revision of failed fusion attempts. Pseudarthrosis. Spinal stenosis. When used in the anterior indication the ISSYS LP Spinal Fixation System are indicated for use in the thoracic and lumbar spine.
Device Story
ISSYS LP Spinal Fixation System provides spinal immobilization and stabilization as an adjunct to fusion. System components include 5.5 mm, 6.0 mm, and 6.35 mm diameter rods used with polyaxial and monoaxial screws. Device implanted by surgeons in clinical settings to treat spinal pathologies including degenerative disc disease, spondylolisthesis, fractures, and deformities. Mechanical fixation achieved via pedicular or non-pedicular (anterior/posterior sacral/iliac) attachment to lumbar/sacral spine. Implants provide structural support to facilitate bone fusion; removed after solid fusion mass development. Benefits include stabilization of spinal segments and correction of deformities.
Clinical Evidence
Bench testing only. Performance data per ASTM F 1717 submitted to characterize system components.
Technological Characteristics
Materials: ASTM F-136 implant grade titanium alloy. Components: 5.5 mm, 6.0 mm, 6.35 mm diameter rods; polyaxial and monoaxial screws. Mechanical fixation system. No software or energy source.
Indications for Use
Indicated for skeletally mature patients requiring spinal immobilization/stabilization as an adjunct to fusion in lumbar/sacral spine. Pedicular use: degenerative spondylolisthesis with neurological impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, pseudarthrosis, or severe spondylolisthesis (grade 3-4) at L5-S1. Non-pedicular use (anterior/posterior sacral/iliac): degenerative disc disease, spondylolisthesis, fracture, spinal deformities (scoliosis, kyphosis, lordosis), tumor, revision of failed fusion, pseudarthrosis, or spinal stenosis.
Regulatory Classification
Identification
A spinal interlaminal fixation orthosis is a device intended to be implanted made of an alloy, such as stainless steel, that consists of various hooks and a posteriorly placed compression or distraction rod. The device is implanted, usually across three adjacent vertebrae, to straighten and immobilize the spine to allow bone grafts to unite and fuse the vertebrae together. The device is used primarily in the treatment of scoliosis (a lateral curvature of the spine), but it also may be used in the treatment of fracture or dislocation of the spine, grades 3 and 4 of spondylolisthesis (a dislocation of the spinal column), and lower back syndrome.
K070281 — ISSYS LP SPINAL FIXATION SYSTEM · Custom Spine, Inc. · Jun 11, 2007
K023438 — MODIFICATION TO MONARCH SPINE SYSTEM · Depuyacromed · Nov 13, 2002
K022949 — SYNTHES USS, CLICK'X, USS VAS, DUAL-OPENING USS AND SMALL STATURE USS · Synthes (Usa) · Mar 24, 2003
K061524 — ISS INTEGRATED SPINE SYSTEM · Ortho Development Corp. · Sep 26, 2006
K013444 — ISOBAR SPINAL SYSTEM · Scien'Tx USA, Inc. · Nov 16, 2001
Submission Summary (Full Text)
{0}------------------------------------------------
# 7. 510(k) Summary
## As Required By Section 807.92 (c)
| Submitter: | Custom Spine, Inc.<br>1140 Parsippany Blvd<br>Suite 201<br>Parsippany, NJ 07054 | DEC 1 3 2007 |
|-----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|
| Contact Person: | Saad Attiyah<br>Manager of Regulatory Affairs and Quality Assurance<br>Voice : (973) 265-5036<br>Fax : (973) 508-0707<br>E-Mail: saad@customspine.com | |
| Date Prepared: | September 27, 2007 | |
| Device Class: | III | |
| Classification Name: | Pedicle Screw Fixation System, Spinal Interlaminal Fixation Orthosis<br>Per 21 CFR §888.3070, 21 CFR §888.3050 | |
| Classification Panel: | Orthopedics | |
| Product Code: | NKB, KWP, MNI, MNH | |
| Proprietary Name: | ISSYS LP Spinal Fixation System | |
| Predicate Devices: | ISSYSTM LP Spinal Fixation System (K070281)<br>ISSYSTM Pedicle Screw System (K043522)<br>Moss® Miami Spinal System (K933881, K955348, K964024, K983583, K022623)<br>Optima® Spinal System (K031585) | |
| Device Description: | The subject ISSYS LP Polyaxial Spinal Fixation System includes 5.5<br>mm diameter rods (preformed rod or straight rod configuration) in<br>addition to the selection of the cleared 6.0 mm and 6.35 mm diameter<br>rods. It is to be used with both the Polyaxial and Monoaxial screws | |
| Intended Use: | The ISSYS LP Spinal Fixation System is intended to help provide<br>immobilization and stabilization of spinal segments as an adjunct to<br>fusion of the lumbar and/or sacral spine, specifically as follows: | |
| | For pedicular use: When used as pedicle screw fixation system of the<br>non cervical posterior spine in skeletally mature patients, these systems<br>are indicated for one or more of the following: degenerative | |
{1}------------------------------------------------
Special 510(k) Submission - Addition to ISSYS LP Spinal Fixation System
spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis).
In addition. this system is indicated for pedicle screw fixation in skeletally mature patients with severe spondylolisthesis (grade 3 & 4) at the L5-S1 joint having fusion with autogenous bone graft, having the device fixed or attached to the lumbar and sacral spine (with pedicle placement at L3 and below) with removal of the implants after the development of a solid fusion mass.
### When used as non pedicular fixation system:
The ISSYS LP Spinal Fixation Systems, when used as an anterior screw fixation system and posterior sacral/iliac screw fixation system are indicated for the following:
- Degenerative disc disease of the thoracic and lumbar spine (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies).
- 제 Spondylolisthesis
- 트 Fracture
- E Spinal deformities such as scoliosis, kyphosis, lordosis
- E Tumor
- 트 Revision of failed fusion attempts
- 트 Pseudarthrosis
- 피 Spinal stenosis
When used in the anterior indication the ISSYS LP Spinal Fixation Systems are indicated for use in the thoracic and lumbar spine.
- Materials : Manufactured from ASTM F-136 implant grade titanium alloy.
- Performance Data: Performance data per ASTM F 1717 were submitted to characterize the subject ISSYS™ LP Spinal Fixation System components address in this notification.
Summary of Technological Characteristics
Documentation is provided which demonstrates the additional 5.5 mm diameter rod of the ISSYS LP Spinal Fixation System to be substantially equivalent to the predicate devices in terms of material, design, and indications for use. Engineering analysis and testing to demonstrate compliance with FDA's Guidance For Spinal System 510(k) May 3, 2004 was completed for the ISSYS LP Spinal Fixation System, included the subject component.
pg 2 of 2
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" is arranged in a circular fashion around the caduceus symbol. The logo is black and white.
Public Health Service
DEC 1 3 2007
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Custom Spine, Inc. % Mr. Saad Attiyah Manager, Regulatory Affairs & Quality Assurance 1140 Parsippany Blvd., Suite 201 Parsippany, NJ 07054
Re: K072866
Trade/Device Name: ISSYS LP Spinal Fixation System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: III Product Code: NKB, KWP, MNI, MNH Dated: December 8, 2007 Received: December 13, 2007
Dear Mr. Attiyah:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 – Mr. Saad Attiyah
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Mark N Millerson
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
### 5. Indications for Use
510(k) Number (if known):K072866
Device Name: ISSYS LP Spinal Fixation System
Indications For Use:
The ISSYS LP Spinal Fixation System is intended to help provide immobilization and stabilization of the spinal segments as an adjunct to fusion of the lumbar and/or sacral spine, specifically as follows:
For pedicular use: When used as pedicle screw fixation system of the non cervical posterior spine in skeletally mature patients, these systems are indicated for one or more of the following: degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis).
In addition, this system is indicated for pedicle screw fixation in skeletally mature patients with severe spondylolisthesis (grade 3 & 4) at the L5-S1 joint having fusion with autogenous bone graft, having the device fixed or attached to the lumbar and sacral spine (with pedicle placement at L3 and below) with removal of the implants after the development of a solid fusion mass.
#### When used as non pedicular fixation system:
The ISSYS LP Spinal Fixation System, when used as an anterior screw fixation system and posterior sacral/iliac screw fixation system are indicated for the following:
- Degenerative disc disease of the thoracic and lumbar spine (defined as back pain of discogenic . origin with degeneration of the disc confirmed by history and radiographic studies).
- Spondylolisthesis ●
- Fracture .
- . Spinal deformities such as scoliosis, kyphosis, lordosis
- Tumor .
- . Revision of failed fusion attempts
- Pseudarthrosis t
- Spinal stenosis ●
When used in the anterior indication the ISSYS LP Spinal Fixation System are indicated for use in the thoracic and lumbar spine.
Prescription Use _ X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
barbare Bneur
510(k) Number Ko72906
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.